FIRMap Catheter

K153093 · Abbott Electrophysiology · MTD · Apr 7, 2016 · Cardiovascular

Device Facts

Record IDK153093
Device NameFIRMap Catheter
ApplicantAbbott Electrophysiology
Product CodeMTD · Cardiovascular
Decision DateApr 7, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Indications for Use

For use in cardiac electrophysiology procedures to assist in the diagnosis of complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). The FIRMap Catheter may also be used for delivery of externally generated pacing stimuli.

Device Story

FIRMap Catheter is an intravascular, basket-style electrode recording and pacing catheter. It is delivered to the heart chamber via an intravascular sheath; once positioned, the sheath is retracted to allow the basket to expand and contact the heart wall. The device features 64 electrodes mounted on eight nitinol splines. It is used by electrophysiologists in clinical settings to record cardiac electrical signals and deliver pacing stimuli to assist in diagnosing complex arrhythmias. The device provides high-density mapping data that supplements conventional linear mapping catheters, aiding clinicians in identifying arrhythmia substrates. The output is used by the physician to guide clinical decision-making during electrophysiology procedures, potentially improving diagnostic accuracy for complex cardiac conditions.

Clinical Evidence

No clinical data. Evidence consists of bench testing and in-vivo simulated use testing. Bench tests included basket radial force, lateral spline stiffness, and introduction/withdrawal cycling, all meeting 95/95 confidence/reliability criteria. In-vivo testing evaluated the modified basket assembly during simulated use, confirming no clinically significant differences in functionality compared to the predicate.

Technological Characteristics

Distal basket assembly with 8 nitinol splines; 64 gold-plated copper electrodes. Spline tubing: Pebax. Radiopaque markers: platinum-iridium alloy. Tip dome/cap: 316 stainless steel. Catheter shaft diameter: 2.8mm. Working length: 128cm. Sterilization: EO gas. Single-use. Standards: ISO 14971:2012 (risk management), ISO 10555-1:2012 (intravascular catheters).

Indications for Use

Indicated for patients undergoing cardiac electrophysiology procedures for the diagnosis of complex arrhythmias that are difficult to identify with conventional linear mapping catheters; also indicated for delivery of pacing stimuli.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 7, 2016 Abbott Electrophysiology Dennis Pozzo Sr. Regulatory Affairs Specialist 3368 S. Geyer Road, Suite 365 St. Louis, MO 63127 Re: K153093 Trade/Device Name: FIRMap Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD Dated: March 4, 2016 Received: March 7, 2016 Dear Dennis Pozzo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mitchell Stein for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K153093 Device Name FIRMap Catheter Indications for Use (Describe) For use in cardiac electrophysiology procedures to assist in the diagnosis of complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). The FIRMap Catheter may also be used for delivery of externally generated pacing stimuli. Type of Use (Select one or both, as applicable) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------------------------------| | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Appendix 1: 510(k) Summary per 21CFR §807.92 | Submitter's<br>information | Abbott Electrophysiology<br>3668 S. Geyer Road, Suite 365<br>St. Louis, MO 63127<br>Contact: Dennis Pozzo/Phone 314-300-6580 | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device/<br>classification<br>name | Device Name: FIRMap Catheter Classification/Common name: Electrode recording catheter or electrode<br>recording probe Product Code/Classification No.: MTD/21 CFR 870.1220 The marketed device(s) to which substantial equivalence is claimed:<br>K130827, cleared October. 31, 2013 | | Device<br>description | The catheter is used in cardiac electrophysiology procedures to assist in the<br>diagnosis of complex arrhythmias that may be difficult to identify using<br>conventional mapping systems alone (i.e., linear mapping catheters). The<br>FIRMap Multiple Electrode Recording and Pacing Catheter may also be used<br>for delivery of externally generated pacing stimuli.<br><br>The FIRMap catheter is delivered to the heart chamber via an intravascular<br>sheath. After the catheter is positioned the sheath is withdrawn enough to<br>allow the basket to expand and the electrodes to contact the heart wall. | | Indications for<br>use | For use in cardiac electrophysiology procedures to assist in the diagnosis of<br>complex arrhythmias that may be difficult to identify using conventional<br>mapping systems alone (i.e., linear mapping catheters). The FIRMap Catheter<br>may also be used for delivery of externally generated pacing stimuli. | | Technological<br>characteristics | Both the proposed and predicate FIRMap catheters contain a distal basket<br>assembly. Arranged into a three dimensional shape that expands, Contain an array of 64 electrodes mounted onto eight support structures<br>called splines. Each spline has 8 electrodes spaced at an equal distance along the<br>length of the spline. The splines are fixed at an equal distance labeled A through H in a<br>clockwise direction viewed from the catheter/basket's distal tip. The array orientation is displayed by use of radiopaque markers. Continued on next page | {4}------------------------------------------------ # Appendix 1: 510(k) Summary per 21CFR §807.92, Continued The proposed device incorporates a design change to the distal spline which is Design change intended to improve performance and manufacturability compared with the predicate device. | Device Characteristic<br>(Technological Comparison) | Predicate FIRMap<br>Catheter (K130827) | Proposed FIRMap<br>Catheter | |-----------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------| | Intended Use | Intracardiac<br>electrophysiological<br>mapping and pacing. | Intracardiac<br>electrophysiological<br>mapping and pacing. | | Splines | | | | No. of splines | 8 (expanding) | 8 (expanding) | | Spline Material | Nitinol | Nitinol | | Spline Tube Material | Pebax | Pebax | | Spline radiopaque material | platinum-iridium alloy | platinum-iridium alloy | | Distal spline connecting tip<br>(tip dome and cap) | 316 Stainless steel | 316 Stainless steel | | Distal Catheter Tubing | Polyester Shrink Tubing | Polyester Shrink Tubing | | Distal Spline Tubing | Polyester Shrink Tubing | NA | | Method for limiting spline<br>movement | EDM "bowtie" feature | Extend Pebax tubing | | Electrodes | | | | No. of electrodes | 64 | 64 | | Electrode material | Gold plated copper | Gold plated copper | | Electrode configuration | Unipolar or bipolar | Unipolar or bipolar | | Dimensions | | | | Basket diameter/ Electrode<br>spacing | 40mm/ 6.8mm<br>50mm/ 9.0mm<br>60mm/ 11.2mm<br>70mm/13.5mm | 40mm/ 6.8mm<br>50mm/ 9.0mm<br>60mm/ 11.2mm<br>70mm/13.5mm | | Device Characteristic<br>(Technological Comparison) | Predicate FIRMap<br>Catheter (K130827) | Proposed FIRMap<br>Catheter | | Catheter shaft diameter | 2.8mm | 2.8mm | | Working length | 128cm +/-1cm | 128cm +/-1cm | | Recommended guide catheter | ID = 8.5 Fr. | ID = 8.5 Fr. | | Electrical Rating | | | | Electrical rating - Typical | +/- 27V, 25mA | +/- 27V, 25mA | | Dielectric strength | 500V DC | 500V DC | | Sterilization | | | | Method of sterilization | EO gas | EO gas | | Single use | Yes | Yes | Continued on next page {5}------------------------------------------------ ## Appendix 1:510(k) Summary per 21CFR §807.92, Continued ### Applicable standards - ISO 14971: 2012, Medical Devices Application of risk management to ● medical devices - ISO 10555-1: 2012 Intravascular catheters Sterile and single-use ● catheters - Part 1: General requirements Performance Bench Testing data The modification only impacts the basket subassembly of the catheter; the scope of testing was limited to confirming the mechanical integrity of the basket subassembly and its functionality. | Test Description | Acceptance Criteria | Conclusion | |----------------------------------------------|---------------------------------------------------------------------|------------| | Basket Radial force | Variable - 95/95<br>Confidence/Reliability | PASS | | Basket Lateral Spline<br>Stiffness | Variable - 95/95<br>Confidence/Reliability | PASS | | Introduction/ Withdrawal<br>Cycling | Attribute - 95/95<br>Confidence/reliability<br>per Weibull analysis | PASS | | FIRMap Catheter Testing in<br>Pancake Atrium | Attribute - 95/95<br>Confidence/reliability | PASS | Continued on next page {6}------------------------------------------------ ## Appendix 1: 510(k) Summary per 21CFR §807.92, Continued | Performance | In-Vivo Testing | |-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | data, continued | Evaluated the modification of the basket assembly during simulated use. No<br>clinically significant differences in use or functionality were noted between<br>the proposed and predicate catheters. | ### Conclusion Performance testing has demonstrated that the modified FIRMap Catheter provides reasonable assurance of conformance to predicate device requirements. Therefore, it is substantially equivalent to the predicate device, safe and effective for its intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...