MOBICATH TRANSSEPTAL NEEDLE

K122832 · Great Batch Medical · DRC · Dec 19, 2012 · Cardiovascular

Device Facts

Record IDK122832
Device NameMOBICATH TRANSSEPTAL NEEDLE
ApplicantGreat Batch Medical
Product CodeDRC · Cardiovascular
Decision DateDec 19, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1390
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MobiCath Transseptal Needles are used to create the primary puncture in the interatrial septum to facilitate the passing of an introducer or catheter through the septum from the right side of the heart to the left side.

Device Story

MobiCath Transseptal Needle consists of thin-walled stainless steel cannula, ergonomic handle, two-way stopcock, and removable stylet. Used in surgical settings by physicians to create primary puncture in interatrial septum; facilitates passage of introducer or catheter from right to left heart. Handle includes pointer for curve orientation; stopcock enables air aspiration, fluid infusion, blood sampling, and pressure monitoring. Device is manual; no electronic or automated components. Modifications in this submission involve replacement of discontinued stopcock and associated handle interface adjustments, plus manufacturing improvements. Safety and efficacy verified via bench testing and biocompatibility assessment.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included visual/dimensional inspection, functional testing (tensile, torque, stopcock gauging, leakage, separation force, resistance to override, ease of assembly, stress cracking, stylet retention/insertion), and comprehensive biocompatibility testing (cytotoxicity, sensitization, reactivity, systemic injection, pyrogenicity, hemocompatibility, hemolysis, PTT, platelet/leukocyte counts, complement activation).

Technological Characteristics

Stainless steel cannula; ergonomic handle; two-way stopcock. Manual operation. Biocompatible materials. Sterilization via EtO. No software, electronics, or energy source.

Indications for Use

Indicated for patients requiring transseptal puncture to facilitate passage of an introducer or catheter from the right to the left side of the heart. For prescription use by trained physicians in a surgical setting.

Regulatory Classification

Identification

A trocar is a sharp-pointed instrument used with a cannula for piercing a vessel or chamber to facilitate insertion of the cannula.

Special Controls

*Classification.* Class II (special controls). Except for trocars that are reprocessed for multiple use, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/21 description: The image shows the logo for Greatbatch Medical. The logo consists of a circular arrangement of six hexagons, with the text "Greatbatch" on the right side of the hexagons. Below the word "Greatbatch" is the word "Medical." #### 510(k) Summary ## DEC 1 9 2012 Submitter's Name/Address: Contact Person: Greatbatch Medical 2300 Berkshire Lane North Minneapolis, MN 55441 K122832 Denise Thompson Sr. Regulatory Affairs Specialist Tel: 763-951-8254 Fax: 716-759-5040 dthompson@greatbatchmedical.com Date Prepared: Trade Name: Common Name: Classification Name: Product Code: Regulation: Classification Panel: September 14th, 2012 MobiCath™ Transseptal Needle Transseptal Needle Trocar DRC Class II 21 CFR 870.1390 Cardiovascular Predicate Device: MobiCath™ Transseptal Needle (K111644) #### Device Description: The MobiCath Transseptal Needles consist of a thin-walled stainless steel cannula bonded to an ergonomic handle and stopcock, and a removable stylet. The cannula is curved in the distal section to facilitate positioning in the cardiac anatomy when used in conjunction with an introducer or catheter. The handle is integrated with a pointer to show the orientation of the curve. The two-way stopcock facilitates air aspiration, fluid infusion, blood sampling, and pressure monitoring. The stylet attaches to the proximal end of the needle assembly. #### Intended Use: The MobiCath Transseptal Needles are used to create the primary puncture in the interatrial septum to facilitate the passing of an introducer or catheter through the septum from the right side of the heart to the left side. #### Comparison to Predicate Device: Greatbatch Medical proposes using an alternate stopcock for the MobiCath Trasseptal Needles. The current stopcock has been discontinued by the supplier. A similar stopcock has been tested as a replacement. A few minor design changes are required for the stopcock to interface with the handle components. In addition, a few manufacturing improvements have been proposed. {1}------------------------------------------------ The Intended Use and Indications for Use for the MobiCath Transseptal Needles have not changed. The fundamental scientific technology of the MobiCath Transseptal Needles has not changed. The proposed modifications do not change the needle cannula or stylet, or the manner in which the needle punctures the inter-atrial septum. The surgical approach and procedure have not changed. The device use environment has not changed. The device is still intended to be used under prescription use by trained physicians in a surgical setting. The overall device design has not changed. The device still consists of a cannula, handle, stopcock and stylet and these components continue to interact in the same manner. The proposed device changes only impact the dimensions and the material formulations of the stopcock and interfacing handle components. #### Summary of Testing: The following non-clinical tests were performed for the design changes to the MobiCath Transseptal Needle: - Design Verification Bench: ● - Visual & Dimensional Inspection O - Functional Testing o - Tensile o - o Torque - Stopcock gauging ಂ - Air and liquid leakage 0 - Stopcock separation force ಂ - Stopcock resistance to override 0 - Stopcock ease of assembly o - Stopcock stress cracking o - Stylet to stopcock retention force o - Stylet insertion through needle ಂ - Biocompatibility Testing . - Cytotoxicity o - Sensitization o - Intracutaneous Reactivity ೦ - Acute Systemic Injection ಂ - Materials Mediated Rabbit Pyrogenicity 0 - Hemocompatibility 0 - Hemolysis o - Partial Thromboplastin Time o - Platelet and Leukocyte Counts 0 - Complement Activation C3a and SC 5b-9 0 - Package Distribution Testing - Shelf Life Image /page/1/Picture/34 description: The image shows the logo for Greatbatch Medical. The logo consists of a circular arrangement of six smaller, solid shapes, resembling a stylized flower or gear. To the right of this symbol is the text "Greatbatch" in a bold, sans-serif font, with the word "Medical" appearing directly below it in a smaller font size. Appendix D - Page 2 of 3 {2}------------------------------------------------ - Sterilization Rationale . - EtO Residual Testing #### Conclusion: Failure Modes and Effect Analysis was used to identify the performance parameters potentially affected by the design modifications. The risks were addressed through verification and validation activities. The results from these activities demonstrate that the risks have been mitigated and that no new safety or efficacy issues were raised. · Greatbatch Medical considers the modified device to be substantially equivalent to the predicate device. Image /page/2/Picture/6 description: The image contains the logo for Greatbatch Medical. The logo consists of a circular arrangement of smaller, stylized shapes on the left, followed by the text "Greatbatch" in a bold, sans-serif font. Below "Greatbatch" is the word "Medical" in a smaller font size. The overall design is simple and professional, likely representing a company in the medical field. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines extending from its body, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 DEC 1 9 2012 Greatbatch Medical Attention: Denise Thompson Sr. Regulatory Affairs Specialist 2300 Berkshire Lane North Minneapolis, MN 55441 #### Re: K122832 Trade/Device Name: MobiCath Transseptal Needle Regulation Number: 21 CFR 870.1390 Regulation Name: Cardiovascular devices - Trocar Regulatory Class: Class II Product Code: DRC Dated: November 20, 2012 Received: November 28, 2012 Dear Ms. Thompson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Ms. Denise Thompson forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ycm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm Sincerely yours, # Matthew G. Hillebrenner Pro for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### 1.4 Indications for Use Statement 510(k) Number (if known): K12283 Device Name: MobiCath Transseptal Needle Indications For Use: The MobiCath Transseptal Needle is used to create the primary puncture in the interatrial septum to facilitate the passing of an introducer or catheter through the septum from the right side of the heart to the left side. Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR Over-The Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) urs Heele (Division Sign-Off) Division of Cardiovascular Devices K122832 510(k) Number Page 1 of 1 Page 15 of 42
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...