OptiCrossTM HD Coronary Imaging Catheter, OptiCrossTM 6 HD Coronary Imaging Catheter
K230453 · Boston Scientific Corporation · OBJ · May 18, 2023 · Cardiovascular
Device Facts
Record ID
K230453
Device Name
OptiCrossTM HD Coronary Imaging Catheter, OptiCrossTM 6 HD Coronary Imaging Catheter
Applicant
Boston Scientific Corporation
Product Code
OBJ · Cardiovascular
Decision Date
May 18, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
This catheter is intended for ultrasound examination of coronary intravascular pathology only. Intrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Device Story
Sterile, short-rail intravascular ultrasound (IVUS) imaging catheter; 60 MHz ultrasonic transducer at distal tip; rotating drive cable; electro-mechanical connector interface to MDU5 PLUS™ Motordrive Unit. Used in clinical settings by physicians during coronary interventions. Catheter body includes distal imaging window, proximal shaft, and telescoping section for 15 cm linear transducer movement. Requires heparinized saline flush for acoustic coupling. Provides real-time ultrasound images of coronary vessels to clinicians; aids in assessment of coronary pathology and lesion length; facilitates procedural decision-making during interventions.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing evaluating imaging quality robustness and packaging integrity.
Technological Characteristics
60 MHz ultrasonic transducer; rotating drive cable; 5F and 6F sizes; radiopaque marker; heparinized saline acoustic coupling; electro-mechanical interface to MDU5 PLUS system; sterile, single-use.
Indications for Use
Indicated for patients who are candidates for transluminal coronary interventional procedures requiring ultrasound examination of coronary intravascular pathology.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
OptiCross HD and OptiCross 6 HD Coronary Imaging Catheter (K213593)
Related Devices
K213593 — OptiCross, OptiCross HD, OptiCross 6, OptiCross 6 HD · Boston Scientific Corporation · Jan 14, 2022
K050577 — ATLANTIS SR PRO2 CORONARY IMAGING CATHETER, MODEL 39014 · Boston Scientific Corp · Mar 30, 2005
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
May 18, 2023
Boston Scientific Corporation Cassie Clark Sr. Regulatory Affairs Specialist 1 Scimed Place Maple Grove, Minnesota 55311
Re: K230453
Trade/Device Name: OptiCross™ HD Coronary Imaging Catheter, OptiCross™ 6 HD Coronary Imaging Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ, ITX Dated: April 20, 2023 Received: April 21, 2023
Dear Cassie Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230453
Device Name
OptiCross™ HD Coronary Imaging Catheter, OptiCross™ 6 HD Coronary Imaging Catheter.
Indications for Use (Describe)
This catheter is intended for ultrasound examination of coronary intravascular pathology only. Intrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Boston Scientific. The text is in a serif font and is a dark blue color. The word "Boston" is on the top line, and the word "Scientific" is on the bottom line.
# 510(k) Summary
per 21 CFR §807.92
| Sponsor | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA | | |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------------|
| Contact Name and<br>Information | Cassie Clark<br>One Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-1425<br>e-mail: Cassie.Clark@bsci.com | | |
| Prepared by | Cassie Clark<br>February 17, 2023 | | |
| Proprietary Name | OptiCross™ HD Coronary Imaging Catheter, OptiCross™ 6 HD<br>Coronary Imaging Catheter | | |
| Common Name | Diagnostic Intravascular Catheter, Ultrasound Transducer | | |
| Product Code | OBJ, ITX | | |
| Classification | Catheter, Ultrasound, Intravascular (OBJ) has been classified as Class<br>II per 21 CFR 870.1200<br>Transducer Ultrasonic (ITX) has been classified as Class II per 21<br>CFR 892.1570 | | |
| Predicate Device | OptiCross HD and OptiCross<br>6 HD Coronary Imaging<br>Catheter | K213593 | January 14, 2022 |
{4}------------------------------------------------
### Device Description
The OptiCross HD and OptiCross 6 HD 60 MHz Coronary Imaging Catheters are sterile, short rail imaging catheters. Available in 5F and 6F sizes.
All catheters consist of two main assemblies:
- 1. Imaging Core
- 2. Catheter Body
The imaging core is composed of a hi-torque, flexible, rotating drive cable with a radial looking 60 MHz ultrasonic transducer at the distal tip. An electro-mechanical connector interface at the proximal end of the catheter makes the connection to the Motordrive Unit (MDU5 PLUS™) Instrument. The MDU5 PLUS-catheter interface consists of an integrated mechanical drive socket and electrical connection.
The catheter body is comprised of three sections:
- 1. Distal Imaging Window Lumen
- 2. Proximal Shaft Lumen
- 3. Telescope Section
The distal imaging window lumen and proximal shaft lumen sections comprise the "working length" of the catheter, and the telescoping section remains outside of the quiding catheter.
The catheter body has a distal imaging window lumen with proximal exit at 1.6 cm from the distal end. A radiopaque (RQ) marker is embedded in the catheter body at 0.5 cm from the distal tip. In addition, two insertion depth markers are located on the proximal shaft lumen at 90 cm and 100 cm from the distal tip to aid in estimating catheter position relative to the distal quide catheter tip. The proximal shaft lumen is attached to the telescoping section via a strain relief connection.
The telescoping shaft (section) allows the imaging core to be advanced and retracted for 15 cm of linear movement. The corresponding movement of the transducer occurs from the proximal end of the quidewire exit port to the proximal end of the distal imaging window lumen. The telescope section has proximal markers for lesion length assessment, consisting of a series of marks spaced 1 cm apart on the telescope body.
A flush port with a one-way check valve is used to flush the interior of the catheter body and maintain a flushed condition. The catheter must be flushed with heparinized saline prior to use, as this provides the acoustic coupling media required for ultrasonic imaging. The one-way check valve helps retain saline in the catheter during use.
The Coronary Imaging Catheter, 17.78 cm (7 in) extension tube, 3 cm³ (3 cc) and 10 cm³ (10 cc) syringes, and 4-way stopcock are provided in a single use sterilized package. The MDU5 PLUS Sterile Bag is provided in a separate sterile pouch within the finished carton along with the catheter package.
#### Indications for Use / Intended Use
This catheter is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.
#### Comparison of Technological Characteristics
The modified device includes a protective packaging component. There are no other design changes with this 510(k).
The OptiCross™ HD, and OptiCross™ 6 HD Coronary Imaging Catheters incorporate substantially equivalent design, materials, fundamental technology, interface with the IVUS imaging system, sterilization process, and intended use as those featured in the predicate; OptiCross HD and OptiCross6 HD Coronary Imaging Catheter K213593. All catheters interface with the same IVUS imaging system in the same way as the predicate.
{5}------------------------------------------------
#### Non-clinical Performance Data
Determination of substantial equivalence is based on an assessment of non-clinical performance bench testing data.
#### Bench Testing
Bench testing was performed to evaluate imaging quality robustness and packaging integrity of the OptiCross HD and OptiCross 6 HD Coronary Imaging Catheters.
#### Clinical Testing
Performance testing from clinical studies is not required to demonstrate substantial equivalence of the OptiCross HD and OptiCross 6 HD Coronary Imaging Catheters.
#### Conclusion
Based on the indications for use, technological characteristics, and performance testing, the OptiCross HD and OptiCross 6 HD Coronary Imaging Catheters have been shown to be appropriate for their intended use and are considered to be substantially equivalent to the predicate devices; OptiCross HD and OptiCross6 HD Coronary Imaging Catheter K213593.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.