Software as a Medical Device, Therapeutic, PCCP, Pediatric
Indications for Use
SmartGuard technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGMs), and alternate controller enabled (ACE) pumps to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose values. SmartGuard technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older, and of Type 2 diabetes mellitus in persons 18 years of age and older requiring insulin. SmartGuard technology is intended for single patient use and requires a prescription. Predictive Low Glucose technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGMs), and alternate controller enabled (ACE) pumps to automatically suspend delivery of insulin when the sensor glucose value falls below or is predicted to fall below predefined threshold values. Predictive Low Glucose technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older, and of Type 2 diabetes mellitus in persons 18 years of age and older requiring insulin. Predictive Low Glucose technology is intended for single patient use and requires a prescription.
Device Story
Software-only interoperable automated glycemic controller (iAGC) residing as embedded firmware on compatible ACE pumps; receives sensor glucose (SG) values via Bluetooth Low Energy (BLE) from iCGMs; AHCL algorithm (SmartGuard) automatically adjusts basal insulin and delivers correction boluses; PLGM algorithm (Predictive Low Glucose) automatically suspends/resumes insulin delivery based on current/predicted glucose thresholds; operated by patients with Type 1 or Type 2 diabetes; provides automated insulin management to improve glycemic control; alerts displayed on host ACE pump; supports firmware-over-the-air updates; requires mandatory user training.
Clinical Evidence
No new clinical studies conducted. Evidence leveraged from previously approved PMA submissions (P160017, P160017/S031, P160017/S118, P160017/S124, P160017/S125) which established safety and effectiveness for the AHCL and PLGM algorithms in Type 1 and Type 2 diabetes populations.
Technological Characteristics
Software-only iAGC algorithms embedded in ACE pump firmware. Connectivity: BLE for CGM data. Updates: Firmware over the Air. Risk management: ISO 14971:2019. Software lifecycle: IEC 62304. Human factors: IEC 62366-1. No physical materials, sterilization, or electrical components as it is a software-only device.
Indications for Use
Indicated for management of Type 1 diabetes (age 7+) and Type 2 diabetes (age 18+) requiring insulin. Requires prescription. For single patient use.
Regulatory Classification
Identification
An interoperable automated glycemic controller is a device intended to automatically calculate drug doses based on inputs such as glucose and other relevant physiological parameters, and to command the delivery of such drug doses from a connected infusion pump. Interoperable automated glycemic controllers are designed to reliably and securely communicate with digitally connected devices to allow drug delivery commands to be sent, received, executed, and confirmed. Interoperable automated glycemic controllers are intended to be used in conjunction with digitally connected devices for the purpose of maintaining glycemic control.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) An appropriate, as determined by FDA, clinical implementation strategy, including data demonstrating appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(A) The clinical data must be representative of the performance of the device in the intended use population and in clinically relevant use scenarios and sufficient to demonstrate appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(B) For devices indicated for use with multiple therapeutic agents for the same therapeutic effect (
*e.g.,* more than one type of insulin), data demonstrating performance with each product or, alternatively, an appropriate, as determined by FDA, clinical justification for why such data are not needed.(C) When determined to be necessary by FDA, the strategy must include postmarket data collection to confirm safe real-world use and monitor for rare adverse events.
(ii) Results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(iii) A detailed and appropriate, as determined by FDA, strategy to ensure secure and reliable means of data transmission with other intended connected devices.
(iv) Specifications that are appropriate, as determined by FDA, for connected devices that shall be eligible to provide input to (
*e.g.,* specification of glucose sensor performance) or accept commands from (*e.g.,* specifications for drug infusion pump performance) the controller, and a detailed strategy for ensuring that connected devices meet these specifications.(v) Specifications for devices responsible for hosting the controller, and a detailed and appropriate, as determined by FDA, strategy for ensuring that the specifications are met by the hosting devices.
(vi) Documentation demonstrating that appropriate, as determined by FDA, measures are in place (
*e.g.,* validated device design features) to ensure that safe therapy is maintained when communication with digitally connected devices is interrupted, lost, or re-established after an interruption. Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled and logged appropriately during and after the interruption to maintain patient safety.(vii) A detailed plan and procedure for assigning postmarket responsibilities including adverse event reporting, complaint handling, and investigations with the manufacturers of devices that are digitally connected to the controller.
(2) Design verification and validation documentation must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures to address device system hazards;
(ii) Design decisions related to how the risk control measures impact essential performance; and
(iii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include appropriate, as determined by FDA, and validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to connected devices;
(ii) Secure, accurate, and reliable means of data transmission between the controller and connected devices;
(iii) Sharing of necessary state information between the controller and any connected devices (
*e.g.,* battery level, reservoir level, sensor use life, pump status, error conditions);(iv) Ensuring that the controller continues to operate safely when data is received in a manner outside the bounds of the parameters specified;
(v) A detailed process and procedures for sharing the controller's interface specification with connected devices and for validating the correct implementation of that protocol; and
(vi) A mechanism for updating the controller software, including any software that is required for operation of the controller in a manner that ensures its safety and performance.
(4) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices, and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the controller must, at a minimum, include:
(i) Commands issued by the controller, and associated confirmations the controller receives from digitally connected devices;
(ii) Malfunctions of the controller and malfunctions reported to the controller by digitally connected devices (
*e.g.,* infusion pump occlusion, glucose sensor shut down);(iii) Alarms and alerts and associated acknowledgements from the controller as well as those reported to the controller by digitally connected devices; and
(iv) Connectivity events (
*e.g.,* establishment or loss of communications).(5) The device must only receive glucose input from devices cleared under § 862.1355 (integrated continuous glucose monitoring system), unless FDA determines an alternate type of glucose input device is designed appropriately to allow the controller to meet the special controls contained within this section.
(6) The device must only command drug delivery from devices cleared under § 880.5730 of this chapter (alternate controller enabled infusion pump), unless FDA determines an alternate type of drug infusion pump device is designed appropriately to allow the controller to meet the special controls contained within this section.
(7) An appropriate, as determined by FDA, training plan must be established for users and healthcare providers to assure the safety and performance of the device when used. This may include, but not be limited to, training on device contraindications, situations in which the device should not be used, notable differences in device functionality or features compared to similar alternative therapies, and information to help prescribers identify suitable candidate patients, as applicable.
(8) The labeling required under § 809.10(b) of this chapter must include:
(i) A contraindication for use in pediatric populations except to the extent clinical performance data or other available information demonstrates that it can be safely used in pediatric populations in whole or in part.
(ii) A prominent statement identifying any populations for which use of this device has been determined to be unsafe.
(iii) A prominent statement identifying by name the therapeutic agents that are compatible with the controller, including their identity and concentration, as appropriate.
(iv) The identity of those digitally connected devices with which the controller can be used, including descriptions of the specific system configurations that can be used, per the detailed strategy submitted under paragraph (b)(1)(iii) of this section.
(v) A comprehensive description of representative clinical performance in the hands of the intended user, including information specific to use in the pediatric use population, as appropriate.
(vi) A comprehensive description of safety of the device, including, for example, the incidence of severe hypoglycemia, diabetic ketoacidosis, and other relevant adverse events observed in a study conducted to satisfy paragraph (b)(1)(i) of this section.
(vii) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(viii) For any controller with hardware components intended for multiple patient reuse, instructions for safely reprocessing the hardware components between uses.
K251217 — SmartGuard technology; Predictive Low Glucose technology · Medtronic Minimed, Inc. · Aug 29, 2025
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K253585
B Applicant
Medtronic MiniMed Inc.
C Proprietary and Established Names
SmartGuard technology
Predictive Low Glucose technology
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QJI | Class II | 21 CFR 862.1356 - Interoperable Automated Glycemic Controller | CH - Clinical Chemistry |
| QJS | Class II | 21 CFR 862.1356 - Interoperable Automated Glycemic Controller | CH - Clinical Chemistry |
## E Purpose for Submission:
Modifications to a cleared device, including:
- Expansion of the intended use population to include people with Type 2 diabetes
- Expansion of compatible U-100 insulins to include Lyumjev® and Fiasp®
- Modification of a Predetermined Change Control Plan (PCCP) for qualifying and integrating with compatible FDA-cleared integrated continuous glucose monitors (iCGMs) and compatible FDA-cleared alternate controller enabled (ACE) pumps to cover the addition of a new patient population (Type 2 diabetes), additional compatible U-100 insulins, along with a new ACE Pump and new iCGM.
## II Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Indication(s) for Use:
SmartGuard Technology
SmartGuard technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGMs), and alternate controller enabled (ACE) pumps to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose values.
SmartGuard technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older, and by people with Type 2 diabetes, ages 18 years or older requiring insulin.
SmartGuard technology is intended for single patient use and requires a prescription.
Predictive Low Glucose Technology
Predictive Low Glucose technology is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible Medtronic continuous glucose monitors (CGMs), and alternate controller enabled (ACE) pumps to automatically suspend delivery of insulin when the sensor glucose value falls below or is predicted to fall below predefined threshold values.
Predictive Low Glucose technology is intended for the management of Type 1 diabetes mellitus in persons 7 years of age and older, and of Type 2 diabetes mellitus in persons 18 years of age and older requiring insulin.
Predictive Low Glucose technology is intended for single patient use and requires a prescription.
C Special Conditions for Use Statement(s):
SmartGuard Technology
Rx - For Prescription Use Only
- SmartGuard technology is contraindicated for use in persons under age 7.
- SmartGuard technology is not recommended for people with a significant cognitive or physical impairment that affects their ability to safely operate the pump, including a lack of physical dexterity.
- SmartGuard technology is not recommended for children who are not under the care of a parent or caregiver who is capable of safely operating the pump for the patient.
- SmartGuard technology is not recommended for persons who are unwilling or unable to perform BG meter readings.
- SmartGuard technology is not recommended for people who are unwilling or unable to maintain contact with their healthcare professional.
Predictive Low Glucose Technology
Rx - For Prescription Use Only
- Predictive Low Glucose technology is contraindicated for use in persons under age 7.
K253585 - Page 2 of 10
{2}
- Predictive Low Glucose technology is not recommended for people with a significant cognitive or physical impairment that affects their ability to safely operate the pump, including a lack of physical dexterity.
- Predictive Low Glucose technology is not recommended for children who are not under the care of a parent or caregiver who is capable of safely operating the pump for the patient.
- Predictive Low Glucose technology is not recommended for persons who are unwilling or unable to perform BG meter readings.
- Predictive Low Glucose technology is not recommended for people who are unwilling or unable to maintain contact with their healthcare professional.
## III Device Description
### SmartGuard Technology
SmartGuard Technology, also referred to as the Advanced Hybrid Closed Loop (AHCL) algorithm, is a software-only device intended for use by people with Type 1 diabetes for ages 7 years or older, and by people with Type 2 diabetes, ages 18 years or older. It is an interoperable automated glycemic controller (iAGC) that is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible interoperable Medtronic continuous glucose monitors (CGM) and compatible alternate controller enabled (ACE) pumps to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose (SG) values.
The AHCL algorithm resides on the compatible ACE pump, which serves as the host device. It is meant to be integrated in a compatible ACE pump and is an embedded part of the ACE pump firmware.
Inputs to AHCL algorithm (e.g. SG values, user inputs) are received from the ACE pump (host device), and outputs from AHCL algorithm (e.g. insulin delivery commands) are sent by the algorithm to the ACE pump via software interface. As an embedded part of the ACE pump firmware, the AHCL algorithm does not connect to or receive data from compatible CGMs; instead, sensor glucose (SG) values or other inputs received by ACE pump from compatible CGMs via Bluetooth Low Energy (BLE) technology are transmitted to the embedded AHCL algorithm.
The AHCL algorithm works in conjunction with the ACE pump and is responsible for controlling insulin delivery when the ACE pump is in Auto Mode. It includes adaptive control algorithms that autonomously and continually adapt to the ever-changing insulin requirements of each individual.
### Predictive Low Glucose Technology
Predictive Low Glucose Technology, also referred to as the Predictive Low Glucose Management (PLGM) algorithm is a software-only device intended for use by people with Type 1 diabetes for ages 7 years or older, and by people with Type 2 diabetes, ages 18 years or older. It is an interoperable automated glycemic controller (iAGC) that is intended for use with compatible integrated continuous glucose monitors (iCGM), compatible interoperable Medtronic
K253585 - Page 3 of 10
{3}
continuous glucose monitors (CGM) and compatible alternate controller enabled (ACE) pumps to automatically suspend delivery of insulin when the sensor glucose value falls below or is predicted to fall below predefined threshold values.
The PLGM algorithm resides on the compatible ACE Pump, which serves as the host device. It is meant to be integrated in a compatible ACE pump and is an embedded part of the ACE pump firmware.
Inputs to PLGM algorithm (e.g., sensor glucose values, user inputs) are received from the ACE pump (host device), and outputs from PLGM algorithm (e.g., suspend/resume commands) are sent by the algorithm to the ACE pump via software interface. As an embedded part of the ACE pump firmware, the PLGM algorithm does not connect to or receive data from compatible CGMs; instead, sensor glucose (SG) values or other inputs are received by the ACE pump from compatible CGMs via Bluetooth Low Energy (BLE) technology are transmitted to the embedded PLGM algorithm.
The PLGM algorithm works in conjunction with the ACE pump. When enabled, the PLGM algorithm is able to suspend insulin delivery for a minimum of 30 minutes and for a maximum of 2 hours based on current or predicted sensor glucose values. It will automatically resume insulin delivery when maximum suspend time of 2 hours has elapsed or when underlying conditions resolve. The user is also able to manually resume insulin at any time.
## IV Substantial Equivalence Information:
A Predicate Device Name(s):
SmartGuard technology
Predictive Low Glucose technology
B Predicate 510(k) Number(s):
K251217
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K253585 | K251217 |
| --- | --- | --- |
| Device Trade Name | SmartGuard Technology | SAME |
| General Device Characteristic Similarities | | |
| Intended Use/Indications for Use | SmartGuard technology is intended for use with compatible continuous glucose monitors (CGM) | SAME |
K253585 - Page 4 of 10
{4}
| | and alternate controller enabled (ACE) pumps to automatically adjust the delivery of basal insulin and to automatically deliver correction boluses based on sensor glucose values.
SmartGuard technology is intended for single patient use and requires a prescription. | |
| --- | --- | --- |
| Principles of Operation | Algorithmic software device intended to automatically increase, decrease, suspend and resume delivery of insulin based on sensor glucose values | SAME |
| Device Hosting Controller | ACE Pump | SAME |
| Communication with ACE Pump | Communicates with an ACE Pump via software interface | SAME |
| Number of Users | Single User | SAME |
| Alarms/Alerts | ACE pump will display algorithm-related alerts to the user | SAME |
| Auto Mode Exit | Algorithm automatically de-activates and exits auto mode when conditions arise | SAME |
| Manual Algorithm Deactivation | User can manually turn off closed loop therapy | SAME |
| Meal/Food Bolus | Users must manually deliver meal boluses they can calculate using the integrated bolus calculator. An SG or BG value is used for meal boluses. | SAME |
| Mechanism of Software Update | Firmware over the Air | SAME |
| Training | There is mandatory user | SAME |
K253585 - Page 5 of 10
{5}
K253585 - Page 6 of 10
| | training before the user can use AHCL Algorithm | |
| --- | --- | --- |
| **General Device Characteristic Differences** | | |
| Intended Use Population | Type 1 diabetes mellitus in persons 7 years of age and greater. Type 2 diabetes mellitus in persons 18 years of age and greater. | Type 1 diabetes mellitus in persons 7 years of age and greater. |
| Specific Drug/Biological Use | U-100 insulin: Novolog® Lyumjev® Humalog® Fiasp® Admelog® | U-100 insulin: Novolog® Humalog® Admelog® |
| Device & Predicate Device(s): | K253585 | K251217 |
| --- | --- | --- |
| Device Trade Name | Predictive Low Glucose Technology | SAME |
| General Device Characteristic Similarities | | |
| Intended Use/Indications for Use | Predictive Low Glucose Technology is intended for use with compatible continuous glucose monitors (CGM) and alternate controller enabled (ACE) pumps to automatically suspend delivery of insulin when the sensor glucose value is predicted to fall below predefined threshold values.
Predictive Low Glucose Technology is intended for single patient use and requires a prescription. | SAME |
| Principles of Operation | Algorithmic software device that utilizes CGM sensor readings to | SAME |
{6}
K253585 - Page 7 of 10
| | suspend and resume insulin based on the current and predicted sensor values. | |
| --- | --- | --- |
| Device Hosting Controller | ACE Pump | SAME |
| Communication with ACE Pump | Communicates with an ACE Pump via software interface | SAME |
| Number of Users | Single User | SAME |
| Alarms/Alerts | ACE pump will display algorithm-related alerts to the user | SAME |
| Manual Algorithm Deactivation | User can manually turn off the therapy/algorithm | SAME |
| Mechanism of Software Update | Firmware over the Air | SAME |
| Training | There is mandatory user training before the user can use PLGM Algorithm | SAME |
| General Device Characteristic Differences | | |
| Intended Use Population | Type 1 diabetes mellitus in persons 7 years of age and greater. Type 2 diabetes mellitus in persons 18 years of age and greater. | Type 1 diabetes mellitus in persons 7 years of age and greater. |
| Specific Drug/Biological Use | U-100 insulin: Novolog® Lyumjev® Humalog® Fiasp® Admelog® | U-100 insulin: Novolog® Humalog® Admelog® |
V Standards/Guidance Documents Referenced:
- Special controls established under 21 CFR 862.1356
- ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices
- IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION 13-79 Medical device software - Software life cycle processes
{7}
- IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION 15-129 Medical devices
- Part 1: Application of usability engineering to medical devices
## VI Performance Characteristics (if/when applicable):
### A. Analytical Performance
Not applicable.
### B. Other Supportive Instrument Performance Characteristics Data
#### Clinical Performance Data
The subject SmartGuard Technology algorithm was previously approved for use, including by people with type 2 diabetes and for use with Fiasp and Lyumjev, as part of the MiniMed 780G System under P160017/S118, P160017/S124, and P160017/S125. The subject Predictive Low Glucose Technology algorithm was previously approved as part of the MiniMed 670G System under P160017, including P160017/S031, as well as the MiniMed 780G System under P160017/S118. SmartGuard Technology and Predictive Low Glucose Technology are the same algorithms as reviewed in these respective PMAs, with the exception of the updated intended uses to include interoperability and the PCCP described below. Therefore, prior pivotal clinical data were leveraged from P160017, P160017/S031, P160017/S118, P160017/S124, and P160017/S125. Refer to the Summary of Safety and Effectiveness Data (SSED) for those submissions for a summary of those clinical studies. No new clinical studies were conducted to support substantial equivalence of the subject devices.
#### Non-Clinical Performance Data
There have been no changes to the SmartGuard technology's design, architecture, features, logical flows, safeguards or principles of operation from the AHCL algorithm approved under P160017/S118, P160017/S124, and P160017/S125 (version 5.3A) or PLGM algorithm approved under in P160017 (version 1.0.0). Both algorithms were subsequently cleared in K251217 as Medtronic iAGCs. Therefore, non-clinical studies were leveraged from previous submissions.
#### Interoperability
A plan and approach for interoperability were provided and determined to be adequate to support and clearly specify expectations, requirements, and interface specifications for potential interoperable devices. Provided interoperability documentation included a description of the respective device interface and its specifications, interoperability design and architecture, and the strategy for the subject iAGCs to be interoperable with compatible connected devices, including the PCCP described below. The SmartGuard technology and Predictive Low Glucose technology algorithms are designed to be interoperable with the qualified Medtronic Simplera Sync and Guardian 4 Sensor, qualified Medtronic ACE pumps, and qualified iCGMs. In addition, their plan provided Medtronic's approach to working with third-party connected iCGM device companies regarding contractual issues, quality agreement, data communication and exchange, and post-market reporting procedures and responsibilities.
K253585 - Page 8 of 10
{8}
# Hazard Analysis
Hazard analyses were provided for the SmartGuard technology and Predictive Low Glucose technology algorithms, respectively, in which risks and risk mitigations for software and interoperable hardware components associated with the safe and effective functioning of the devices were provided, as well as verification and validation activities for implemented risk control measures. In particular, these hazard analyses accounted for the risks associated with interoperability between the software devices developed by the sponsor and other third-party interoperable devices (i.e., iCGMs).
# PCCP
This submission included an updated version of the PCCP originally cleared in K251217 that Medtronic will follow (1) for qualifying and integrating new compatible iCGMs and ACE pumps and (2) for continued maintenance of previously marketed and qualified connected iCGMs and ACE pumps. The PCCP includes a description of the modifications, a modification protocol, and an impact assessment. The PCCP specifies the testing methods, validation activities, and performance requirements to implement the specified modifications such that SmartGuard technology and Predictive Low Glucose technology remain as safe and as effective as their predicate device and continue to comply with 21 CFR 862.1356(b). The sponsor's PCCP comprises the following for each subject iAGC:
- A new candidate iCGM must have demonstrated compatibility to established minimum iCGM performance, compatibility, and sensor specifications FDA found appropriate during this review before proceeding with qualification.
- Medtronic will conduct virtual clinical trials with virtual patients, with pre-defined acceptance criteria based on virtual patient equivalency testing and glycemic outcomes from the iAGC clinical studies. Software design integration of the new iCGM with the subject iAGC will then be completed, and integration V&V testing will be conducted according to the established PCCP. If integration V&V testing meets the PCCP's prespecified acceptance criteria, the iCGM may be added as an interoperable component to the subject iAGC. As part of this 510(k), the scope of these virtual clinical studies (previously described in the PCCP for K251217) was expanded to address device use in type 2 diabetes, as well as the use of Fiasp and Lyumjev insulin products. Specific test cases and pre-specified acceptance criteria were added to address these new uses.
- A new candidate ACE pump must have demonstrated compatibility to established minimum ACE pump compatibility and performance specifications defined in the PCCP before proceeding to software design integration with the subject iAGC. Integration V&V testing will then be conducted according to the established PCCP. If integration V&V testing meets the PCCP's prespecified acceptance criteria, the ACE pump may be added as an interoperable component to the subject iAGC.
- Modifications to previously marketed and qualified iCGM and ACE pump devices will be evaluated and, as needed, re-qualified according to the PCCP. For iCGMs, re-qualification requires demonstrated continued compatibility to the minimum iCGM specifications, virtual clinical trial testing as needed, and integration V&V testing. Modified ACE pumps are re-qualified through demonstrated continued compatibility to the minimum ACE pump specifications defined in the PCCP and successful integration V&V testing.
- Related iAGC labeling will be revised as needed to identify new qualified and integrated compatible devices or to remove previously compatible devices found to
K253585 - Page 9 of 10
{9}
no longer be compatible. Communication will be provided through various channels to notify users about a new compatible device or previously compatible but now incompatible device.
## VII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device and met the special controls for 21 CFR 862.1356.
## VIII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K253585 - Page 10 of 10
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.