ECG Mini System Continuous ECG Monitor and Arrhythmia Detector
Applicant
Lifewatch Technologies , Ltd.
Product Code
DSI · Cardiovascular
Decision Date
Jan 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The ECG Mini System Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center provides the ECG data to the medical practitioner for evaluation.
Device Story
System records 1-lead ECG via disposable 3-electrode patch; includes reusable 'Brain' processing/transmitter module and cellular 'Gateway' device. Used in home environment by patients. Device continuously monitors ECG; uses arrhythmia detection algorithm to trigger automatic alarms; supports patient-triggered manual alarms. Data transmitted via cellular network to monitoring center for medical practitioner evaluation. Facilitates remote cardiac rhythm monitoring; aids in diagnosis of transient symptoms; enables timely clinical intervention.
Clinical Evidence
Bench testing only. Performance validated per ANSI/AAMI/IEC 60601-2-47 (ambulatory ECG systems) and ANSI/AAMI EC57 (arrhythmia detection algorithm). Usability validation conducted per IEC 62366. Electrode performance tested per ANSI/AAMI EC12. Alarm system compliance verified per IEC 60601-1-8. Wireless co-existence assessed per FDA guidance. No clinical trial data presented.
Technological Characteristics
System comprises disposable 1-lead patch (3 electrodes), reusable 'Brain' module, and cellular gateway. Materials tested for cytotoxicity (ISO 10993-5), sensitization, and irritation (ISO 10993-10). Connectivity via cellular network. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Non-sterile. Software level of concern: Major.
Indications for Use
Indicated for adult patients experiencing transient symptoms suggestive of cardiac arrhythmia. Not intended for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
K081444 — BRAEMAR FUSION WIRELESS - AMBULATORY ECG ARRHYTHMIA MONITORING SYSTEM · Braemar, Inc. · Jul 31, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, with three curved lines above them, possibly representing the flow of information or services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2016
Lifewatch Technologies Ltd. % Donna-Bea Tillman Senior Consultant Biologics Consulting Group, Inc. 400 N. Washington St. Suite 100 Alexandria, Virginia 22314
Re: K151269
Trade/Device Name: ECG Mini System Continuous ECG Monitor and Arrhythmia Detector Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: DSI, DXH Dated: December 10, 2015 Received: December 14, 2015
Dear Donna-Bea Tillman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) N/A K151269
#### Device Name
ECG Mini System Continuous ECG Monitor and Arrhythmia Detector
Indications for Use (Describe)
The ECG Mini System Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arthythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arthythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center provides the ECG data to the medical practitioner for evaluation.
The device has not been tested for and it is not intended for pediatric use.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for LifeWatch. The logo consists of a series of overlapping sine waves in shades of green and blue. To the right of the sine waves is the word "LifeWatch" in a sans-serif font, with the "L" in a slightly larger font size than the rest of the letters.
# 510(k) Summary
# Introduction
This document contains the 510(k) summary for the ECG Mini System Continuous ECG Monitor and Arrhythmia Detector. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).
| Applicant: | LifeWatch Technologies Ltd. |
|------------------|----------------------------------------------------------------|
| | 2 Pekeris St., P.O.B. 527, Rehovot, 7610303, Israel |
| Contact Person: | Asher Kassel, Director of RA & QA, LifeWatch Technologies Ltd. |
| | LifeWatch Technologies Ltd. |
| | 2 Pekeris St., P.O.B. 527, Rehovot, 7610303, Israel |
| | Phone: 972-8-948-4010 (direct) |
| | Fax: 972-8-948-4044 |
| | Email: akassel@lifewatch.com |
| Date Prepared: | May 11, 2015 |
| Prepared By: | Donna-Bea Tillman, Ph.D. |
| | Senior Consultant |
| | Biologics Consulting Group, Inc. |
| | 400 N. Washington St. Suite 100 |
| | Alexandria, VA 22314 |
| | dtillman@bcg-usa.com |
| | Phone: 410-531-6542 |
| | Fax: 703-548-7457 |
| Predicate device | CG 6108-ACT1Lead system (K101639) |
| Trade Name: | ECG Mini System Continuous ECG Monitor and Arrhythmia Detector |
| Classification: | Detector and Alarm, Arrhythmia / |
| | Telephone Electrocardiograph Transmitters and Receivers |
| Product Code: | DSI / DXH |
| Regulation No: | 870.1025 / 870.2920 |
| Class: | II |
# Device Description
The ECG Mini System Continuous ECG Monitor and Arrhythmia Detector ("ECG Mini") is a system that records ECG data and detects defined arrhythmias. It is comprised of a disposable 1-lead ECG Patch which includes a processing/transmitting module called a "Brain", and a cellular device (also called "Gateway"). The ECG Patch has 3 electrodes which are used to obtain a 1-lead ECG that is used for arrhythmia detection. The ECG Patch is placed on the upper left side of the chest according to the instructions and guidance in the Patient User Guide (UG-00105).
The ECG Mini system consists of following three components:
- . Patch: Disposable 1-lead Patch with ECG electrodes
- Brain: Reusable processing and transmitter device
- . Gateway Cellular Device: Communicates data to the Monitoring Center
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for LifeWatch. The logo consists of a series of four overlapping wave-like lines in shades of green and blue. To the right of the wave-like lines is the word "LifeWatch" in a sans-serif font. The word is in a dark blue color.
## Indications for Use
The ECG Mini System Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.
The device has not been tested for and it is not intended for pediatric use.
#### Sterilization and Shelf Life
The ECG Mini is not provided sterile and is non-sterile when used.
The ECG Mini has been tested for accelerated aging and real time shelf life. The shelf life of the ECG Mini is 6 months from the date of manufacture.
#### Biocompatibility
The ECG Mini System patch is intended for single patient use only, for about 5-7 days, after which it can be replaced by a new ECG Mini System patch. According to the classification of "duration category" in Table 1 in ISO 10993-1, the ECG Mini System patch is classified as a surface device, in contact with the skin for a prolonged exposure (B), i.e. a device whose single, multiple or longterm use or contact is likely to exceed 24 hours but not 30 days.
The patient contacting materials of the ECG Mini patch have been tested for cytoxicity,
sensitization and skin irritation per the following standards:
- Cytotoxicity ISO 10993-5 ●
- Sensitization ISO 10993-10 ●
- Irritation ISO 10993-10 ●
#### Software
Similar to the predicate device, the software level of concern for the ECG Mini is MAJOR.
This determination was reached by a careful review of the FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff" (5/11/2005).
#### Electromagnetic Compatibility and Electrical Safety
The following Electrical Safety and EMC Testing was performed on the ECG Mini:
- IEC 60601-1:2005, 3rd edition, plus CORR. 1 (2006) + CORR. 2 (2007) Medical electrical ● equipment - General requirements for safety
- . IEC 60601-1-2:2014. General requirements for safety. Collateral Standard - Medical electrical equipment, Part 1-2; Electromagnetic compatibility - Requirements and tests
- IEC 60601-1-11 Edition 1.0 2010-04, Medical electrical equipment Part 1-11: General ● requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . CFR 47 FCC: Rules and Regulations; Part 15. "Radio frequency devices"; Subpart C: "Intentional radiators" (2015). Section 15.209. "Radiated emission limits, general requirements". "Radiated Emission Limits, Additional Provisions"; Section 15.249. "Operation within the bands 902 – 928 MHz, 2400 –2483.5 MHz, 5725 – 5875 MHz and 24.0 - 24.25 GHz"
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for LifeWatch. The logo features a series of overlapping, curved lines in shades of green and blue, resembling a wave or a pulse. To the right of the lines is the word "LifeWatch" in a sans-serif font, with the "L" in blue and the rest of the word in a slightly lighter shade of blue.
# Performance Testing
The following performance testing was conducted on the ECG Mini:
- ECG Performance - The ECG Mini has been tested and was shown to comply with the requirements of ANSI/AAMI/IEC 60601-2-47: 2012 (60601-2-47 in short); Medical electrical equipment - Part 2-47: Particular requirements for the safety, including essential performance, of ambulatory electrocardiographic systems.
- . Arrhythmia Detection - The algorithm used by the ECG Mini for arrhythmia detection has been tested in accordance with the requirements of the FDA recognized standard ANSI/AAMI EC57: 2012 - Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms.
- . Human Factors - LifeWatch conducted a Usability Validation Study of the ECG Mini in accordance with Section 5.9 of IEC 62366:2007. This study focused on the ability of a test population that meets the device's indicated user population to initiate the use of the device and to understand the user feedback signals while using the device. The study simulated Lifewatch's existing user support model, which is to provide instructions in the form of a Quick Start Guide and a User Guide, coupled with a Monitoring call center, which the user must contact in order to start signal collection and monitoring.
- ECG Electrode - In accordance with FDA's Guidance Document for ECG Electrodes, the ECG electrode assembly of the ECG Mini was tested in accordance with PART 4.2.2 (Electrical performance) of ANSI/AAMI EC12:2000/(R)2010: Disposable ECG electrodes.
- Alarms - Lifewatch has conducted an assessment demonstrating that the ECG Mini complies with the applicable requirements of IEC 60601-1-8, ed. 2.1:2012: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems.
- ICD Use LifeWatch had conducted testing to demonstrate that the ECG Mini is compatible ● with Implantable Cardioverter Defibrillators (ICDs).
- . Wireless Co-Existence - Lifewatch has assessed the ECG Mini for Wireless Coexistence in accordance with the FDA guidance document Radio Frequency Wireless Technology in Medical Devices (August 14, 2013).
## Performance Standards and Guidance Documents
This 510(k) submission was written in accordance with the following FDA Guidance documents:
- Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm, October 28, 2003
- Guidance for the Content of Premarket Submissions for Software Contained in Medical ● Devices, May 11 2005
- Class II Special Controls Guidance Document: Electrocardiograph Electrodes, July 21, 2011
The design of the ECG Mini System Continuous ECG Monitor and Arrhythmia Detector conforms to the following voluntary standards:
- IEC 60601 1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012, Medical electrical ● equipment - General requirements for safety
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for LifeWatch. The logo consists of a series of overlapping, curved lines in shades of green and blue, resembling a wave pattern. To the right of the wave pattern is the word "LifeWatch" in a sans-serif font, with the letters in a dark blue color.
- . IEC 60601-1-2 Edition 4: 2014, Part 1: Medical electrical equipment, Part 1-2; Electromagnetic compatibility - Requirements and tests
- . IEC 60601-1-11 Edition 1.0 2010-04, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-1-8 Edition 2.1 2012-11 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- AAMI/ANSI/ISO 60601-2-47:2012, Medical electrical equipment -- Part 2-47: Particular ● requirements for the basic safety and essential performance of ambulatory electrocardiographic systems
- ISO 14971 Second edition 2007-03-01, Medical devices Application of risk management . to medical devices
- IEC 62304 First edition 2006-05, Medical Device Software Software Lifecycle Processes .
- ANSI/AAMI/ISO EC57:2012, Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms
- AAMI ANSI EC12:2000/(R)2010 Disposable ECG electrodes .
- ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and ● testing within a risk management process.
## Substantial Equivalence Discussion
The predicate device is the CG 6108-ACT1Lead system (ACT or CG 6108 in short) cleared in K101639 on June 25, 2010, as an arrhythmia detector and alarm system.
The only difference in the Indications for use is that the subject ECG Mini explicitly states that it is not intended for pediatric use. The difference does not raise any issues of safety or effectiveness.
The subject device and the predicate device are both intended to be used by patients who experience transient symptoms that may suggest cardiac arrhythmia. Both devices continuously monitor patient ECG, and either automatically generate an event triggered by an arrhythmia detection algorithm or generates an event manually triggered by the patient. Both devices transmit the event and the recorded data over a cellular network to a monitoring center, where the data is provided to a medical practitioner for evaluation. The device therefore have the same intended use.
The subject device and the predicate device both use three electrodes to obtain a 1-lead ECG that is stored on a smartphone for transmission to a remote monitoring system. Both devices include an arrhythmia detection algorithm that is capable of detecting the following types of arrhythmias:
- . Pause
- . Tachycardia
- Bradycardia
- . Atrial Fibrillation
Both devices also allow the user to manually trigger an event.
The two significant differences between the subject ECG Mini and the predicate ACT1 system are the design of the electrode component and the mobile platform used as the Gateway device.
The differences between the CG 6108-ACT1Lead (K101639) and the subject device are minimal and pose no new questions to safety and effectiveness. The Indications for Use, and system
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for LifeWatch. The logo features a series of curved lines in shades of blue and green, resembling a waveform or a stylized heartbeat. To the right of the waveform is the word "LifeWatch" in a sans-serif font, with the "L" slightly larger than the other letters. The overall design is clean and modern, suggesting a focus on health, technology, or monitoring.
components (monitoring device, data transmission device and Monitoring Center) of both devices are substantially equivalent.
#### Substantial Equivalence Conclusion
The ECG Mini System Continuous ECG Monitor and Arrhythmia Detector is substantially equivalent with respect to the indications for use, technological characteristics and performance characteristics to the identified legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.