K961183 · Columbia Bioscience, Inc. · JJQ · Aug 2, 1996 · Clinical Chemistry
Device Facts
Record ID
K961183
Device Name
CBI 7520 READER
Applicant
Columbia Bioscience, Inc.
Product Code
JJQ · Clinical Chemistry
Decision Date
Aug 2, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2300
Device Class
Class 1
Indications for Use
The CBI 7520 Microplate Reader is intended to be used as a general purpose microplate photometer for clinical use.
Device Story
CBI 7520 Microplate Reader; general purpose photometer for clinical laboratory use. Input: light transmitted through microtiter plate wells. Operation: Tungsten halogen lamp light source; silicon-photodiode sensors measure light absorbance per well; automatic transport mechanism positions wells. Output: absorbance values processed by software to yield qualitative or quantitative assay results. Used in clinical settings by laboratory personnel. Benefits: enables automated, high-throughput measurement of clinical assays.
Clinical Evidence
Bench testing only; no clinical data presented. Equivalence established via comparison of technical characteristics and principles of operation.
Technological Characteristics
Tungsten halogen lamp light source; silicon-photodiode detectors; automatic microplate transport mechanism. Standalone benchtop photometer. Data reduction software performs absorbance calculations.
Indications for Use
Indicated for clinical use as a general purpose microplate photometer for measuring light absorbance in microtiter plate wells.
Regulatory Classification
Identification
A colorimeter, a photometer, or a spectrophotometer for clinical use is an instrument intended to measure radiant energy emitted, transmitted, absorbed, or reflected under controlled conditions. The device may include a monochromator to produce light of a specific wavelength.
{0}
K961183
AUG - 2 1996
Attachment 7
# 510(K) SUMMARY
## CBI 7520 Microplate Reader
### Submitter's Name, Address, Telephone and Fax Number, And Contact Person
Columbia Bioscience, Inc.
182 Thomas Johnson Drive
Suite 205
Frederick, Maryland 21702
Contact:
- Norman Jenkins
- Columbia Bioscience, Inc.
- 182 Thomas Johnson Drive
- Suite 205
- Frederick, Maryland 21702
Phone: (301) 941-7971
Facsimile: (410) 995-0508
### Name of the Device
CBI 7520 Microplate Reader
### Common or Usual Name
Microplate reader
### Predicate Devices
1. Anthos 2001 Microplate Reader (K894270);
2. Anthos ht2 Microplate Reader (K931907)
3. Anthos 2010 Microplate Reader (K9xxxxx);
{1}
15
# Intended Use
The CBI 7520 Microplate Reader is intended to be used as a general purpose microplate photometer for clinical use.
# Principles of Operation
The CBI 7520 and its predicate devices all share the same principle of operation. Briefly, each device provides a light source which is focused to provide illumination of wells in a microtiter plate. A corresponding silicon-photodiodes measures the amount of light absorbed by the sample as the light passes through the microplate well. The respective well absorbance measurements is used by the data reduction software to yield a specific well absorbance value which in turn can yield a qualitative or quantitative assay result. Thus, the CBI 7520 and its predicate devices have similar principles of operation.
# Technical Characteristics
The CBI 7520 microplate reader and the predicate readers employ an automatic transport mechanism to bring the individual microplate wells into position for reading. The devices all employ Tungsten halogen lamp as the light source. All three devices use silicon-photodiodes to measure the light absorbance.
# Summary of the Basis for the Finding of Substantial Equivalence
The CBI 7520 for the measurement of light absorbance is equivalent to the predicate devices for its intended use. There is no substantial difference between the CBI 7520 and its predicate devices in performance or technical characteristics. The CBI 7520 has the same intended use, indications for use, and the same principles of operation for its indications for use as the predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.