Reprocessed Webster Duo-Decapolar Diagnostic EP Catheter
K233825 · Innovative Health · NLH · Jun 7, 2024 · Cardiovascular
Device Facts
Record ID
K233825
Device Name
Reprocessed Webster Duo-Decapolar Diagnostic EP Catheter
Applicant
Innovative Health
Product Code
NLH · Cardiovascular
Decision Date
Jun 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
The Reprocessed Webster Duo-Decapolar Diagnostic EP Catheter is indicated for electrophysiological mapping of cardiac structures, i.e. stimulation and recording only. In addition, the Webster Duo-Decapolar catheter is designed to facilitate electrogram mapping in the atrial region of the heart and the coronary sinus.
Device Story
Diagnostic EP catheter; high-torque shaft with deflectable tip; platinum electrode array. Used for cardiac electrophysiological mapping; stimulation; recording. Tip deflection controlled by proximal tubular handpiece with sliding piston; thumbknob mechanism. High-torque shaft allows rotation for accurate positioning. Used in clinical settings by physicians. Device reprocessed by Innovative Health; maximum 3 cycles; serialized and tracked. Provides electrogram data to clinicians for cardiac structure mapping; aids in diagnosis of cardiac electrical activity.
Clinical Evidence
Bench testing only. Testing included biocompatibility, cleaning validation, sterilization validation, functional testing (visual inspection, dimensional verification, electrical continuity/resistance, simulated use, mechanical characteristics), and electrical safety testing (dielectric and current leakage).
Technological Characteristics
High-torque shaft; deflectable tip; platinum electrodes. 7F French size; 110cm usable length; 20 electrodes. Manual deflection mechanism via proximal handpiece piston. Reprocessed device; cleaned and sterilized. No changes to materials or design compared to predicate.
Indications for Use
Indicated for electrophysiological mapping of cardiac structures (stimulation and recording) and electrogram mapping in the atrial region of the heart and coronary sinus.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
K061045 — REPROCESSED ELECTROPHYSIOLOGY CATHETERS · Alliance Medical Corp. · Aug 31, 2006
K043513 — STERILMED REPROCESSED DEFLECTABLE ELECTROPHYSIOLOGY DIAGNOSTIC CATHETER · Sterilmed, Inc. · Mar 1, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2024
Innovative Health Meerna Muradvich Regulatory Affairs Engineer 1435 N. Hayden Rd, Suite 100 Scottsdale, Arizona 85257
Re: K233825
Trade/Device Name: Reprocessed Webster Duo-Decapolar Diagnostic EP Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode recording catheter or electrode recording probe Regulatory Class: Class II Product Code: NLH Dated: May 8, 2024 Received: May 9, 2024
Dear Meerna Muradvich:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality
{2}------------------------------------------------
Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
The item numbers included in the scope of this submission are as follows:
| Item Number | Description | Sheath<br>Usable<br>Length<br>(cm) | French<br>Size | Curve | Number of<br>Electrodes | Spacing<br>(mm) |
|-------------|--------------------------------------------------------|------------------------------------|----------------|-------|-------------------------|--------------------|
| D728260RT | Webster Duo-<br>Decapolar<br>Diagnostic EP<br>Catheter | 110 | 7F | Large | 20 | 2-9-3-61-<br>3-9-3 |
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233825
Device Name
Reprocessed Webster Duo-Decapolar Diagnostic EP Catheter
Indications for Use (Describe)
The Reprocessed Webster Duo-Decapolar Diagnostic EP Catheter is indicated for electrophysiological mapping of cardiac structures, i.e. stimulation and recording only. In addition, the Webster Duo-Decapolar catheter is designed to facilitate electrogram mapping in the atrial reqion of the heart and the coronary sinus.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
As required by 21 CFR 807.92
#### Submitter's Name and Address:
Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257
#### Contact Name and Information:
Meerna Muradvich Regulatory Affairs Engineer Innovative Health, LLC. (480) 692-7176 (office) (888) 965-7706 (fax) mmuradvich@innovative-health.com
#### Date prepared:
November 30, 2023
#### Device Information:
| Trade/Proprietary Name: | Reprocessed Webster Duo-Decapolar Diagnostic<br>Electrophysiology Catheter |
|-------------------------|----------------------------------------------------------------------------|
| Common Name: | Diagnostic Electrophysiology Catheter |
| Classification Name: | Catheter, Recording, Electrode, Reprocessed |
| Classification Number: | Class II, 21 CFR 870.1220 |
| Product Code: | NLH |
#### Predicate Device:
| 510(k) Number | 510(k) Device | Manufacturer |
|---------------|----------------------------------------------------------------------------|-------------------------|
| K101991 | Webster Duo-Decapolar Catheter | Biosense Webster |
| K190980 | Reprocessed Webster Duo-Decapolar<br>Diagnostic Electrophysiology Catheter | Innovative Health, LLC. |
#### Reference Device:
| 510(k) Number | 510(k) Device | Manufacturer |
|---------------|-----------------------------------------------------------------|-------------------------|
| K170922 | Reprocessed Webster CS Bi-Directional<br>Diagnostic EP Catheter | Innovative Health, LLC. |
| K171503 | Reprocessed CristaCath Diagnostic EP<br>Catheter | Innovative Health, LLC. |
#### Device Description:
The Reprocessed Webster Duo-Decapolar Diagnostic Electrophysiology (EP) Catheter has been designed for electrophysiological mapping of heart. The catheters have a hightorque shaft with a deflectable tip section containing an array of platinum electrodes that can be used for stimulation and recording.
The tip deflection is controlled at the proximal end by a tubular handpiece in which a piston slides. When the piston is pushed forward with the thumbknob, the tip is deflected (curved). When the piston is pulled back, the tip straightens. The high-torque shaft allows the plan of the curved tip to be rotated to facilitate accurate positioning of the catheter tip at the desired site.
{6}------------------------------------------------
### Indications for Use:
The Reprocessed Webster Duo-Decapolar Diagnostic EP Catheter is indicated for electrophysiological mapping of cardiac structures, i.e. stimulation and recording only. In addition, the Webster Duo-Decapolar catheter is designed to facilitate electrogram mapping in the atrial region of the heart and the coronary sinus.
The item numbers included in the scope of this submission are as follows:
| Item Number | Description | Sheath<br>Usable<br>Length<br>(cm) | French<br>Size | Curve | Number of<br>Electrodes | Spacing<br>(mm) |
|-------------|-----------------------------------------------------------------------|------------------------------------|----------------|-------|-------------------------|--------------------|
| D728260RT | Reprocessed<br>Webster Duo-<br>Decapolar<br>Diagnostic EP<br>Catheter | 110 | 7F | Large | 20 | 2-9-3-61-<br>3-9-3 |
Table 1: Item Numbers
## Technological Characteristics:
The purpose, design, materials, function, and intended use of the Webster Duo-Decapolar Diagnostic EP Catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of these devices includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.
## Functional and Safety Testing:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Webster Duo-Decapolar Diagnostic EP Catheters.
This included the following:
- Biocompatibilitv ●
- Cleaning Validation ●
- Sterilization Validation ●
- Functional Testing ●
- Visual Inspection ●
- . Dimensional Verification
- Electrical Continuity and Resistance .
- Simulated Use .
- Mechanical Characteristics ●
- . Electrical Safety Testing
- Dielectric and Current Leakage ●
- Packaging Validation ●
{7}------------------------------------------------
The Reprocessed Webster Duo-Decapolar Diagnostic Electrophysiology (EP) Catheter is reprocessed no more than three (3) times. Each device is marked, serialized, and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.
#### Conclusion:
Innovative Health concludes that the Reprocessed Webster Duo-Decapolar Diagnostic EP Catheters are as safe and effective as the predicate devices described herein.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.