SAT CenterFlow Molding Balloon Catheter (IN20-00313)
Applicant
Strait Access Technologies Holdings
Product Code
DQY · Cardiovascular
Decision Date
Jun 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CenterFlow Molding Balloon Catheter is intended to assist in the dilatation of self-expanding endoprostheses in large-diameter vessels.
Device Story
The CenterFlow Molding Balloon Catheter is a single-use, over-the-wire device designed to assist in the dilatation of self-expanding endoprostheses within large-diameter vessels. It features two elongated, non-compliant nylon balloons constrained in a zig-zag arrangement within a nitinol support frame. Upon inflation, the balloons straighten, expanding the device head while creating a central orifice that maintains hemodynamic flow during the molding procedure. The device is inserted over a 0.035" guidewire through a transfemoral introducer sheath. Platinum-iridium markers allow for angiographic visualization for precise placement. The catheter connects to a Y-connector for inflation/deflation and guidewire access. It is intended for use by clinicians in a procedural setting to facilitate endoprosthesis expansion without completely obstructing blood flow, potentially reducing hemodynamic instability during the procedure.
Clinical Evidence
No clinical data provided. Evidence consists of benchtop performance testing, biological safety evaluation, biocompatibility testing, in vivo safety evaluation, packaging validation, and shelf-life validation.
Technological Characteristics
Materials: Nylon (balloons), Nitinol (support frame), PEBAX 72D (shaft/lumens), Polycarbonate (tip), Platinum-iridium (markers). Dimensions: 110 cm working length, 18 Fr sheath compatibility, 20-46 mm balloon diameter. Connectivity: None (mechanical). Standards: ISO 10555-4, ISO 10555-14. Principle: Mechanical balloon dilatation with flow-permissive central orifice.
Indications for Use
Indicated for patients requiring dilatation of self-expanding endoprostheses in large-diameter vessels.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K123531 — UNBALLOON NON-OCCLUSIVE MODELING CATHETER · LeMaitre Vascular, Inc. · Dec 14, 2012
K133026 — UNBALLOON NON-OCCLUSION MODELING CATHETER · LeMaitre Vascular, Inc. · Dec 31, 2013
K250410 — GORE® Tri-Lobe Balloon Catheter · W. L. Gore & Associates, Inc. · Jun 2, 2025
K121839 — UNBALLOON NON-OCCLUSIVE MODELING CATHETER · LeMaitre Vascular, Inc. · Aug 8, 2012
K033670 — GORE TRI-LOBE BALLOON CATHETER · W. L. Gore & Associates, Inc. · May 5, 2004
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 25, 2025
Strait Access Technologies Holdings
% Robert Packard
Regulatory Consultant
Medical Device Academy
345 Lincoln Hill Rd
Shrewsbury, Vermont 05738
Re: K243184
Trade/Device Name: CenterFlow Molding Balloon Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: DQY
Dated: September 30, 2024
Received: May 23, 2025
Dear Robert Packard:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243184 - Robert Packard
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243184 - Robert Packard
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rohini Retarekar -S
for Carmen Gacchina Johnson, Ph.D.
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243184 | |
| Device Name CenterFlow Molding Balloon Catheter | |
| Indications for Use (Describe) The CenterFlow Molding Balloon Catheter is intended to assist in the dilatation of self-expanding endoprostheses in large diameter vessels. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
# SAT CenterFlow Molding Balloon Catheter; 510(k) Summary K243184
Date: 19 June 2025
Applicant Name: Strait Access Technologies Holdings (Pty) Ltd
Applicant Address: 313 Chris Barnard Building
Anzio Road
University of Cape Town Medical Campus
Observatory
Cape Town
7925
Correspondent Name: Medical Device Academy
Correspondent Address: 345 Lincoln Hill Rd
Shrewsbury, VT
05738
United States
Correspondent Tel: +1 802 281 4381
Correspondent Contact: Mr. Rob Packard
Correspondent Email: rob@fdaestar.com
{5}
Page 2 of 3
1. Device Name
Device Trade Name: CenterFlow Molding Balloon Catheter
Common Name: Percutaneous Catheter, Balloon Catheter
Classification Name: Catheter, Percutaneous
Product Code: DQY
Regulation Number: 870.1250
2. Legally Marketed Predicate Device(s)
| Device Name | Manufacturer | 510(k) Number | Product Code |
| --- | --- | --- | --- |
| Tri-Lobe Balloon Catheter | W.L Gore & Associates | K081799 | DQY |
3. Device Description Summary
The CenterFlow Molding Balloon Catheter is intended to assist in the dilatation of self-expanding endoprostheses in large diameter vessels. When used in its indications for use, it is a low-permissive, over-the-wire, single-use device comprised of two elongated non-compliant nylon balloons that are attached to and within a nitinol support frame. The two elongated balloons are constrained in a zig-zag arrangement by the nitinol support frame. Upon inflation, this configuration enables expansion of the device head through balloon straightening, to assist the dilatation of self-expanding endoprostheses in large diameter vessels. Expansion of the zig-zag balloon configuration creates a central orifice permitting maintained hemodynamic flow during the molding procedure.
The inflation lumen is 9 Fr and is made of PEBAX 72D. A co-axial PEBAX 72D guidewire lumen runs through the inflation lumen and device head and terminates into a distal polycarbonate tip. Two platinum-iridium markers provide angiographic visualization of the distal and proximal balloon edges to facilitate intravascular balloon placement prior to inflation. The catheter has a working length of 110 cm and it connects proximally to a standard Y-connector at the proximal end. The Y-connector allows for separation of two lumens - an angled luer-lock hub to inflate and deflate the balloon and a straight luer-lock hub which accepts a 0.035" guidewire. The device is deflated and inserted over the guidewire through the hemostasis valve of a transfemoral introducer sheath.
4. Intended Use/Indications for Use
The SAT CenterFlow Molding Balloon Catheter is intended to assist in the dilatation of self-expanding endoprostheses in large-diameter vessels.
5. Comparison of Essential Design Features
| Design Features - Summary | CenterFlow Molding Balloon Catheter | Gore Tri-Lobe Balloon Catheter |
| --- | --- | --- |
| Device Design Description - Overall | A flow-permissive balloon catheter that assists in the dilatation of self-expanding endoprostheses in large vessels. | A lobed balloon catheter that assists in the dilatation of self-expanding endoprostheses in large vessels without complete blockage of aortic blood flow. |
| | Consists of two elongated nylon balloons attached to, and | Consists of three polyurethane balloons mounted on the leading |
{6}
Page 3 of 3
| | contained within a nitinol support frame. | edge on a multi-lumen catheter shaft. |
| --- | --- | --- |
| | The elongated balloons, constrained in a zig-zag arrangement within the frame, expand through balloon straightening, creating a central orifice that maintains hemodynamic flow. | Each of the three inflation lumens communicates with one of the balloons, and an inflation port communicates with all the lumens. |
| | A Y-connector allows for the separation of two lumens and a straight luer-lock hub accepts a 0.035” guidewire. | The guidewire lumen allows for a 0.035” guidewire. |
| Materials of Use | The device uses radiopaque marker bands. | Same |
| | The balloons are made of nylon. | The balloons are made of polyurethane. |
| Dimensions | Balloon diameter, 20-46 mm | Balloon diameter for BCM1634, 16-34 mm for BCL2645, 26-42 mm |
| | Catheter shaft – 113 cm | Catheter shaft – 108 cm |
| | Introducer sheath compatible, 18 Fr | Introducer sheath compatible, 18 Fr |
| Relevant Standards | ISO 10555-4 Intravascular catheters – sterile and single use catheters, part 4 Balloon dilatation catheters, 2023-11. | Same |
| | ISO 10555-14 Intravascular catheters – sterile and single use catheters, part 1 General requirements, 2023-11. | |
## 6. Non-Clinical Tests Summary & Conclusions
The non-clinical testing conducted showed that the subject device is substantially equivalent to the predicate device. Below is an indication of the tests performed:
- Benchtop performance testing
- Biological Safety Evaluation and Biocompatibility
- In vivo safety evaluation
- Packaging validation
- Shelf-life validation
- Usability Engineering evaluation
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.