Serranator PTA Serration Balloon Catheter

K220704 · Cagent Vascular, LLC · PNO · May 4, 2022 · Cardiovascular

Device Facts

Record IDK220704
Device NameSerranator PTA Serration Balloon Catheter
ApplicantCagent Vascular, LLC
Product CodePNO · Cardiovascular
Decision DateMay 4, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Serranator® PTA Serration Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, iliofemoral, popliteal, infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

Device Story

Serranator PTA Serration Balloon Catheter is an over-the-wire (OTW) balloon dilatation catheter; used for percutaneous transluminal angioplasty (PTA). Device features nylon semicompliant balloon with embedded external metal serrated scoring elements. During inflation, elements create linear, interrupted scoring along endoluminal surface to modify plaque and aid arterial expansion. Used in clinical settings by physicians for peripheral vascular interventions. Output is mechanical dilation of target vessel. Benefits include improved vessel patency through controlled plaque modification.

Clinical Evidence

No clinical data provided. Bench and pre-clinical testing leveraged from previously cleared submissions K213728 and K203437.

Technological Characteristics

Over-the-wire (OTW) balloon dilatation catheter; nylon semicompliant balloon; embedded external metal serrated scoring elements for plaque modification. Mechanical energy source. Dimensions vary by balloon diameter (2.5mm to 6.0mm) and length (40mm to 120mm).

Indications for Use

Indicated for patients requiring dilatation of lesions in iliac, femoral, iliofemoral, popliteal, and infrapopliteal arteries, or treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Contraindicated for use in coronary or neuro-vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 4, 2022 Cagent Vascular, LLC Ms. Carol Burns President and CEO 150 Strafford Avenue #315 Wayne, Pennsylvania 19087 Re: K220704 Trade/Device Name: Serranator PTA Serration Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PNO Dated: March 8, 2022 Received: March 10, 2022 Dear Ms. Burns: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K220704 Device Name Serranator® PTA Serration Balloon Catheter Indications for Use (Describe) The Serranator® PTA Serration Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, illofemoral, popliteal, infrapopliteal arteries and for the treatment of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------|----------------------------------------------------------| | <div><span style="font-size:14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>□ Over-The-Counter Use (21 CFR 801 Subpart C)</div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 1 K220704 Image /page/3/Picture/1 description: The image shows the logo for Cagent Vascular. The logo features the word "CAGENT" in a bold, blue font, with the exclamation point in "CAGENT" being red. Below "CAGENT" is the word "VASCULAR" in a smaller, blue font. To the right of "CAGENT" is the registered trademark symbol. The logo is surrounded by a broken circle made of small, light green dashes. Cagent Vascular, Inc. 150 Strafford Avenue #315 Wayne, PA 19087 # 510(k) Summary # Serranator® PTA Serration Balloon Catheter Cagent Vascular, Inc. ## K220704 | DATE PREPARED | May 4, 2022 | |-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | APPLICANT INFORMATION | Carol Burns/President & CEO<br>150 Strafford Avenue #315<br>Wayne, PA 19087 | | CONTACT INFORMATION | Carol A. Burns, President and CEO<br>Phone: (610) 688-2006<br>Fax: (610) 688-2667<br>Email: cburns@cagentvascular.com | | TRADE NAME | Serranator® PTA Serration Balloon Catheter | | DEVICE CLASSIFICATION | Class 2 per 21 CFR §870.1250 | | CLASSIFICATION NAME | Percutaneous Catheter | | PRODUCT CODE | PNO | | PREDICATE DEVICE | Predicate Device: Serranator® PTA Serration Balloon Catheter (K213728)<br>Reference Device: Serranator® PTA Serration Balloon Catheter (K203437) | # 5.1 Intended Use/Indications for Use The Serranator® PTA Serration Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, iliofemoral, popliteal, infrapopliteal arteries and for the treatment of obstructive {4}------------------------------------------------ lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature. # 5.2 Device Description The Serranator® PTA Serration Balloon Catheter is an over-the-wire (OTW) balloon dilatation catheter designed to perform percutaneous transluminal angioplasty (PTA) for peripheral indications as described in the Indication for Use statement. The Serranator® has a nylon semicompliant balloon with multiple embedded external metal strips or scoring elements. The unique scoring elements are serrated, designed to modify the plaque by creating linear, interrupted scoring along the endoluminal surface. This occurs during balloon inflation and is designed to aid arterial expansion. This 510(k) Summary represents a consolidation (i.e., bundling) of previously cleared 510(k) indication statements and labeling. Bundled devices are identical to previously cleared devices, the predicate device Serranator® PTA Serration Balloon Catheter (K213728) and reference device Serranator® PTA Serration Balloon Catheter (K203437). The bundled device sizes are shown in Table 5.1. | | Balloon<br>Nominal OD<br>[mm] | Balloon Length<br>[mm] | | | | |--------------------------------------------------------------|-------------------------------|------------------------|-------|-----|--------------------------------------------------------------------------------| | | | 40 | 80 | 120 | | | Subject Device Serranator® PTA<br>Serration Balloon Catheter | 2.5 | ✓ | ✓ | ✓ | Reference Device<br>Serranator® PTA<br>Serration Balloon<br>Catheter (K203437) | | | 3.0 | ✓ | ✓ | ✓ | | | | 3.5 | ✓ | ✓ | ✓ | | | | 4.0 | ✓ | | ✓ | Predicate Device | | | 5.0 | ✓ | N.A.* | ✓ | Serranator® PTA<br>Serration Balloon<br>Catheter (K213728) | | | 6.0 | ✓ | | ✓ | | | .1<br>> | Table<br>U | | |---------|------------|--| |---------|------------|--| *The 80mm length size is commercially available for the below the knee sizes only. ### 5.3 Comparison to Predicate and Reference Devices A comparison of the Serranator® PTA Serration Balloon Catheter (Serranator) to the predicate device shows that the technological characteristics such as the components, design, materials, {5}------------------------------------------------ sterilization method, and operating principle of the Serranator are identical to the currently marketed Serranator® PTA Serration Balloon Catheter (Serranator) (K213728) and the Serranator® PTA Serration Balloon Catheter (Serranator) (K203437). The intended use of the subject device is identical to the combination of the predicate device, Serranator® PTA Serration Balloon Catheter (K213728) and the Serranator® PTA Serration Balloon Catheter (Serranator) (K203437), as shown in Table 5.2. | Table 5.2 | |--------------------------------------------------------------------| | Indications for Use statements of the Subject and Predicate Device | | | Device Name | Indications For Use | |---------------------|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Subject Device | Serranator®<br>PTA Serration<br>Balloon Catheter | The Serranator® PTA Serration Balloon Catheter is<br>intended for dilatation of lesions in the iliac, femoral,<br>iliofemoral, popliteal, infrapopliteal arteries and for the<br>treatment of obstructive lesions of native or synthetic<br>arteriovenous dialysis fistulae. Not for use in the<br>coronary or neuro-vasculature. | | Predicate<br>Device | Serranator®<br>PTA Serration<br>Balloon Catheter<br>(K213728) | The Serranator® PTA Serration Balloon Catheter is<br>intended for dilatation of lesions in the iliac, femoral,<br>iliofemoral, and popliteal arteries and for the treatment of<br>obstructive lesions of native or synthetic arteriovenous<br>dialysis fistulae. Not for use in the coronary or neuro-<br>vasculature. | | Reference<br>Device | Serranator®<br>PTA Serration<br>Balloon Catheter<br>(K203437) | The Serranator® PTA Serration Balloon Catheter is<br>intended for dilatation of lesions in infrapopliteal arteries.<br>Not for use in the coronary or neuro-vasculature. | ### 5.4 Previous Non-Clinical and Pre-Clinical Testing The devices represented herein are identical to the predicate and reference devices (Serranator® PTA Serration Balloon Catheter (Serranator) (K213728) and the Serranator® PTA Serration Balloon Catheter (Serranator) (K203437). All non-clinical and pre-clinical testing has been leveraged from those submissions. ### 5.5 Conclusion No new data are included in this bundled 510(K) application. The Serranator® PTA Serration Balloon Catheters represented in this bundled submission do not differ in purpose, design, materials, energy source, function, or any other feature. Performance data previously provided in K213728 and K203437 demonstrate that the devices function as intended.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...