← Product Code [JHI](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHI) · K132834

# FASTEP S10 HCG SERUM/URINE COMBO TEST (K132834)

_Polymed Therapeutics, Inc. · JHI · Jan 9, 2014 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHI/K132834

## Device Facts

- **Applicant:** Polymed Therapeutics, Inc.
- **Product Code:** [JHI](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHI.md)
- **Decision Date:** Jan 9, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1155
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Fastep S10 hCG Serum/Urine Combo Test is a rapid visual immunoassay for the qualitative, presumptive detection of human chorionic gonadotropin in human urine or serum specimens. This kit is intended for use as an aid in early detection of pregnancy. This product is only intended for prescription use in clinical laboratories and is not intended for point-of-care use settings.

## Device Story

The Fastep S10 hCG Serum/Urine Combo Test is a lateral flow chromatographic immunoassay for the qualitative detection of hCG in human urine or serum. The device utilizes a strip or cassette format containing a membrane pre-coated with goat anti-hCG polyclonal antibody and an antibody-dye conjugate (mouse anti-beta hCG monoclonal antibody). When the absorbent end is immersed in a specimen, capillary action draws the sample across the membrane. If hCG is present at or above the cutoff (10 mIU/mL for serum, 20 mIU/mL for urine), a colored test line forms. A control line indicates valid test performance. The device is intended for use by clinical laboratory professionals. Results are visually interpreted by the clinician to aid in the early detection of pregnancy. The test provides a rapid, presumptive diagnosis, assisting in clinical decision-making regarding pregnancy status.

## Clinical Evidence

Bench testing only. Method comparison study performed with 100 urine and 100 serum samples from 100 women, comparing candidate device to predicate. Results showed high agreement across all formats (strip/cassette). Analytical performance included precision/reproducibility studies across 3 lots, 3 operators, and 5 days. Specificity testing confirmed no cross-reactivity with LH, FSH, or TSH. Interference studies evaluated various exogenous compounds, pH, and specific gravity. High-dose hook effect testing showed no interference up to 2,000,000 mIU/mL hCG.

## Technological Characteristics

Lateral flow sandwich immunochromatographic assay. Formats: strip and cassette. Detection reagent: colloidal gold. Cut-off: 10 mIU/mL (serum), 20 mIU/mL (urine). Storage: 4-30°C. Shelf life: 24 months. No electronic components or software.

## Regulatory Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

## Predicate Devices

- QuickVue One-Step hCG Combo test ([K020801](/device/K020801.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k132834

B. Purpose for Submission:
New device

C. Measurand:
Human chorionic gonadotropin (hCG)

D. Type of Test:
Qualitative chromatographic immunoassay

E. Applicant:
Polymed Therapeutics, Inc.

F. Proprietary and Established Names:
Fastep S10 hCG Serum/Urine Combo Test

G. Regulatory Information:
|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JHI | Class II | 21 CFR 862.1155, Human Chorionic Gonadotropin (HCG) test system | 75-Chemistry  |

H. Intended Use:

1. Intended use(s):
See Indications for Use below.

2. Indication(s) for use:
The Fastep S10 hCG Serum/Urine Combo Test is a rapid visual immunoassay for the qualitative, presumptive detection of human chorionic gonadotropin in human urine or

{1}

serum specimens. This kit is intended for use as an aid in early detection of pregnancy.

This product is only intended for prescription use in clinical laboratories and is not intended for point-of-care use settings.

3. Special conditions for use statement(s):

For prescription use only in clinical laboratories

4. Special instrument requirements:

Not applicable,

I. Device Description:

The Fastep S10 hCG Serum/Urine Combo Test measures the presence of the hormone Human Chorionic Gonadotropin (hCG) in human urine or serum to aid in the early detection of pregnancy. The test devices are in two different formats: Strip and Cassette. Results are read visually by the end user. For a positive result, a colored line appears in the test and control regions of the device. For a negative result, a colored line appears in the control region, but no colored line appears in the test region of the device.

J. Substantial Equivalence Information:

1. Predicate device name(s):

QuickVue One-Step hCG Combo Test

2. Predicate 510(k) number(s):

k020801

3. Comparison with predicate:

Similarities

|  Item | Candidate Device: Fastep S10 hCG Serum/Urine Combo Test (k132834) | Predicate: QuickVue One-Step hCG Combo Test (k020801)  |
| --- | --- | --- |
|  Intended Use | Rapid qualitative detection of hCG to aid in the early detection of pregnancy. | Same  |
|  Specimen | Urine or serum | Same  |
|  Principle | Lateral flow Sandwich Immunochromatographic Assay | Same  |
|  Detection reagent | Colloidal gold | Same  |
|  Read time | Serum: 5 minutes
Urine: 3 minutes | Same  |

{2}

3

|  Cut-Off Values | 10 mIU/mL for serum and 20 mIU/mL for urine | Same  |
| --- | --- | --- |
|  Configurations | Strip and cassette | Same  |

## Differences

|   | Candidate Device: Fastep S10 hCG Serum/Urine Combo Test (k132834) | Predicate: QuickVue One-Step hCG Combo Test (k020801)  |
| --- | --- | --- |
|  Intended Population | For prescription use in clinical laboratories only | For prescription use in point of care sites  |

## K. Standard/Guidance Document Referenced (if applicable):

None referenced

## L. Test Principle:

The Fastep S10 hCG Serum/Urine Combo Test is a lateral flow chromatographic immunoassay. When the absorbent end of the test is immersed into a serum or urine sample, the sample is absorbed into the device by capillary action and mixes with the antibody-dye conjugate (mouse anti-beta hCG monoclonal antibody), flowing across the pre-coated (goat anti hCG polyclonal antibody) membrane. At analyte concentrations above the target cut off, it produces a colored test line that indicates a positive result. When analyte concentrations are below the cutoff, no colored band shows in the test region, indicating a negative result.

One colored line appears in the control region “C” regardless of the presence of hCG. The presence of this colored line in the “C” region serves as verification for sufficient volume and proper flow and as a control for the reagents. No line in the “C” region indicates that the test is invalid.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

Negative serum and urine specimens were spiked with varying hCG (commercially available and traceable to the 4th WHO international Standard) concentrations. The spiked samples were measured in 10 replicates using 3 different lots for each format. Tests were performed by three different operators for each lot in 2 runs per day for 5 days. Results are shown in the following tables:

{3}

Serum Strip format

|  hCG Concentration (mIU/mL) | Overall % agreement | Lot 1 |   | Lot 2 |   | Lot 3  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  - | + | - | + | - | +  |
|  0 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  2 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  4 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  6 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  8 | 91% | 28 | 2 | 27 | 3 | 27 | 3  |
|  9 | 71% | 23 | 7 | 20 | 10 | 21 | 9  |
|  10 | 97% | 1 | 29 | 2 | 28 | 0 | 30  |
|  12 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  14 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  16 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  18 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  20 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  50 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  100 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  150 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |

Serum Cassette format

|  hCG Concentration (mIU/mL) | Overall % agreement | Lot 1 |   | Lot 2 |   | Lot 3  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  - | + | - | + | - | +  |
|  0 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  2 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  4 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  6 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  8 | 90% | 27 | 3 | 28 | 2 | 26 | 4  |
|  9 | 74% | 24 | 6 | 21 | 9 | 22 | 8  |
|  10 | 96% | 1 | 29 | 1 | 29 | 2 | 28  |
|  12 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  14 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  16 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  18 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |

{4}

5

|  20 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  50 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  100 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  150 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |

## Urine Strip format

|  hCG Concentration (mIU/mL) | Overall % agreement | Lot 1 |   | Lot 2 |   | Lot 3  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  - | + | - | - | + | -  |
|  0 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  5 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  10 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  12 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  14 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  16 | 88% | 25 | 5 | 28 | 2 | 26 | 4  |
|  18 | 66% | 21 | 9 | 19 | 11 | 19 | 11  |
|  20 | 96% | 1 | 29 | 2 | 28 | 1 | 29  |
|  22 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  25 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  30 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  35 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  50 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  75 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  100 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  250 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |

## Urine Casette format

|  hCG Concentration (mIU/mL) | Overall % agreement | Lot 1 |   | Lot 2 |   | Lot 3  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  - | + | - | - | + | -  |
|  0 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  5 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  10 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  12 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  14 | 100% | 30 | 0 | 30 | 0 | 30 | 0  |
|  16 | 80% | 24 | 6 | 26 | 4 | 22 | 8  |

{5}

6

|  18 | 57% | 19 | 11 | 17 | 13 | 15 | 15  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  20 | 98% | 1 | 29 | 0 | 30 | 1 | 29  |
|  22 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  25 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  30 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  35 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  50 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  75 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  100 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |
|  250 | 100% | 0 | 30 | 0 | 30 | 0 | 30  |

b. Linearity/assay reportable range:

Not applicable. This is a qualitative test.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

This test is calibrated with reference material traceable to the World Health Organization (WHO) 4th International Standard for Chorionic Gonadotropin (75/589).

The stability testing protocol and acceptance criteria used to support the shelf life were reviewed and found to be acceptable. The sponsor claims a 24-month shelf life for the test when stored at 39-86°F (4-30°C).

d. Detection limit

See section M.1.a above. The sponsor claims that the cut-off values of serum and urine are 10 mIU/mL and 20 mIU/mL hCG, respectively.

e. Analytical specificity:

A cross-reactivity study was performed by adding known amounts of LH, FSH and TSH to negative and positive samples using three different lots by three operators. Results are shown in the following table:

|  Negative Sample Spiked with | Operator A | Operator | Operator C  |
| --- | --- | --- | --- |
|  LH (300mIU/mL) | — | — | —  |
|  FSH (1000mIU/mL) | — | — | —  |
|  TSH (1000μIU/mL) | — | — | —  |

{6}

|  Positive Sample Spiked with | Operator A | Operator | Operator C  |
| --- | --- | --- | --- |
|  LH (300mIU/mL) | + | + | +  |
|  FSH (1000mIU/mL) | + | + | +  |
|  TSH (1000μIU/mL) | + | + | +  |

No cross-reactivity was observed for either urine or serum samples at the concentrations that were tested for LH, FSH and TSH.

To evaluate potential interference from certain exogenous compounds, each interferent was made at 100X concentrate bulk and spiked into both hCG free and hCG positive samples. Each spiked sample was mixed for 5 minutes to ensure a homogeneous solution before testing. Each sample was tested using 3 different lots of the testing kit. The sponsor states that none of the following substances at the stated concentrations interfere with the assay.

|  Interferents | Concentration | Interferents | Concentration  |
| --- | --- | --- | --- |
|  Acetaminophen | 20mg/dL | Methanol | 10%  |
|  Acetoacetic Acid | 2000mg/dL | Albumin | 2000mg/dL  |
|  Asorbic Acid | 20mg/dL | Glucose | 2000mg/dL  |
|  B-hydroxybutyrate | 2000mg/dL | Bilirubin | 2mg/dL  |
|  Caffeine | 20mg/dL | Atropine | 20mg/dL  |
|  Ephedrine | 20mg/dL | Estriol-17-beta | 1400ug/dL  |
|  Gentisic Acid | 20mg/dL | Hemoglobin | 500mg/dL  |
|  Phenylpropanolamine | 20mg/dL | Pregnanediol | 1500ug/dL  |
|  Salicylic Acid | 20mg/dL | Thiophene | 20mg/dl  |
|  Phenothiazine | 20mg/dL | Ampicillin | 20mg/dl  |
|  EDTA | 80mg/dL | Tetracycline | 20mg/dl  |
|  Acetylsalicylic Acid | 20mg/dL | Ketone | 20mg/dl  |
|  Benzoylecgonine | 10mg/dL | Codeine | 6ug/dL  |
|  Cannabinol | 10mg/dL | Ethanol | 1.00%  |

A pH study was performed to evaluate the device. Results demonstrate that urine and serum samples with pH 4.0 - 9.0, do not affect the results.

A specific gravity study was performed to evaluate the device. Results demonstrate that urine samples are not affected when the specific gravity range is from 1.00 to 1.035.

Interference studies to hCG  $\beta$ -core fragment were also performed in samples containing intact hCG in the presence of  $\beta$ -core fragment at concentrations of  $1.0 \times 10^{6}$  picomoles. The results indicate that high levels of hCG  $\beta$ -core fragment ( $\geq 500$  pmol/L) can lead to a positive result for hCG-free samples (in both serum and urine

{7}

formats). No false negative results were identified for the sample with 1.0 x 10⁶ picomoles of beta core.

A high dose hook effect study was performed by spiking high levels of hCG concentrations 62,500-2,000,000 mIU/mL (62,500, 125,000, 250,000, 500,000, 1,000,000 and 2,000,000) into negative urine and serum samples and evaluating the test result lines. No hook effect was observed up to 2,000,000 mIU/mL hCG.

f. Assay cut-off:

See section M.1.a above. The sponsor claims that the cut-off values of serum and urine are 10 mIU/mL and 20 mIU/mL hCG, respectively.

2. Comparison studies:

a. Method comparison with predicate device:

A method comparison study was performed, comparing the results obtained from the Fastep S10 hCG Serum/Urine Combo Test to the results from the predicate device. 100 urine and 100 serum samples were collected from 100 women (about half of whom were less than 5 weeks pregnant). All samples were randomly masked prior to analysis. Samples were tested by three different health professionals with the candidate device (using three different lots) and the predicate device. Testing was performed according to the labeling. The results are shown in the tables below.

*All discrepant results were from samples from women subjects who were early in pregnancy (less than 5 weeks pregnant based on day of last menstrual period) or were from samples with hCG concentrations around the cut-off of the device.

Summary Results for Urine Strip
Predicate Device

|  New Device |  | + | -  |
| --- | --- | --- | --- |
|   |  + | 47 | 0  |
|   |  - | 2* | 51  |

Summary Results for Urine Cassette
Predicate Device

|  New Device |  | + | -  |
| --- | --- | --- | --- |
|   |  + | 48 | 0  |
|   |  - | 1* | 51  |

Summary Results for Serum Strip
Predicate Device

|  New Device |  | + | -  |
| --- | --- | --- | --- |
|   |  + | 48 | 0  |
|   |  - | 1* | 51  |

{8}

9

Summary Results for Serum Cassette
Predicate Device

|  New Device | Cleared device | + | -  |
| --- | --- | --- | --- |
|   |  + | 48 | 0  |
|   |  - | 1* | 51  |

b. Matrix comparison:
Not applicable.

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
The labeling states “hCG concentration in pregnant women rises very rapidly after implantation, reaching a peak concentration in excess of 200IU/mL about 2-3 months after the last menstrual period. The Fastep S10 hCG Serum/Urine Combo Test Device (Urine/Serum) has a cut-off level of 10mIU/ml in serum and 20mIU/ml in urine. Reportedly, a level of 10-25 mIU/mL or more, is present 7-10 days after conception. Patients suspected to be pregnant but showing negative test results should be re-tested with first morning specimens obtained 48-72 hours later.”

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHI/K132834](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/JHI/K132834)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
