This product is to be used for the quantitative determination of carbon dioxide in human serum by spectrophotometric analysis. The determination of the level of carbon dioxide in serum is commonly performed as an indicator of acid-base balance disturbances.
Device Story
Carbon Dioxide Liquid Stable Reagent is an in vitro diagnostic reagent for clinical laboratory use. It utilizes spectrophotometric analysis to measure carbon dioxide levels in human serum samples. The reagent reacts with the sample to produce a measurable change in absorbance, which correlates to the concentration of carbon dioxide. This quantitative measurement assists clinicians in evaluating patient acid-base balance. The device is intended for professional use in clinical settings.
Clinical Evidence
No clinical studies were performed. Evidence is based on bench testing. Method comparison study (n=136 samples, range 14-40 mmol/L) against predicate device yielded a correlation coefficient of 0.986 and linear regression y = 0.965x + 1.2. Precision evaluated on Hitachi 917 analyzer showed within-day CVs of 2.2-3.1% and day-to-day CVs of 2.9-3.1%. Analytical specificity testing showed no interference from bilirubin (up to 20 mg/dL), hemoglobin (up to 500 mg/dL), or lipemia (up to 1000 mg/dL).
Technological Characteristics
Quantitative enzymatic assay. Reagents: Phosphoenolpyruvate (8.0 mM), Magnesium Ions (20 mM), NADH analog (0.5 mM), PEPC (>200 U/L), MDH (>1200 U/L), buffer (pH 7.5). Form factor: liquid, ready-to-use. Energy source: spectrophotometric (405-415 nm). Performance demonstrated on Hitachi 917 chemistry analyzer. Stability: 15-month shelf life at 2-8°C.
Indications for Use
Indicated for quantitative determination of carbon dioxide in human serum to assess acid-base balance disturbances.
Regulatory Classification
Identification
A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
Related Devices
K012332 — JAS CARBON DIOXIDE LIQUID REAGENT · Jas Diagnostics, Inc. · Sep 6, 2001
K991108 — CARBON DIOXIDE (CO2), PRODUCT NO. CC0107-01 · A.P. Total Care, Inc. · May 25, 1999
K073223 — S-TEST C02 · Alfa Wassermann Diagnostic Technologies, Inc. · Jun 26, 2008
K963541 — SIGMA DIAGNOSTICS CO2 ACID REAGENT · Sigma Diagnostics, Inc. · Oct 24, 1996
K071340 — CARBON DIOXIDE (CO2), SCAL, NORTROL AND ABTROL · Thermofisher Scientific OY · Oct 5, 2007
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k070251
B. Purpose for Submission:
New device
C. Measurand:
Carbon Dioxide
D. Type of Test:
Quantitative enzymatic assay
E. Applicant:
Pointe Scientific, Inc.
F. Proprietary and Established Names:
Carbon Dioxide Liquid Stable Reagent
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1160, Bicarbonate/carbon dioxide test system
2. Classification:
Class II
3. Product code:
KHS - enzymatic, carbon-dioxide
4. Panel:
75 (Clinical Chemistry)
H. Intended Use:
1. Intended use(s):
See Indications for use.
2. Indication(s) for use:
This product is to be used for the quantitative determination of carbon dioxide in human serum by spectrophotometric analysis. The determination of the level of carbon dioxide in serum is commonly performed as an indicator of acid-base balance disturbances.
3. Special conditions for use statement(s):
For Prescription use only
4. Special instrument requirements:
Assay performance was demonstrated on the Hitachi 917 chemistry analyzer.
{1}
I. Device Description:
The carbon dioxide reagent is supplied as a liquid, ready-to-use, single reagent kit. It contains Phosphoenolpyruvate 8.0 mM, Magnesium Ions (20 mM), NADH analog 0.5 mM, Phosphoenolpyruvate carboxylase (PEPC) (Microbial) >200 U/L, Malate Dehydrogenase (Porcine) >1200 U/L, and Buffer, pH 7.5 ± 0.1 non-reactive stabilizers with surfactants and preservative. The reagent is supplied in three sizes: 120 mL, 500 mL, and 1000 mL volumes.
J. Substantial Equivalence Information:
1. Predicate device name(s): Carbon Dioxide - L3k Assay (Diagnostic Chemicals Ltd)
2. Predicate 510(k) number(s): k990754
3. Comparison with predicate:
| Characteristics | Carbon Dioxide Liquid Stable Reagent (Proposed Device) | Diagnostic Chemicals Limited (DCL) (Predicate Device) |
| --- | --- | --- |
| Intended Use | This product is to be used for the quantitative determination of carbon dioxide in human serum by spectrophotometric analysis. For in vitro diagnostic use only. | For the quantitative determination of carbon dioxide in serum. For IN VITRO diagnostic use. |
| Reagent | PEP 8.0mM, Magnesium Ions 20mM, NADH analog, MDH (porcine) ≥ 1200U/L, PEPC (microbial) ≥ 200U/L, Buffer, pH 7.5 ± 0.1 non-reactive stabilizers with surfactants and preservative. | PEP 12.5 mmol/L, NADH analog, MDH (mammalian) > 4100 U/L, PEPC (microbial) > 400 U/L, Buffer, pH 7.5 at 25°C, activators, stabilizers, surfactant and a preservative. |
| Format | Reagent provided as a ready to use liquid. | Reagent is provided in a ready to use format. |
| Stability | • Reagent is stable until expiration date indicated on vial label when stored tightly capped at 2-8°C.
• Shelf life is 15 months when stored tightly capped at 2-8°C. | Reagent is stable until the expiration date stated on the label at 2-8°C. Specific Shelf life not indicated. |
| Linearity / Assay range | 2.0 – 40.0 mmol/L | 2.9 – 50.0 mmol/L |
| Limit of Detection | 2.0 mmol/L | 2.9 mmol/L |
| Interference | No interference was observed by bilirubin up to 20.0 mg/dL, hemoglobin up to 400 mg/dL and | No significant (> 10.0%) lipemic interference found at Intralipid levels from 1-1000 |
{2}
3
| | lipemia (intralipid) up to 1000 mg/dl. (using a criteria of >10% variance from control) | mg/dl (0-3000 mg/dl Triglyceride). No significant (> 10.0%) icteric interference at Bilirubin levels of 0-40 mg/dl. Hemoglobin levels of 600 mg/dl gave a positive bias of 10.3% in a 26.3 mmol/L carbon Dioxide sample. |
| --- | --- | --- |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A: Evaluation of Precision Performance of Clinical Chemistry Devices
# L. Test Principle:
Carbon Dioxide Reagent is a quantitative enzymatic assay based on the PEP Carboxylase methodology. Carbon Dioxide (in the form of bicarbonate ions) reacts with phosphoenolpyruvate (PEP), in the presence of phosphoenolpyruvate carboxylase (PEPC) to form oxaloacetate. Malate dehydrogenase (MDH) catalyzes the reduction of oxalacetate to malate with the concomitant oxidation of reduced nicotinamide adenine dinucleotide (NADH) analog. Spectrophotometric determination of the decrease in absorbance monitored between 405 and 415 nm is proportional to the amount of CO₂ in the sample.
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The precision was evaluated using Pointe Scientific chemistry controls containing low and normal levels of CO₂ on the Hitachi 917 chemistry analyzer. Within-day precision was evaluated by running the two specimens in replicates of 20 on the same day. Day-to-day precision was evaluated by performing one run per day in duplicate over 20 days using the same control material. The results are tabulated below.
Within-day Precision
| Description | Control 1 | Control 2 |
| --- | --- | --- |
| Number of data points | 40 | 40 |
| Mean (mmol/L) | 16.5 | 27.4 |
| SD (mmol/L) | 0.5 | 0.6 |
| CV | 3.1% | 2.2% |
{3}
Day-to-day Precision
| Description | Control 1 | Control 2 |
| --- | --- | --- |
| Number of data points | 40 | 40 |
| Mean (mmol/L) | 18.2 | 27.1 |
| SD (mmol/L) | 0.5 | 0.9 |
| CV | 2.9% | 3.1% |
# b. Linearity/assay reportable range:
To determine the linearity range, the sponsor prepared eleven $\mathrm{CO}_{2}$ test samples through serial dilution of a $100\mathrm{mmol / L}$ $\mathrm{CO}_{2}$ standard. The samples ranged in concentration from 0.67 to $95.3\mathrm{mmol / L}$ . All eleven $\mathrm{CO}_{2}$ levels were run in triplicate on the Hitachi 917 analyzer. The reagent showed the assay is linear up to $47.7\mathrm{mmol / L}$ . Up to this level, the linear regression analysis demonstrated a linear regression equation, $\mathrm{Y} = 0.999\mathrm{X} - 0.3$ . The assay range claimed by the sponsor is 2.0 to $40\mathrm{mmol / L}$ .
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
No information was provided for traceability. The device does not include calibrators and controls. The sponsor recommends using an aqueous $\mathrm{CO}_{2}$ standard (30 mmol/L) or an appropriate serum calibrator. To monitor the reliability of results, the sponsor also recommends two levels of control sera with known $\mathrm{CO}_{2}$ values to be run along with patient samples.
Accelerated stability of the $\mathrm{CO}_{2}$ reagent was determined based on Arrehnius equations from J. Pharmaceutical Sciences, 53:7, 8158-8180, 1964. Accelerated stability of regent for 12, 18, and 24 month duration was tested using New England Reagent Laboratory (NERL) standards (0, 10, 20, 30, and 40 mmol/L) and two Pointe Scientific controls (14-26 mmol/L and 22-42 mmol/L). Based on this study, accelerated stability of the $\mathrm{CO}_{2}$ reagent was demonstrated to be at least 18 months. The sponsor has established a 15-month shelf life for the $\mathrm{CO}_{2}$ reagent when tightly capped and stored at $2 - 8^{\circ}\mathrm{C}$ .
# d. Detection limit:
To demonstrate the limit of detection, a known NERL standard of $\mathrm{CO}_{2}$ (10 mmol/L) and a blank sample (0 mmol/L) were tested in five replicates. The sponsor defined the limit of zero-concentration (blank) sample as $\pm 2\mathrm{SD}$ , which was demonstrated to be $0.5\mathrm{mmol/L}$ . Based on these results, the sponsor claimed LOD of $2.0\mathrm{mmol/L}$ .
# e. Analytical specificity:
The sponsor evaluated the effect of bilirubin (0-20 mg/dL), hemoglobin (0-500 mg/dL), and lipemia (intralipid) (0-1000 mg/dL) on a sample at $\mathrm{CO}_{2}$ concentration of 19 mmol/L. Based on the sponsor-defined interference limit of $\pm 10\%$ of control, results indicated that there was no interference up to the concentrations tested above for any of the substances.
{4}
f. Assay cut-off:
Not Applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Performance of the Carbon Dioxide Liquid Stable Reagent was compared with performance of the predicate device, Carbon Dioxide - L3k Assay (Diagnostic Chemicals Ltd) (k990754). The study was performed on Hitachi 917 chemistry analyzer using 136 samples ranging from 14 to 40 mmol/L of CO₂. Comparison of the data based on 2 methods gave a correlation coefficient of 0.986. Linear regression analysis resulted in the equation, y = 0.965x + 1.2.
b. Matrix comparison:
Carbon Dioxide Liquid Stable Reagent is to be used with serum only.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not Applicable.
5. Expected values/Reference range:
The expected values for CO₂ were based on published literature, Henry, R.J., Clinical Chemistry: Principles and Technics, Harper & Row, Publishers, New York, N.Y., 1974. They are stated as 23-34 mmol/L. In the labeling, the sponsor recommends that each laboratory determine its own reference range.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.