The SMART M-CELL Bone Marrow Concentration System is intended to be used in the clinical laboratory or intraoperatively at point-of-care for the safe and rapid preparation of a cell concentrate from bone marrow.
Device Story
System comprises centrifuge (SMART M-CELL 2 or 4) and disposable Procedure Pack. Input: patient bone marrow. Operation: two-phase centrifugation cycle (primary 2300 RPM; secondary 2200 RPM) separates marrow into erythrocytes, platelet-poor plasma, and platelet-rich plasma (PRP) based on density. Procedure packs (BmSC+③/⑨) include centrifuge barrels and blunt cannulas for manual extraction of PRP and removal of PPP. Used in clinical labs or intraoperatively by clinicians. Output: concentrated bone marrow components. Clinicians use output for therapeutic applications; device facilitates rapid, safe preparation of autologous cell concentrates.
Clinical Evidence
Bench testing only. Comparative study (n=not specified) using healthy donor bone marrow compared subject device to predicate (K052925). Metrics included cell count profiles and platelet recovery rates, showing equivalence. Rotor-speed performance testing verified operational parameters across all procedure pack volumes. Disinfection study confirmed 83% ethyl alcohol efficacy against non-enveloped viruses. Robustness study confirmed device integrity over 6,000 cleaning/disinfection cycles (simulated 5-year life).
Technological Characteristics
Centrifuge-based density gradient separation. Fixed-speed/time cycles (2300 RPM/6 min primary; 2200 RPM/11-14 min secondary). Procedure packs include centrifuge barrels, blunt cannulas, and spacers. Sterilization: Ethylene Oxide (ISO 11135). Electrical safety: IEC 61010-1. Factory calibrated. Standalone operation.
Indications for Use
Indicated for clinical laboratory or intraoperative point-of-care preparation of bone marrow cell concentrate. For prescription use only. In vitro use only.
Regulatory Classification
Identification
General purpose laboratory equipment labeled or promoted for a specific medical use is a device that is intended to prepare or examine specimens from the human body and that is labeled or promoted for a specific medical use.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
INSTRUMENT ONLY
## I Background Information:
A 510(k) Number
K250518
B Applicant
Miracell Co., Ltd
C Proprietary and Established Names
SMART M-CELL Bone Marrow Concentration System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QBV | Class I | 21 CFR 862.2050 - General Purpose Laboratory Equipment Labeled Or Promoted For A Specific Medical Use | Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
Clearance of a new device
B Type of Test:
Not applicable
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K250518 - Page 2 of 6
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The SMART M-CELL Bone Marrow Concentration System is intended to be used in the clinical laboratory or intraoperatively at point-of-care for the safe and rapid preparation of a cell concentrate from bone marrow.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
It is for in vitro use only.
The safety and effectiveness of this device for in vivo indications for use has not been established
## IV Device/System Characteristics:
### A Device Description:
The SMART M-CELL Bone Marrow Concentration System consists of a centrifuge component (i.e., SMART M-CELL 2 or SMART M-CELL 4) and a Procedure Pack.
The SMART M-CELL 2/SMART M-CELL 4 centrifuges provide a two-phase centrifugation cycle consisting of a primary centrifugation and secondary centrifugation. The centrifuge speed (2300 RPM for primary separation; 2200 RPM for secondary separation) and time (6 minutes for primary separation; 11 minutes (SMART M-CELL 2) or 14 minutes (SMART M-CELL 4) for secondary separation) are fixed without any operator adjustment.
#### PRINCIPLE:
Bone marrow collected in BmSC Transfer Bag from the patient is loaded into the separation compartment of the Bone marrow separating Kit. The centrifugation cycle will separate the bone marrow into erythrocytes, platelet poor plasma, and platelet rich plasma (PRP). The process involves a two-phase centrifugation cycle consisting of primary centrifugation and secondary centrifugation. Once the two-phase separation process is complete, allowing for collection of the bone marrow component layers by separating based on density of liquids.
The Procedure Pack has two models: BmSC+③ and BmSC+⑨, each with three volumes (60, 120 and 240 mL). The process kit includes the following components:
{2}
| Process kit model | Components | Description |
| --- | --- | --- |
| BmSC+③ 60 Process Kit /BmSC+⑨ 60 Process Kit | BmSC+③/BmSC+⑨ 60 Process Disposable (PD) (1 each) | Centrifuge barrel to separate bone marrow into layers
In BmSC+③ and BmSC+⑨, Chamber 1: where bone marrow is injected
Chamber 2: where cell concentrate is collected |
| | Blunt Cannula and Cap (1 each) | Extract platelet rich plasma (PRP) in the PD barrel after centrifugation. |
| | Blunt Cannula, Spacer and Cap (1 each) | Remove platelet poor plasma (PPP) from the PD after centrifugation to leave approximately 10 mL PRP. |
| | IN Blunt Cannula | Attach to the syringe and transfer the filtered bone marrow into the PD. |
| BmSC+③ 120 Process Kit /BmSC+⑨ 120 Process Kit | BmSC+③/BmSC+⑨ 60 Process Disposable (PD) (2 each) | Same as BmSC+③ 60 /BmSC+⑨ 60 PD. |
| | Blunt Cannula and Cap (2 each) | |
| | Blunt Cannula and Spacer (2 each) | |
| | IN Blunt Cannula | Same as BmSC+③ 60 /BmSC+⑨ 60 PD. |
| BmSC+③ 240 Process Kit /BmSC+⑨ 240 Process Kit | BmSC+③/BmSC+⑨ 60 Process Disposable (PD) (4 each) | Same as BmSC+③ 60 Process Kit /BmSC+⑨ 60 PD. |
| | Blunt Cannula and Cap (4 each) | |
| | Blunt Cannula, Spacer and Cap (4 each) | |
| | IN Blunt Cannula | Same as BmSC+③ 60 /BmSC+⑨ 60 PD. |
## B Instrument Description Information:
1. **Instrument Name:**
SMART M-CELL 2 and SMART M-CELL 4
2. **Specimen Identification:**
Sample identification label
3. **Specimen Sampling and Handling:**
Bone Marrow collection should be performed in accordance with the standard operating procedures (SOPs) established at the bone marrow collection site.
4. **Calibration:**
Factory Calibrated
K250518 - Page 3 of 6
{3}
K250518 - Page 4 of 6
5. Quality Control:
Not applicable
V Substantial Equivalence Information:
A Predicate Device Name(s):
Smartprep2 Centrifuge System
B Predicate 510(k) Number(s):
K052925
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K250518 | K052925 |
| --- | --- | --- |
| Device Trade Name | SMART M-CELL Bone Marrow Concentration System | SmartPReP2 Centrifuge System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The SMART M-CELL Bone Marrow Concentration System is intended to be used in the clinical laboratory or intraoperatively at point-of-care for the safe and rapid preparation of a cell concentrate from bone marrow. | The SmartPReP2 Centrifuge System is intended to be used in the clinical laboratory or intraoperatively at point-of-care for the safe and rapid preparation of platelet poor plasma and platelet concentrate from a small sample of blood and for preparation of a cell concentrate from bone marrow. |
| Principle of Operation | Density gradient separation by a two-phase centrifugation cycle | Same |
| Device Components | Centrifuge, disposable procedure pack | Centrifuge, disposable processing procedure pack |
| Sterilization Method for Processing Kit | Ethylene Oxide | Same |
| General Device Characteristic Differences | | |
| Centrifuge | SMART M-CELL 2 and SMART M-CELL 4 | SMART PReP2 |
{4}
| Centrifugation Time and Speed | Centrifugation Time: 17 minutes for SMART M-CELL 2 and 20 minutes for SMART M-CELL 4
Primary centrifugation: 2300 RPM
Second centrifugation: 2200 RPM | Centrifugation Time: 14 minutes
Primary centrifugation: 2500 ± 150 RPM
Second centrifugation: 2300 ± 140 RPM |
| --- | --- | --- |
VI Standards/Guidance Documents Referenced:
- ISO 11135 Second edition (2014): Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
- IEC 61010-1 Edition 3.1 (2017): Safety requirements for electrical equipment for measurement control and laboratory use - Part 1: General requirements [Including: Corrigendum 1 (2019)]
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Not applicable
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Accuracy (Instrument):
Comparative studies were performed using BmSC+③ 60 and BmSC+⑨ 60 kits with the SMART M-CELL 2 device at a single site. Bone marrow samples were collected from healthy donors, and baseline cell counts were assessed prior to centrifugation. Each sample was then divided, with half processed using the subject device and the other half processed using the predicate SmartPReP2 Centrifuge System (K052925). Cell counts from the processed samples were compared to the initial baseline samples and evaluated side-by-side between the subject and predicate devices. The results demonstrate that the SMART M-CELL Bone Marrow Concentration System is equivalent to the predicate device in preparing bone marrow concentrate, with comparable cell count profiles and platelet recovery rates.
K250518 - Page 5 of 6
{5}
A separate rotor-speed performance test was conducted to verify the operational parameters of the SMART M-CELL 2 and SMART M-CELL 4 centrifuge systems. Testing was performed using multiple procedure packs including BmSC+③ 60, BmSC+③ 120, BmSC+③ 240, under specified centrifugation conditions for SMART M-CELL 2 and SMART M-CELL 4. Both models met all acceptance criteria, with appropriate centrifugation results and no instances of product deviation or plasma separation observed. The results demonstrate that both device models perform consistently according to their design parameters with respect to centrifugation performance and operational functionality.
5. Carry-Over:
Not applicable
B Other Supportive Instrument Performance Characteristics Data:
Disinfection Study
A study was conducted to evaluate the antimicrobial activity of 83% ethyl alcohol against non-enveloped viruses. The study used standardized test methods to determine the log reduction in viral titer following exposure to the 83% ethyl alcohol for 5 min at room temperature. The results demonstrated that the subject device achieved significant viral inactivation, with log reduction values meeting established performance criteria for virucidal efficacy.
Robustness Study
The study evaluated the device's ability to withstand repeated exposure to cleaning and disinfection over its expected 5-year service life. The study employed a worst-case scenario of 6,000 cleaning and disinfection cycles on the SMART M-CELL 4 model. Testing was performed according to the cleaning and disinfection procedures specified in the user manual (83% ethyl alcohol), with comprehensive functional, display, and visual inspections conducted at 1,000-cycle intervals. All acceptance criteria were met throughout the validation, with no evidence of functional degradation, display malfunction, material damage, or corrosion observed at any inspection point. The results demonstrate that the device maintains its structural integrity and functional performance after repeated exposure to the instructed cleaning and disinfection process.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K250518 - Page 6 of 6
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.