← Product Code [JJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY) · K974826

# BM PRECIPATH(R) HDL/LDL-C CONTROL (K974826)

_Boehringer Mannheim Corp. · JJY · Jan 22, 1998 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K974826

## Device Facts

- **Applicant:** Boehringer Mannheim Corp.
- **Product Code:** [JJY](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY.md)
- **Decision Date:** Jan 22, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

For the monitoring and control of processes for measuring the lipoproteins, HDL-C and LDL-C.

## Device Story

Boehringer Mannheim Precipath® HDL/LDL-C Control is a lyophilized human serum-based quality control product. It is used in clinical laboratory settings to monitor the accuracy and precision of test systems measuring HDL-C and LDL-C. The control is processed alongside patient samples; healthcare providers compare the measured values of the control against established ranges to verify the performance of the analytical process. This ensures the reliability of patient test results, aiding in the clinical management of lipid-related conditions.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Lyophilized human serum base with added HDL-Cholesterol and LDL-Cholesterol. The device functions as an assayed quality control material for in vitro diagnostic use. No software, electronic components, or complex algorithms are involved.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Sigma Diagnostics Cardiolipid™ Controls

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary

## JAN 22 1998 K9774826

According to the requirements of 21 CFR 807.92, the following information Introduction provides sufficient detail to understand the basis for a determination of substantial equivalence.

Boehringer Mannheim Corporation 1) Submitter name, address, 9115 Hague Rd contact Indianapolis, IN 46250 (317) 845-2386

Contact person: Edward R. Kimmelman

Date prepared: Dec. 23, 1997

2) Device name Proprietary name: Boehringer Mannheim Precipath® HDL/LDL-C

Common name: BM Precipath® HDL/LDL-C

Classification name: Multi-Analyte Controls, All Kinds (Assayed and Unassaved)

3) Predicate We claim substantial equivalence to the Sigma Diagostics Cardiolipid™ device Controls. The intended use of the above controls is monitoring and control of processes for measuring the lipoproteins. HDL-C and LDL-C.

The Boehringer Mannheim Precipath® HDL/LDL-C Control consists of 4) Device description lyophilized human serum with added HDL-Cholesterol and LDL-Cholesterol.

The Boehringer Mannheim Precipath® HDL/LDL-C Control is intended to be 5) Intended use used in the monitoring and control of processes for measuring HDL-C and LDL-C.

Continued on next page

11:43

:

{1}------------------------------------------------

## 510(k) Summary, Continued

The Boehringer Mannheim Precipath® HDL/LDL-C control is substantially 6) Comparison equivalent to other products in commercial distribution intended for similar to the predicate device use. Most notably, it is substantially equivalent to the currently marketed Sigma Diagnostics Cardiolipid™ Controls.

> The intended use of this BM control and the predicate devices is the same in that they are intended to be used for the monitoring and control of test systems for the measurement of their labeled analytes.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design resembling an eagle or bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 2 2 1998

Edward R. Kimmelman · Program Director, Regulatory Affairs and Compliance Boehringer Mannheim Corporation 9115 Haque Road Indianapolis, Indiana 46250

K974826 Re: Precipath® HDL/LDL-C Control Requlatory Class: I JJY, LBS Product Code: Dated: December 23, 1997 Received: December 24, 1997__ _ - -

Dear Mr. Kimmelman:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note:

this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact Also, please note the Office of Compliance at (301) 594-4639. the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97).- Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

K974826

Device Name: Boehringer Mannheim Precipath® HDL/LDL-C Control

Indications for Use: For the monitoring and control of processes for measuring the lipoproteins, HDL-C and LDL-C.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-the-Counter Use __

(Optional format 1-2-96)

Division Sign-Off)
vision of Clinical Laboratory Device:

510(k) Number. k974826

16

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K974826](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JJY/K974826)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
