ADVIA Chemistry Enzymatic Hemoglobin Ale (Alc_E) Assay
Device Facts
| Record ID | K200256 |
|---|---|
| Device Name | ADVIA Chemistry Enzymatic Hemoglobin Ale (Alc_E) Assay |
| Applicant | Siemens Healthcare Diagnostics, Inc. |
| Product Code | PDJ · Clinical Chemistry |
| Decision Date | Jul 2, 2021 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.1373 |
| Device Class | Class 2 |
Indications for Use
The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay is an in vitro diagnostic assay for the quantitative determination of mmol/mol HbA1c (IFCC) and % HbA1c (DCCT/NGSP) in human anticoagulated venous whole blood and hemolysate for use on the ADVIA® Chemistry Systems. Measurement of Hemoglobin A 1 c is used as an aid in the diagnosis and monitoring of long-term blood glucose control in patients with diabetes mellitus, and as an aid in the identification of patients at risk for developing diabetes mellitus.
Device Story
ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay is an in vitro diagnostic test for quantitative determination of HbA1c in human anticoagulated venous whole blood and hemolysate. Device operates on ADVIA® Chemistry Systems; utilizes enzymatic methodology to measure HbA1c levels. Assay consists of three liquid, ready-to-use reagents (R1, R2, and Pretreatment Solution). Automated application lyses whole blood specimens on-system; manual application uses Pretreatment Solution to create hemolysate. Healthcare professionals use results to aid in diagnosis and monitoring of long-term blood glucose control in patients with diabetes mellitus and identification of at-risk patients. Modification extends low end of analytical measuring range for total hemoglobin (tHb).
Clinical Evidence
Bench testing only. Verification and validation activities performed based on risk analysis to confirm performance following the modification of the total hemoglobin measurement range.
Technological Characteristics
Quantitative enzymatic assay. Reagents: R1, R2, and Pretreatment Solution. Measuring range: 3.8-14.0% HbA1c (DCCT/NGSP); 18.03-129.50 mmol/mol HbA1c (IFCC). Total hemoglobin analytical measuring range: 74.74-320.15 μmol/L. Instrument platform: ADVIA® 1800 Clinical Chemistry System. Standardization: Traceable to IFCC reference calibrators; NGSP certified.
Indications for Use
Indicated for patients with diabetes mellitus or those at risk for developing diabetes mellitus. Used for diagnosis and monitoring of long-term blood glucose control.
Regulatory Classification
Identification
A hemoglobin A1c test system is a device used to measure the percentage concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c is used as an aid in the diagnosis of diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.
Special Controls
(b) Classification. Class II (special controls). Hemoglobin A1c test systems must comply with the following special controls: 1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. 2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity and interference, including the following: i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0%, 6.5%, 8.0% and 12% hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least 3 lots of the device and 3 instruments, as applicable. ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the new device and compare results of the new device to results of the standardized test method. Results must demonstrate little or no bias versus the standardized method. iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6%. iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2 and Hemoglobin S. 3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must have initial and annual standardization verification by a certifying glycohemoglobin standardization organization deemed acceptable by FDA. (2) The premarket notification submission must include performance testing to evaluate precision, accuracy, linearity, and interference, including the following: (i) Performance testing of device precision must, at a minimum, use blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0 percent, and 12 percent hemoglobin A1c. This testing must evaluate precision over a minimum of 20 days using at least three lots of the device and three instruments, as applicable. (ii) Performance testing of device accuracy must include a minimum of 120 blood samples that span the measuring interval of the device and compare results of the new device to results of a standardized test method. Results must demonstrate little or no bias versus the standardized method. (iii) Total error of the new device must be evaluated using single measurements by the new device compared to results of the standardized test method, and this evaluation must demonstrate a total error less than or equal to 6 percent. (iv) Performance testing must demonstrate that there is little to no interference from common hemoglobin variants, including Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S. (3) When assay interference from Hemoglobin F or interference with other hemoglobin variants with low frequency in the population is observed, a warning statement must be placed in a black box and must appear in all labeling material for these devices describing the interference and any affected populations.
Predicate Devices
- ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c_E) Assay (K171771)
Related Devices
- K171771 — ADVIA Chemistry Enzymatic Hemoglobin A1c (A1c_E) Assay · Siemens Healthcare Diagnostics, Inc. · Dec 4, 2017
- K173206 — SEKURE HbA1c Assay · Sekisui Diagnostics P.E.I., Inc. · Jul 12, 2018
- K130255 — HEMOGLOBIN A1C, HEMOGLOBIN A1C CALIBRATORS, HEMOGLOBIN A1C CONTROLS · Abbott Laboratories · Feb 28, 2014
- K081895 — ADVIA CHEMISTRY HEMOGLOBIN A1C, ADVIA CHEMISTRY A1C CALIBRATORS · Siemens Medical Solutions Diagnostics · Jan 8, 2009