QDH · Menopause Test System

Clinical Chemistry · 21 CFR 862.1093 · Class 2

Overview

Product CodeQDH
Device NameMenopause Test System
Regulation21 CFR 862.1093
Device ClassClass 2
Review PanelClinical Chemistry
3rd-Party ReviewableYes

Identification

A menopause test system is an in vitro diagnostic device intended to measure hormones or other analytes in human clinical specimens as an aid in the determination of menopausal status in women.

Classification Rationale

Class II (special controls). A menopause test system must comply with the following special controls:

Special Controls

A menopause test system must comply with the following special controls: (1) Design verification and validation must include the following: (i) An appropriate traceability plan to minimize the risk of drift in the menopause test system results over time. (ii) Detailed documentation of a clinical study to demonstrate clinical performance or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information: (A) Results must demonstrate appropriate clinical performance relative to a well-accepted and appropriate comparator. (B) Data must demonstrate accuracy of device output for each indicated specimen type throughout the device measuring range as appropriate for the intended use population. (2) The 21 CFR 809.10 labeling must include the following: (i) A statement in the intended use that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions. (ii) A limiting statement that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions. (iii) A limiting statement appropriately describing the risks of false test results and that test results should not be relied upon in clinical decision making (e.g., to discontinue contraceptive use and/or to evaluate patients for the presence of endometrial cancer) without other clinical and laboratory findings.

*Classification.* Class II (special controls). A menopause test system must comply with the following special controls:(1) Design verification and validation must include the following: (i) An appropriate traceability plan to minimize the risk of drift in the menopause test system results over time. (ii) Detailed documentation of a clinical study to demonstrate clinical performance or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information: (A) Results must demonstrate appropriate clinical performance relative to a well-accepted and appropriate comparator. (B) Data must demonstrate accuracy of device output for each indicated specimen type throughout the device measuring range as appropriate for the intended use population. (2) The labeling required under § 809.10 of this chapter must include the following: (i) A statement in the indications for use that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions. (ii) A limiting statement that the device is intended to be used for the determination of menopausal status only in conjunction with other clinical and laboratory findings prior to any diagnostic or treatment decisions. (iii) A limiting statement appropriately describing the risks of false test results and that test results should not be relied upon in clinical decision making ( *e.g.,* to discontinue contraceptive use and/or to evaluate patients for the presence of endometrial cancer) without other clinical and laboratory findings.

Recent Cleared Devices (1 of 1)

RecordDevice NameApplicantDecision DateDecision
DEN180004picoAMH ELISAAnsh Labs, LLCOct 24, 2018DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...