← Product Code [GIM](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/GIM) · K971449

# VACUTAINER BRAND PLUS TUBE WITH EDTA ANTICOAGULANT(MULTIPLE) (K971449)

_Becton Dickinson Vacutainer Systems · GIM · Jun 17, 1997 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/GIM/K971449

## Device Facts

- **Applicant:** Becton Dickinson Vacutainer Systems
- **Product Code:** [GIM](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/GIM.md)
- **Decision Date:** Jun 17, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1675
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The VACUTAINER® Brand PLUS (plastic) Tube with EDTA is an evacuated blood collection tube which provide a means of collecting, transporting, separating and processing blood in a plastic tube. When the tube is used together with VACUTAINER® Brand Needles and Holders, it is a closed system for the collection of venous blood with the same indications as described herein. Blood collected in a tube containing EDTA Anticoagulant is primarily used for clinical laboratory testing-hematology study using whole blood but may be used for other studies including such studies as testing for lead and FEP analyses, requiring whole blood as determined by the laboratory.

## Device Story

Evacuated blood collection tube; PET plastic construction; spray-coated with dipotassium (K2) EDTA anticoagulant to prevent coagulation. Used in clinical laboratory settings by healthcare professionals for venous blood collection. Provides closed system when used with VACUTAINER needles/holders; reduces risk of tube breakage and exposure to blood compared to glass tubes. Facilitates transport and processing of whole blood specimens for hematology, lead, and FEP analysis. Enhances safety during disposal via incineration.

## Clinical Evidence

Clinical evaluation compared VACUTAINER® Brand Plus plastic tubes to glass tubes for FEP and blood lead analysis. Results demonstrated analytical and statistical equivalence in performance and sample stability. Bench testing confirmed tube lead levels were <2 ug/L, supporting a 2.5 ug/L limit claim, well below CDC toxicity levels.

## Technological Characteristics

PET (Polyethylene terephthalate) plastic evacuated tube; spray-coated with dipotassium (K2) EDTA; rubber stopper or HEMOGARD™ closure assembly. Designed for venous blood collection. No electronic components or software.

## Regulatory Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

## Predicate Devices

- VACUTAINER® Brand Tube with EDTA Anticoagulant (Pre-Amendment Device)

## Reference Devices

- VACUTAINER® Brand Plus Tube with EDTA Anticoagulant ([K901449](/device/K901449.md)/A)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K911449

JUN 17 1997

Image /page/0/Picture/2 description: The image shows the number 100 with a globe in the middle. The globe is black and white and shows the continents of North America, Europe, and Asia. The number 1 is on the left side of the globe, and the two zeros are on the right side of the globe. The numbers are black and white and have a striped pattern.

CELEBRATING THE FIRST ONE HUNDRED: 1897-1997

Becton Dickinson VACUTAINER Systems 1 Becton Drive Franklin Lakes, New Jersey 07417 (201) 847-4500

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

### I. General Information

This Summary of 510(k) Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA of 1990 and 21 CFR 807.92

Establishment:

- Becton Dickinson VACUTAINER Systems · Address: 1 Becton Drive Franklin Lakes, NJ 07417-1885
- · Registration Number: 2243072
- Contact Person: John A. Schalago . Regulatory Affairs Specialist Telephone no .: 201 - 847 - 6280 Facsimile no .: 201 - 847 - 4858
	- April 16, 1997 · Date of Summary:

#### Device Name:

- VACUTAINER® Brand Plus Tube · Trade Name: with EDTA Anticoagulant
- Blood Specimen Collection Device . Classification Name :
- · Classification: Class II
- · Performance Standards: None Established under 514 of the Food, Drug and Cosmetic Act

## II. Safety and Effectiveness Information Supporting the Substantial Equivalence Determination · Device Description

The VACUTAINER® Brand PLUS Tube with EDTA is an evacuated plastic tube for collecting, transmitting and processing blood in a closed plastic tube. The blood collection tube consists of closure assembly, a plastic tube and EDTA coating (dipotassium, K2).

Image /page/0/Picture/20 description: The image shows the logo for Becton Dickinson. The logo is in black and white. The words "BECTON" and "DICKINSON" are stacked on top of each other. 

{1}------------------------------------------------

The plastic tube is manufactured from PET (Polyethylene terepthalate Plastic and enhances user safety and disposal because of the reduced risk of tube breakage and incineration as a method of disposal. The standard closure assembly is a basic rubber stopper. The tube is also available with VACUTAINER Systems Hemogard™ Closure assembly which is designed to reduce user exposure to blood. The EDTA anticoagulant tube coating is spray coated in a dipotassium (K2) form. The EDTA prevents specimen coagulation.

### ● Intended Use

The VACUTAINER® Brand PLUS Tube with EDTA is an evacuated blood collection tube which provides a means of collecting, transporting and processing blood in a plastic tube. Blood collected in a tube containing EDTA anticoagulant is used primarily for clinical laboratory hematology studies but may be used for other whole blood specimens, including but not limited to lead level testing and FEP level testing, as determined by the laboratory.

### · Synopsis of Test Methods and Results

The VACUTAINER® Brand Plus Tube with EDTA anticoagulant, as described above, is a plastic tube with closure assembly (rubber stopper or HEMOGARD™ Closure). The plastic tube provides enhanced user safety and disposal because of the reduced risk of tube breakage and the use of incineration as a method of disposal. Breakage reduction and enhanced disposal studies and test results were described 510(k) Premarket Notification K901449/A which received FDA clearance on August 9, 1990.

Clinical evaluation of the VACUTAINER® Brand Plus tube with EDTA Anticoagulant was performed to demonstrate device equivalence for FEP and Blood Lead analyses. The VACUTAINER® Brand Plus Plastic Tube with EDTA was compared to VACUTAINER® Brand Glass Tube with EDTA, at initial time and extended tube storage. The VACUTAINER® Brand PLUS Tube with EDTA demonstrated analytically and statistically equivalent results and sample stability. Therefore, Clinical evaluation supports the Use of the VACUTAINER® Brand PLUS Tube with EDTA for FEP and Blood Lead analyses.

Additionally, bench testing of three different lots of VACUTAINER® Brand Plus Tube with EDTA was performed to determine the tube lead levels. As identified in the draft product labeling, the maximum limit is 2.5 ug/L (ppb). This maximum lead limit for the Plus plastic tube is significantly lower than the current Center for Disease Control and Prevention decision level of 10 ug/dL or 100 ug/L for toxicity in children. Twenty tubes from each of the three lots were tested according to internal test procedures. The Lead levels were evaluated using Atomic Absorption (AA). The calibration standards used to calibrate the AA had lead concentration levels of 2 and 4 ug/L. The results demonstrated that lead level of all tubes from the three different lots were below 2 ug/L and therefore, support the labeling claim of 2.5 ue/L. The test data and results are maintained on file at Becton Dickinson VACUTAINER Systems, Franklin Lakes, NJ.

{2}------------------------------------------------

### · Substantial Equivalence

Becton Dickinson VACUTAINER Systems believes that the VACUTAINER® Brand PLUS Tube with EDTA with the expanded Indications for Use is substantially equivalent to a commercially available blood collection tube. Clinical testing, as described in this premarket notification, demonstrates equivalent performance and effectiveness and supports the determination of substantial equivalence. The predicate devices, manufacturer, K number and clearance date are identified below:

| Manufacturer          | Predicate Device                                     | K-Number          | Clearance Date                                                                                                              |
|-----------------------|------------------------------------------------------|-------------------|-----------------------------------------------------------------------------------------------------------------------------|
| VACUTAINER<br>Systems | VACUTAINER®<br>Brand Tube with<br>EDTA Anticoagulant | Not<br>Applicable | Pre-Amendment Device and,<br>therefore, exempt from premarket<br>notification requirements according to<br>the MDA of 1976. |

Also included in this premarket notification, is a description of an incremental packaging modification which occurred prior to the submission of this premarket notification. VACUTAINER Systems determined that the incremental modification did not significantly affect the safety or efficacy of the VACUTAINER® Brand Tube and Brand PLUS Tube, and therefore, did not require a 510(k) premarket notification.

Oshkoshlago

ohn A. Schalago Regulatory Affairs Specialist Regulatory Affairs Department April 16, 1997 Date

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle's head and neck. The eagle's head is facing to the right.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 17

John A. Schalago Requlatory Affairs Specialist Becton Dickinson VACUTIANER Systems 1 Becton Drive Franklin Lakes, New Jersey 07417

Re: K971449 VACUTAINER® Brand PLUS (Plastic) Blood Collection Tube with EDTA Anticoagulant Requlatory Class: II Product Code: GIM Dated: April 16, 1997 Received: April 21, 1997

Dear Mr. Schalaqo:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

* If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".

Sincerely yours,
Steven Gutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: VACUTAINER® Brand PLUS Tube with EDTA Anticoagulant

Indications for Use:

The VACUTAINER® Brand PLUS (plastic) Tube with EDTA is an evacuated blood collection tube which provide a means of collecting, transporting, separating and processing blood in a plastic tube. When the tube is used together with VACUTAINER® Brand Needles and Holders, it is a closed system for the collection of venous blood with the same indications as described herein.

Blood collected in a tube containing EDTA Anticoagulant is primarily used for clinical laboratory testing-hematology study using whole blood but may be used for other studies including such studies as testing for lead and FEP analyses, requiring whole blood as determined by the laboratory.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use <span style="text-decoration: underline;">✓</span> OR | Over-the-Counter Use <span style="text-decoration: underline;">_____</span> |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| (Per 21 CFR 801.109)                                                   | (Optional Format 1-2-96)                                                    |
|                                                                        | (Division Sign-Off)                                                         |
|                                                                        | Division of Clinical Laboratory Devices                                     |
| 510(k) Number                                                          | K971449                                                                     |

510(K) Premarket Notification April 16, 1997 Page 45Page 45

(K) Premarket Notification VACUTAINER® Brand PLUS Tube with EDTA

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/GIM/K971449](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/GIM/K971449)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
