← Product Code [DPS](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS) · K082967

# MORTARA STRESS ECHO BED (K082967)

_Medical Positioning, Inc. · DPS · Mar 26, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K082967

## Device Facts

- **Applicant:** Medical Positioning, Inc.
- **Product Code:** [DPS](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS.md)
- **Decision Date:** Mar 26, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The M-Stress Echo™ Bed is intended for use in stress echocardiography examination. The M-Stress Echo™ Bed provides an exercise source that delivers programmable, controlled variable resistance. The M-Stress Echo™ Bed incorporates an electrocardiograph that records either normal conditions or patterns of arrhythmia and/or rate abnormalities in patients. In addition, the stress echo workstation provides "QRS" complex to a cardiac ultrasound device to be used to capture images (heart beats), either digitally or on videotape, such that each image begins at the time systole begins.

## Device Story

Integrated stress echocardiography system; combines supine bicycle exercise table with Mortara X-Scribe II electrocardiograph. Exercise table provides programmable, controlled variable resistance for patient exertion. ECG monitors cardiac electrical activity; detects arrhythmias and rate abnormalities. System outputs QRS complex signal to external cardiac ultrasound device; enables image capture synchronized with onset of systole. Used in clinical settings; operated by healthcare professionals. Benefits include standardized exercise testing and improved temporal alignment of ultrasound images with cardiac cycle.

## Clinical Evidence

Bench testing only. Physical performance studies and software evaluation conducted to verify intended performance. No clinical data provided.

## Technological Characteristics

Integrated system comprising a supine bicycle exercise table and Mortara X-Scribe II electrocardiograph. Features include height adjustability, Trendelenburg, dual/lateral tilt, and computer controllers. Connectivity via standard patient electrodes/leads. ECG provides reports, trends, averages, and ST segment analysis. Software performs ECG monitoring and QRS complex signal generation.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- Vertex System ([K002822](/device/K002822.md))
- Q-Stress Echo Bed (no number provided)
- Mortara X-Scribe II ([K022618](/device/K022618.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KO82967

# SECTION 16: SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.

MAR 2 6 2009

#### 16.1 SUBMITTER INFORMATION

| a. Company Name:                        | Medical Positioning, Inc.                                    |
|-----------------------------------------|--------------------------------------------------------------|
| b. Company Address:                     | 1717 Washington<br>Kansas City, MO 64108                     |
| c. Company Phone:<br>Company Facsimile: | (816) 474-1555<br>(816) 474-7755                             |
| d. Contact Person:                      | Michael G. Falbo, Sr.<br>Vice President, Product Development |
| e. Date Summary Prepared:               | September 30, 2008                                           |

#### DEVICE IDENTIFICATION 16.2.

| a. Trade/Proprietary Name: | Mortara Stress Echo™ Bed              |
|----------------------------|---------------------------------------|
| b. Classification Name:    | Electrocardiograph<br>21 CFR 870.2340 |

#### 16.3 IDENTIFICATION OF PREDICATE DEVICE

| Company             | Device              | 510(k) No. | Date Cleared |
|---------------------|---------------------|------------|--------------|
| Medical Positioning | Vertex System       | K002822    | 01/30/01     |
| Medical Positioning | Q-Stress Echo Bed   | -          | 05/08/02     |
| Mortara Instruments | Mortara X-Scribe II | K022618    | 09/06/02     |

{1}------------------------------------------------

#### 16.4 DEVICE DESCRIPTION

The Mortara Stress Echo™ Bed is a complete integrated stress echocardiography system. The Mortara Stress Echo™ Bed combines the Stress Echo Bed / Table with an electrocardiograph. The Stress Echo Bed provides an exercise source that delivers programmable, controlled variable resistance, while the ECG provides the patient monitoring and recording. Several models of the Stress Echo Bed/Table are available with features that include height adjustability, Trendelenburg, dual, lateral tilt, and computer controllers.

#### SUBSTANTIAL EQUIVALENCE 16.5

The Mortara Stress Echo™ Bed is substantially equivalent to the Vertex System and the Q-Stress Echo™ currently in commercial distribution by Medical Positioning. The Mortara Stress Echo™ Bed and the predicate devices incorporate the same Stress Echo™ Bed and an electrocardiograph. The Mortara Stress Echo™ Bed incorporates the Mortara X-Scribe II electrocardiograph manufactured by Mortara Instruments.

The fundamental technical characteristics of the Mortara Stress Echo™ Bed and the Vertex System and Q-Stress Echo Bed are equivalent and are listed on the comparison charts provided in this 510(k) submission. The Mortara Stress Echo™ Bed and the predicate devices function by providing the user with an integrated exercise source and electrocardiograph for use during cardiovascular monitoring.

#### INTENDED USE 16.6

The Mortara Stress Echo™ Bed is intended for use in stress echocardiography examination. The Mortara Stress Echo™ Bed provides an exercise source that delivers programmable, controlled variable resistance.

The Mortara Stress Echo™ Bed incorporates an electrocardiograph that records either normal conditions or patterns of arrhythmia and/or rate abnormalities in

{2}------------------------------------------------

patients. In addition, the Mortara Stress Echo™ Bed provides "ORS" complex to a cardiac ultrasound device to be used to capture images (heart beats), either digitally or on videotape, such that each image begins at the time systole begins.

#### TECHNOLOGICAL CHARACTERISTICS 16.7

The technological characteristics of the Mortara Stress Echo™ Bed are equivalent to those of the Vertex System and the Q-Stress Echo™ Bed. The Mortara Stress Echo™ Bed utilizes a supine bicycle for the exercise source. Preprogrammed exercise protocols are run for purposes of electrocardiographic monitoring. The ECG used in the Mortara Stress Echo™ Bed is the Mortara X-Scribe II electrocardiograph that has been cleared for commercial distribution under K022618. ECG reports, trends, averages and ST segments are printed by the Mortara Stress Echo™ Bed. The Mortara Stress Echo™ Bed is connected using standard patient electrodes and leads that are not included in the system.

#### 16.8 PERFORMANCE DATA

The Mortara Stress Echo™ Bed was subjected to performance bench testing. Physical performance studies and software evaluation were conducted to verify that the Mortara Stress Echo™ Bed performed as intended.

#### CONCLUSIONS 16.9

This notification contains all information required by 21 CFR 807.87. The Mortara Stress Echo™ Bed was found to perform as intended during verification and validation testing. The Mortara Stress Echo™ Bed is substantially equivalent to the current Vertex System and the Q-Stress Echo™ Bed in commercial distribution. The Mortara Stress Echo™ Bed is intended for use in stress echocardiography examination. The Mortara Stress Echo™ Bed provides an exercise source that delivers programmable, controlled variable resistance,

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 6 2009

Medical Positioning, Inc. c/o Ms. Carol White Consultant 21521 Hummingbird St. Trabuco Canyon, CA 92679

Re: K082967

Trade Name: M-Stress Echo Bed Regulation Number: 21 CFR 870. 2340 Regulation Name: Electrocariograph Regulatory Class: Class II (two) Product Code: DPS Dated: February 23, 2009 Received: March 2, 2009

Dear Ms. White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

{4}------------------------------------------------

### Page 2 - Ms. Carol White

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours

Bram D. Zukerman, M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name:

K082967

M-Stress Echo Bed TM

Indications For Use:

The M-Stress Echo™ Bed is intended for use in stress echocardiography examination. The M-Stress Echo™ Bed provides an exercise source that delivers programmable, controlled variable resistance.

The M-Stress Echo™ Bed incorporates an electrocardiograph that records either normal conditions or patterns of arrhythmia and/or rate abnormalities in patients. In addition, the stress echo workstation provides "QRS" complex to a cardiac ultrasound device to be used to capture images (heart beats). either digitally or on videotape, such that each image begins at the time systole begins.

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Offm
Division of Cardiovascular Devices
510(k) Number K082967-
- Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K082967](https://fda.innolitics.com/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/DPS/K082967)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
