The LifeOutcomes C-Quest Blood Culture Sampling Device intended to collect a patient blood sample and sequester the first portion of the sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion. It is used in conjunction with either a scalp vein set for venipuncture or access to an existing patient IV using an extension set. C-Quest may be used with a blood culture tube or a standard blood collection syringe.
Device Story
Device collects patient blood samples; sequesters initial 1.5-2.0 mL of blood to reduce contamination risk. Operates via user-controlled negative pressure; initial blood drawn into diversion syringe; unidirectional check valve in tee body assembly isolates initial sample; prevents backflow into subsequent non-contaminated sample. Subsequent blood flow directed to blood culture bottle or collection syringe. Used by medical professionals in clinical settings; venipuncture or existing IV access. Output is uncontaminated blood sample for diagnostic testing; improves accuracy of blood culture results; reduces false-positive contamination rates.
Clinical Evidence
No clinical data was submitted in this submission. Bench testing only.
Technological Characteristics
Materials: stainless steel, silicone, polypropylene. Principle: user-controlled negative pressure with unidirectional check valve for blood sequestration. Form factor: tee body assembly with universal culture tube holder and diversion syringe. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ASTM F1980-16 (aging), ISO 10993-1 (biocompatibility), ISO 80369-7 (connectors), ISO 8536-12 (check valves).
Indications for Use
Indicated for medical professionals to collect blood samples and sequester the initial 1.5-2.0 mL of blood to reduce blood culture contamination. Used with scalp vein sets or existing IV extension sets; compatible with blood culture tubes or standard collection syringes.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
K222299 — Steripath Micro Blood Collection System · Magnolia Medical Technologies · Nov 30, 2022
K251812 — Steripath® Flow Blood Collection System · Magnolia Medical Technologies · Sep 25, 2025
K162233 — Kurin Blood Culture Collection Set · Pathway, LLC · Dec 23, 2016
K192247 — Steripath Gen2 Blood Collection System · Magnolia Medical Technologies, Inc. · Feb 28, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 26, 2024
LifeOutcomes LLC % Thomas Kroenke Principal Consultant Speed To Market, Inc. P.O. Box 3018 Nederland, Colorado 80466
Re: K231282
Trade/Device Name: LifeOutcomes C-Quest™ Blood Culture Sampling Device Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: December 22, 2023 Received: December 26, 2023
Dear Thomas Kroenke:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ely, Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231282
#### Device Name
LifeOutcomes C-Quest™ Blood Culture Sampling Device
#### Indications for Use (Describe)
The LifeOutcomes C-Quest Blood Culture Sampling Device intended to collect a patient blood sample and sequester the first portion of the sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion. It is used in conjunction with either a scalp vein set for venipuncture or access to an existing patient IV using an extension set. C-Quest may be used with a blood culture tube or a standard blood collection syringe.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> |
|------------------------------------------------------------|---------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Preparation Date: | 26 January 2024 |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | LifeOutcomes LLC<br>88 Inverness Circle East, Unit E-102<br>Englewood, CO 80112 USA |
| Submitter<br>Correspondent | Mr. Robert J. McKinnon<br>Phone: +1 (720) 797-8070<br>Email: jon@lifeoutcomes.us |
| Application<br>Correspondent: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>+1 (303) 956 4232 |
| Trade Name: | LifeOutcomes C-Quest™ Blood Culture Sampling Device |
| Common Name: | Tubes, Vials, Systems, Serum Separators, Blood Collection |
| Regulation Name: | Blood Specimen Collection Device |
| Regulation<br>Number: | 21 CFR §862.1675 |
| Product Code: | JKA |
| Predicate Device: | Steripath® Gen2 Blood Collection System (K192247) |
| Device Description: | The LifeOutcomes LLC (LO) C-Quest™ Blood Culture Sampling<br>Device (C-Quest) is a blood collection system that diverts and sequesters<br>the initial specimen prior to collection of a subsequent test sample. This<br>reduces the frequency of blood culture contamination when contaminants<br>are present in the initial blood sample compared to blood cultures drawn<br>with standard procedure without manual diversion. C-Quest is intended<br>for use by medical professionals.<br>The C-Quest consists of three (3) major components:<br>• the tee body assembly.<br>• a universal culture tube holder, and |
- a third-party, FDA cleared, 5 ml diversion syringe.
{4}------------------------------------------------
The LifeOutcomes C-Quest Blood Culture Sampling Device (C-Quest) is Indications for Use: a device intended to collect a patient blood sample and sequester the first portion of the sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion. It is used in conjunction with either a scalp vein set for venipuncture or access to an existing patient IV using an extension set. C-Quest may be used with a blood culture tube or a standard blood collection syringe.
> The C-Ouest employs the same or similar technological characteristics as the predicate device.
| Characteristic | Magnolia Medical<br>Technologies, Inc.<br>Steripath Gen2 Blood<br>Collection System<br>(Predicate Device -<br>K192247) | LifeOutcomes LLC<br>C-Quest Blood Culture<br>Sampling Device<br>(Subject Device -<br>K231282) | Discussion of Differences |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Steripath® Gen2 Blood<br>Collection System is a<br>system to draw blood for in<br>vitro diagnostic testing.<br>The Steripath® Gen2 Blood<br>Collection System is<br>indicated for use as a blood<br>collection system that<br>diverts and sequesters the<br>initial specimen prior to<br>collection of a subsequent<br>test sample to reduce the<br>frequency of blood culture<br>contamination when<br>contaminants are present in<br>the initial blood sample<br>compared to blood cultures<br>drawn with standard<br>procedure without manual<br>diversion.<br>Additionally, components of<br>the system may be used for<br>infusion following sample<br>collection after<br>disconnection of the Initial<br>Specimen Diversion<br>Device® (ISDD®).<br>Venipuncture needles are<br>indicated for short term<br>infusion (less than 2 hours). | The LifeOutcomes C-<br>Quest Blood Culture<br>Sampling Device (C-<br>Quest) is a device intended<br>to collect a patient blood<br>sample and sequester the<br>first portion of the sample<br>to reduce the frequency of<br>blood culture<br>contamination when<br>contaminants are present in<br>the initial blood sample<br>compared to blood cultures<br>drawn with standard<br>procedure without manual<br>diversion. It is used in<br>conjunction with either a<br>scalp vein set for<br>venipuncture or access to<br>an existing patient IV<br>using an extension set. C-<br>Quest may be used with a<br>blood culture tube or a<br>standard blood collection<br>syringe. | Both C-Quest and the<br>predicate are indicated<br>for the collection of blood<br>samples, and<br>sequestration of the initial<br>1.5 ml of blood.<br>C-Quest components are<br>not indicated for infusion<br>whereas the predicate<br>device components are<br>indicated for infusion<br>after removal of the<br>ISDD. The subject device<br>is indicated for a subset<br>of the predicate device's<br>indication. The difference<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Product Code<br>and Regulation | JKA, 21 CFR §862.1675.<br>FPA, 21 CFR §880.5440. | JKA, 21 CFR §862.1675. | C-Quest components are<br>not indicated for infusion<br>whereas the predicate<br>device components are<br>indicated for infusion<br>after removal of the<br>ISDD. The subject device<br>is indicated for a subset<br>of the predicate device's<br>indication. The difference<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Characteristic | Magnolia Medical<br>Technologies, Inc.<br>Steripath Gen2 Blood<br>Collection System<br>(Predicate Device -<br>K192247) | LifeOutcomes LLC<br>C-Quest Blood Culture<br>Sampling Device<br>(Subject Device -<br>K231282) | Discussion of Differences |
| Initial Blood<br>Volume<br>Sequestered | 1.5 to 2.0 ml. | A minimum of 1.5 ml, not<br>to exceed 2.0 ml. | Same; a slight difference<br>in how the sequestration<br>amount is stated. |
| Single Use | Yes | Yes | Same. |
| Sterilization<br>Method and<br>Sterility<br>Assurance Level | Ethylene Oxide (EO), SAL<br>10-6. | EO, SAL 10-6. | Same. |
| Labeled Non-<br>pyrogenic | Yes. | Yes. | Same. |
| Shelf Life | 1 year. | 18 months. | C-Quest has an 18 month<br>shelf-life supported by<br>testing post-accelerated<br>aging. Therefore, the<br>difference does not raise<br>different questions of<br>safety and effectiveness. |
| Components | The Steripath Gen2 Blood<br>Collection System includes:<br>· Initial specimen<br>diversion device<br>(ISDD®).<br>Inlet accessory<br>(venipuncture needle or<br>extension set).<br>· Outlet accessory<br>(culture blood transfer<br>adapter or blood<br>collection syringes). | The C-Quest Blood<br>Culture Sampling Device<br>includes:<br>· C-Quest diversion<br>device.<br>Inlet accessory<br>(venipuncture needle or<br>extension set).<br>· Outlet accessory<br>(blood culture bottle or<br>blood collection<br>syringe). | Both the predicate and<br>subject devices are<br>systems that incorporate a<br>blood diversion device,<br>and inlet and outlet<br>accessories. Both blood<br>diversion devices initially<br>sequester 1.5 to 2.0 ml of<br>potentially contaminated<br>blood and prevent<br>backflow of the<br>potentially contaminated<br>blood into the<br>uncontaminated blood<br>sample. Both systems<br>utilize inlet and outlet<br>accessories, which allow<br>venous access through<br>either a venipuncture or<br>extension line connected<br>to a patient intravenous<br>(IV) line. The C-Quest<br>device was tested to<br>demonstrate sequestration<br>performance, backflow<br>prevention, and syringe<br>activation force.<br>Therefore, the differences<br>do not raise different<br>questions of safety and<br>effectiveness. |
| Characteristic | Magnolia Medical<br>Technologies, Inc.<br>Steripath Gen2 Blood<br>Collection System<br>(Predicate Device -<br>K192247) | LifeOutcomes LLC<br>C-Quest Blood Culture<br>Sampling Device<br>(Subject Device -<br>K231282) | Discussion of Differences |
| Materials | Medical grade materials<br>(stainless steel, PVC tubing,<br>medical grade adhesives<br>polycarbonate) | Medical grade materials<br>(stainless steel, silicone,<br>polypropylene) | Similar. The subject and<br>predicate devices are<br>comprised of medical<br>grade materials; the<br>difference in the specific<br>component materials<br>were tested to FDA<br>recognized consensus<br>standards to demonstrate<br>biocompatibility<br>compliance. Therefore,<br>the difference does not<br>raise different questions<br>of safety and<br>effectiveness. |
| Packaging | Chevron Pouch, 12" x 6"<br>TPT-0270 to TPF-0524a | Chevron Pouch, 9" x 6" 60<br>gm/m2 Arjo paper, glue<br>coated heated sealed to<br>52µm PET/PE film. | While different materials,<br>both pouches are<br>comprised of materials<br>routinely used for<br>sterilized and sterilizable<br>medical device pouches;<br>the difference in the<br>specific packaging<br>materials were tested to<br>FDA recognized<br>consensus standards to<br>demonstrate package<br>integrity compliance.<br>Therefore, the difference<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Mechanism of<br>Action | User-controlled negative<br>pressure to sequester 1.5 ml<br>to 2 ml of the initial<br>contaminated blood drawn<br>into a chamber.<br>A second, independent<br>blood flow path, after the<br>sequestration of this initial<br>contaminated blood, is<br>isolated from this chamber<br>which prevents the<br>contaminated blood from<br>entering the non-<br>contaminated blood sample. | User-controlled negative<br>pressure to sequester a<br>minimum of 1.5 ml of the<br>initial contaminated blood<br>drawn. Once the initial<br>contaminated blood is<br>drawn into the syringe, it is<br>isolated in the syringe by<br>the unidirectional check<br>value in the C-Quest tee<br>body assembly which<br>prevents backflow of the<br>initial contaminated blood<br>from entering the non-<br>contaminated blood<br>sample.<br>The subsequent blood flow<br>is directed into a second,<br>independent blood flow<br>path, which could be either<br>a blood culture bottle or<br>into a blood collection<br>syringe activated by the<br>user pulling the plunger on<br>the blood collection<br>syringe. | Same. |
### Technology Comparison:
{5}------------------------------------------------
### Technology Comparison (continued):
{6}------------------------------------------------
### Technology Comparison (continued):
{7}------------------------------------------------
### Summary of Verification and Validation Activities:
| Sterilization | The C-Quest sterilization process and sterilized product was validated<br>in accordance with the following Standards and guidance document: |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • ANSI AAMI ST72: 2019, Bacterial endotoxins - Test methods,<br>routine monitoring, and alternatives to batch testing. |
| | • ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile<br>Barrier Systems for Medical Devices. |
| | • ISO 10993-7: 2008, Cor1: 2009, Am1: 2019, Biological evaluation<br>of medical devices – Part 7: Ethylene oxide sterilization residuals. |
| | • ISO 11135: 2014 – Sterilization of health-care products – Ethylene<br>oxide – Requirements for the development, validation and routine<br>control of a sterilization process for medical devices. |
| | • Submission and Review of Sterility Information in Premarket<br>Notification (510(k)) Submissions for Devices Labeled as Sterile, 2<br>Jan 16 |
An analysis and use of adoption methods indicated that the C-Quest sterilization process and sterilized product complies with the applicable Standards and guidance document.
{8}------------------------------------------------
### Summary of Verification and Validation Activities (continued):
| Shelf-Life | The C-Quest sterile packaging has a shelf life of 18 months.<br>The C-Quest sterile packaging was tested for packaging integrity in<br>accordance with the following Standards and guidance document:<br>ASTM F88-21, Standard Test Method for Seal Strength of Flexible<br>Barrier Materials. ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile<br>Barrier Systems for Medical Devices. ISO 10993-7: 2008, Am1: 2019, Biological evaluation of medical<br>devices – Part 7: Ethylene oxide sterilization residuals (including<br>Amendment 1: Applicability of allowable limits for neonates and<br>infants). ISO 11607-1: 2019 – Packaging for terminally sterilized medical<br>devices - Part 1: Requirements for materials, sterile barrier systems<br>and packaging systems. ISTA 3A – 18, Packaged-Products for Parcel Delivery System<br>Shipment 70kg or Less. Submission and Review of Sterility Information in Premarket<br>Notification (510(k)) Submissions for Devices Labeled as Sterile, 21<br>Jan 16 Shelf Life of Medical Devices, Apr 91 Test results indicated that the C-Quest sterile packaging complies with<br>the applicable Standards and guidance document. |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{9}------------------------------------------------
### Summary of Verification and Validation Activities (continued):
Biocompatibility C-Quest is considered a blood path indirect contact, external communicating device for limited contact (less than 24 hours).
> The indirect patient-contact materials in C-Quest were tested for biocompatibility compliance in accordance with the following Standard and guidance document:
- . ISO 10993-1: 2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
- Use of International Standard ISO 10993-1, "Biological evaluation . of medical devices - Part 1: Evaluation and testing within a risk management process," 04 Sep 20.
The following biocompatibility tests were performed:
- Cytotoxicity.
- Sensitization.
- Intracutaneous reactivity.
- Acute system toxicity. ●
- Material-mediated pyrogenicity. ●
- Hemolysis.
- . Particulate testing per USP <788>.
Test results indicated that the C-Quest component materials comply with the applicable Standard and guidance document.
{10}------------------------------------------------
| Performance Testing<br>– Bench | C-Quest was tested for performance in accordance with internal<br>requirements, and the following internal requirements, applicable<br>Standards, and guidance document: |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Sequestration Verification. Actuation Force Verification. IEC 62366: 2007, Medical devices – Application of usability<br>engineering to medical devices. ISO 8536-12: 2021, Infusion equipment for medical use – Part 12:<br>Check valves for single use. ISO 14971: 2019, Medical devices – Application of risk<br>management to medical devices. ISO 80369-7:2021 Small-bore connectors for liquids and gases in<br>healthcare applications-Part 7: Connectors for intravascular or<br>hypodermic applications. ISO 80639-20: 2015, Small-bore connectors for liquids and gases in<br>healthcare applications – Part 20: Common test methods. |
| | Test results indicated that C-Quest complies with internal requirements,<br>applicable Standards, and the guidance document. |
| Clinical Testing | No clinical data was submitted in this submission. |
| Conclusion | The differences between the predicate and the subject device do not<br>raise new or different questions of safety or effectiveness. The<br>LifeOutcomes C-Quest™ Blood Culture Sampling Device is<br>substantially equivalent to the Steripath® Gen2 Blood Collection<br>System with respect to the indications and technological characteristics. |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.