K160495 · Ortho-Clinical Diagnostics, Inc. · KLI · Oct 18, 2016 · Clinical Chemistry
Device Facts
Record ID
K160495
Device Name
VITROS® Automation Solutions
Applicant
Ortho-Clinical Diagnostics, Inc.
Product Code
KLI · Clinical Chemistry
Decision Date
Oct 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1700
Device Class
Class 2
Indications for Use
VITROS® Automation Solutions is intended to automate pre-analytical and post-analytical sample processing in the clinical laboratory. VITROS® Automation Solutions allows the consolidation of software, automation modules and clinical analyzers, such as VITROS® Systems into a unified workstation to perform a variety of assays such as total T4, carbamazepine and gentamicin. Carbamazepine measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose. Gentamicin measurements are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy. Total thyroxine (T4) measurements are used to aid in the differential diagnosis of thyroid disease. For prescription use only. For in vitro diagnostic use.
Device Story
VITROS® Automation Solutions is a configurable laboratory automation system (LAS) for clinical labs; streamlines pre- and post-analytical sample processing. Inputs: patient sample tubes loaded onto an automation track. Components: PC kit, sample conveyors, centrifuge, de-capper, storage modules, and clinical analyzers (e.g., VITROS® Systems). Operation: system performs bar code reading, sample identification, and routing; uses point-in-space pipetting or robotic interface modules (RIM) to transfer samples to analyzers. Analyzers perform assay processing and result generation. Output: processed samples and diagnostic results for assays like carbamazepine, gentamicin, and total T4. Used in clinical laboratories by lab personnel. Benefits: consolidates workflow, automates sample handling, and improves laboratory efficiency by integrating multiple modules and analyzers into a unified workstation.
Clinical Evidence
Bench testing only. Method comparison study performed comparing automated on-track sample processing versus manual off-track loading using VITROS® CRBM (N=70), GENT (N=55), and Total T4 (N=57) assays. Linear regression analysis showed high correlation (R² 0.9796–0.9989) and slopes near 1.0 (1.00–1.04), confirming no clinically significant difference in assay performance.
Technological Characteristics
Modular LAS comprising PC kits, sample conveyors, bypasses, centrifuges, and de-capper modules. Connectivity via LIS integration and TCAutomation software. Sample tracking via barcode and RFID. Operates as an accessory to clinical chemistry/immunodiagnostic analyzers. Software includes Instrument Manager and TCAutomation drivers.
Indications for Use
Indicated for clinical laboratory use to automate pre- and post-analytical sample processing, including sample handling, routing, and data management for patients requiring diagnostic testing such as carbamazepine monitoring, gentamicin monitoring, and total thyroxine (T4) thyroid disease diagnosis.
Regulatory Classification
Identification
A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
enGen™ Laboratory Automation System with VITROS® 5,1 FS Chemistry System (K063144)
Related Devices
K230937 — Alinity i Total ß-hCG Reagent Kit, GLP systems Track · Abbott Laboratories · Jun 5, 2023
K213486 — GLP systems Track · Abbott Laboratories · Mar 10, 2022
K140496 — Access Vitamin B12 Assay, Access Ferritin Assay, Access Folate Assay, Access HYPERsensitive hTSH Assay, Power Express Sample Processing System Generic Connection Module and Access Immunoassay System Reagents · Beckman Coulter, Inc. · Sep 16, 2014
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
k160495
B. Purpose for Submission:
The submission is to obtain clearance for the VITROS Automation Solution, a laboratory automation system used as an accessory to clinical laboratory analyzers such as the VITROS 5, 1 FS Chemistry System, VITROS 5600 Chemistry System, VITROS 4600 Chemistry System and 3600 Immunodiagnostics System.
The manufacturer uses the performance of VITROS® Chemistry Products CRBM Reagent, VITROS® Chemistry Products GENT Reagent and the VITROS® Immunodiagnostic Products Total T4 Reagent to demonstrate the VITROS Automation Solution is equivalent to the manual sample introduction on the analyzer.
C. Manufacturer and Instrument Name:
Ortho-Clinical Diagnostics, VITROS® Automation Solutions
D. Type of Test or Tests Performed:
Quantitative, immunoassay
E. System Descriptions:
1. Device Description:
VITROS® Automation Solutions is a configurable, scalable laboratory automation system (LAS) designed to streamline pre and post analytical processes in the clinical laboratory. VITROS® Automation Solutions is comprised of Personal Computer (PC) Kit(s) (including software and hardware), sample conveyors with turns, parallel and perpendicular bypasses, storage module, single-tube entry, rack entry and exit, centrifuge, de-capper modules and clinical analyzers.
The connection between the automation software of the VITROS® Automation Solutions allows the establishment of a connection with the laboratory information system (LIS) and the clinical analyzers such as VITROS® Systems to enable sample routing based on reagent and calibration status.
In the basic configuration, patient sample tubes are loaded onto the automation track to be centrifuged, de-capped, and sorted for further processing on clinical analyzers such as
{1}
the VITROS® Systems. Additional modules may be added to enable aliquot capability, sample capping, and refrigerated storage.
2. Principles of Operation:
The Carbamazepine assay is based on an enzyme heterogeneous, competitive immunoassay.
The Gentamicin assay is a homogenous enzyme immunoassay.
The Total T4 assay is a competitive immunoassay.
3. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X ☐ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X ☐
4. Specimen Identification:
Barcode identification of patient samples. Sample bar code read by VITROS Automation Solutions linked to Transport carriers identified on the system by Radio-frequency identification (RFID) tags, presented to instrument via Bypass Module.
5. Specimen Sampling and Handling:
The patient’s sample tubes are loaded onto the VITROS Automation Solutions automated track to be centrifuged, de-capped and sorted for processing on clinical analyzers. The sample bar codes are read to direct the sample to a specific analyzer.
6. Calibration:
Provided in VITROS CRBM (carbamazepine) k980283, VITROS GENT (Gentamicin) k042479, and VITROS TT4 (Total T4) k964721
7. Quality Control:
Provided in VITROS CRBM (carbamazepine) k980283, VITROS GENT (Gentamicin) k042479, and VITROS TT4 (Total T4) k964721
{2}
3
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐
F. Regulatory Information:
| Classification name | Product Code | Classification | Regulatory Section | Panel |
| --- | --- | --- | --- | --- |
| Carbamazepine test system | KLT | class II | 21 CFR 862.3645 | Toxicology 91 |
| Gentamicin test system | LDC | class II | 21 CFR 862.3450 | Toxicology 91 |
| Total thyroxine test system | KLI | class II | 21 CFR 862.1700 | Chemistry 75 |
| Discrete photometric chemistry analyzers for clinical use | JJE | class I | 21 CFR 862.2160 | Chemistry 75 |
G. Intended Use:
1. Indication(s) for Use:
VITROS® Automation Solutions is intended to automate pre-analytical and post-analytical sample processing in the clinical laboratory. VITROS® Automation Solutions allows the consolidation of software, automation modules and clinical analyzers, such as VITROS® Systems into a unified workstation to perform a variety of assays such as total T4, carbamazepine and gentamicin.
Carbamazepine measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose.
Gentamicin measurements are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.
Total thyroxine (T4) measurements are used to aid in the differential diagnosis of thyroid disease.
2. Special Conditions for Use Statement(s):
For prescription use only
For in vitro diagnostic use
{3}
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
enGen™ Laboratory Automation System with VITROS 5, 1 FS Chemistry System – k063144
2. Comparison with Predicate Device:
| Similarities / Differences | | |
| --- | --- | --- |
| Item | VITROS® Automation Solution Candidate Device (k160495) | enGen™ Laboratory Automation System Predicate Device (k063144) |
| Intended Use | Automate pre-analytical and post-analytical sample processing in the clinical laboratory | Same |
| Sample transfer method | Supports both “Point-in-space” and the physical transfer of the sample tube/rack from the automation track to the analyzer’s sample load and identification area using automation interface modules. | Same |
| Fundamental Technology | Outboard analyzer connections accessed by track bypass modules and/or robotic interface | Same |
| System Modules | Instrument Manager Hardware/software TCAutomation Software TCAutomation Modules including but not limited to: Rack Entry/Exit Modules, Single Tube Entry Modules Conveyors and turns, Centrifuges(operating as a single cluster), Decappers Aliquoter, Recappers Bypasses for analyzers such as VITROS® Systems and 3rd party analyzers, ES Flex Module | Same |
{4}
| Similarities / Differences | | |
| --- | --- | --- |
| Item | VITROS® Automation Solution Candidate Device (k160495) | enGen™ Laboratory Automation System Predicate Device (k063144) |
| Sample Identification | Sample bar code read by automation linked to Transport carriers identified on the system by Radio-frequency identification (RFID) tags, presented to instrument via Bypass Module | Same |
| Sample Handling | The ability to interface with a LIS device to receive patient identification and test requests via a communications protocol to provide sample tracking via bar code labeling and RFID. Processes multiple tube sizes/types simultaneously; sorts and maps samples for routing, storage, and retrieval; performs intelligent sample routing based on system status | Same |
| Software Components | Data Innovations
• Instrument Manager™ Core with drivers Thermo Fisher Scientific
• TCAutomation software communicates to IM and track | Same |
| Software Components | Developed and provided by Data Innovations
• Incorporated into Automation Driver, Instrument Driver, Universal LIS Driver and Out of Service driver
• Optional VITROS Autoverification Rules set – developed by Data Innovations | Developed and provided by Ortho enGen configuration file (gsb)
• sample routing and reprocessing
• includes Autoverification |
5
{5}
6
I. Special Control/Guidance Document Referenced (if applicable):
- EN ISO 14971: 2009 Medical devices – Application of risk management to medical devices
- EN 13612:2002 Performance evaluation of in vitro diagnostic medical devices
- EN ISO 18113 – 1:2011 In vitro diagnostic medical devices-Information supplied by the manufacturer (labeling)-Part1: Terms, definitions and general requirements
- EN ISO 18113-3:2011 In vitro diagnostics medical devices-Information supplied b the manufacturer (labeling)-Part 3: Instruments for professional use.
- EN ISO 13485:2012 Medical devices – Quality management systems – Requirements for regulatory purposes
- EN ISO 15223-1:2012 Medical devices – Symbols to be used with medical device labels, labelling and information to be supplied – Part 1: General requirements
- EN 61326-2-6 Electrical equipment for measurement, control and laboratory use- EMC requirements
- EMC: IEC 61010-2-101 Medical Electrical Equipment – Part 1: General Requirements for Safety
- EC 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
Provided in VITROS CRBM (carbamazepine) k980283, VITROS GENT (Gentamicin) k042479, and VITROS TT4 (Total T4) k964721
b. Precision/Reproducibility:
Provided in VITROS CRBM (carbamazepine) k980283, VITROS GENT (Gentamicin) k042479, and VITROS TT4 (Total T4) k964721
c. Linearity:
Provided in VITROS CRBM (carbamazepine) k980283, VITROS GENT (Gentamicin) k042479, and VITROS TT4 (Total T4) k964721
d. Carryover:
Not applicable
e. Interfering Substances:
Provided in VITROS CRBM (carbamazepine) k980283, VITROS GENT
{6}
(Gentamicin) k042479, and VITROS TT4 (Total T4) k964721
## 2. Other Supportive Instrument Performance Data Not Covered Above:
### a. Method comparison with predicate device:
A comparison study was performed to demonstrate equivalence between results for samples delivered to the VITROS System analyzers via the VITROS Automation Solutions (on-track) and samples manually placing on the VITROS System analyzers by the operator (off-tack). A total of 70 samples were tested for carbamazepine, 13 samples were spiked (18%) with a stock solution of carbamazepine using the VITROS® 5,1 FS Chemistry System. A total of 55 native samples were tested for gentamicin using the 5600 Integrated System and a total of 51 native samples and an additional 6 diluted samples (10% spiked) were tested for Total T4 using the 3600 Immunodiagnostic System. The samples were split into pairs: one sample processed through the VITROS Automation Solutions (on-track) and delivered to the VITROS chemistry analyzer and the other was processed by manually loading directly to the VITROS chemistry analyzers systems (off-track). The samples were tested in three replicates per sample both on and off the track. The first replicate from each sample was used to generate the regression statistics. The regression analysis for each assay are summarized in the tables below:
| Assay | Sample Range | N | Slope | Intercept | R² |
| --- | --- | --- | --- | --- | --- |
| CRBM (μg/mL)
(5,1 FS Chemistry System) | 3.09 – 17.12 | 70 | 1.04 | -0.0905 | 0.9796 |
| GENT (μg/mL)
(5600 Integrated System) | 0.63 – 9.72 | 55 | 1.00 | 0.0075 | 0.0989 |
| Total T4 (nmol/L)
(3600Immunodiagnostic) | 12.70 – 288.70 | 57 | 1.01 | -1.1936 | 0.9969 |
### b. Software/Hardware Verficaiton and Validation
FDA has reviewed the applicant’s software validation and verification documentation for this product and found it to be acceptable.
{7}
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.