← Product Code [JII](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/JII) · K033360

# DIAZYME LITHIUM ENZYMATIC ASSAY KIT (K033360)

_General Atomics · JII · Dec 23, 2003 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-d%E2%80%94clinical-toxicology-test-systems/JII/K033360

## Device Facts

- **Applicant:** General Atomics
- **Product Code:** [JII](/submissions/TX/subpart-d%E2%80%94clinical-toxicology-test-systems/JII.md)
- **Decision Date:** Dec 23, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3560
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

Diazyme Lithium Enzymatic Assay Kit is for quantitative in vitro determination of lithium in human serum. Measurements of lithium are carried out essentially to ensure that the proper drug dosage is administered in the treatment of patients suffering from bipolar disorder and to avoid toxicity

## Device Story

Diazyme Lithium Enzymatic Assay is a spectrophotometric, kinetic coupling assay for quantitative determination of lithium in human serum and plasma. Input: patient serum/plasma samples. Principle: proprietary phosphatase activity sensitive to lithium (IC50=0.1mM) converts adenosine biphosphate (PAP) to hypoxanthine; subsequent enzymatic reactions generate uric acid and hydrogen peroxide (H2O2); H2O2 reacts with EHSPT and 4-aminoantipyrine in presence of peroxidase to form a quinone dye (max absorbance 556nm). Output: rate of dye formation, inversely proportional to lithium concentration. Used in clinical chemistry analyzers; operated by laboratory personnel. Results assist clinicians in adjusting lithium dosage for bipolar disorder treatment to maintain therapeutic levels and prevent toxicity.

## Clinical Evidence

Bench testing only. Precision evaluated at two labs (n=80 per level); total precision CV 5.5-6.9%. Linearity established 0.1-3.0 mM. Analytical recovery 94-107%. Specificity tested against common serum cations/substances with <10% deviation. Method comparison against predicate (n=61) showed slope 0.94, r=0.97, intercept 0.1. No clinical studies performed.

## Technological Characteristics

Two-reagent (R1 and R2) kinetic spectrophotometric assay. Proprietary phosphatase-based enzymatic coupling system. Measures absorbance at 550 nm. Compatible with automated clinical chemistry analyzers. Measuring range: 0-3 mmol/L.

## Regulatory Identification

A lithium test system is a device intended to measure lithium (from the drug lithium carbonate) in serum or plasma. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).

## Predicate Devices

- Trace Lithium Test System (k003583)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number: k033360
B. Analyte: Lithium
C. Type of Test: Quantitative Enzymatic Assay
D. Applicant: General Atomics
E. Proprietary and Established Names: Proprietary - Diazyme Lithium Enzymatic Assay, Established – Lithium enzymatic assay
F. Regulatory Information:
1. Regulation section: 21 CFR 862.3560
2. Classification: Class II
3. Product Code: JII
4. Panel: 91

G. Intended Use:
1. Intended use(s): Diazyme Lithium Enzymatic Assay Kit is for quantitative in vitro determination of lithium in human serum.
2. Indication(s) for use: Diazyme Lithium Enzymatic Assay Kit is for quantitative in vitro determination of lithium in human serum. Measurements of lithium are carried out essentially to ensure that the proper drug dosage is administered in the treatment of patients suffering from bipolar disorder and to avoid toxicity
3. Special condition for use statement(s): No special conditions were stated.
4. Special instrument Requirements: Roche Cobas Mira and Hitachi 717

H. Device Description: The Diazyme lithium assay is determined spectrophotometrically through a kinetic coupling assay system involving a proprietary Phosphatase. Through enzymatic coupling, the phosphate is converted to hypoxanthine and generates hydrogen peroxide which in the presence of peroxidase forms a quinine dye with an absorbance of 556nm.

{1}

Page 2 of 3

# I. Substantial Equivalence Information:

1. Predicate device name(s): Trace Lithium Test System
2. Predicate K number(s): k003583
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Principle | Diazyme Enzymatic Colorimetric | Trace colorimetric  |
|  Type of test | Quantitative | Quantitative  |
|  Specimen type | Human Serum | Human Serum and Plasma  |
|  Reagent | Two reagent system | One reagent system  |

J. Standard/Guidance Document Referenced (if applicable): NCCLS EP5-A
K. Test Principle: Enzymatic colorimetric

# L. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility: The Diazyme Lithium Enzymatic assay was evaluated at two laboratories. In each run, low control serum sample containing 1.0 mM lithium and high control containing 2.2 mM lithium were analyzed. Within run precision @ 1mM Lithium 20 days, n=80, mean = 1.0mM, C.V. = 4.7%. Within run precision @ 2.3mM Lithium 20 days n=80, mean = 2.2, C.V = 3.3%. Total precision @ 1mM Lithium 20 days, n=80, mean = 1.0mM, C.V = 6.9%. Total precision @ 2.2 mM Lithium 20 days n=80, mean = 2.2, C.V = 5.5%.
b. Linearity/assay reportable range: Lithium free serum samples spiked with 1 M lithium acetate ranging from 0mM to 5mM were tested on UV-visible spectrophotometer and automated analyzers. The linearity of the Diazyme Lithium assay is 0.1 mM to 3.0 mM.
c. Traceability (controls, calibrators, or method): Serum calibrators 0mM, 1mM and 3mM are prepared by spiking lithium free serum with a 1mM lithium acetate stock solution
d. Detection limit: When testing serum samples between 0.1 mM to 3 mM lithium levels analytical recovery range ranged from 94 – 107%.

{2}

Page 3 of 3

e. Analytical specificity: To determine the level of interference from other cations and substances normally present in the serum, the Diazyme Lithium Assay was tested with 1.0 mM lithium serum spiked with a variety of concentration of substances. The following cations and substances normally present in the serum reduced less than 10% deviation when tested at levels equal to the following concentrations – NH4 – 0.5 mM, Pi – 1.5 mM, Ca – 5 mM, Na – 200 mM, K – 10 mM, Cu – 0.25 mM, Fe – 0.25 mM, Zn – 0.25 mM, Triglyceride 250 mg/dl, Ascorbic Acid – 5 mM, Bilirubin – 45 mg/dl.

f. Assay cut-off: The assay cutoff is established as the linear range of the Diazyme Lithium Assay (0.1 mM – 3.0 mM).

2. Comparison studies:

a. Method comparison with predicate device: The correlation of the Diazyme Lithium Assay was evaluated against the predicate. N=61 slope = 0.94 and r=0.97 intercept = 0.1

b. Matrix comparison:

3. Clinical studies:

a. Clinical sensitivity:

b. Clinical specificity:

c. Other clinical supportive data (when a and b is not applicable):

4. Clinical cut-off:

5. Expected values/Reference range:

A trough concentration for 12 hours post dose is expected to be 1.0 -1.2 mM. Levels higher than 1.5 nM, 12 hours after a dose indicates a significant risk of intoxication. Referenced to N. Tietz. Textbook of Clinical Chemistry, p. 1841. W.B. Saunders Company, Philadelphia (1986).

The recommendation is made in the labeling that the above values should serve as reference only. Each laboratory should verify this range for the population it serves.

M. Conclusion: Based upon the information provided, I recommend that the Diazyme Lithium Enzymatic Assay be found substantially equivalent with the predicate devices as defined in 21 CFR 862.3560.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-d%E2%80%94clinical-toxicology-test-systems/JII/K033360](https://fda.innolitics.com/submissions/CH/subpart-d%E2%80%94clinical-toxicology-test-systems/JII/K033360)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
