← Product Code [KHS](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/KHS) · K073223

# S-TEST C02 (K073223)

_Alfa Wassermann Diagnostic Technologies, Inc. · KHS · Jun 26, 2008 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/KHS/K073223

## Device Facts

- **Applicant:** Alfa Wassermann Diagnostic Technologies, Inc.
- **Product Code:** [KHS](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/KHS.md)
- **Decision Date:** Jun 26, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1160
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The S-Test Carbon Dioxide Reagent is intended for the quantitative determination of carbon dioxide concentration in serum or heparin plasma using the S40 Clinical Analyzer. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

## Device Story

Single-use reagent cartridge for S40 Clinical Analyzer; contains PEP, PEPC, MD, and NADH analog. Input: serum or heparin plasma sample (5 μL). Principle: enzymatic endpoint reaction; CO2 reacts with PEP via PEPC to yield oxaloacetic acid; MD reduces cofactor; absorbance measured bichromatically at 405/508 nm. Output: quantitative CO2 concentration (mEq/L). Used in clinical labs or physician's office labs by trained operators. 2-D barcode on cartridge provides lot-specific calibration data. Results assist clinicians in diagnosing/treating acid-base balance disorders.

## Clinical Evidence

Bench testing only. Precision studies (21 days, 3 levels) showed within-run CV 1.7-3.0%, total CV 5.7-6.7%. POL site precision showed within-run CV 1.3-6.6%, total CV 2.0-6.6%. Accuracy correlation study (n=94) yielded correlation coefficient 0.984, standard error 1.1. POL site correlation studies showed correlation coefficients 0.937-0.980. Detection limit 5 mEq/L.

## Technological Characteristics

Photometric enzymatic endpoint assay. Reagents: PEP, PEPC (microbial), MD (mammalian), NADH analog, buffer (pH 7.2), activators, stabilizers, surfactant. Form factor: single-use cartridge with two reagent cells, cuvette, film seal, 2-D barcode. Connectivity: 2-D barcode for lot-specific calibration data. Standards: NIST SRM351 traceability; CLSI EP5-A2, EP10-A, EP6-A, EP7-A, EP17-A, EP9-A2, C28-A2.

## Regulatory Identification

A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.

## Predicate Devices

- ACE plus ISE/Clinical Chemistry System ACE Carbon Dioxide Reagent ([K931786](/device/K931786.md))
- Ortho Clinical Diagnostics Fusion Clinical Chemistry Analyzer Carbon Dioxide Reagent ([K946090](/device/K946090.md))
- Piccolo® xpress Chemistry Analyzer Carbon Dioxide Reagent ([K942782](/device/K942782.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k073223

B. Purpose for Submission:
New device

C. Measurand:
Total Carbon Dioxide

D. Type of Test:
Quantitative, Photometric

E. Applicant:
Alfa Wassermann Diagnostic Technologies, Inc.

F. Proprietary and Established Names:
S-Test Carbon Dioxide (CO2) Reagent Cartridge

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  KHS – Bicarbonate/carbon dioxide test system | Class II | 862.1160 | 75, Chemistry  |

H. Intended Use:

1. Intended use(s):
See Indications for use below

{1}

2. Indication(s) for use:

The S-Test Carbon Dioxide Reagent is intended for the quantitative determination of carbon dioxide concentration in serum or heparin plasma using the S40 Clinical Analyzer. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. The test is intended for use in clinical laboratories or physician's office laboratories. For in vitro diagnostic use only.

3. Special conditions for use statement(s):

For Prescription Use only

4. Special instrument requirements:

S40 Clinical Chemistry Analyzer

I. Device Description:

The Carbon Dioxide (CO2) is a single use reagent cartridge having two reagent cells, Photometric reaction cuvette, film seal and a 2-D code label. The reagent cells contain the following reagents; Reagent 1 – water and preservative, Reagent 2 - Phosphoenolpyruvate (PEP), Nicotinamide adenine dinucleotide (NADH) analog, reduced, Phosphoenol pyruvate carboxylase (PEPC) (Microbial), Malate dehydrogenase (MD) (Mammalian), Buffer (pH 7.2, 25°C), Activators, Stabilizers, Surfactant and Preservative.

J. Substantial Equivalence Information:

1. Predicate device name(s):

ACE plus ISE/Clinical Chemistry System, Alfa Wassermann Vitros Fusion Clinical Chemistry Analyzer, Ortho Clinical Diagnostics Piccolo xpress Chemistry Analyzer, Abaxis Inc.

2. Predicate 510(k) number(s):

k930104, k001133 and k950164 respectively

3. Comparison with predicate:

The device and the predicate devices share a similar intended use, analytes measured, analysis temperature, reactive ingredients and sample type.

{2}

|  Differences  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Item | S-Test Carbon dioxide reagent on the S40 Chemistry Analyzer | ACE plus ISE/ clinical chemistry analyzer | Vitros Fusion Clinical Chemistry Analyzer | Piccolo xpress Chemistry Analyzer  |
|  Sample volume | 5 μL | 6 μL | 6 μL | 100 μL  |
|  Linearity range | 5-45 mEq/L | 2-50 mEq/L | 5-40 mEq/L | 5-40 mEq/L  |
|  Detection limit | 5 mEq/L | 2 mEq/L | 5 mEq/L | 5 mEq/L  |
|  Reaction Type | Enzymatic endpoint | Kinetic | Enzymatic endpoint | Kinetic  |

## K. Standard/Guidance Document Referenced (if applicable):

CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition (2004)

CLSI EP10-A: Preliminary Evaluation of Quantitative Clinical Laboratory Methods; Approved Guideline –Second Edition (2002)

CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach: Approved Guideline (2003)

CLSI EP7-A: Interference Testing in Clinical Chemistry; Approved Guideline (2002)

CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline (2004)

CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (2002)

CLSI C28-A2: How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline-Second Edition (2000), Section 8.2: Transference and Validation

## L. Test Principle:

Carbon dioxide in sample reacts with PEP in the presence of PEPC and magnesium to yield oxaloacetic acid and phosphate. In the presence of MD the reduced cofactor absorbs strongly at 405 nm. The difference in absorbance between the final and first reading which is measured bichromatically at 405/508 nm is directly proportional to the carbon dioxide concentration in the sample.

{3}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Precision studies were conducted in-house and at three Physician Office Laboratories (POL) (with three trained operators typically found in these settings) by testing three serum samples. The samples were run once a day, three times per run for five days using one instrument at each site. The results are presented below:

|   | CO2 mEq/L  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  |  | % CV  |   |
|  Lab | Sample | Mean | Within Run | Total  |
|  In-House | 1 | 18 | 2.2% | 2.8%  |
|  POL 1 | 1 | 21 | 3.9% | 5.7%  |
|  POL 2 | 1 | 19 | 3.8% | 3.7%  |
|  POL 3 | 1 | 19 | 3.6% | 3.5%  |
|  |   |   |   |   |
|  In-House | 2 | 23 | 1.3% | 3.0%  |
|  POL 1 | 2 | 25 | 2.8% | 2.8%  |
|  POL 2 | 2 | 23 | 1.7% | 2.0%  |
|  POL 3 | 2 | 25 | 2.4% | 5.3%  |
|  |   |   |   |   |
|  In-house | 3 | 27 | 6.6% | 6.6%  |
|  POL 1 | 3 | 32 | 4.0% | 5.6%  |
|  POL 2 | 3 | 33 | 2.8% | 6.4%  |
|  POL 3 | 3 | 33 | 4.2% | 5.9%  |

b. Linearity/assay reportable range:

Linearity across the assay range was confirmed by testing commercial linearity standards with 9 levels with know concentrations of CO2. Each level was tested in replicates of four. Results are presented below:

{4}

|  Total Carbon Dioxide  |   |   |   |
| --- | --- | --- | --- |
|  Sample | Assigned Value mg/dL | Measured Value mg/dL | % Recovery  |
|  1 | 3 | 4 | +1 mEq/L  |
|  2 | 7 | 6 | -1 mEq/L  |
|  3 | 13 | 14 | 104%  |
|  4 | 18 | 17 | 96%  |
|  5 | 24 | 24 | 101%  |
|  6 | 28 | 27 | 96%  |
|  7 | 33 | 33 | 101%  |
|  8 | 38 | 37 | 97%  |
|  9 | 43 | 43 | 100%  |
|  Linear regression y = 0.988x + 0.02, r² = .9964  |   |   |   |

The reportable range of 5-43 mEq/L is based on linearity, detection limit and method comparison.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The S Test Carbon dioxide cartridge is factory calibrated by using standard material with an assigned value which is traceable to the NIST standard Reference material SRM351. The 2-D barcode printed on each cartridge provides the analyzer with lot-specific calibration data.

Real time stability studies have been conducted. Protocols and acceptance criteria were described and found to be acceptable. The manufacturer claims the following expiration date:

When stored at 2-8 °C the assay reagent is good until the expiration date.

d. Detection limit:

The Limit of Blank and Limit of Detection was determined by running a low sample (five different samples) and a BSA sample, (7.5% in saline) for 3 days, 4 replicates/day, for each low sample and 20 replicates/day, for the true blank sample, for a total of 60 results. The testing was split between two instruments. The detection limit is 5.0 mEq/L CO2.

e. Analytical specificity:

Interference studies to determine the effects of Unconjugated Bilirubin, Hemolysis were performed using seven serum pools containing approximately 24.0 mg/dl CO2 which were spiked with varies concentrations of unconjugated bilirubin (0-50 mg/dL), hemoglobin (0-1000 mg/dL). To test Lipemia interference three lipemic serum samples were used.

{5}

Sponsor states that interference is considered to be significant if the analyte result is different from the control by $\pm 10\%$. There was no significant interference from bilirubin, hemoglobin or Lipemia.

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

Clinical correlation studies were performed comparing the S-Test CO2 results generated on the S40 Clinical analyzer against the results from the ACE Clinical analyzer using a total of 94 serum sample. Of the 94 serum samples (9 were diluted and 8 were spiked samples). All the samples were measured in singlet and only one specimen per patient was obtained.

The correlation study between the device and the predicate yielded the following results.

|  Test | n | Slope | Intercept | r | Sample range (mEq/L)  |
| --- | --- | --- | --- | --- | --- |
|  S Test Carbon Dioxide | 94 | 0.985 | 0.42 | 0.984 | 6-41.8  |

Performance for the S Test Carbon Dioxide was evaluated at three Physician Office Laboratories with a total of three operators who are typical operators at these sites. Operators ran from 39 or more unaltered clinical serum samples obtained from each site as well as 8 diluted and 4 spiked samples. The S Test Carbon dioxide test results were compared to the ACE results. The correlation study between the device and the predicate for serum yielded the following results.

|   |  | n | Slope | Intercept | r | Sample range (mEq/L)  |
| --- | --- | --- | --- | --- | --- | --- |
|  Carbon Dioxide | Lab A | 57 | 0.995 | .15 | 0.948 | 5.3 – 38.9  |
|   |  Lab B | 47 | 0.941 | -0.08 | 0.937 | 5.3 – 38.9  |
|   |  Lab C | 57 | 1.000 | -0.76 | 0.980 | 5.3 – 38.9  |

{6}

b. Matrix comparison:

A serum / plasma comparison test was performed for the S-Test CO2 assay. Thirty-five paired samples were assayed on the S40 System. Seven of the samples where spiked and four samples where diluted to help cover the assay range. The correlation is as follows:

$$
y = 0.998x - 0.40, r = 0.962, \text{range } 10-42
$$

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

Forty-five (45) normal serum samples for CO2 were evaluated on the S40 Clinical Analyzer to determine if the reference ranges of the predicate (Ortho Clinical Fusion) could be transferred to the new assay. The sponsors' acceptance criterion is 90% of the assay results for the normal samples are within the predicate range. Analysis confirmed sufficient agreement (6.7% non-congruent results for CO2, sponsor specification ≤10%) to transfer the reference range.

CO2 – 22-30 mEq/L

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

{7}

8

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/KHS/K073223](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/KHS/K073223)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
