K210608 · Boston Scientific Corporation · MXD · May 25, 2021 · Cardiovascular
Device Facts
Record ID
K210608
Device Name
LUX-Dx Insertable Cardiac Monitor
Applicant
Boston Scientific Corporation
Product Code
MXD · Cardiovascular
Decision Date
May 25, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECC). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx has not been tested specifically for pediatric use.
Device Story
LUX-Dx is a subcutaneously implanted sensor (Model M301) for continuous S-ECG monitoring; detects cardiac arrhythmias; marks S-ECG signals when algorithm criteria are met. System includes sensor, mobile apps (myLUX Patient app, LUX-Dx Clinic Assistant app) on OTS mobile devices, and LATITUDE Clarity server. Sensor communicates via Bluetooth Low Energy (BLE) to mobile monitor; mobile monitor transmits data via cellular/Wi-Fi to server. Clinicians access server for remote monitoring, workflow management, and arrhythmia evaluation. Device aids diagnosis of arrhythmias; benefits patients through long-term monitoring and timely clinical intervention. Implanted in fourth intercostal space of left chest wall; powered by integrated LiMnO2 battery; 3-year longevity.
Clinical Evidence
Bench testing only. Evidence includes software verification, system verification, and confirmatory validation. No clinical data presented.
Technological Characteristics
Subcutaneously implanted sensor; 1.2 cc volume; LiMnO2 battery powered; 3-year longevity; MR Conditional (1.5T and 3T). Connectivity via BLE to mobile device and cellular/Wi-Fi to server. Supplied sterile. Software-based arrhythmia detection.
Indications for Use
Indicated for patients with known heart conditions at risk of abnormal heart rhythms or symptoms suggesting cardiac arrhythmia (dizziness, palpitations, syncope, chest pain, shortness of breath). Not tested for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
K231328 — LUX-Dx II (M302); LUX-Dx II+ (M312) · Boston Scientific Corp · Aug 19, 2023
K150614 — Reveal LINQ · Medtronic, Inc. · Aug 6, 2015
K200795 — LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device Command Library, Instrument Command Library, LINQ Tool Kit · Medtronic, Inc. · Jul 3, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
May 25, 2021
Boston Scientific Corporation Melissa Klamerus Principal Regulatory Affairs Specialist 4100 Hamline Ave North St. Paul, Minnesota 55112
Re: K210608
Trade/Device Name: LUX-Dx Insertable Cardiac Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: February 24, 2021 Received: March 1, 2021
Dear Melissa Klamerus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210608
Device Name LUX-Dx™ Insertable Cardiac Monitor
#### Indications for Use (Describe)
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECC). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath.
The LUX-Dx has not been tested specifically for pediatric use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
4100 Hamline Avenue North St Paul, MN 55112-5798 651.582.4000 Tel 651.582.5134 Fax www.bostonscientific.com
# 510(k) Summary
#### 1. Submitter
Boston Scientific Corporation 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798
- Contact: Melissa Klamerus Principal Regulatory Specialist Phone: (612) 247-6878 Email: melissa.klamerus@bsci.com
Date Prepared: 23 February 2021
## 2. Device
Trade Name: LUX-Dx™ Insertable Cardiac Monitor Common Name: Arrythmia detector and alarm Product Code/Panel: MXD, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025
#### 3. Predicate Device
Trade Name: LUX-Dx™ Insertable Cardiac Monitor Manufacturer: Boston Scientific Corp Clearance Number: K193473, 26 June 2020 Common Name: Arrythmia detector and alarm Product Code/Panel: MXD, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025
## 4. Indication for Use
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous electrocardiogram (S-ECG). The recorded S-ECG is used for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition and are at risk of
{4}------------------------------------------------
developing an abnormal heart rhythm, or have symptoms that may suggest a cardiac arrhythmia such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath.
The LUX-Dx has not been tested specifically for pediatric use.
#### 5. Device Description
The LUX-Dx ICM sensor device (Model M301) evaluates S-ECG waveform data for indications of cardiac arrhythmias and "marks" the S-ECG signal for clinical presentation when the algorithm criteria are met. The device is inserted into the subcutaneous layer of the fourth intercostal space of the left chest wall. The sensor device is powered by an integrated battery. The electrodes used for detecting the S-ECG signal are located on each end of the sensor device, in the header and at the base of the battery. The LUX-Dx system includes the following main components:
- LUX-Dx™ sensor device a subcutaneously-implanted sensor device for cardiac arrhythmia event ● data collection and transmission. In addition, symptom events are collected and transmitted from the sensor device.
- . Mobile Monitor (MM) - mobile applications (myLUX™ Patient app and LUX-Dx™ Clinic Assistant app) running on an OTS mobile device that communicates with the LUX-Dx sensor device (using Bluetooth Low Energy (BLE)) and the LATITUDE Clarity™ server (using cellular/Wi-Fi) for collection and transmission of event, patient, and device data.
- . LATITUDE Clarity™ server - a server that communicates with the Mobile Monitor for bidirectional data transmission and provides web access for clinicians to perform remote monitoring activities and manage general patient and system parameters and workflow activities.
- . System Accessories- for insertion of the sensor device, an insertion tool and incision tool are provided. In addition, a magnet is provided to initiate sensor/MM app communication.
| Characteristic | Model M301 LUX-Dx ICM | Model M301 LUX-Dx ICM<br>Patient Patch v1.00 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Indications for use | The LUX-Dx Insertable Cardiac Monitor (ICM) is<br>intended to monitor and record subcutaneous<br>electrocardiogram (S-ECG). The recorded S-ECG<br>is used for the clinical evaluation and diagnosis<br>of cardiac arrhythmias. The LUX-Dx is indicated<br>for use in patients that have a known heart<br>condition and are at risk of developing an<br>abnormal heart rhythm, or have symptoms that<br>may suggest a cardiac arrhythmia such as<br>dizziness, palpitations, syncope, chest pain,<br>and/or shortness of breath.<br>The LUX-Dx has not been tested specifically for<br>pediatric use. | Identical |
| Usage | Single use | Identical |
# 6. Substantial Equivalence
{5}------------------------------------------------
| Characteristic | Model M301 LUX-Dx ICM | Model M301 LUX-Dx ICM<br>Patient Patch v1.00 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Sterility | Supplied sterile | Identical |
| Volume Dimension | 1.2 cc | Identical |
| MR Conditional | 1.5 T and 3T | Identical |
| Energy Source | Battery powered - LiMnO2 | Identical |
| Longevity | 3 years | Identical |
| Device Software<br>Version | FW v 1.00.00 | FW v 1.00.00 Patient Patch v1.00 |
| Remote Monitoring | Yes | Identical |
| Packaging | Sterile kit including ICM device and custom<br>implant tools.<br>Accessories: External, non-sterile devices are<br>packaged separately. | Identical |
# 7. Summary of Performance Testing
Boston Scientific performed safety risk management activities and design verification and validation testing to demonstrate that the LUX-Dx sensor device software is substantially equivalent to the predicate device software. The system conforms to user needs and intended use and software requirements and design specifications.
Performance Testing - Bench:
- . Software Verification
- . System Verification
- Confirmatory Validation ●
## 8. Conclusion
Based on the intended use, fundamental technological characteristics, and performance testing, the proposed LUX-Dx sensor device software has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the predicate device software.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.