SABER .014 PTA Dilatation Catheter; SABERX .014 PTA Dilatation Catheter

K221832 · Cordis US Corp · LIT · Aug 22, 2022 · Cardiovascular

Device Facts

Record IDK221832
Device NameSABER .014 PTA Dilatation Catheter; SABERX .014 PTA Dilatation Catheter
ApplicantCordis US Corp
Product CodeLIT · Cardiovascular
Decision DateAug 22, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters are intended to dilate stenoses in the peripheral vasculature, including femoral, popliteal and infra popliteal arteries. The device is also indicated for postdilation of balloon-expandable and self-expanding stents in the peripheral vasculature.

Device Story

SABER™ .014 (OTW) and SABERX™ .014 (Rapid Exchange) are PTA dilatation catheters; used to dilate peripheral artery stenoses and post-dilate stents. Catheters feature semi-compliant balloons mounted on distal ends; compatible with 0.014" guidewires. Radiopaque marker bands aid placement; tapered tips facilitate crossing tight stenoses; dual-layer hydrophilic coating increases lubricity. Operated by trained medical professionals in catheterization labs or hospitals. Inflation/deflation of balloon via inflation lumen restores vessel patency. Differences from predicates include silicone-coated guidewire lumen and inclusion of flushing needle for SABERX™.

Clinical Evidence

No clinical studies were performed. Substantial equivalence supported by non-clinical bench testing (dimensional verification, balloon burst pressure, fatigue, compliance, deflation time, bond strength, flexibility, torque, radiopacity) and animal studies (insertion/withdrawal force, pushability, crossability, fluoroscopy). Biocompatibility testing per ISO 10993-1 and sterilization validation per ISO 11135 confirm safety.

Technological Characteristics

OTW and Rapid Exchange PTA catheters; 0.014" guidewire compatibility. Materials: dual-layer hydrophilic coating, silicone-coated guidewire lumen. Sterilization: Ethylene Oxide (ISO 11135). Dimensions: 90-200 cm lengths, 1.25-6.0 mm diameters. Mechanical operation: balloon inflation/deflation via inflation lumen. No electronic components or software.

Indications for Use

Indicated for patients requiring dilation of stenoses in peripheral vasculature (femoral, popliteal, infra-popliteal arteries) and post-dilation of balloon-expandable or self-expanding stents. Contraindicated for use in coronary arteries.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". August 22, 2022 Cordis US Corp. Gina Flores Principal Specialist, Regulatory Affairs 5452 Betsy Ross Drive Santa Clara, California 95054 # Re: K221832 Trade/Device Name: SABER™ .014 PTA Dilatation Catheter; SABERX™ .014 PTA Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: June 22, 2022 Received: June 23, 2022 Dear Gina Flores: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221832 Device Name SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters #### Indications for Use (Describe) The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters are intended to dilate stenoses in the peripheral vasculature, including femoral, popliteal and infra popliteal arteries. The device is also indicated for postdilation of balloon-expandable and self-expanding stents in the peripheral vasculature. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summarv for K221832 #### I. SUBMITTER Applicant: Cordis US Corp 14201 North West 60th Avenue Miami Lakes, Florida, 33014 USA Establishment Registration: 1016427 Contact: Luis Davila Cordis US Corp 5452 Betsv Ross Drive Santa Clara, CA 95054, USA. Date Prepared: August 15, 2022 #### II. DEVICE Name of Device: SABER™ .014 and SABERX™ .014 Percutaneous Transluminal Angioplasty Dilatation Catheter Common Name: Percutaneous Transluminal Angioplasty Catheter Classification Name: Percutaneous Catheter (21 CFR 870.1250), Class II Product Code: LIT #### III. PREDICATE DEVICES SLEEK® OTW PTA Balloon Dilatation Catheter cleared on 12/01/2010 under K102035 and SLEEK® PTA Balloon Dilatation Catheter cleared on 02/08/2008 under K072947. #### Additional Reference Devices NANOCROSS™ ELITE OTW PTA Dilatation Catheter .014 cleared on 7/18/201 under K141118 RAPIDCROSS™ PTA Balloon Dilatation Catheter RX .014 cleared on 5/24/2013 under K130911 ULTRAVERSE® RX PTA Dilatation Catheter cleared on 5/30/2013 under K131199 Predicate/reference devices cited above have not been the subject of a recall. #### IV. DEVICE DESCRIPTION The SABER™ .014 and SABERX™ .014 Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter are single-use sterile devices, sterilized by ethylene oxide. The SABER™ .014 Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter is an over-the-wire (OTW) coaxial lumen catheter with a semi-compliant balloon mounted on its distal end. It is compatible with 0.014" (0.3556 mm) guidewire. The catheter is available in working {4}------------------------------------------------ lengths of 90 cm and 150 cm with balloon diameters ranging from 1.25 mm to 5.0 mm, and balloon lengths ranging from 15 mm to 300 mm. Radiopaque marker bands indicate the dilating section of the balloon and aid in balloon placement. For balloon lengths greater than or equal to 100 mm, the distal section will have two (2) adjacent marker bands and the proximal section will have one (1) marker band. For balloon lengths less than 100 mm, the distal and proximal section will each have one (1) marker band. The catheter tip is tapered to ease entry into peripheral arteries and to facilitate the crossing of tight stenoses. SABER™ .014 balloons are coated with a dual-layer hydrophilic material designed to increase lubricity throughout the lifetime of the device. The working pressure range for the balloon is between the nominal pressure and the rated burst pressure. All balloons distend to sizes above the nominal size at pressures greater than the nominal pressure. Consult the compliance chart on the tray label for typical diameters of the balloons at qiven pressures. The balloon lumen is used to inflate and deflate the balloon. The nominal balloon size in printed on the hub. The guidewire lumen is used to track the catheter over a prepositioned guidewire. The radiopaque marker bands (the length between the outer edge of most proximal to the outer edge of most distal marker band) indicate the stated nominal length of the balloon. The SABERX™ .014 Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter is a catheter with a distal inflatable balloon. The catheter is available in working lengths of 90 cm, 150 cm and 200 cm with balloon diameters ranging from 1.25 mm to 6.0 mm, and balloon lengths ranging from 15 mm to 300 mm. The catheter utilizes a Rapid Exchange design, consisting of a single inflation lumen and a distal guidewire lumen. The guidewire lumen begins at the distal tip and terminates at the guidewire exit port. The proximal hub is used as a balloon inflation port. Radiopaque marker bands indicate the dilating section of the balloon and aid in balloon placement. For balloon lengths greater than or equal to 100 mm, the distal section will have two (2) adjacent marker bands and the proximal section will have one (1) marker band. For balloon lengths less than 100 mm, the distal and proximal section will each have one (1) marker band. The catheter tip is tapered to ease entry into peripheral arteries and to facilitate the crossing of tight stenoses. SABERX™ .014 balloons are coated with a dual-layer hydrophilic material designed to increase lubricity throughout the lifetime of the device. The working pressure range for the balloon is between the nominal pressure and the rated burst pressure. All balloons distend to sizes above the nominal size at pressures greater than the nominal pressure. Consult the compliance chart on the tray label for typical diameters of the balloons at given pressures. The radiopaque marker bands (the length between the outer edge of most proximal to the outer edge of most distal marker band) indicate the stated nominal length of the balloon. The associated accessories for SABERX™ .014 PTA Dilatation Catheter include: - · Flushing Needle The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheter are for use in a catherization lab, hospital or other suitable healthcare facility by appropriately trained medical professionals only. {5}------------------------------------------------ # V. INDICATIONS FOR USE The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters are intended to dilate stenoses in the peripheral vasculature, including femoral, popliteal and infra-popliteal arteries. The device is also indicated for post-dilation of balloon-expandable and self-expanding stents in the peripheral vasculature. The Indications for Use statement for SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters is similar to the predicate devices. The subject and predicate devices have the same fundamental intended use, which is to dilate stenoses in femoral, popliteal. Minor differences in the Indications for Use statements do not affect the safety and effectiveness of the device relative to the predicate. As the predicate devices, The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters are not to be used in the coronary arteries. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE AND REFERENCE DEVICES The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters and the predicate devices facilitate the dilation of stenoses in femoral, popliteal and infra-popliteal in the peripheral vasculature using the same fundamental mechanism of action. The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters have the additional indication for stent post dilation, as do the reference devices RapidCross™ 0.14 RX PTA and NanoCross™ Elite 0.14 RX PTA. The SABER™ .014, SABERX™ .014 PTA Dilatation Catheters and the predicate devices have a guidewire lumen, that after gaining access, permits the use of a guidewire to facilitate advancement of the catheter to and through the stenosis to be dilated. The subject and predicate devices have a tapered tip to facilitate advancement of the catheter to and through the stenosis to be dilated. The subject and predicate devices include a lubricious coating, which is applied from the distal tip and over the balloon and a portion of the distal shaft for increased lubricity. The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters have the following similarities to the predicate and reference devices: - Same intended use ● - Same principle of operation ● - Same mechanism of action ● - Same method of sterilization and sterility assurance level ● - Same biocompatibility classification - Biocompatible for intended use - Labeled non-pyrogenic . - Similar materials ● - . Similar components - . Similar device dimensions - Similar packaging configuration - Similar compatibility with other devices (i.e., Guidewire, Catheter Sheath and Guiding Catheter). {6}------------------------------------------------ The following technological differences exist between the subject and primary predicate devices: - . SABER™ .014 and SABERX™ PTA Dilatation Catheters have the guidewire lumen silicone coated. - . SABERX™ .014 PTA Dilatation Catheter is provided with a flushing needle while the predicate devices do not include a flushing needle. - . The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters have the additional indication for stent post dilation. Based on a thorough analysis of technological characteristics, including design, materials, dimensions, mechanism of action and clinical use, the differences between the SABER™. 014 and SABERX™ .014 PTA Dilatation Catheters and the predicate devices do not raise any additional questions of safety and effectiveness. No new questions of safety and effectiveness are raised by the SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters for their intended use. # VII. PERFORMANCE DATA The performance data described below are provided in support of the substantial equivalence determination. # Biocompatibility Testing The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters, as the predicates, are externally communicating devices with limited contact duration (≤ 24 hours) with circulating blood. Biocompatibility testing was performed for SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA and ISO 10993-1:2009/Cor 1:2010, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. Based on the testing results, SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters are biocompatible for their intended use. Biocompatibility testing performed to the subject device is listed below: - . MEM Elution Using L-929 Mouse Fibroblast Cells - . Hemolysis Assay - Direct Contact and Extract Method - . Complement Activation SC5b-9 Assay with Sponsor-Supplied Comparison Article - . In Vitro Mouse Lymphoma with Extended Treatment - Bacterial Mutagenicity Test - Ames Assay - . Guinea Pig Maximization Sensitization Test - . Acute Systemic Injection Test - . Intracutaneous Irritation Test - . Materials Mediated Rabbit Pyrogen - Thromboresistance Evaluation {7}------------------------------------------------ # Sterilization The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters are sterilized using Ethylene Oxide. The sterilization cycle used to sterilize SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters, was validated per ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices to provide a sterility assurance level (SAL) of 10-6. Ethylene oxide and ethylene chlorohydrin residuals meet requirements for limited exposure devices (contact < 24 hours) in accordance with ISO 10993-7:2018, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals. ### Bench Testing The safety and effectiveness of the SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters and their substantial equivalence to the predicate devices, has been demonstrated through data collected during non-clinical design verification and validation tests and analyses. The following testing was successfully completed for the SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters per applicable sections of the indicated standards and/or validated internal test methods: - Catheter - FDA PTCA Guidance, USP-788 and internal test methods - Dimensional verification O - Balloon Preparation o - Balloon Rated Burst Pressure (with and without stent) O - Balloon Fatigue (Repeat Balloon Inflations with and without stent) O - Balloon Compliance O - O Balloon Deflation Time - Catheter Bond Strength O - Flexibility and Kink Test о - Torque Strength O - Particulate Evaluation / Coating Uniformity-Integrity O - Radiopacity o - Packaging integrity ISO 11607-1:2019 and ISO 11607-2:2019 ● - Visual inspection O - Component position O - Peel strength O - Bubble test O - Particulate о The passing results for the above tests provide reasonable assurance that the subject device has been designed to meet its intended use. No different issues of safety and effectiveness relative to the predicate were raised by this testing. ### Electrical safety and electromagnetic compatibility (EMC) This section is not applicable. The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters contain no electronic components, and thus, no electrical safety evaluation is required. {8}------------------------------------------------ # Software Verification and Validation Testing This section is not applicable. The SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters contain no software, and thus, no information regarding software/firmware is required for this device. #### Animal Study The safety and effectiveness of the SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters have been demonstrated through data collected during Design Validation (Animal Study) tests and analyses. The following testing was successfully completed for the SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters per applicable sections of the indicated standards and/or validated internal test methods. - Insertion and withdrawal force . - Kinking - Pushability - Crossability . - Atraumatic Tip ● - Guidewire compatibility ● - Fluoroscopy ● - . Balloon visibility - . Radiopacity - . Balloon inflation and deflation - Luer lock compatibility - Aseptic presentation and product removal - . Torque and kink resistance #### Clinical Studies No clinical studies were deemed necessary to support substantial equivalence. Appropriate verification and validation of the device requirements were achieved based on the similarities of the subject device to the predicate and from the results bench testing. #### VIII. CONCLUSIONS The information presented in this Premarket Notification demonstrates the following for the SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters - SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters have a legally marketed . predicate - SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters have the same Intended Use ● as the predicate - SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters have the same principle of . operation and mechanism of action - SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters incorporate the same . fundamental technology as the predicate {9}------------------------------------------------ - . Accepted scientific methods and international standards were used to evaluate safety and effectiveness of the SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters relative to the predicate - Safety and performance characteristics of the SABER™ .014 and SABERX™ .014 PTA . Dilatation Catheters are equivalent to the predicate devices and do not raise any additional questions of safety and effectiveness On the basis of the intended use, design, performance characteristics and non-clinical performance testing, and of detailed comparisons to the legally marketed predicate devices, it is concluded that the SABER™ .014 and SABERX™ .014 PTA Dilatation Catheters are appropriate for their intended use and are substantially equivalent to SLEEK® OTW and SLEEK® PTA Dilatation Catheters.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...