K971203 · Beckman Instruments, Inc. · KLI · May 16, 1997 · Clinical Chemistry
Device Facts
Record ID
K971203
Device Name
SYNCHRON SYSTEMS THYROXINE REAGENT
Applicant
Beckman Instruments, Inc.
Product Code
KLI · Clinical Chemistry
Decision Date
May 16, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1700
Device Class
Class 2
Indications for Use
The SYNCHRON Systems Thyroxine (T4) Reagent is intended for the quantitative determination of total thyroxine concentrations in serum on SYNCHRON Clinical Systems.
Device Story
The SYNCHRON Systems Thyroxine (T4) Reagent is an in vitro diagnostic reagent kit used on Beckman SYNCHRON Clinical Systems. It quantitatively measures total thyroxine (T4) concentrations in serum samples. The device utilizes a chemical reaction principle identical to the predicate CEDIA T4 MAb Reagent. The system processes serum samples to provide T4 concentration values, which clinicians use to assess thyroid function. The device is intended for use in clinical laboratory settings by trained laboratory personnel. It supports the use of Sodium Heparin and Lithium Heparin as anticoagulants.
Clinical Evidence
Bench testing only. Method comparison study (n=97) against the predicate CEDIA MAb Thyroxine Reagent yielded a slope of 1.018, intercept of 0.12 μg/dL, and correlation coefficient (r) of 0.992. Imprecision studies (n=80 per level) showed within-run %C.V. ranging from 1.7% to 4.3% and total imprecision %C.V. ranging from 2.4% to 8.8% across three concentration levels (4.2–17.1 μg/dL). On-board stability is validated for 30 days.
Technological Characteristics
In vitro diagnostic reagent kit for quantitative T4 determination. Utilizes chemical reaction principle identical to predicate. Compatible with SYNCHRON Clinical Systems. Reagent on-board stability: 30 days. Specimen types: serum, Sodium Heparin, Lithium Heparin.
Indications for Use
Indicated for the quantitative determination of total thyroxine (T4) concentrations in human serum to aid in the assessment of thyroid function.
Regulatory Classification
Identification
A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.