PARAMAX AMMONIA AND AMMONIA BLANK REAGENT
Device Facts
| Record ID | K964565 |
|---|---|
| Device Name | PARAMAX AMMONIA AND AMMONIA BLANK REAGENT |
| Applicant | Dade Intl., Inc. |
| Product Code | JIF · Clinical Chemistry |
| Decision Date | Nov 22, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1065 |
| Device Class | Class 1 |
Indications for Use
The Paramax® Ammonia and Ammonia Blank Reagent are intended for use as quantitative determination of Ammonia and Ammonia Blank in plasma.
Device Story
In vitro diagnostic reagent; utilizes GLDH enzymatic method for quantitative ammonia measurement in plasma; used on Paramax® clinical chemistry systems; provides quantitative results to clinicians for assessment of ammonia levels; aids in diagnosis and monitoring of metabolic conditions; performance validated via correlation with predicate system.
Clinical Evidence
Bench testing only. Comparison study of 197 plasma samples between Paramax® and DuPont aca systems. Correlation coefficient 0.986; regression equation: Paramax® = (0.980 * aca) - 8.8.
Technological Characteristics
GLDH enzymatic reagent system for quantitative ammonia determination. Designed for use on automated clinical chemistry analyzers.
Indications for Use
Indicated for the quantitative determination of ammonia and ammonia blank in human plasma.
Regulatory Classification
Identification
An ammonia test system is a device intended to measure ammonia levels in blood, serum, and plasma, Ammonia measurements are used in the diagnosis and treatment of severe liver disorders, such as cirrhosis, hepatitis, and Reye's syndrome.
Predicate Devices
- DuPont aca Ammonia and Ammonia Blank Reagent (K770363)