CoraForce Microcatheter, CoraFlex Microcatheter

K251277 · Reflow Medical, Inc. · DQY · Aug 22, 2025 · Cardiovascular

Device Facts

Record IDK251277
Device NameCoraForce Microcatheter, CoraFlex Microcatheter
ApplicantReflow Medical, Inc.
Product CodeDQY · Cardiovascular
Decision DateAug 22, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Cora Microcatheters are intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature. They may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions of diagnostic contrast.

Device Story

Single-lumen percutaneous microcatheter; hydrophilic-coated coiled shaft with braided support matrix; radiopaque distal tip; proximal luer; spin-friendly strain relief. Used in coronary and peripheral vasculature; operated by interventional physicians. Delivered over-the-wire; provides conduit for guidewire exchange and diagnostic/therapeutic agent delivery. Enhances trackability and pushability via distal floppy zone and proximal spin zone. Benefits patient by facilitating access to discrete vascular regions for interventional procedures.

Clinical Evidence

No clinical data submitted. Substantial equivalence supported by bench testing including simulated use, particulate, dimensional, torque, radiopacity, burst, leak, lubricity, kink resistance, design validation, corrosion, component integrity, and sterility testing.

Technological Characteristics

Single-lumen catheter; stainless steel with PTFE coating and gold plating; polymer shaft; hydrophilic coating. Dimensions: 135cm/150cm length; OD tapers from 0.050" (proximal) to 0.029" (distal). 0.014" guidewire compatibility; 4F sheath compatibility. Sterilization: Ethylene Oxide (EO).

Indications for Use

Indicated for patients requiring access to coronary and peripheral vasculature for guidewire placement/exchange and delivery of diagnostic contrast agents.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 22, 2025 Reflow Medical, Inc. Angela Lamprey Director, Regulatory Affairs 208 Avenida Fabricante Suite 100 San Clemente, California 92672 Re: K251277 Trade/Device Name: CoraForce Microcatheter, CoraFlex Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 17, 2025 Received: July 18, 2025 Dear Angela Lamprey: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251277 - Angela Lamprey Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251277 - Angela Lamprey Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jenny R. Katsnelson -S Digitally signed by Jenny R. Katsnelson -S Date: 2025.08.22 11:19:03 -04'00' for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251277 | | | Device Name CoraForce® and CoraFlex® Microcatheters (Cora Microcatheters) | | | Indications for Use (Describe) The Cora Microcatheters are intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature. They may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions of diagnostic contrast. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K251277 Page 1 of 2 REFLOW MEDICAL CoraForce and CoraFlex Microcatheters (Cora Microcatheters) SPECIAL 510(K) SUMMARY | Submitter | Reflow Medical, Inc. 208 Avenida Fabricante, Suite 100 San Clemente, CA 92672 Contact person: Lori Grace Email: lgrace@reflowmedical.com Phone: (949)-481-0399 | | --- | --- | | Date Prepared | April 23, 2025 | | Device | Name of the Device: CoraForce® and CoraFlex® Microcatheters (Cora Microcatheters) Common Name: Percutaneous Catheter Classification Name: Percutaneous Catheter Regulatory Class: 2 Product Code: DQY | | Predicate Device | Predicate: coraForce and coraFlex Support Catheters (Cora Catheters) – K201811 No reference devices were used in this submission. | | Description of the device | The CoraFlex and CoraForce Microcatheters (Cora Microcatheters) are single lumen catheters designed to access the peripheral and coronary vasculature. Each configuration has a hydrophilic coated coiled catheter with a braided support matrix, radiopaque distal tip, proximal luer, and “spin-friendly” strain relief. The Cora Microcatheters will also allow for the exchange of guidewires and provide a conduit for delivery diagnostic or therapeutic agents. | | Indications for Use | The Cora Microcatheters are intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature. They may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions of diagnostic contrast. | | Summary of the technological characteristics of your device compared to the predicate device | | | The technological characteristics of the subject Cora Microcatheters are similar to the technological characteristics of the Cora Catheters previously cleared under K201811. At a high level, the subject and predicate devices are based on the following same technological elements: • Both are delivered to the target site using a over-the-wire percutaneous technique • Both have a through lumen to allow passage and exchange of guidewires • Both have smooth inner lumens to provide reduced friction for guidewire movement • Both have polymer catheter shafts with specific geometries to control the torque and push movements associated with lesion crossing • Both use a specialized distal tip to facilitate the crossing of the lesion • Both have coiled catheter shafts to promote additional flexibility The following technological differences exist between the subject and predicate devices: • The subject device has a proximal “spin zone” intended to aid in pushability • The subject device has an increased distal “floppy zone” intended to aid in trackability • The subject device has an “spin-friendly” strain relief intended to allow users to grip and torque the device. Additionally, the spiral design of the strain relief gives the user a direction to torque the device | | Special 510(k) Summary {5} K251277 Page 2 of 2 REFLOW MEDICAL CoraForce and CoraFlex Microcatheters (Cora Microcatheters) | | Predicate – coraForce and coraFlex Support Catheter (K201811) | Subject - CoraForce and CoraFlex Microcatheters (Cora Microcatheters) | | --- | --- | --- | | Indications for Use | The Cora Catheters are intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature. They may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions of diagnostic contrast. | Same | | Guidewire Compatibility | 0.014” | Same | | Sheath Compatibility | 4F | Same | | Catheter Length | 135cm/150cm | Same | | Catheter Shaft OD | Max 0.032” (nominal 0.029”) | The catheter OD tapers from Max 0.051” (nominal 0.050”) at the proximal end to Max 0.032” (nominal 0.029”) at the distal end. | | Component Materials | Stainless Steel with PTFE Coating Stainless Steel with gold plating (coraForce) Polymers | Same | | Tip | coraForce: rigid coraFlex: flexible | coraForce: rigid coraFlex: flexible | | Coating Material | Lubricious | Same | | Coating Length | 60cm | Same | | Packaging Configuration | HDPE backer card and coil in single poly/Tyvek Pouch | Same | | Sterilization Method | Ethylene Oxide (EO) | Same | | A brief discussion of the nonclinical tests submitted | | | | The following bench testing was completed: | | | | • Simulated Use • Particulate Testing • Dimensional Verification • Torque Testing • Radiopacity • Burst Testing • Leak Testing • Lubricity and Coating Integrity Testing • Kink Resistance • Design Validation/Usability • Corrosion Resistance • Sterility Testing • Component Integrity • Biocompatibility Assessment • Bond Integrity | | | | The Cora Microcatheters met all specified criteria and did not raise new questions regarding safety and effectiveness compared to the predicate. Based on the performance testing, the Cora Microcatheters were found to be substantially equivalent to the predicate. | | | | A brief discussion of the clinical data submitted | | | | No clinical data is submitted. | | | | Conclusions | | | | The conclusions drawn from the non-clinical testing demonstrate that the CoraForce and CoraFlex Microcatheters are substantially equivalent to the legally marketed predicate devices. | | | Special 510(k) Summary
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...