C.F.A.S. PROTEINS, C.F.A.S. PAC, PRECICONTROL CLINCHEM MULTI 1 & 2, PRECINORM PROTEIN & PRECIPATH PROTEIN

K133330 · Roche Diagnostics · JIX · Nov 26, 2013 · Clinical Chemistry

Device Facts

Record IDK133330
Device NameC.F.A.S. PROTEINS, C.F.A.S. PAC, PRECICONTROL CLINCHEM MULTI 1 & 2, PRECINORM PROTEIN & PRECIPATH PROTEIN
ApplicantRoche Diagnostics
Product CodeJIX · Clinical Chemistry
Decision DateNov 26, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

C.f.a.s. (Calibrator for automated systems) PAC (Prealbumin-ASLO-Ceruloplasmin) is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the value sheets. C.f.a.s. (Calibrator for automated systems) Proteins is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the value sheets. PreciControl ClinChem Multi 1 is for the use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets. PreciControl ClinChem Multi 2 is for the use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets. Precinorm Protein and Precipath Protein are for the use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets.

Device Story

Device suite includes two calibrators (C.f.a.s. Proteins, C.f.a.s. PAC) and two control sets (PreciControl ClinChem Multi 1/2, Precinorm Protein/Precipath Protein). These are liquid or lyophilized serum-based products used on Roche/Hitachi and COBAS INTEGRA clinical chemistry analyzers. Primary modification involves replacing human-derived Antistreptolysin O source material with sheep-derived material to mitigate human-sourcing conflicts. Calibrators establish target values for quantitative assays; controls monitor accuracy and precision of these assays. Healthcare providers use analyzer-generated results to assess patient samples. Benefits include standardized calibration and reliable quality control for clinical chemistry testing, ensuring consistent diagnostic performance.

Clinical Evidence

Bench testing only. Submitter provided design control activities summary, risk analysis, and verification/validation data demonstrating that modified devices meet predetermined acceptance criteria.

Technological Characteristics

Serum-based calibrators and controls (liquid or lyophilized). Antistreptolysin O derived from sheep serum. Designed for use with Roche/Hitachi and COBAS INTEGRA clinical chemistry analyzers. Traceable to reference methods/materials via master lots. Stability maintained at 2-8°C.

Indications for Use

Indicated for use as calibrators (C.f.a.s. Proteins, C.f.a.s. PAC) or quality control materials (PreciControl ClinChem Multi 1/2, Precinorm Protein, Precipath Protein) for quantitative Roche clinical chemistry assays. Prescription use only.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER k133330 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared devices. | Predicate Device | 510(k) number | | --- | --- | | C.f.a.s. Proteins | k080607 | | C.f.a.s. PAC | k040245 | | PreciControl ClinChem Multi 1 and 2 | k102016 | | Precinorm Protein and Precipath Protein | k981401 | 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. The changes were: 1) The source of the Antistreptolysin O was changed from human to sheep for all four devices: C.f.a.s. (Calibrator for automated systems) Proteins, C.f.a.s. PAC (Prealbumin- Antistreptolysin-Ceruloplasmin), PreciControl ClinChem Multi 1 and 2 (PCCC), and Precinorm Protein (PNP) and Precipath Protein (PPP). 2) For the C.f.a.s. PAC device, Ceruloplasmin and Prealbumin were removed. 3) For the PCCC device, Ferritin was added. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, constituents, and performance characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. {1} Page 2 of 2 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...