21 CFR 870.1415 — Coronary Vascular Physiologic Simulation Software

Cardiovascular (CV) · Part 870 Subpart B—Cardiovascular Diagnostic Devices · § 870.1415

Identification

A coronary vascular physiologic simulation software device is a prescription device that provides simulated functional assessment of blood flow in the coronary vascular system using data extracted from medical device imaging to solve algorithms and yield simulated metrics of physiologic information (e.g., blood flow, coronary flow reserve, fractional flow reserve, myocardial perfusion). A coronary vascular physiologic simulation software device is intended to generate results for use and review by a qualified clinician.

Classification Rationale

Class II (special controls). The special controls for this device are:

Product Codes

Product CodeDevice NameClassDevicesAttributes
PJACoronary Vascular Physiologic Simulation Software29SaMD
QEKAngiographic Coronary Vascular Physiologic Simulation Software22SaMD

Special Controls

PJA — Coronary Vascular Physiologic Simulation Software

In combination with the general controls of the FD&C Act. the Coronary Physiologic Simulation Software Device is subject to the following special controls:

De Novo Order DEN130045

PJA — Coronary Vascular Physiologic Simulation Software

*Classification.* Class II (special controls). The special controls for this device are:(1) Adequate software verification and validation based on comprehensive hazard analysis, with identification of appropriate mitigations, must be performed, including: (i) Full characterization of the technical parameters of the software, including: (A) Any proprietary algorithm(s) used to model the vascular anatomy; and (B) Adequate description of the expected impact of all applicable image acquisition hardware features and characteristics on performance and any associated minimum specifications; (ii) Adequate consideration of privacy and security issues in the system design; and (iii) Adequate mitigation of the impact of failure of any subsystem components ( *e.g.,* signal detection and analysis, data storage, system communications and cybersecurity) with respect to incorrect patient reports and operator failures.(2) Adequate non-clinical performance testing must be provided to demonstrate the validity of computational modeling methods for flow measurement; and (3) Clinical data supporting the proposed intended use must be provided, including the following: (i) Output measure(s) must be compared to a clinically acceptable method and must adequately represent the simulated measure(s) the device provides in an accurate and reproducible manner; (ii) Clinical utility of the device measurement accuracy must be demonstrated by comparison to that of other available diagnostic tests ( *e.g.,* from literature analysis);(iii) Statistical performance of the device within clinical risk strata ( *e.g.,* age, relevant comorbidities, disease stability) must be reported;(iv) The dataset must be adequately representative of the intended use population for the device ( *e.g.,* patients, range of vessel sizes, imaging device models). Any selection criteria or limitations of the samples must be fully described and justified;(v) Statistical methods must consider the predefined endpoints: (A) Estimates of probabilities of incorrect results must be provided for each endpoint, (B) Where multiple samples from the same patient are used, statistical analysis must not assume statistical independence without adequate justification, and (C) The report must provide appropriate confidence intervals for each performance metric; (vi) Sensitivity and specificity must be characterized across the range of available measurements; (vii) Agreement of the simulated measure(s) with clinically acceptable measure(s) must be assessed across the full range of measurements; (viii) Comparison of the measurement performance must be provided across the range of intended image acquisition hardware; and (ix) If the device uses a cutoff threshold or operates across a spectrum of disease, it must be established prior to validation, and it must be justified as to how it was determined and clinically validated; (4) Adequate validation must be performed and controls implemented to characterize and ensure consistency ( *i.e.,* repeatability and reproducibility) of measurement outputs:(i) Acceptable incoming image quality control measures and the resulting image rejection rate for the clinical data must be specified, and (ii) Data must be provided within the clinical validation study or using equivalent datasets demonstrating the consistency ( *i.e.,* repeatability and reproducibility) of the output that is representative of the range of data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment;(A) Testing must be performed using multiple operators meeting planned qualification criteria and using the procedure that will be implemented in the production use of the device, and (B) The factors ( *e.g.,* medical imaging dataset, operator) must be identified regarding which were held constant and which were varied during the evaluation, and a description must be provided for the computations and statistical analyses used to evaluate the data;(5) Human factors evaluation and validation must be provided to demonstrate adequate performance of the user interface to allow for users to accurately measure intended parameters, particularly where parameter settings that have impact on measurements require significant user intervention; and (6) Device labeling must be provided that adequately describes the following: (i) The device's intended use, including the type of imaging data used, what the device measures and outputs to the user, whether the measure is qualitative or quantitative, the clinical indications for which it is to be used, and the specific population for which the device use is intended; (ii) Appropriate warnings specifying the intended patient population, identifying anatomy and image acquisition factors that may impact measurement results, and providing cautionary guidance for interpretation of the provided measurements; (iii) Key assumptions made in the calculation and determination of simulated measurements; (iv) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance. Per-vessel clinical performance, including where applicable localized performance according to vessel and segment, must be included as well as a characterization of the measurement error across the expected range of measurement for key parameters based on the clinical data; (v) A detailed description of the patients studied in the clinical validation ( *e.g.,* age, gender, race or ethnicity, clinical stability, current treatment regimen) as well as procedural details of the clinical study (*e.g.,* scanner representation, calcium scores, use of beta-blockers or nitrates); and(vi) Where significant human interface is necessary for accurate analysis, adequately detailed description of the analysis procedure using the device and any data features that could affect accuracy of results.

eCFR

PJA — Coronary Vascular Physiologic Simulation Software

(1) Adequate software verification and validation based on comprehensive hazard analysis, with identification of appropriate mitigations, must be performed, including: (i) Full characterization of the technical parameters of the software, including: (A) Any proprietary algorithm(s) used to model the vascular anatomy; and (B) Adequate description of the expected impact of all applicable image acquisition hardware features and characteristics on performance and any associated minimum specifications; (ii) Adequate consideration of privacy and security issues in the system design; and (iii) Adequate mitigation of the impact of failure of any subsystem components ( *e.g.,* signal detection and analysis, data storage, system communications and cybersecurity) with respect to incorrect patient reports and operator failures.(2) Adequate non-clinical performance testing must be provided to demonstrate the validity of computational modeling methods for flow measurement; and (3) Clinical data supporting the proposed intended use must be provided, including the following: (i) Output measure(s) must be compared to a clinically acceptable method and must adequately represent the simulated measure(s) the device provides in an accurate and reproducible manner; (ii) Clinical utility of the device measurement accuracy must be demonstrated by comparison to that of other available diagnostic tests ( *e.g.,* from literature analysis);(iii) Statistical performance of the device within clinical risk strata ( *e.g.,* age, relevant comorbidities, disease stability) must be reported;(iv) The dataset must be adequately representative of the intended use population for the device ( *e.g.,* patients, range of vessel sizes, imaging device models). Any selection criteria or limitations of the samples must be fully described and justified;(v) Statistical methods must consider the predefined endpoints: (A) Estimates of probabilities of incorrect results must be provided for each endpoint, (B) Where multiple samples from the same patient are used, statistical analysis must not assume statistical independence without adequate justification, and (C) The report must provide appropriate confidence intervals for each performance metric; (vi) Sensitivity and specificity must be characterized across the range of available measurements; (vii) Agreement of the simulated measure(s) with clinically acceptable measure(s) must be assessed across the full range of measurements; (viii) Comparison of the measurement performance must be provided across the range of intended image acquisition hardware; and (ix) If the device uses a cutoff threshold or operates across a spectrum of disease, it must be established prior to validation, and it must be justified as to how it was determined and clinically validated; (4) Adequate validation must be performed and controls implemented to characterize and ensure consistency ( *i.e.,* repeatability and reproducibility) of measurement outputs:(i) Acceptable incoming image quality control measures and the resulting image rejection rate for the clinical data must be specified, and (ii) Data must be provided within the clinical validation study or using equivalent datasets demonstrating the consistency ( *i.e.,* repeatability and reproducibility) of the output that is representative of the range of data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment;(A) Testing must be performed using multiple operators meeting planned qualification criteria and using the procedure that will be implemented in the production use of the device, and (B) The factors ( *e.g.,* medical imaging dataset, operator) must be identified regarding which were held constant and which were varied during the evaluation, and a description must be provided for the computations and statistical analyses used to evaluate the data;(5) Human factors evaluation and validation must be provided to demonstrate adequate performance of the user interface to allow for users to accurately measure intended parameters, particularly where parameter settings that have impact on measurements require significant user intervention; and (6) Device labeling must be provided that adequately describes the following: (i) The device's intended use, including the type of imaging data used, what the device measures and outputs to the user, whether the measure is qualitative or quantitative, the clinical indications for which it is to be used, and the specific population for which the device use is intended; (ii) Appropriate warnings specifying the intended patient population, identifying anatomy and image acquisition factors that may impact measurement results, and providing cautionary guidance for interpretation of the provided measurements; (iii) Key assumptions made in the calculation and determination of simulated measurements; (iv) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance. Per-vessel clinical performance, including where applicable localized performance according to vessel and segment, must be included as well as a characterization of the measurement error across the expected range of measurement for key parameters based on the clinical data; (v) A detailed description of the patients studied in the clinical validation ( *e.g.,* age, gender, race or ethnicity, clinical stability, current treatment regimen) as well as procedural details of the clinical study (*e.g.,* scanner representation, calcium scores, use of beta-blockers or nitrates); and(vi) Where significant human interface is necessary for accurate analysis, adequately detailed description of the analysis procedure using the device and any data features that could affect accuracy of results.

Ecfr Llm

QEK — Angiographic Coronary Vascular Physiologic Simulation Software

*Classification.* Class II (special controls). The special controls for this device are:(1) Adequate software verification and validation based on comprehensive hazard analysis, with identification of appropriate mitigations, must be performed, including: (i) Full characterization of the technical parameters of the software, including: (A) Any proprietary algorithm(s) used to model the vascular anatomy; and (B) Adequate description of the expected impact of all applicable image acquisition hardware features and characteristics on performance and any associated minimum specifications; (ii) Adequate consideration of privacy and security issues in the system design; and (iii) Adequate mitigation of the impact of failure of any subsystem components ( *e.g.,* signal detection and analysis, data storage, system communications and cybersecurity) with respect to incorrect patient reports and operator failures.(2) Adequate non-clinical performance testing must be provided to demonstrate the validity of computational modeling methods for flow measurement; and (3) Clinical data supporting the proposed intended use must be provided, including the following: (i) Output measure(s) must be compared to a clinically acceptable method and must adequately represent the simulated measure(s) the device provides in an accurate and reproducible manner; (ii) Clinical utility of the device measurement accuracy must be demonstrated by comparison to that of other available diagnostic tests ( *e.g.,* from literature analysis);(iii) Statistical performance of the device within clinical risk strata ( *e.g.,* age, relevant comorbidities, disease stability) must be reported;(iv) The dataset must be adequately representative of the intended use population for the device ( *e.g.,* patients, range of vessel sizes, imaging device models). Any selection criteria or limitations of the samples must be fully described and justified;(v) Statistical methods must consider the predefined endpoints: (A) Estimates of probabilities of incorrect results must be provided for each endpoint, (B) Where multiple samples from the same patient are used, statistical analysis must not assume statistical independence without adequate justification, and (C) The report must provide appropriate confidence intervals for each performance metric; (vi) Sensitivity and specificity must be characterized across the range of available measurements; (vii) Agreement of the simulated measure(s) with clinically acceptable measure(s) must be assessed across the full range of measurements; (viii) Comparison of the measurement performance must be provided across the range of intended image acquisition hardware; and (ix) If the device uses a cutoff threshold or operates across a spectrum of disease, it must be established prior to validation, and it must be justified as to how it was determined and clinically validated; (4) Adequate validation must be performed and controls implemented to characterize and ensure consistency ( *i.e.,* repeatability and reproducibility) of measurement outputs:(i) Acceptable incoming image quality control measures and the resulting image rejection rate for the clinical data must be specified, and (ii) Data must be provided within the clinical validation study or using equivalent datasets demonstrating the consistency ( *i.e.,* repeatability and reproducibility) of the output that is representative of the range of data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment;(A) Testing must be performed using multiple operators meeting planned qualification criteria and using the procedure that will be implemented in the production use of the device, and (B) The factors ( *e.g.,* medical imaging dataset, operator) must be identified regarding which were held constant and which were varied during the evaluation, and a description must be provided for the computations and statistical analyses used to evaluate the data;(5) Human factors evaluation and validation must be provided to demonstrate adequate performance of the user interface to allow for users to accurately measure intended parameters, particularly where parameter settings that have impact on measurements require significant user intervention; and (6) Device labeling must be provided that adequately describes the following: (i) The device's intended use, including the type of imaging data used, what the device measures and outputs to the user, whether the measure is qualitative or quantitative, the clinical indications for which it is to be used, and the specific population for which the device use is intended; (ii) Appropriate warnings specifying the intended patient population, identifying anatomy and image acquisition factors that may impact measurement results, and providing cautionary guidance for interpretation of the provided measurements; (iii) Key assumptions made in the calculation and determination of simulated measurements; (iv) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance. Per-vessel clinical performance, including where applicable localized performance according to vessel and segment, must be included as well as a characterization of the measurement error across the expected range of measurement for key parameters based on the clinical data; (v) A detailed description of the patients studied in the clinical validation ( *e.g.,* age, gender, race or ethnicity, clinical stability, current treatment regimen) as well as procedural details of the clinical study (*e.g.,* scanner representation, calcium scores, use of beta-blockers or nitrates); and(vi) Where significant human interface is necessary for accurate analysis, adequately detailed description of the analysis procedure using the device and any data features that could affect accuracy of results.

eCFR

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...