K163420 · Bard Peripheral Vascular, Inc. · PNO · Jun 14, 2017 · Cardiovascular
Device Facts
Record ID
K163420
Device Name
UltraScore Focused Force PTA Balloon
Applicant
Bard Peripheral Vascular, Inc.
Product Code
PNO · Cardiovascular
Decision Date
Jun 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UltraScore Focused Force PTA Balloon is intended to dilate stenoses in the iliac, femoral, popliteal, infra-popliteal and renal arteries and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also recommended for post dilatation of balloon expanding stents, and stent grafts in the peripheral vasculature.
Device Story
Over-the-wire (OTW) PTA catheter with semi-compliant balloon; features two scoring wires oriented 180° apart to provide focused force during dilatation. Used by physicians in clinical settings to treat peripheral arterial stenoses and AV fistulae; also used for post-dilatation of stents/stent grafts. Input: manual inflation via luer lock hub. Transformation: balloon expansion with scoring wires concentrating force on lesion. Output: mechanical dilation of vessel/fistula. Benefits: improved vessel patency; reduced risk of vessel injury through focused force. Radiopaque markers aid placement; atraumatic tip facilitates navigation. Compatible with .014" or .035" guidewires.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated via in vitro testing including dimensional analysis, balloon compliance, trackability, sheath/guidewire compatibility, inflation/deflation times, rated burst pressure, fatigue testing, and tensile/torque strength. Biocompatibility testing performed per ISO 10993-1:2009, including cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility, and material-mediated pyrogenicity.
Technological Characteristics
OTW catheter with semi-compliant balloon; two scoring wires (180° apart). Materials: hydrophilic coating on distal portion (.014" version). Dimensions: 2.0-8.0mm diameter; 20-300mm length. Connectivity: N/A. Sterilization: Not specified. Software: None. Mechanical principle: Focused force balloon angioplasty.
Indications for Use
Indicated for patients requiring dilation of stenoses in iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, or treatment of obstructive lesions in native/synthetic arteriovenous dialysis fistulae. Also indicated for post-dilatation of balloon-expandable stents, self-expanding stents, and stent grafts in peripheral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 14, 2017
Bard Peripheral Vascular, Inc. Mr. Timothy Wade Regulatory Affairs Project Manager 1625 West 3rd St Tempe, AZ 85281
Re: K163420
Trade/Device Name: UltraScore Focused Force PTA Balloon Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: PNO Dated: May 4, 2017 Received: May 5, 2017
Dear Mr. Wade:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163420
Device Name UltraScore Focused Force PTA Balloon
#### Indications for Use (Describe)
The UltraScore Focused Force PTA Balloon is intended to dilate stenoses in the iliac, femoral, popliteal, infra-popliteal and renal arteries and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. This device is also recommended for post dilatation of balloon expanding stents, and stent grafts in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# ULTRASCORE™ Focused Force PTA Balloon 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
## Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|--------------------------------------------------------------------------------|
| Phone: | 480 638 2939 |
| Fax: | 480 449 2546 |
| Contact: | Timothy Wade |
- Date 05 Dec 2016
## Subject Device Name:
| Device Trade Name: | ULTRASCORE™ Focused Force PTA Balloon |
|-----------------------|-----------------------------------------|
| Regulation Number: | 21 CFR 870.1250 |
| Product Code: | PNO |
| Classification: | Class II |
| Classification Name: | Catheter, Percutaneous, Cutting/Scoring |
| Classification Panel: | Cardiovascular |
### Predicate Device:
- . VASCUTRAK® PTA Dilatation Catheter (K103459; cleared December 13, 2010)
### Reference Device:
- ULTRAVERSE® 035 PTA Dilatation Catheter (K142261; cleared September 24, ● 2014)
Image /page/3/Picture/13 description: The image shows the word "BARD" in a stylized, bold, sans-serif font. The letters are interconnected, with the "A" having a unique triangular shape. The overall design is clean and modern, with a focus on the geometric forms of the letters.
{4}------------------------------------------------
## Device Description:
The ULTRASCORE ™ Focused Force PTA Balloon consists of a flexible, over-the-wire (OTW) catheter shaft with a semi-compliant balloon fixed at the distal end. For all balloon lengths, radiopaque markers delineate the working length of the balloon and aid in balloon placement. For balloon lengths of 100 mm and greater, two radiopaque markers are positioned on the distal portion of the balloon and one radiopaque marker is positioned on the proximal portion of the balloon to differentiate between the distal and proximal ends of the balloon. The catheter includes an atraumatic tip to facilitate advancement of the catheter to and through the stenosis. Two scoring wires, oriented 180° apart, provide focused force upon dilatation. The ULTRASCORE ™ Focused Force PTA Balloon is compatible with .014" or .035" quidewires, as denoted by the product labeling. The distal portion of the .014" guidewire compatible catheters is hydrophilically coated. The proximal portion of the catheter includes a female luer lock hub connected to the catheter with a guidewire lumen and an inflation lumen. Packaged with every product is a protective sheath that is positioned over the balloon and must be removed prior to use. A stylet is placed into the tip of the catheter. These products are not made with natural rubber latex.
| Attribute | ULTRASCORE™ Focused Force PTA Balloon |
|---------------------------------------------------|----------------------------------------------|
| Balloon Diameter (mm) | 2.0, 2.5, 3.0, 3.5, 4.0, 5.0, 6.0, 7.0, 8.0. |
| Balloon Length (mm) | 20, 40, 60, 80, 100, 120, 150, 200, 250, 300 |
| Catheter Shaft Length (cm) | 40, 75, 100, 130, 150 |
| .014" Platform Introducer Sheath<br>Compatibility | 5F |
| .035" Platform Introducer Sheath<br>Compatibility | 6F |
{5}------------------------------------------------
### Indications for Use of Device:
The ULTRASCORE™ Focused Force PTA Balloon is intended to dilate stenoses in the illac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post dilatation of balloon expandable stents, self-expanding stents, and stent grafts in the peripheral vasculature.
## Comparison of Indications for Use to Predicate Device:
The indications for use statement for the ULTRASCORE™ Focused Force PTA Balloon does not raise any new issues of safety and effectiveness based on the proposed indications for use statement as compared to the predicate device. Therefore, the subject device, the ULTRASCORE™ Focused Force PTA Balloon, is substantially equivalent to the predicate device.
## Technological Comparison to Predicate Devices:
The ULTRASCORE™ Focused Force PTA Balloon has the following similarities to the predicate device, the VASCUTRAK® PTA Dilatation Catheter (clearance to market via K103459 on December 13, 2010):
- . Same intended use
- . Same indications for use
- Same target population .
- Similar operating principle 0
- o Same fundamental scientific technology
- Same sterility assurance level and method of sterilization .
The ULTRASCORE™ Focused Force PTA Balloon has the following differences when compared to the VASCUTRAK® PTA Dilatation Catheter predicate device:
- Over-the-wire platform o
- Additional 8.0mm balloon diameter offered o
- New 0.035" quidewire compatibility offered o
- Modifications to marker bands o
- Material differences .
- Addition of packaging stylet o
{6}------------------------------------------------
### Performance Data:
To demonstrate substantial equivalence of the subject device to the predicate device, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device:
- . Dimensional
- Balloon Outer Diameter o
- Balloon Working Length o
- o Catheter Shaft Length
- Balloon Compliance ●
- GEOALIGN® Gradient Marking o Position
- GEOALIGN® Gradient Marking . Legibility
- GEOALIGN® Gradient Marking ● Durability
- Trackability .
- Sheath Compatibility ●
- Guidewire Compatibility ●
- Flushability ●
- Reinsertion ●
- Inflation Time ●
- Deflation Time o
- Rated Burst Pressure ●
- . Balloon Burst Mode
- . Balloon Burst in Stent/Stent Graft
- Balloon Fatique .
- Balloon Fatigue in Stent/Stent . Graft
- Balloon Removal from . Stent/Stent Graft
- Tip Taper .
- Device Compatibility ●
- Hub Stress ●
- Hub Torque ●
- Balloon to Shaft Tensile ●
- Tip to Wire Tensile ●
- Catheter Elongation ●
- Marker Band Radiopacity ●
- Marker Band Alignment ●
- Proximal Assembly Tensile .
- Packaging Pouch Tensile ●
- . Packaging Visual Inspection
- Packaging Bubble Emission ●
The following in vitro biocompatibility testing was conducted in accordance with ISO 10993-1: 2009
- Cytotoxicity
- Sensitization
- o Intracutaneous Reactivity
- Acute Systemic Toxicity 0
- Hemocompatibility
Image /page/6/Picture/44 description: The image shows the word "BARD" in a stylized, outlined font. The letters are bold and geometric, with the "A" resembling a triangle without a horizontal crossbar. The overall design is clean and modern, with a focus on the shape and form of the letters.
{7}------------------------------------------------
- Material Mediated Pyrogenicity
The results from these tests demonstrate that the technological characteristics and performance criteria of the ULTRASCORE™ Focused Force PTA Balloon are substantially equivalent to the predicate device, and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
# Conclusions:
The subject device, the ULTRASCORE™ Focused Force PTA Balloon, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The ULTRASCORE™ Focused Force PTA Balloon is substantially equivalent to the legally marketed predicate device, VASCUTRAK® PTA Dilatation Catheter.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.