OPIATE ORAL FLUID HOMOGENEOUS ENZYME IMMUNOASSAY, CALIBRATORS AND CONTROLS

K050988 · Lin-Zhi International, Inc. · DLG · Apr 6, 2006 · Clinical Chemistry

Device Facts

Record IDK050988
Device NameOPIATE ORAL FLUID HOMOGENEOUS ENZYME IMMUNOASSAY, CALIBRATORS AND CONTROLS
ApplicantLin-Zhi International, Inc.
Product CodeDLG · Clinical Chemistry
Decision DateApr 6, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2250
Device ClassClass 1

Indications for Use

The Opiate Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay system to detect opiates in human saliva with a cutoff of 30 ng/mL when testing oral fluid specimen collected with Salivette collector (manufactured by Sarstedt) and diluted with 1 mL of buffer. The calibrators and controls of the analyte Morphine are prepared with oral fluid buffer so that it can be used to verify and validate the assay. The assay is intended for use in the qualitative determination for opiate drugs. The assay is designed for professional use with a number of automated clinical chemistry analyzers. The Opiate Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay system provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Device Story

Liquid reagent homogeneous enzyme immunoassay; detects opiates in human saliva. Principle: competition between G6PDH-labeled drug and free drug in saliva sample for fixed antibody binding sites; absence of free drug allows antibody to bind labeled enzyme, decreasing activity. Enzyme activity measured spectrophotometrically at 340 nm via conversion of NAD to NADH; drug concentration proportional to enzyme activity. Used in clinical laboratories; operated by professional staff on automated clinical chemistry analyzers. Provides preliminary qualitative results; requires GC/MS confirmation. Assists clinicians in drug-of-abuse screening.

Clinical Evidence

No clinical data provided. Substantial equivalence established via bench testing of assay performance, including sensitivity, specificity, and cross-reactivity profiles.

Technological Characteristics

Homogeneous enzyme immunoassay; liquid reagents (antibody/substrate and enzyme/drug conjugate). Sensing principle: competitive binding measured via absorbance change at 340 nm. Requires automated clinical analyzer (Roche Hitachi 717). Calibrators/controls based on oral fluid buffer. Stability: 18 months at 2-8°C.

Indications for Use

Indicated for the qualitative detection of opiates in human saliva specimens collected via Salivette collector for professional use in clinical chemistry laboratories. Not for point-of-care settings.

Regulatory Classification

Identification

A gas liquid chromatography system for clinical use is a device intended to separate one or more drugs or compounds from a mixture. Each of the constituents in a vaporized mixture of compounds is separated according to its vapor pressure. The device may include accessories such as columns, gases, column supports, and liquid coating.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k050988 B. Purpose for Submission: Clearance of new device C. Measurand: Opiates (Morphine) D. Type of Test: Qualitative Immunoassay E. Applicant: Lin-Zhi International Inc. F. Proprietary and Established Names: Opiate Oral Fluid Enzyme Immunoassay, Calibrators, and Controls G. Regulatory Information: 1. Regulation section: 21 CFR §862.3650, Opiate test system 21 CFR §862.3200, Clinical toxicology calibrator 21 CFR §862.3280, Clinical toxicology control material 2. Classification: Class II 3. Product code: DJG, enzyme immunoassay, opiates DLJ, calibrators, drug specific LAS, drug specific control materials 4. Panel: Toxicology (91) H. Intended Use: 1. Intended use(s): See Indications for Use {1} 2. Indication(s) for use: "The Opiate Enzyme Immunoassay is a homogeneous enzyme immunoassay system to detect opiates in human saliva with a cutoff of 30 ng/mL when testing oral fluid specimen collected with Salivette collector (manufactured by Sarstedt) and diluted with 1 mL of buffer. The calibrators and controls of the analyte Morphine are prepared with oral fluid buffer so that it can be used to verify and validate the assay. The assay is intended for use in the qualitative determination of methadone drugs. The assay is designed for professional use with a number of automated clinical analyzers. The Opiate Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay system provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used." 3. Special conditions for use statement(s): For prescription, professional use only in clinical chemistry laboratories; the assay is not designated for use in point-of-care settings. 4. Special instrument requirements: Roche Hitachi 717 Clinical Chemistry Analyzer I. Device Description: The device consists of two liquid reagents; an antibody/substrate reagent and an enzyme/drug conjugate. The antibody/substrate reagent contains mouse monoclonal anti-morphine antibody, glucose-6-phosphate (G6P), NAD, stabilizers, and preservative. The enzyme/drug conjugate contains morphine-labeled glucose-6-phosphate dehydrogenase (G6PDH) in buffer with preservatives. Calibrators (3 levels) and controls (2 levels) are based on oral fluid buffer with preservatives and contain varying levels of morphine. J. Substantial Equivalence Information: 1. Predicate device name(s): LZI Opiate EIA OraSure Opiates Intercept Micro-Plate EIA Dade Behring Emit II Plus Opiate Assay 2. Predicate 510(k) number(s): k020368 k981341 k011289 {2} 3. Comparison with predicate: The device and the predicate devices share a similar intended use. The device uses similar reagents and test principle as the LZI urine assay, and the device is intended for the same matrix as the OraSure assay. The cutoff differs between the assays. K. Standard/Guidance Document Referenced (if applicable): None referenced by the sponsor. L. Test Principle: Enzyme-labeled drug and drug present in the sample compete for limited monoclonal mouse anti-morphine antibody binding sites. Binding of the enzyme-labeled drug inhibits its reaction with the substrate, thereby influencing the rate of absorbance change measured by the instrument. The rate of absorbance change is proportional to the concentration of drug in the sample. Concentrations of controls and unknowns are calculated from the standard curve. Results are read at 340 nm. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The negative, cutoff, and high calibrators and two levels of controls were assayed with the Hitachi 717 analyzer in replicates of 6, twice a day for 10 days. The within-run % CV ranged from 0.54 to 0.70% and the total precision %CV ranged from 0.73 to 0.87%. A recovery study showed that the assay correctly identified six spiked samples at approximately 50% above the cutoff as positive and six spiked samples at approximately 50% below the cutoff as negative. b. Linearity/assay reportable range: Not applicable. This is a qualitative assay. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: A commercially available standard solution is made into a secondary (lower concentration) stock solution. The secondary stock solution is then spiked into the calibrators and controls to the desired concentration. The concentrations are confirmed by GC/MS. Stability Studies: Real time and accelerated studies meet the manufacture's acceptance criteria and support a claim of 18 months when the product is stored at 2-8 °C Stability of morphine in the Salivette collection device was determined by taking a pool of negative oral fluid samples spiked at three different concentrations and split into three hundred and thirty-three samples. On day 3 {3} one, 10 samples were run to determine the baseline for comparison to subsequent runs. Of the remaining one hundred samples for each concentration half were stored at 2-8 °C and the other half were stored frozen at -20 °C. The study was conducted over 22 days; samples were run on days 1, 2, 5, 8, 14 and 22. The results support the manufacturer's claim that the sample is stable for 22 days when stored at refrigerator 2-8 °C or frozen at -20 °C. d. Detection limit: See the Precision/Reproducibility section above for performance around the stated cutoff concentration. e. Analytical specificity: Various morphine-related and -unrelated compounds were tested for cross-reactivity with the assay. Test compounds were spiked into drug-free negative oral fluid calibrator matrix and evaluated against the cutoff calibrator. The following table lists the lowest concentration that gave a response equivalent to that of the cutoff calibrator. Cross-reactivity was calculated by dividing the concentration at which the compound yielded a positive result by the designated cut-off concentration. Opiate Oral Fluid Assay: Cross-reactivity of Related Compounds | Compound | Conc (ng/mL) | | --- | --- | | Morphine | 20 | | Codeine | 10 | | Oxycodone | 100 | | Hydroxycodone | 100 | | Meperidine | 2000 | Potential interference caused by other substances or drugs was evaluated by adding various concentrations of each drug to oral fluid calibrator matrix and testing with the Opiate Oral Fluid Assay: all compounds below gave a response less than the cutoff calibrator at the concentration specified. Compounds that Do Not Cross React with the Opiate Oral Fluid Assay | Compound Tested | Conc (ug/mL) | | Compound Tested | Conc (ug/mL) | | --- | --- | --- | --- | --- | | Acetaminophen | 100 | | d-Methamphetamine | 25 | | Acetylsalicyclic acid | 250 | | p-Methoxymethamphetamine | 25 | | Amobarbital | 100 | | p-Methoxyamphetamine | 100 | | d-Amphetamine | 5 | | Oxazepam | 100 | | l-Amphetamine | 20 | | Phencyclidine | 25 | | Benzoylecgonine | 100 | | Phenethylamine | 5 | | Benzphetamine | 200 | | Phenmetrazine | 15 | | Bromophemiramine | 25 | | Phenobarbital | 200 | | Bupropion | 250 | | Phentermine | 5 | {4} 5 | Compound Tested | Conc (ug/mL) | Compound Tested | Conc (ug/mL) | | --- | --- | --- | --- | | Buspirone | 250 | d-Phenylpropanolamine | 250 | | Caffeine | 300 | d,l-Phenylpropanolamine | 25 | | Chlorpheniramine | 12.5 | l-Phenylpropanolamine | 12.5 | | Chlorpromazine | 35 | Procainamide | 125 | | Dextromethorphan | 2 | Promethazone | 50 | | Doxepine | 200 | Propoxyphene | 10 | | Ecgonine Methyl Ester | 5 | Propanol | 150 | | d-Ephedrine | 300 | d-Pseudoephedrine | 25 | | d,l-Ephedrine | 40 | l-Pseudoephedrine | 100 | | l-Ephedrine | 15 | Rantidine | 2 | | 3-OH Tryamine | 250 | Secobarbital | 200 | | Isoxsuprine | 250 | Trazodone | 250 | | Mephentermine | 25 | Trifluroperazine | 250 | | Methadone | 100 | Tyramine | 50 | | l-Methamphetamine | 100 | Valproic acid | 300 | f. Assay cut-off: The stated cutoff of this assay, which includes the dilution of the samples with the Salivette collection device, is 30 ng/mL. Characterization of how the device performs analytically around the claimed cutoff concentration appears in the detection limit section, above. 2. Comparison studies: a. Method comparison with predicate device: Ninety-seven (97) clinical oral fluid specimens were collected. The oral fluid specimens were tested with LZI Oral Fluid Opiate Enzyme Immunoassay and compared to a Gas Chromatography/ Mass Spectrophotometer (GC/MS). Results from the study are presented below. The table describes the agreement between the device and the GC/MS. | LCI EIA | Positive | GC/MS | | | --- | --- | --- | --- | | | | Positive | Negative | | LZI EIA | Positive | 41 | 5* | | | Negative | 0 | 51 | *Five samples contain less than 30 ng/mL morphine, but all contain codeine at a concentration that produces a result higher than the cutoff calibrator. $$ \% \text{Agreement} = 94.8\% $$ b. Matrix comparison: Not applicable. This device is intended for use with only saliva. {5} 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: Not applicable. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...