K203833 · Shanghai Genext Medical Technology Co., Ltd. · MLM · Jan 27, 2023 · Clinical Chemistry
Device Facts
Record ID
K203833
Device Name
Tacrolimus Assay Kit
Applicant
Shanghai Genext Medical Technology Co., Ltd.
Product Code
MLM · Clinical Chemistry
Decision Date
Jan 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1678
Device Class
Class 2
Indications for Use
The Tacrolimus Assay Kit is used for quantitative determination of the tacrolimus concentration in human whole blood on the Beckman Coulter AU480. It is to be used as an aid in the management of liver and kidney allograft patients receiving tacrolimus therapy. For In Vitro Diagnostic Use.
Device Story
In vitro diagnostic assay for quantitative measurement of tacrolimus in human whole blood; used to manage liver and kidney allograft patients. Input: whole blood sample requiring manual pretreatment with extractant (methanol/inorganic salt denaturant). Principle: latex-enhanced competitive immunoturbidimetry; FK506 in sample competes with FK506-protein complex for anti-FK506 antibody on latex particles; turbidity measured at 546 nm. Output: tacrolimus concentration (1.5–30 ng/mL). Used in clinical laboratory settings on Beckman Coulter AU480 analyzer. Results interpreted by clinicians to adjust tacrolimus therapy, ensuring therapeutic drug monitoring for transplant patients.
Clinical Evidence
Bench testing only. Analytical studies performed to evaluate precision, linearity, assay reportable range, recovery, specificity, and accuracy.
Indicated for quantitative determination of tacrolimus in human whole blood to aid in the management of liver and kidney allograft patients receiving tacrolimus therapy. Prescription use only.
Regulatory Classification
Identification
A tacrolimus test system is a device intended to quantitatively determine tacrolimus concentrations as an aid in the management of transplant patients receiving therapy with this drug. This generic type of device includes immunoassays and chromatographic assays for tacrolimus.
Special Controls
*Classification.* Class II (special controls). The special control is “Class II Special Controls Guidance Document: Cyclosporine and Tacrolimus Assays; Guidance for Industry and FDA.” See § 862.1(d) for the availability of this guidance document.
K060385 — EMIT 2000 TACROLIMUS ASSAY AND SAMPLE PRETREATMENT REAGENT · Dade Behring, Inc. · Apr 6, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 27, 2023
Shanghai Genext Medical Technology Co., Ltd. % Giselle Zhang Regulatory Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746
Re: K203833
Trade/Device Name: Tacrolimus Assay Kit Regulation Number: 21 CFR 862.1678 Regulation Name: Tacrolimus Test System Regulatory Class: Class II Product Code: MLM Dated: August 19, 2022 Received: August 19, 2022
Dear Giselle Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Paula Caposino -S | Digitally signed by Paula Caposino -S |
|-------------------|---------------------------------------|
| | Date: 2023.01.27 |
| | 15:31:02 -05'00' |
Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) k203833
Device Name Tacrolimus Assay Kit
Indications for Use (Describe)
The Tacrolimus Assay Kit is used for quantitative determination of the tacrolimus concentration in human whole blood on the Beckman Coulter AU480. It is to be used as an aid in the management of liver and kidney allograft patients receiving tacrolimus therapy.
For In Vitro Diagnostic Use.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Submission Number: k203833
The following information is provided in accordance with 21 CFR 807.92:
#### Submitter Information
| Company: | Shanghai Genext Medical Technology<br>Co., Ltd. | |
|--------------------------|----------------------------------------------------------------------------------------------------------------------|--|
| Address: | Rml101, 11th Floor, Bldg 2,<br>Pujiang Hi-Tech Plaza, No.2388<br>Chenhang Road, Minhang District,<br>Shanghai, China | |
| Contact Name: | Kimmy Zong | |
| Contact Title: | RA Associate Director | |
| Tel: | + 86-13918832180 | |
| Fax: | N/A | |
| Email: | wenwen.zong@genext.com.cn | |
| Regulatory Correspondent | Giselle Zhang | |
| Email: | LST.AUS.ProjectManagement@ul.com | |
| Date Summary Prepared: | January 24, 2023 | |
#### Name of the Device
| Device Trade Name: | Tacrolimus Assay Kit |
|----------------------|-------------------------|
| Common Name: | Tacrolimus Assay Kit |
| Classification Name: | Toxicology |
| Review Panel | Clinical Chemistry (TX) |
| Regulation Number: | 862.1678 |
| Classification: | II |
| Product Code: | MLM |
#### Intended Use:
Indications for Use:
The Tacrolimus Assay Kit is used for quantitative determination of the tacrolimus concentration in human whole blood on the Beckman Coulter AU480. It is to be used as an aid in the management of liver and kidney allograft patients receiving tacrolimus therapy.
For In Vitro Diagnostic Use.
Reagent:
- Reagent 1 (R1): FK506-protein complex, stored in MES buffer; Preservative: sodium azide. .
- Reagent 2(R2): Latex particles sensitized by anti-FK506 monoclonal antibody, 4-(2-. hydroxyerhyl) piperazine-1-erhaesulfonic acid; Preservative: sodium azide.
## Pretreatment Reagent:
- Sample extractant: Containing methanol and inorganic salt protein denaturant. ●
Calibrator:
- Calibrator: Bovine whole blood solution containing different concentrations of FK506 (the . calibration target list please see the following table); preservative.: sodium azide.
{4}------------------------------------------------
- The calibrators yield the following concentrations: ●
| Calibrator name | Target value (ng/mL) |
|-----------------|----------------------|
| Calibrator A | 0.0 |
| Calibrator B | 3.0 |
| Calibrator C | 6.0 |
| Calibrator D | 10.0 |
| Calibrator E | 20.0 |
| Calibrator F | 30.0 |
Special Conditions for use statement(s):
- For prescription use only. ●
Special Instrument Requirements:
- Beckman Coulter AU480. ●
# Device Description
The Tacrolimus Assay Kit utilizes the latex-enhanced competitive immunoturbidimetry method for quantitative determination of the tacrolimus concentration in human whole blood. The assay consists of reagents, sample extract and calibrator.
# Test Principle
The kit is based on the principle of latex-enhanced competitive immunoturbidimetry method; The sample should be pretreated, and the protein-bound tacrolimus in the whole blood should be extracted for detection before the tacrolimus concentration is determined in the detection process. FK506 in the sample competes with FK506 in the FK506- protein complex in the reagent to bind the anti-FK506 antibody on the surface of the sensitized latex particles. The higher the concentration of FK506 in the sample, the greater the degree to which the latex surface antibody is bound by it, and the lower the latex aggregation extent induced by FK506- protein complex. The turbidity of the reaction system is negatively correlated with the concentration of FK506 in the sample. Therefore, a calibration curve can be established and the concentration of the whole blood sample FK506 can be calculated by measuring the absorbance at 546 nm.
## Substantial Equivalence Information
Predicate Device Identification: ARCHITECT TACROLIMUS: MODEL IL 77 (K070820)
Comparison with Predicate:
| Device & Predicate<br>Device(s): | Candidate Device<br>K203833 | Predicate Device<br>K070820 |
|-----------------------------------------------|-----------------------------|---------------------------------------------------------------------------------|
| Device Trade Name | Tacrolimus Assay Kit | ARCHITECT<br>Tacrolimus Assay |
| General Device<br>Characteristic Similarities | | |
| Intended Use/Indications<br>For Use | Same | For the quantitative<br>determination of<br>tacrolimus in human<br>whole blood. |
| Sample pre-treatment | Same | Manual |
{5}------------------------------------------------
| General Device<br>Characteristic Differences | Candidate Device<br>K203833 | Candidate Device<br>K070820 |
|----------------------------------------------|---------------------------------------------------------------------------|--------------------------------------------------|
| Measuring Range | 1.5 - 30 ng/mL | 2 - 30 ng/mL |
| Methodology | Particle-enhanced<br>turbidimetric inhibition<br>immunoassay<br>(PETINIA) | Chemiluminescent<br>microparticle<br>immunoassay |
| Analyzer platform | Beckman Coulter<br>AU480 analyzer | ARCHITECT i System. |
# Summary of Testing:
Analytical studies were performed to evaluate the precision, linearity, assay reportable range, recovery, specificity, and accuracy of the candidate device.
## Conclusion:
Tacrolimus Assay Kit has the same indications for use as the predicate device. Any minor differences in the technological characteristics of the subject device when compared to the predicate device have been evaluated through performance testing which demonstrates that the Tacrolimus Assay Kit is substantially equivalent.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.