K211304 · Medtronic, Inc. · MXD · May 28, 2021 · Cardiovascular
Device Facts
Record ID
K211304
Device Name
LINQ II Insertable Cardiac Monitor
Applicant
Medtronic, Inc.
Product Code
MXD · Cardiovascular
Decision Date
May 28, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The LINQ II ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in the following cases: patients with clinical syndromes or situations at increased risk of cardiac arrhythmias patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia The device has not been tested specifically for pediatric use.
Device Story
LINQ II ICM is an insertable, programmable device for continuous subcutaneous ECG monitoring; records cardiac rhythms automatically upon arrhythmia detection or via patient-initiated activation. Patients use MyCareLink Heart App or Patient Assistant (PA97000) to manually record symptoms. Device communicates with Reveal LINQ Mobile Manager (LMM) via telemetry head. Data processed through CareLink Network using Device Command Library (DCL) and Instrument Command Library (ICL) software for remote programming and state management. Clinicians review recorded ECG data to diagnose arrhythmias (tachyarrhythmia, bradyarrhythmia, pause, atrial fibrillation). Enables long-term cardiac monitoring to correlate transient symptoms with rhythm disturbances, facilitating clinical decision-making for arrhythmia management.
Clinical Evidence
Bench testing and system validation only. No clinical data presented. Verification and validation activities confirmed the RAMware update successfully mitigated battery reporting and rejection parameter setting issues.
Technological Characteristics
Insertable cardiac monitor; subcutaneous ECG sensing. Communicates via telemetry head to mobile app/programmer. Software-based arrhythmia detection. Firmware includes RAMware for battery reporting and rejection parameter management. Substantially equivalent to predicate in materials, design, and packaging.
Indications for Use
Indicated for patients at increased risk of cardiac arrhythmias or experiencing transient symptoms (dizziness, palpitation, syncope, chest pain) suggestive of cardiac arrhythmia. Not tested for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
LINQ II Insertable Cardiac Monitor (Model LNQ22) (K200795)
Related Devices
K233320 — LINQ II Insertable Cardiac Monitor (LNQ22) · Medtronic, Inc. · Oct 31, 2023
K200795 — LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device Command Library, Instrument Command Library, LINQ Tool Kit · Medtronic, Inc. · Jul 3, 2020
K240693 — LINQ II Insertable Cardiac Monitor (ICM) · Medtronic, Inc. · Mar 28, 2024
K221962 — LINQ II Insertable Cardiac Monitor · Medtronic, Inc. · Aug 25, 2022
K150614 — Reveal LINQ · Medtronic, Inc. · Aug 6, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
May 28, 2021
Medtronic, Inc. Andrea Artman Principal Regulatory Affairs Specialist 8200 Coral Sea Street NE Mounds View, Minnesota 55112
Re: K211304
Trade/Device Name: LINQ II Insertable Cardiac Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: April 28, 2021 Received: April 29, 2021
Dear Andrea Artman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211034
Device Name LINQ II Insertable Cardiac Monitor
#### Indications for Use (Describe)
The LINQ II ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:
· patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
· patients who experience transent symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia
The device has not been tested specifically for pediatric use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
LINQ II POR RAMware Update Medtronic Confidential
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
### 510(k) Summary
| Date Prepared: | April 28, 2021 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Medtronic, Inc.<br>Medtronic Cardiovascular Diagnostics & Services<br>8200 Coral Sea Street N.E.<br>Mounds View, MN 55112<br>Establishment Registration Number: 2182208 |
| Contact Person: | Andrea Artman<br>Principal Regulatory Affairs Specialist<br>Medtronic Cardiovascular Diagnostics & Services<br>Phone: 419.708.4172<br>Fax: 651.367.0603<br>Email: andrea.p.artman@medtronic.com |
| Alternate Contact: | Ryan Calabrese<br>Senior Director Regulatory Affairs<br>Medtronic Cardiovascular Diagnostics & Services<br>Phone: 763.234.3574<br>Fax: 651.367.0603<br>Email: ryan.s.calabrese@medtronic.com |
## General Information
| Trade Name: | LINQ™ II Insertable Cardiac Monitor |
|-----------------------|-------------------------------------------------------------------------------|
| Common Name: | Insertable Cardiac Monitor |
| Regulation Number: | 21 CFR 870.1025 |
| Product Code: | MXD |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
| Special Controls: | Class II Special Controls Guidance Document: Arrhythmia Detector and<br>Alarm |
| Predicate Device: | LINQ™ II Insertable Cardiac Monitor (Model LNQ22) K200795 |
## Device Description
The LINQ II Insertable Cardiac Monitor (ICM) Model LNQ22 is a programmable device that continuously monitors a patient's ECG and other physiological parameters. The device records cardiac information in response to automatically detected arrhythmias and patient-initiated activation or markings. The device is designed to automatically record the occurrence of an episode of arrhythmia in a patient. Note: Arrhythmias are classified as tachyarrhythmia, bradyarrhythmia, pause, atrial tachyarrhythmia, or atrial fibrillation. Patients may also manually
{4}------------------------------------------------
record symptoms. In order to manually record symptoms, the patient will also need either the MyCareLink Heart App (patient app on mobile device) or the Patient Assistant Model PA97000. The patient can use the MyCareLink Heart App or the Patient to manually record his or her cardiac rhythm while experiencing or immediately after a symptomatic event.
#### Accessories
- The LINQ Tool Kit Model LNQ22TK includes two implant tools: The Incision Tool is . used to make a small incision through the patient's skin: and the Insertion Tool is used to insert the device through the incision and into the patient's body at the desired location.
- o The Reveal LINO™ Mobile Manager (LMM) Model MSW002 is the programmer designed as a mobile app that communicates with the Reveal LINQ and LINQ II ICM devices via the existing model 24967 telemetry head.
- The 2692 Device Command Library (DCL) software is a component of the Data Transformation Services subsystem of the CareLink Network. This software is responsible for understanding command status, implant device state within CareLink, and the initiation of commands for the implanted device.
- o The 2691 Instrument Command Library (ICL) software is the logic and data files required for remote programming.
## Indications for Use
There are no changes to the LINQ II ICM Indications for Use as a result of this submission. The Indications for Use are provided below:
The LINQ II ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:
- patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia
The device has not been tested specifically for pediatric use.
# Technological Characteristics
Intended use, design, materials, performance and technological characteristics are substantially equivalent to the predicate devices referenced.
When compared to the predicate device (K200795), the modified LINQTM II Insertable Cardiac Monitor presented in this submission has the same:
- Intended use/indications for use
- Operating principle ●
- o Design features
- Device functionality
- Biological safety
- . Packaging materials
- Shelf life
{5}------------------------------------------------
The modified Model LNQ22 ICM differs from the predicate in that the modified device contains RAMware that has been added to augment the firmware ensuring appropriate functionality of the device with respect to battery reporting and rejection parameter settings.
# Substantial Equivalence and Summary of Studies
Technological differences between the subject and predicate devices have been evaluated with device verification and system validation testing. The objective evidence from the verification and validation testing confirm that the change adequately mitigates incorrect battery reporting and rejection parameter settings.
The modified LINQ II Insertable Cardiac Monitor is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics. and indications for use. The modifications to the subject device were verified and validated through design verification and validation activities. All design verification and design validation activities were completed successfully and demonstrated there was no adverse impact to the functioning of the modified LINQ II ICM device.
## Conclusion
The results of the above verification and validation testing met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, the modifications made to the LINQ II Insertable Cardiac Monitor described in this submission result in a device that is substantially equivalent to the predicate.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.