ATTAIN 6215 VENOGRAM BALLOON CATHETER

K012225 · Medtronic Vascular · DYG · Aug 28, 2001 · Cardiovascular

Device Facts

Record IDK012225
Device NameATTAIN 6215 VENOGRAM BALLOON CATHETER
ApplicantMedtronic Vascular
Product CodeDYG · Cardiovascular
Decision DateAug 28, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1240
Device ClassClass 2

Indications for Use

The Attain 6215 Venogram balloon catheter is indicated for use within the coronary sinus; it is intended for infusing contrast solutions into the coronary vasculature for venogram imaging. The Attain 6215 Venogram balloon catheter is intended for single use only.

Device Story

Attain 6215 Venogram balloon catheter; dual-lumen extruded polymeric tubing; inflation lumen with luer lock/stopcock; infusion lumen with luer lock; distal latex balloon. Used in coronary sinus to infuse contrast media for venogram imaging; facilitates left ventricular lead placement. Operated by physicians in clinical settings. Device provides controlled contrast delivery via 1.25 cc syringe. Benefits include improved visualization of coronary vasculature to assist lead implantation.

Clinical Evidence

No formal clinical trial data provided; device performance was evaluated in two Medtronic-sponsored IDE clinical studies for left ventricular lead implantation. Device performed as expected in the clinical environment for venogram imaging and lead placement.

Technological Characteristics

Extruded polymeric tubing; dual-lumen (inflation/infusion); distal latex balloon; 1.25 cc syringe. Sterilized via 100% Ethylene Oxide (ETO).

Indications for Use

Indicated for use in the coronary sinus to infuse contrast solutions into the coronary vasculature for venogram imaging. Intended for single use.

Regulatory Classification

Identification

A flow-directed catheter is a device that incorporates a gas-filled balloon to help direct the catheter to the desired position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) Summary of Substantial Equivalence | Date prepared | July 13, 2001 | | |------------------------|----------------------------------------------------------------------|--| | Submitter: | Medtronic, Inc.<br>7000 Central Avenue N.E.<br>Minneapolis, MN 55432 | | | Contact: | Karen Reidt, RAC<br>Sr. Product Regulation Manager | | | Telephone: | (763) 514-3914 | | | Fax: | (763) 514-6424 | | | E-Mail: | karen.reidt@medtronic.com | | | Proprietary Name: | Attain™ 6215 Venogram Balloon Catheter | | | Common Name: | Catheter, Flow Directed | | | Device Classification: | Class II, 21 CFR § 870.1240 | | | Product Code: | 74 DYG | | ### Device Description The Medtronic Attain 6215 Venogram balloon catheter is designed for use within the coronary sinus for infusing contrast solutions for venogram imaging. It consists of extruded polymeric tubing with two lumens (inflation and infusion) in the main body of the catheter and a 1.25 cc controlled stroke volume syringe. The inflation lumen features a luer lock and stopcock at the proximal hub. Its distal end opens into a latex balloon, which is located near the catheter tip. The infusion lumen has a luer lock at its proximal hub. #### Indications for Use The Attain™ 6215 Venogram balloon catheter is intended for use within the coronary sinus; it is intended for infusing contrast solutions into the coronary vasculature for venogram imaging. The Attain 6215 Venogram balloon catheter is intended for single use only. **Medtronic COMMERCIAL** {1}------------------------------------------------ ## Substantially Equivalent Device The predicate device for the Attain™ 6215 Venogram balloon catheter is outlined in the following table. | Predicate Device | Predicate Device<br>Manufacturer | Predicate 510(k) | |----------------------------------------------------------------|------------------------------------------------|-----------------------------| | Balloon Wedge Pressure, Angiographic | B. Braun Medical, Inc.*<br>Allentown, PA 18103 | K822806<br>cleared 11/1/82 | | Accent DG ™ Balloon Angioplasty<br>Catheter | Cook, Inc<br>Bloomington, IN 47402 | K900677<br>cleared 02/22/91 | | Vueport ™ Coronary Sinus Balloon<br>Occlusion Guiding Catheter | Cardima, Inc<br>Freemont, CA 94538 | K973298<br>cleared 06/26/98 | ### Summary of Studies Compatibility testing and incoming inspection of final packaged device was performed to support substantial equivalence to the predicate device. The Attain 6215 Venogram balloon catheter passed all of the in vitro specified requirements, and ensures that the Attain 6215 Venogram balloon catheter meets all of its design and performance requirements. ### Clinical Use Experience The Attain 6215 Venogram balloon catheter has been used in two Medtronic sponsored IDE clinical studies to facilitate the implantation of left ventricular leads. The Attain 6215 Venogram balloon catheter has performed as expected in the clinical environment during venogram imaging, and left ventricular lead placement via the coronary sinus. ### Biocompatibility Information Because there are no new materials or material composition changes made to the Attain 6215 Venogram balloon catheter from those used in the predicate Model WP balloon catheter manufactured by B. Braun Medical, Inc., biocompatibility testing was not repeated for the Attain 6215 Venogram balloon catheter. ### Sterilization Validation The Attain 6215 Venogram balloon catheter is sterilized using a 100% Ethylene Oxide (ETO) sterilization process. Processes appropriate for sterilizing the devices were validated. {2}------------------------------------------------ ## Conclusion Through the data and information presented, as well as similarities to legally marketed devices, Medtronic Inc, considers the Attain 6215 Venogram balloon catheter to be substantially equivalent to the previously discussed legally marketed predicate devices. **Medtronic CONFIDENTIAL** {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the department's name around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them, resembling a symbol or emblem. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## AUG 2 8 2001 Ms. Karen Reidt Medtronic, Inc. Cardiac Rhythm Management 7000 Central Avenue NE Minneapolis, MN 55432-3576 Re: K012225 Trade Name: Attain 6215 Venogram Balloon Catheter Regulation Number: 21 CFR 870.1240 Regulatory Class: Class II (two) Product Code: 74 DYG Dated: July 13, 2001 Received: July 16, 2001 #### Dear Ms. Reidt: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this {4}------------------------------------------------ #### Page 2 - Ms. Karen Reidt response to your premarket notification submission does not affect any obligation you might It spoulse to your promations of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premater notification. The FDA finding of substantial equivalence of your device to a legally marketed nouncation. The I Dri imanig of easing for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific arriver diagnostic devices), please contact the Office of Compliance at additionally 607.10 for in This allestions on the promotion and advertising of your device, (201) 594-4640. Raditionally, For quice at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mormation on your responsibility and its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Dak Tiller James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use | 510(k) Number (if known): | N/A K012225 | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Medtronic® Attain™ 6215 Venogram balloon catheter | | Indications For Use: | The Attain 6215 Venogram balloon catheter is indicated for use within the coronary sinus; it is intended for infusing contrast solutions into the coronary vasculature for venogram imaging. The Attain 6215 Venogram balloon catheter is intended for single use only. | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use_ (Optional Format 1-2-96) Division of Cardiovascular & Respiratory Devices 510(k) Number K012225 CONNOCHTAL
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...