Liquichek Cardiac Markers Plus Control LT is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert. For prescription use only.
Device Story
Liquichek Cardiac Markers Plus Control LT is an assayed quality control serum used in clinical laboratories to monitor the precision of testing procedures for cardiac markers. The device is prepared from human serum with added constituents of human and animal origin, stabilizers, and preservatives. It is provided in liquid form in various configurations. The control is used by laboratory personnel to verify the performance of diagnostic assays. By comparing observed results against the provided mean values and ranges, clinicians can assess the reliability of their testing procedures, ensuring accurate diagnostic information for patient management.
Clinical Evidence
No clinical data. Bench testing only. Stability studies (real-time and accelerated) were performed to establish shelf life and open-vial stability for the added analytes. Value assignment was derived from replicate analyses using manufacturer-supported reagents.
Technological Characteristics
Liquid human serum-based control containing stabilizers and preservatives. Analytes include BNP, CK-MB, CRP, CK, D-dimer, Digitoxin, Homocysteine, MPO, Myoglobin, NT-proBNP, Troponin I, and Troponin T. Storage requirements: -20°C to -70°C for shelf life; 2°C to 8°C for open-vial use. No electronic or software components.
Indications for Use
Indicated for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for cardiac markers including BNP, CK-MB, CRP, CK, D-dimer, Digitoxin, Homocysteine, MPO, Myoglobin, NT-proBNP, Troponin I, and Troponin T.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Liquichek Cardiac Markers Plus Control LT (k050537)
Related Devices
K980556 — LIQUICHEK CARDIAC MARKERS CONTROL LT · Bio-Rad · Feb 25, 1998
K123663 — LIQUIDCHEK CARDIAC MARKERS PLUS CONTROL LT; LEVEL 1C, LEVEL 2, LEVEL 3, TRILEVEL MINIPAK · Bio-Rad Laboratories · Dec 27, 2012
K050537 — LIQUICHEK CARDIAC MARKERS PLUS CONTROL, LIQUICHEK CARDIAC MARKERS PLUS CONTROL LT, LIQUICHEK CARDIAC MARKERS PLUS · Bio-Rad · Apr 27, 2005
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k150300
B. Purpose for Submission:
Modification of previously cleared control (k050537) to add additional analytes
C. Measurand:
Multi-analyte quality control materials
D. Type of Test:
Not applicable
E. Applicant:
Bio-Rad Laboratories
F. Proprietary and Established Names:
Liquichek Cardiac Markers Plus Control LT
1. Regulation section:
21 CFR 862.1660
2. Classification:
Class I, Reserved
3. Product code:
JJY
4. Panel:
Clinical Chemistry (75)
{1}
2
H. Intended Use:
1. Intended use(s):
See the indications for use statements below.
2. Indication(s) for use:
Liquichek Cardiac Markers Plus Control LT is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.
The following analytes are listed in the package insert:
B-type Natriuretic Peptide (BNP)
CK-MB Isoenzyme
C-Reactive Protein (CRP)
Creatine Kinase (CK)
D-dimer
Digitoxin
Homocysteine
Myeloperoxidase (MPO)
Myoglobin
N-terminal pro-Brain Natriuretic
Peptide (NT-proBNP)
Troponin I
Troponin T
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
See product insert for instruments requirements.
I. Device Description:
Liquichek Cardiac Markers Plus Control LT is prepared from human serum with added constituents of human and animal origin, stabilizers and preservatives. This product is provided in liquid form for convenience in the following configurations. This product contains the following analytes;
| Description | Configuration |
| --- | --- |
| Liquichek Cardiac Markers Plus Control LT, Level Low | 6 x 3 mL |
{2}
| Description | Configuration |
| --- | --- |
| Liquichek Cardiac Markers Plus Control LT, Level 1 | 6 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 2 | 6 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 3 | 6 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 1A | 6 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 1B | 6 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 1C | 6 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level Low MiniPak | 1 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 1 MiniPak | 1 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 2 MiniPak | 1 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 3 MiniPak | 1 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 1A MiniPak | 1 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 1B MiniPak | 1 x 3 mL |
| Liquichek Cardiac Markers Plus Control LT, Level 1C MiniPak | 1 x 3 mL |
The sponsor has the following caution statement in their labeling: Each human donor unit used to manufacture this control was tested by FDA accepted methods and found non-reactive for Hepatitis B Surface Antigen (HBsAg), antibody to Hepatitis C (HCV) and antibody to HIV-1/HIV-2.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Bio-Rad Laboratories Liquichek Cardiac Markers Plus Control LT
2. Predicate 510(k) number(s):
k050537
{3}
3. Comparison with predicate:
| SIMILARITIES | | |
| --- | --- | --- |
| | Candidate Device | Predicate Device (k050537) |
| Intended Use | Liquichek Cardiac Markers Plus Control LT is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert. | Same |
| Form | Liquid | Same |
| Base Matrix | Human Serum | Same |
| Thawed and Open Stability | N-terminal pro-Brain Natriuretic Peptide (NT-proBNP): 15 days at 2 to 8°C Troponin I: 10 days at 2 to 8°C B-type Natriuretic Peptide (BNP): 8 days at 2 to 8°C Troponin T: 4 days at 2 to 8°C All other analytes: 20 days at 2 to 8°C | Same |
| Frozen Aliquot Stability | 30 days at -20 to -70°C | Same |
| Shelf Life | 36 months at -20 to -70°C | Same |
| Analytes | B-type Natriuretic Peptide (BNP) CK-MB Isoenzyme C-Reactive Protein (CRP) Creatine Kinase (CK) Digitoxin Homocysteine Myoglobin N-terminal pro-Brain Natriuretic Peptide (NT-proBNP) Troponin I Troponin T | Same |
{4}
| DIFFERENCES | | |
| --- | --- | --- |
| | Candidate Device | Predicate Device (k050537) |
| Analytes | Contains: D-dimer Myeloperoxidase (MPO) | Does not contain: D-dimer Myeloperoxidase (MPO) |
K. Standard/Guidance Document Referenced (if applicable):
None was referenced.
L. Test Principle:
Not applicable.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Value Assignment:
The mean values and the corresponding $\pm 3\mathrm{SD}$ ranges printed in the package inserts were derived from replicate analyses and are specific for each lot of product. The tests listed were performed by the manufacturer and/or independent laboratories using manufacturer supported reagents and a representative sampling of this lot of control. It is recommended that each laboratory establish its own means and acceptable ranges and use those provided only as guides.
Stability Studies:
Real time stability studies were performed to establish thawed and opened stability claims. Thawed and open vial stability when stored at $2^{\circ}\mathrm{C}$ to $8^{\circ}\mathrm{C}$ is as follows: N-terminal pro-Brain Natriuretic Peptide (NT-proBNP) is 15 days, Troponin 1 is 10 days, B-type Natriuretic Peptide (BNP) is 8 days, Troponin T is 4 days and all other analytes are 20 days. Frozen aliquot stability is 30 days at when stored at $-20^{\circ}\mathrm{C}$ to
{5}
–70°C. Accelerated stability studies were performed for establishing the shelf life stability. The shelf life of the Liquichek Cardiac Markers Plus Control LT is 36 months from the date of manufacture when stored at –20°C to –70°C. Real time studies are ongoing. Shelf life and open-vial stability studies protocols and acceptance criteria were reviewed and found to be acceptable.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
{6}
5. Expected values/Reference range:
Expected values for the representative analyzers are provided as a guide in the labeling for each specific lot.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
7
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.