K251591 · Edan Instruments, Inc. · DQK · Feb 9, 2026 · Cardiovascular
Device Facts
Record ID
K251591
Device Name
Holter ECG and ABP system (HolterABP)
Applicant
Edan Instruments, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Feb 9, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The Holter ECG and ABP system is intended to acquire, analyze, display, edit and generate report of ambulatory ECG, ambulatory blood pressure and combined Holter ECG and ABP data. The analysis results are offered to doctors on an advisory basis only. Data acquired may be used for the following indications: · Evaluation of symptoms suggesting arrhythmia or myocardial ischemia. · Evaluation of patients for ST segment changes. · Evaluation of drug response in patients taking anti-arrhythmic medications. · Evaluation of patients with pacemakers. The ambulatory blood pressure function is intended for use in patients over 12 years of age from data collected from an FDA cleared ambulatory blood pressure cuff. The ECG Analysis function is intended for use in adults and pediatric, but arrhythmia detection is intended for use in patients over 12 years of age. The Holter ECG and ABP system is intended to be used only in hospital and healthcare facilities by physicians or trained healthcare professionals.
Device Story
Windows-based software application for retrospective analysis of ambulatory ECG and blood pressure (ABP) data; inputs data from compatible recorders; performs automatic QRS detection, classification, arrhythmia detection, ST segment measurement, and ABP statistics; generates summary tables, trends, and reports for physician review; used in hospital/healthcare facilities by trained professionals; provides advisory-only results to support clinical decision-making regarding arrhythmia, ischemia, and pacemaker function; does not perform real-time monitoring or alarming.
Clinical Evidence
No clinical testing performed. Substantial equivalence established via bench testing and software verification/validation. Performance validated using public datasets (AHA, MIT-BIH, NST, ESC, QT) and proprietary datasets; metrics included sensitivity, positive predictive value, false positive rate for QRS/arrhythmia detection, and error rates for ST, QT, and HR measurements.
Technological Characteristics
Windows-based software; performs ECG/ABP data analysis; supports 3/12 lead ST segment measurement, pacemaker analysis, HRV, HRT, VCG, and VLP; complies with IEC 60601-2-47; standalone software application.
Indications for Use
Indicated for patients requiring evaluation of symptoms suggesting arrhythmia or myocardial ischemia, ST segment changes, drug response to anti-arrhythmic medications, or pacemaker function. Ambulatory blood pressure monitoring is indicated for patients >12 years. ECG analysis is indicated for adults and pediatric patients; arrhythmia detection is indicated for patients >12 years.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Edan Instruments, Holter System (SE-2003, SE-2012) (K151787)
K221904 — EK12 V2 Algorithm · Ge Medical Systems Information Technologies, Inc. · Dec 20, 2022
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 9, 2026
EDAN Instruments, Inc.
Tracy Yue
Official Correspondent
15 Jinhui Road, Jinsha Community, Kengzi Sub-district
Pingshan District
Shenzhen, 518122
China
Re: K251591
Trade/Device Name: Holter ECG and ABP system (HolterABP)
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II
Product Code: DQK
Dated: May 23, 2025
Received: January 9, 2026
Dear Tracy Yue:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251591 - Tracy Yue
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251591 - Tracy Yue
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251591 | |
| Device Name Holter ECG and ABP system (HolterABP) | |
| Indications for Use (Describe) The Holter ECG and ABP system is intended to acquire, analyze, display, edit and generate report of ambulatory ECG, ambulatory blood pressure and combined Holter ECG and ABP data. The analysis results are offered to doctors on an advisory basis only. Data acquired may be used for the following indications: · Evaluation of symptoms suggesting arrhythmia or myocardial ischemia. · Evaluation of patients for ST segment changes. · Evaluation of drug response in patients taking anti-arrhythmic medications. · Evaluation of patients with pacemakers. The ambulatory blood pressure function is intended for use in patients over 12 years of age from data collected from an FDA cleared ambulatory blood pressure cuff. The ECG Analysis function is intended for use in adults and pediatric, but arrhythmia detection is intended for use in patients over 12 years of age. The Holter ECG and ABP system is intended to be used only in hospital and healthcare facilities by physicians or trained healthcare professionals. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Traditional 510(k) of Holter ECG and ABP system
K251591
# 510(k) Summary
# Prepared in accordance with the requirements of 21 CFR Part 807.92
1. Submitter: Edan Instruments, Inc.
#15 Jinhui Road, Jinsha Community, Kengzi Sub-District, Pingshan District, Shenzhen, 518122 P.R.China.
Tel: +86(0755) 84513592
Fax: +86(0755) 26882223
2. Contact person: Tracy Yue
Preparing date: February 7, 2026
3. Device name and classification: Trade name: Holter ECG and ABP system, Model: HolterABP
Classification Name/ Product code: 21 CFR 870.1425 Computer, Diagnostic, Programmable
DQK
Regulatory Class: Class II
4. Predicate Device(s): 1) Edan Instruments, Holter System, models SE-2003 and SE-2012, cleared under K151787
2) Edan Instruments, Ambulatory Blood Pressure Monitor, models: SA-10, SA-05, SA-06, SA-08 and SA-09, cleared under K210139
5. Reason for Submission: By submission of the Traditional 510(k), Edan Instruments is requesting clearance for Holter ECG and ABP system, Model: HolterABP
6. Device Description: The Holter ECG and ABP system (Model: HolterABP) is a Windows-based application software. It is intended to acquire, analyze, display, edit and generate report of ambulatory ECG data, ambulatory blood pressure data and combined Holter ECG and ABP data that has been stored by the compatible recorders.
The Holter ECG and ABP system provides automatic QRS detection and classification, automatic arrhythmia detection, ST segment measurement and creates summary tables, trends.
The Holter ECG and ABP system display the ABP data, then form the summary table, and statistics information.
The final report can be previewed, printed and saved after the data analysis
1 / 7
{5}
Traditional 510(k) of Holter ECG and ABP system
K251591
is completed.
All results obtained from automatic analysis and the resulting unconfirmed reports must be considered as suggestions only.
The subject device is intended only for retrospective analysis, and not for use in real-time monitoring or alarming.
# 8. Indication for Use
The Holter ECG and ABP system is intended to acquire, analyze, display, edit and generate report of ambulatory ECG, ambulatory blood pressure and combined Holter ECG and ABP data. The analysis results are offered to doctors on an advisory basis only.
Data acquired may be used for the following indications:
- Evaluation of symptoms suggesting arrhythmia or myocardial ischemia.
- Evaluation of patients for ST segment changes.
- Evaluation of drug response in patients taking anti-arrhythmic medications.
- Evaluation of patients with pacemakers.
The ambulatory blood pressure function is intended for use in patients over 12 years of age from data collected from an FDA cleared ambulatory blood pressure cuff.
The ECG Analysis function is intended for use in adults and pediatric, but arrhythmia detection is intended for use in patients over 12 years of age.
The Holter ECG and ABP system is intended to be used only in hospital and healthcare facilities by physicians or trained healthcare professionals.
# 9. Predicate Device Comparison
The subject device is technologically equivalent to predicate devices. Any differences are explained to demonstrate in this submission that these differences do not raise any new questions of safety and effectiveness. The table below compares the intended use and technological characteristics of the subject and predicate devices:
Table 1: Comparison to Primary Predicate Device
| Item | Primary Predicate Device K151787 Holter System (Model: SE-2003,SE-2012) | Subject Device Holter ECG and ABP system (Model: HolterABP) | Comparison Result |
| --- | --- | --- | --- |
| Manufacturer | Edan Instruments, Inc | Edan Instruments, Inc | Same |
| K# | K151787 | K251591 | - |
| Regulation Number | 21 CFR 870.1425 | 21 CFR 870.1425 | Same |
{6}
Traditional 510(k) of Holter ECG and ABP system
K251591
| Regulation Name | Programmable Diagnostic Computer | Programmable diagnostic computer | Same |
| --- | --- | --- | --- |
| Classification | II | II | Same |
| **Intended Use** | The SE-2003/SE-2012 Holter System (including recorder and analysis software) is intended to record, analyze, display, edit and generate report of ambulatory ECG. The Holter System is intended to be used by trained personnel under the direction of doctors. The analysis results are offered to doctors on an advisory basis only. The Holter System is intended for adult, pediatric patients including infants weighing less than 10 kg.
It can be used for the following indications:
1. Evaluation of symptoms suggesting arrhythmia or myocardial ischemia.
2. Evaluation of patients for ST segment changes.
3. Evaluation of drug response in patients taking anti-arrhythmic medications.
4. Evaluation of patients with pacemakers. | The Holter ECG and ABP system is intended to acquire, analyze, display, edit and generate report of ambulatory ECG, ambulatory blood pressure and combined Holter ECG and ABP data. The analysis results are offered to doctors on an advisory basis only.
Data acquired may be used for the following indications:
1. Evaluation of symptoms suggesting arrhythmia or myocardial ischemia.
2. Evaluation of patients for ST segment changes.
3. Evaluation of drug response in patients taking anti-arrhythmic medications.
4. Evaluation of patients with pacemakers.
The Ambulatory blood pressure function is intended for use in patients over 12 years of age from data collected from an FDA cleared ambulatory blood pressure cuff.
The ECG Analysis function is intended for use in adults and pediatric, but arrhythmia detection is intended for use in patients over 12 years of age.
The Holter ECG and ABP system is intended to be used only in hospital and healthcare facilities by physicians or trained healthcare | Different |
<br>
3 / 7</br>
{7}
Traditional 510(k) of Holter ECG and ABP system
K251591
| | | professionals. | |
| --- | --- | --- | --- |
| **Review Modes** | ECG strip review, ECG template review, ECG Poincare Plot review , ECG Superimposed Graph , ECG Histogram , ECG Event review, ECG Trend review, Tabular and Page review, | ECG strip review, ECG template review, ECG Poincare Plot review , ECG Superimposed Graph , ECG Histogram , ECG Event review, ECG Trend review, Tabular and Page review | same |
| **Rhythm Analysis** | Bradycardia, Tachycardia, Ventricular pairs, Ventricular tachycardia, Supraventricular pairs, Long R-R interval, Pause, Min/max heart rate, ST depression, ST elevation | Bradycardia, Tachycardia, Ventricular pairs, Ventricular tachycardia, Supraventricular pairs, Long R-R interval, Pause, Min/max heart rate, ST depression, ST elevation | same |
| **ST Segment Measurement** | 3/12 lead ST segment measurement, max ST depression, max ST elevation, 3/12 lead trends | 3/12 lead ST segment measurement, max ST depression, max ST elevation, 3/12 lead trends | same |
| **Pacemaker Analysis** | Atrial, ventricular and dual pacemaker detection; capture failure, Output failure, under sensing | Atrial, ventricular and dual pacemaker detection; capture failure, Output failure, under sensing | same |
| **QT analysis** | QT Measurement | QT Measurement | same |
| **HRV analysis** | Frequency domain analysis, Time domain analysis | Frequency domain analysis, Time domain analysis | same |
| **Heart rate calculation** | heart rate of a single beat Maximum and minimum heart rate Average heart rate | heart rate of a single beat Maximum and minimum heart rate Average heart rate | Same |
| **HRT** | Turbulence onset (TO), Turbulence slope(TS) | Turbulence onset (TO), Turbulence slope(TS) | Same |
| **VCG** | Display waveform of three orthogonal leads X, Y and Z, and display QRS loop, P loop, and T loop of frontal plane, horizontal plane, and right sagittal plane. | Display waveform of three orthogonal leads X, Y and Z, and display QRS loop, P loop, and T loop of frontal plane, horizontal plane, and right sagittal plane. | Same |
| **VLP** | Time domain analysis, Frequency domain analysis | Time domain analysis, Frequency domain analysis | Same |
| **ABPM data viewer** | - | Display BP Data Table, Day & Night BP Profile, BP Histogram, Supports displaying BP Pie Chart | Different |
| **Intended population of algorithm** | Adult and pediatric population | Adult and pediatric population | Same |
4 / 7
{8}
Traditional 510(k) of Holter ECG and ABP system
K251591
Table 2: Comparison to Secondary Predicate Device
| Item | Secondary Predicate device K210139 Ambulatory Blood Pressure Monitor Models: SA-10, SA-05, SA-06, SA-08 and SA-09 | Subject Device Holter ECG and ABP system (Model: HolterABP) | Comparison Result |
| --- | --- | --- | --- |
| Manufacturer | Edan Instruments, Inc | Edan Instruments, Inc | Same |
| K# | K210139 | K251591 | -- |
| Intended Use/Indication for use | The Ambulatory Blood Pressure Monitor is capable of measuring systolic and diastolic blood pressures, and pulse of adults and pediatrics (>12 years) over a preprogrammed period of time. These measurements are stored in the monitor and can be transferred to analysis system via wire or wireless transmission. It is intended for use as an aid or adjunct to diagnosis and treatment. | The Holter ECG and ABP system is intended to acquire, analyze, display, edit and generate report of ambulatory ECG, ambulatory blood pressure and combined Holter ECG and ABP data. The analysis results are offered to doctors on an advisory basis only. Data acquired may be used for the following indications: • Evaluation of symptoms suggesting arrhythmia or myocardial ischemia. • Evaluation of patients for ST segment changes. • Evaluation of drug response in patients taking anti-arrhythmic medications. • Evaluation of patients with pacemakers. The ambulatory blood pressure function is intended for use in patients over 12 years of age from data collected from an FDA cleared ambulatory blood | Different |
{9}
Traditional 510(k) of Holter ECG and ABP system
K251591
| | | pressure cuff.
The ECG Analysis function is intended for use in adults and pediatric, but arrhythmia detection is intended for use in patients over 12 years of age.
The Holter ECG and ABP system is intended to be used only in hospital and healthcare facilities by physicians or trained healthcare professionals. | |
| --- | --- | --- | --- |
| **Intended Patient Population** | Ambulatory BP part: adult patients and pediatrics (>12yrs.) | Ambulatory BP part: adult patients and pediatrics (>12yrs.) | Same |
| **ABP data viewer** | Display BP Data Table,
Day & Night BP Profile(including White-Coat phenomenon),
BP Histogram,
Supports displaying BP Pie Chart,
Supports BP thresholds and data elimination thresholds setting
Supports Day/Night Setting | Display BP Data Table,
Day & Night BP Profile(including White-Coat phenomenon),
BP Histogram,
Supports displaying BP Pie Chart,
Supports BP thresholds and data elimination thresholds setting
Supports Day/Night Setting | Same |
| **ABP data statistics show** | Based on the blood pressure values stored in the received record, the following results can be statistically output:
SBP, DBP, PR, PP, Valid Data(%), Morning Surge, BP Load, BPR, Morning BP, Asleep Dip, coefficient of variation (CV) | Based on the blood pressure values stored in the received record, the following results can be statistically output:
SBP, DBP, PR, PP, Valid Data(%), Morning Surge, BP Load, BPR , Morning BP, Asleep Dip, coefficient of variation (CV) | Same |
Although there are differences between the subject device and the predicate device, after verification, the difference between them does not raise different questions of safety and effectiveness of Subject device.
**10. Performance Data:**
{10}
Traditional 510(k) of Holter ECG and ABP system
K251591
## Non-clinical test:
### Biocompatibility Testing
Not applicable.
### Electrical safety and electromagnetic compatibility (EMC)
Not applicable.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.”
## Bench Testing:
Edan has conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets relevant consensus standards.
- IEC 60601-2-47 Edition 2.0 2012-02 Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems
## Performance validation:
EDAN used the public datasets (AHA, MIT-BIH, NST, ESC, QT database) and independent EDAN proprietary datasets to validate the subject device. Performance metrics assessed include sensitivity, positive predictive value, and false positive rate for QRS detection and classification, arrhythmia and event detection, as well as mean error, standard deviation for ST and QT interval measurements and RMS error for HR calculation. All the outcomes were compared against clinically derived acceptance criteria and demonstrated the substantial equivalence of the subject device.
## Clinical testing:
Not applicable. Clinical testing is not required to establish substantial equivalence to the predicate device.
## 11. Conclusion
The bench testing data and software verification and validation demonstrate that Holter ECG and ABP system, Model: HolterABP is substantially equivalent to the predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.