Vein360 Reprocessed Visions PV.014P Rx Digital IVUS Catheter, Vein360 Reprocessed Eagle Eye Platinum Rx Digital IVUS Catheter, Vein360 Reprocessed Eagle Eye Platinum ST Rx Digital IVUS Catheter
Applicant
Vein 360, LLC
Product Code
OWQ · Cardiovascular
Decision Date
Oct 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The Vein360 Reprocessed Visions® PV .014P RX (85910P), Visions® PV.014P RX (014R), Eagle Eye Platinum RX, and Eagle Eve Platinum ST RX Digital IVUS catheters are all designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. These devices are not currently indicated for use in the cerebral vessels. The Vein360 Reprocessed Visions® PV .014P RX (85910P), Visions® PV.014p RX (014R), Eagle Eye Platinum RX, and Eagle Eye Platinum ST RX Digital IVUS catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.
Device Story
Reprocessed single-use intravascular imaging catheters; distal 64-element cylindrical ultrasound transducer; tracks over 0.014" guidewire. Input: acoustic echoes from vessel tissue. Transformation: transducer radiates acoustic energy; detects echoes; generates real-time cross-sectional images of vessel lumen and wall structures. Used in clinical settings (coronary/peripheral procedures) by physicians. Output: real-time IVUS images displayed on Volcano s5 or CORE Series imaging systems. Clinical decision-making: provides visual assessment of vascular morphology to supplement angiography; aids in procedural guidance. Benefits: enables visualization of vessel wall structures; supports diagnostic evaluation of coronary and peripheral vasculature.
Clinical Evidence
Bench testing only. Validation included cleaning, drying, and sterilization validation; biocompatibility; simulated use; dimensional/mechanical integrity; cross/re-cross testing; electrical integrity/safety; hydrophilic coating integrity; acoustic output; image quality; system compatibility; and packaging validation.
Technological Characteristics
Rapid exchange intravascular imaging catheter; 3.5F transducer diameter; 150cm working length; 0.014" guidewire compatibility. 64-element cylindrical ultrasound transducer array. Materials: hydrophilic coated. Sterilization: Ethylene Oxide (EO) gas. Connectivity: compatible with Volcano s5 or CORE Series imaging systems. Reprocessed single-use device.
Indications for Use
Indicated for evaluation of vascular morphology in coronary and peripheral blood vessels; adjunct to conventional angiography to image vessel lumen and wall structures. Not indicated for cerebral vessels.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
K230928 — Vein360 Reprocessed Visions PV.018 Digital IVUS Catheter · Vein 360, LLC · Aug 25, 2023
K253399 — Visions® PV .014P RX Digital IVUS Catheter ; Visions® PV .018 Digital IVUS Catheter · Volcano Corporation (Dba Philips Image Guided Therapy Device · Oct 28, 2025
K200195 — NES Reprocessed Visions PV .014P RX Digital IVUS Catheter · Northeast Scientific, Inc. · Jun 22, 2020
K210655 — Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter · Innovative Health, LLC · May 25, 2022
K250592 — NES Reprocessed Visions PV .018 Digital IVUS Catheter (R-86700) · Northeast Scientific, Inc. · May 12, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
October 24, 2023
Vein 360 LLC Suzanne Meyer, CEO 4460 Lake Forest Dr Suite 230 Blue Ash, Ohio 45242-3741
## Re: K232584
Trade/Device Name: Vein360 Reprocessed Visions PV.014P RX Digital IVUS Catheter (85910P), Vein360 Reprocessed Visions PV.014P RX Digital IVUS Catheter (014R), Vein360 Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter, Vein360 Reprocessed Eagle Eye Platinum ST RX Digital IVUS Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: OWO Dated: August 23, 2023 Received: August 25, 2023
Dear Suzanne Meyer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology,
{2}------------------------------------------------
Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
The item numbers included in the scope of this submission are as follows:
| Subject Device Trade Name | Reference<br>Number | French<br>Size | Guide<br>Wire | Minimum<br>Sheath | Working<br>Length |
|-----------------------------------------------------------------------|---------------------|----------------|---------------|-------------------|-------------------|
| Vein360 Reprocessed Visions PV.014P RX<br>Digital IVUS Catheter | VEN-PV-014P | 3.5F | 0.014" | 5F | 150 cm |
| Vein360 Reprocessed Eagle Eye Platinum<br>RX Digital IVUS Catheter | VEN-PV-EEP | 3.5F | 0.014" | 5F | 150 cm |
| Vein360 Reprocessed Eagle Eye Platinum<br>ST RX Digital IVUS Catheter | VEN-PV-EEPST | 3.5F | 0.014" | 5F | 150 cm |
| Vein360 Reprocessed Visions PV.014P RX<br>Digital IVUS Catheter | VEN-PV-014R | 3.5F | 0.014" | 5F | 150 cm |
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232584
#### Device Name
Vein360 Reprocessed Visions PV.014P RX Digital IVUS Catheter (85910P), Vein360 Reprocessed Visions PV.014P RX Digital IVUS Catheter (014R), Vein360 Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter, Vein360 Reprocessed Eagle Eye Platinum ST RX Digital IVUS Catheter
#### Indications for Use (Describe)
The Vein360 Reprocessed Visions® PV .014P RX (85910P), Visions® PV.014P RX (014R), Eagle Eye Platinum RX, and Eagle Eve Platinum ST RX Digital IVUS catheters are all designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. These devices are not currently indicated for use in the cerebral vessels.
The Vein360 Reprocessed Visions® PV .014P RX (85910P), Visions® PV.014p RX (014R), Eagle Eye Platinum RX, and Eagle Eye Platinum ST RX Digital IVUS catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Vein 360. The word "VEIN" is written in large, bold, black letters above the number "360", which is also written in large, bold, black letters. A blue and green arrow wraps around the number 360 in a semi-circle.
August 23rd, 2023 Date of Preparation:
# Company Name / Contact:
- Company: Vein360, LLC 4460 Lake Forest Drive Suite 230 Blue Ash, OH 45242
- Contact: Suzanne Meyer CEO Phone: (513) 554-1300
# Device Identification:
| Proprietary<br>Names: | Vein360 Reprocessed Visions PV.014P RX Digital IVUS Catheter (85910P)<br>Vein360 Reprocessed Eagle Eye Platinum RX Digital IVUS Catheter<br>Vein360 Reprocessed Eagle Eye Platinum ST RX Digital IVUS Catheter<br>Vein360 Reprocessed Visions PV.014P RX Digital IVUS Catheter (014R) |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common<br>Name: | Diagnostic Intravascular Catheter |
| Classification<br>Reference: | 21 CFR 870.1200 |
| Classification<br>Panel: | Cardiovascular |
| Device<br>Product Code: | OWQ |
| Regulatory<br>Class: | Class II |
## Predicate Devices:
The primary predicate device for this submission is K143701. The secondary predicate device for this submission is K152829.
# Device Description:
The subject devices are reprocessed single use devices. After clinical use of the predicate devices (Manufactured by Philips), the devices are shipped to Vein360 per established Vein360 instructions. Upon receipt, the subject devices are cleaned, inspected, hydrophilic coated, functionally tested, packaged, and sterilized using ethylene oxide (EO) gas.
{6}------------------------------------------------
The subject devices are rapid exchange intravascular imaging catheters containing an ultrasound transducer located at the distal end of the catheter. This transducer utilizes a 64-element cylindrical array that radiates acoustic energy into the surrounding tissue and detects subsequent echoes. The information from the echoes is used to generate real-time images of the coronary and peripheral vessels.
The subject devices utilize an internal lumen that allows the catheter to track over the 0.014" (0.36mm) guide wire. The guide wire exits from the guide wire lumen approximately 24 cm proximal to the catheter tip. The subject devices are introduced percutaneously or via surgical cutdown into the vascular system.
Three 1 mm-long radiopaque markers are incorporated on the internal lumen positioned 10 mm apart from distal edge to distal edge, starting 10 mm from the proximal edge of the portion of the scanner marker tube normally visible under fluoroscopy.
The subject devices may only be used with Volcano s5 Series or CORE Series imaging systems.
The subject device is packaged with a flushing tool that is equivalent to the predicate and is used for flushing the device's lumen with heparinized normal saline.
The subject devices are reprocessed once and permanently marked to indicate it has been reprocessed by Vein360.
| Subject Device Trade Name | Reference Number | French Size | Guide Wire | Minimum Sheath | Working Length |
|--------------------------------------------------------------------------|------------------|-------------|------------|----------------|----------------|
| Vein360 Reprocessed Visions PV.014P<br>RX Digital IVUS Catheter (85910P) | VEN-PV-014P | 3.5F | 0.014" | 5F | 150 cm |
| Vein360 Reprocessed Eagle Eye<br>Platinum RX Digital IVUS Catheter | VEN-PV-EEP | 3.5F | 0.014" | 5F | 150 cm |
| Vein360 Reprocessed Eagle Eye<br>Platinum ST RX Digital IVUS Catheter | VEN-PV-EEPST | 3.5F | 0.014" | 5F | 150 cm |
| Vein360 Reprocessed Visions PV.014P<br>RX Digital IVUS Catheter (014R) | VEN-PV-014R | 3.5F | 0.014" | 5F | 150 cm |
The scope of this submission is as follows:
# Indications for Use:
The Vein360 Reprocessed Visions® PV .014P RX (85910P), Visions® PV.014P RX (014R). Eagle Eve Platinum RX, and Eagle Eve Platinum ST RX Digital IVUS catheters are all designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by
{7}------------------------------------------------
providing a cross-sectional image of such vessels. These devices are not currently indicated for use in the cerebral vessels.
The Vein360 Reprocessed Visions® PV .014P RX (85910P), Visions® PV.014p RX (014R), Eagle Eye Platinum RX, and Eagle Eye Platinum ST RX Digital IVUS catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.
# Substantial Equivalence Information:
The subject devices are substantially equivalent to the predicate devices of the same product currently marketed by the device's original equipment manufacturer (OEM) and described herein with respect to intended use, design, materials, performance, and function. As a reprocessed single use device (SUD), there are no changes to the clinical applications, patient population, or method of operation. A high-level comparison of features between the predicate and subject devices is provided below.
| Feature | Predicate Device<br>(K143701, K152829) | Subject Device |
|-----------------------------|----------------------------------------|----------------|
| Hydrophilic Coating Length | 24cm | Identical |
| Working Length | 150cm | Identical |
| Max Working Length Diameter | 3.3 F | Identical |
| Transducer Diameter | 3.5 F | Identical |
| Guidewire Compatibility | 0.014" (0.36mm) | Identical |
| Sterilization Method | Ethylene Oxide (EO) gas | Identical |
| Markers | 3 radiopaque markers | Identical |
| Shelf Life | 2 years | 13 months |
| Uses | Single patient use | Identical |
| Accessories | Flushing tool | Equivalent |
# Performance Data:
With respect to SUD reprocessing, comprehensive cleaning validation studies were performed ensuring subject devices were clinically used and then soiled with artificial test soil. The cleaning operation was validated with a high degree of confidence by objectively demonstrating removal of all physical soil under minimum operating conditions. The body of this submission includes all data related to the cleaning process and validation.
Performance validation studies were performed after ensuring subject devices were clinically used and then soiled with artificial test soils. The cleaning process was conducted using maximum operating conditions to challenge functional performance of the device. Electro-mechanical performance testing was performed
{8}------------------------------------------------
# 510(k) SUMMARY
to demonstrate that the reprocessing operations did not adversely affect the predicate device's form, fit, or function.
Results of performance testing demonstrate the subject devices are substantially equivalent to the predicate devices which are safe and effective for their intended use. Substantial equivalence determination was concluded through successful completion of bench and laboratory testing, which included:
- Cleaning Validation ●
- Drying Validation ●
- Sterilization Validation ●
- Endotoxin Test Method Validation ●
- Biocompatibility .
- Performance Validation ●
- Simulated Use O
- Dimensional Integrity о
- Mechanical integrity o
- Cross / re-cross Testing O
- Electrical Integrity о
- Electrical Safety o
- Hydrophilic Coating Integrity о
- Acoustic Output o
- Image Quality о
- System Compatibility o
- Packaging Validation
The subject devices are validated for one reprocessing cycle after successful completion of the above performance testing. All subject devices are permanently marked and are tracked via OEM label during reprocessing. Subject devices are taken out of service and rejected from further reprocessing once the maximum number of cycles have been reached. Further, Vein360 restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.
# Conclusion:
Vein360 concludes the subject devices are as safe, as effective, and perform as well as or better than the predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.