iFlash-HCG; Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C)

K223690 · Shenzhen Yhlo Biotech Co., Ltd. · DHA · Dec 11, 2023 · Clinical Chemistry

Device Facts

Record IDK223690
Device NameiFlash-HCG; Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C)
ApplicantShenzhen Yhlo Biotech Co., Ltd.
Product CodeDHA · Clinical Chemistry
Decision DateDec 11, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

iFlash-HCG is a paramagnetic particle chemiluminescent immunoassay (CLIA) for quantitative detection of the intact human chorionic gonadotropin (hCG) molecule and the hCG β-subunit (β-hCG) in human serum and plasma using the automated Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C). The iFlash-HCG assay is to be used by laboratory professionals as an aid in early detection of pregnancy together with other clinical methods. Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) is a fully-automated, chemiluminescence immunoassay analyzer intended for quantitative or qualitative determination of analytes in human body fluids taken from clinical settings. It is used together with its supporting chemiluminescence immunoassay reagents. The Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) is intended for use in clinical laboratories.

Device Story

iFlash-HCG is a quantitative sandwich chemiluminescent immunoassay (CLIA) for human serum/plasma; utilizes anti-HCG antibody-coated paramagnetic microparticles and acridinium-ester-labeled anti-HCG antibody conjugate. Sample hCG forms a sandwich complex with reagents; magnetic separation removes unbound materials; pre-trigger and trigger solutions initiate chemiluminescent reaction. iFlash 3000-C analyzer measures resulting relative light units (RLUs) via optical system; concentration derived from 3-point calibration curve and master curve (QR code). Used in clinical laboratories by professionals; results aid clinical decision-making for pregnancy detection. System includes sample loading, incubation, magnetic separation, and photon detection modules. Barcode reader identifies specimens; automated dilution (1:50 or 1:100) available for high-concentration samples.

Clinical Evidence

Bench testing only. Performance validated via precision (CLSI EP05-A3), linearity (CLSI EP06-Ed2), interference (CLSI EP7-A2), and detection limits (CLSI EP17-A2). Method comparison study (n=110) against predicate showed high correlation (R²=0.9980, slope=0.9860). Matrix comparison (n=97) confirmed suitability of serum, lithium heparin, sodium heparin, and K2-EDTA plasma. No clinical sensitivity/specificity studies performed.

Technological Characteristics

Sandwich immunoassay; paramagnetic microparticles; acridinium-ester label; chemiluminescent detection. Fully automated analyzer (iFlash 3000-C) with incubation, magnetic separation, and photon detection. Reagents contain ProClin 300 preservative. Connectivity via barcode reader; calibration via QR code. Standards: CLSI EP05-A3, EP06, EP07, EP37, EP17-A2, EP09c, EP34, EP28-A3c.

Indications for Use

Indicated for laboratory professionals to aid in early pregnancy detection via quantitative measurement of intact hCG and hCG ß-subunit in human serum and plasma.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K223690 - Page 1 of 11 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT ## I Background Information: A 510(k) Number K223690 B Applicant Shenzhen YHLO Biotech Co., LTD. C Proprietary and Established Names iFlash-HCG; Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | DHA | Class II | 21 CFR 862.1155 - Human Chorionic Gonadotropin (HCG) Test System | CH - Clinical Chemistry | | JJE | Class I | 21 CFR 862.2160 - Discrete photometric chemistry analyzer for clinical use | CH - Clinical Chemistry | ## II Submission/Device Overview: A Purpose for Submission: New Device B Measurand: Total beta human chorionic gonadotropin (intact hCG) and $\beta$-hCG. C Type of Test: Quantitative chemiluminescent immunoassay {1} K223690 - Page 2 of 11 # III Intended Use/Indications for Use: ## A Intended Use(s): See Indications for Use below. ## B Indication(s) for Use: iFlash-HCG is a paramagnetic particle chemiluminescent immunoassay (CLIA) for quantitative detection of the intact human chorionic gonadotropin (hCG) molecule and the hCG β-subunit (β-hCG) in human serum and plasma using the automated Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C). The iFlash-HCG assay is to be used by laboratory professionals as an aid in early detection of pregnancy together with other clinical methods. Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) is a fully-automated, chemiluminescence immunoassay analyzer intended for quantitative or qualitative determination of analytes in human body fluids taken from clinical settings. It is used together with its supporting chemiluminescence immunoassay reagents. The Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) is intended for use in clinical laboratories. ## C Special Conditions for Use Statement(s): Rx - For Prescription Use Only ## D Special Instrument Requirements: Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) # IV Device/System Characteristics: ## A Device Description: The iFlash hCG is provided in kits comprised of 100 tests, 2 packs, 50 tests/pack, Catalog # C86012 contain the following: | R1 (2 x 3.5 mL) | Anti-HCG antibody-coated paramagnetic micro-particles (mouse, 0.1 mg/mL), 0.05% ProClin 300 as preservative. | | --- | --- | | R2 (2 x 4.0 mL) | Acridinium-ester-labeled anti-HCG antibody conjugate (mouse, 0.4 μg/mL), 0.1% ProClin 300 as preservative. | | CAL1 (1 x 1.0 mL) | Calibrator 1, TRIS buffer with protein stabilizer (BSA), 0.05% ProClin 300 as preservative, lyophilized. | | CAL2 (1 x 1.0 mL) | Calibrator 2, HCG protein (human) in TRIS buffer with protein stabilizer (BSA), 0.05% ProClin 300 as preservative, lyophilized. | | CAL3 (1 x 1.0 mL) | Calibrator 3, HCG protein (human) in TRIS buffer with protein stabilizer (BSA), 0.05% ProClin 300 as preservative, lyophilized. | Chemiluminescence immunoassay analyzer (Model: iFlash 3000-C) is a fully- automated, analyzer intended for quantitative or qualitative determination of analytes in human body fluids. It is intended for use in clinical laboratories. The Chemiluminescence immunoassay analyzer (Model: iFlash 3000-C) measures analytes in samples using the reagents, calibrators, quality {2} control materials and other consumables that can be in the iFlash-HCG instruction for use manual. Components of the Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) include the components in the sample/reagent area, components in the consumables area, and components in the measurement area: Components in the sample/reagent area include: - Sample loading unit - Sampling unit - Reagent processing system Components in the consumables area include: - Reaction vessel system - System reagent bin (Pre-trigger Solution and Trigger Solution) - Wash buffer tank - Waste tank - Solid waste container Components in the measurement area include: - Incubation assembly - Magnetic separation washing assembly - Photon detection system ## B Principle of Operation: The iFlash-HCG assay is a sandwich immunoassay using a direct chemiluminometric technique. The hCG in the sample reacts with anti-HCG antibody-coated paramagnetic microparticles and acridinium-labeled anti-HCG antibody conjugate to form a sandwich complex. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then the pre-trigger and trigger solutions are added to the reaction mixture. The resulting chemiluminescent reaction is measured as relative light units (RLUs) by the iFlash optical system. The RLUs detected production is directly proportional to the analyte in the sample. Analyte concentration is determined from a calibration curve, which is generated by 3-point calibration and a master curve provided via the reagent QR code. ## C Instrument Description Information: 1. Instrument Name: Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) 2. Specimen Identification: Human serum and lithium heparin, sodium heparin and K₂-EDTA specimens are identified using the built-in barcode reader. 3. Specimen Sampling and Handling: Specimen sampling and handling procedures are described in the Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) Analyzer’s instruction for use manual. K223690 - Page 3 of 11 {3} Assay specific specimen collection and preparation procedure can be found in the iFlash-HCG instruction for use manual. 4. Calibration: Calibration methods and procedures are described in the Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) Analyzer's instruction for use manual. Assay specific specimen collection and preparation procedure can be found in the iFlash-HCG instruction for use manual. 5. Quality Control: YHLO Immunoassay Multi Control, EVER-BLUE Immunoassay Multi Control are recommended. V Substantial Equivalence Information: A Predicate Device Name(s): Elecsys Hcg And Beta Test System, Elecsys TSH assay, cobas e 801 Immunoassay analyzer B Predicate 510(k) Number(s): K003178, K162606 C Comparison with Predicate(s): Assay | Device & Predicate Device(s): | K223690 | K003178 | | --- | --- | --- | | Device Trade Name | iFlash-HCG | Roche Elecsys HCG and Beta Test System | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | For the quantitative determination of intact human chorionic gonadotropin (hCG) and the hCG β-subunit (β-hCG) in human serum and plasma to aid in the early detection of pregnancy | Same | | Type of Test | Quantitative | Same | | Technology | Sandwich Immunoassay | Same | | Method | Automated | Same | | Specimen Type | Human Serum and Plasma | Same | K223690 - Page 4 of 11 {4} | General Device Characteristic Differences | | | | --- | --- | --- | | Assay Range | 0.5-10000 mIU/mL | 0.2 – 10000 mIU/mL | | Calibrator | Three-level Calibrators | Five-level Calibrators | | Traceability | WHO International Standard 5th WHO IS Chorionic Gonadotrophin 07/364 | 4th International Standard (NIBSC) code 75/589 | Instrument | Device & Predicate Device(s): | K223690 | K162606 | | --- | --- | --- | | Device Trade Name | Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) | Cobas e 801 analyzer | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | For quantitative or qualitative determination of analytes in human body fluids. | Same | | Calibration | Utilizes a stored calibration curve | Same | | General Device Characteristic Differences | | | | Detection System | Chemiluminescent | Electrochemiluminescent | VI Standards/Guidance Documents Referenced: - CLSI EP05- A3 Evaluation of Precision of Quantitative Measurement Methods-Approved Guideline-Third Edition - CLSI EP06 2nd Edition – Evaluation of the Linearity of Quantitative Measurement Procedures - CLSI EP07 3rd Edition – Interference Testing in Clinical Chemistry - CLSI EP37 1st Edition – Supplemental Tables for Interference Testing in Clinical Chemistry - CLSI EP17-A2 – Evaluation of Detection capability for Clinical Laboratory Measurement Procedures; Approved Guideline 2nd Edition - CLSI EP09c 3rd Edition – Measurement Procedure Comparison and Bias Estimation Using Patient Samples - CLSI EP34 1st Edition – Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking - CLSI EP28-A3c – Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, Third Edition K223690 - Page 5 of 11 {5} VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Precision of the iFlash-HCG Assay was evaluated according to CLSI EP05-A3 using nine serum samples and two levels of controls. Each sample was measured in two replicates per run with two runs per day over 20 days by three different operators at three different sites using three lots of the reagents on three different Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) resulting in a total of 240 test results per level. Results from one analyzer are provided in the table below. | Sample | Mean (mIU/mL) | Repeatability | | Between-Run | | Between-Day | | Total imprecision | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | N=240 | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | Sample 1 | 0.57 | 0.028 | 4.9 | 0.015 | 2.6 | 0.013 | 2.3 | 0.041 | 7.2 | | Sample 2 | 4.68 | 0.199 | 4.3 | 0.072 | 1.5 | 0.045 | 1.0 | 0.216 | 4.6 | | Sample 3 | 14.9 | 0.589 | 3.9 | 0.078 | 0.5 | 0.172 | 1.2 | 0.978 | 6.6 | | Sample 4 | 23.5 | 0.838 | 3.6 | 0.269 | 1.1 | 0.116 | 0.5 | 0.914 | 3.9 | | Sample 5 | 91.5 | 1.98 | 2.2 | 0.213 | 0.2 | 0.382 | 0.4 | 4.18 | 4.6 | | Sample 6 | 957 | 13.9 | 1.5 | 6.61 | 0.7 | 6.31 | 0.7 | 22.5 | 2.4 | | Sample 7 | 2823 | 34.8 | 1.2 | 0.000 | 0.0 | 14.3 | 0.5 | 46.6 | 1.6 | | Sample 8 | 5627 | 63.1 | 1.1 | 26.5 | 0.5 | 11.9 | 0.2 | 86.0 | 1.5 | | Sample 9 | 9023 | 100 | 1.1 | 33.8 | 0.4 | 14.1 | 0.2 | 138 | 1.5 | | Control 1 | 2.71 | 0.127 | 4.7 | 0.045 | 1.7 | 0.029 | 1.1 | 0.200 | 7.4 | | Control 2 | 101 | 1.96 | 1.9 | 0.486 | 0.5 | 0.620 | 0.6 | 5.44 | 5.4 | 2. Linearity: Linearity of the iFlash-HCG Assay was evaluated following recommendations in CLSI EP06-Ed2. Three linearity studies were performed using eleven serum samples each. Varying concentrations of hCG were obtained by mixing low level serum samples and high level serum samples. All three studies were performed using three lots of reagents on one Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C). In the first study, linearity was established by testing samples with concentrations ranging from 0.43 to $11786.53\mathrm{mIU/mL}$. Each sample was tested in four replicates. A regression analysis was conducted resulting in the following linear regression equation: $y = 0.9824x + 38.719$, $R^2 = 0.9994$ In the second study, mid-range linearity was established by testing samples with hCG concentrations ranging from 0.45 to $1,050.30\mathrm{mIU/mL}$. Each sample was tested in four K223690 - Page 6 of 11 {6} replicates. A regression analysis was conducted resulting in the following linear regression equation: $y = 0.9994x - 0.545$ , $R2 = 0.9996$ In the third study, low-range linearity was established by testing samples with concentrations ranging from 0.46 to $128.02\mathrm{mIU / mL}$ . Each sample level was tested in four replicates. A regression analysis was conducted resulting in the following linear regression equation: $\mathrm{y} = 1.014\mathrm{x} - 2.857$ , $\mathrm{R2} = 0.9995$ The results from the linearity studies and limit of quantitation studies (LOQ = 0.50 ng/mL, see below), supports the sponsor's claimed analytical measuring interval of 0.50 to 10,000 mIU/mL. # Dilution Studies Dilution studies were performed to determine the sample recovery after a 1:100 dilution either manually or automatically by the Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C). The dilution study results support labeling claims that samples with hCG concentrations above $10,000\mathrm{mIU / mL}$ may be diluted 1:50 or 1:100 with sample diluent either manually or automatically on-onboard the analyzer. # 3. Analytical Specificity/Interference: # Interference The potential interference of certain exogenous and endogenous substances was assessed using female human serum pool samples containing approximately $6\mathrm{mIU / mL}$ and 5000 $\mathrm{mIU / mL}$ ("low" and "high" respectively) hCG spiked with various concentrations of substances based on recommendations in CLSI EP7-A2. The following endogenous substances do not interfere with this assay at the levels tested (less than $10\%$ bias between the test sample and the control sample). The following table lists the concentration of each endogenous substance at which no significant interference was found in the study. | Substance | Concentration | | --- | --- | | Bilirubin (Conjugated) | 40mg/dL | | Bilirubin (Unconjugated) | 40mg/dL | | Hemoglobin | 1000mg/dL | | Triglycerides | 3000 mg/dL | | Total protein | ≤10000 mg/dL | | Rheumatoid factors | 2000 IU/mL | | Human Anti-Mouse Antibodies (HAMA) | 600 ng/mL | | Antinuclear antibody (ANA) | 500 AU/mL | The following table lists the concentration of each exogenous substance at which no significant interference was found in the study. | Substance | Concentration | | --- | --- | | Phenylbutazone | 400 μg/mL | K223690 - Page 7 of 11 {7} | Aspirin (Acetylsalicylic Acid) | 1000 μg/mL | | --- | --- | | Acetaminophen | 200 μg/mL | | Ibuprofen | 500 μg/mL | | N-acetylcysteine | 150 μg/mL | | Methyldopa | 25 μg/mL | | Theophylline | 60 μg/mL | | Metformin | 12 μg/mL | | Isosorbide dinitrate | 6 μg/mL | | Rifampicin | 48 μg/mL | | Tetracycline hydrochloride | 24 μg/mL | | Cefoxitin | 6600 μg/mL | | Cyclosporine | 2 μg/mL | | Metronidazole | 125 μg/mL | | Ascorbic acid | 60 μg/mL | | Ampicillin-Na | 100 μg/mL | | Levodopa | 20 μg/mL | ## Cross-Reactivity Cross-reactivity of the iFlash-HCG assay was determined with three lots of reagent on one Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) using female human serum samples with approximately 0.1, 5, and 4000 mIU/mL ("low", "medium", and "high" respectively) hCG concentrations that were spiked with the following potential cross-reactants: luteinizing hormone (LH), follicle stimulating hormone (FSH), or thyroid stimulating hormone (TSH). Samples were tested in replicates of five using three lots of reagents on one iFlash 3000-C analyzer. The results demonstrated no cross-reactivity from potential cross-reactants up to 500 mIU/mL LH, 200 mIU/mL FSH, and 10000 mIU/mL TSH ## Hook Effect No high dose hook effect was observed for hCG concentrations in female human serum up to 1,250,000 mIU/mL hCG. 4. Assay Reportable Range: The reportable range is 0.50 – 10,000 mIU/mL. 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Traceability: This assay is traceable to the World Health Organization (WHO) 5th International Standard Chorionic Gonadotropin, Human, National Institute for Biological Standards and Control (NIBSC), code 07/364. K223690 - Page 8 of 11 {8} K223690 - Page 9 of 11 6. Detection Limit: The limit of blank (LoB), limit of detection (LoD) and limit of quantitation (LoQ) of the iFlash-HCG Assay were evaluated based upon recommendations in CLSI EP17-A2. Limit of Blank For determination of limit of blank (LoB), five analyte free samples were tested with three lots of reagents over at least three days on one Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) in four runs with five replicates per run over three days. In total 60 measured values of analyte free samples were obtained per lot. The LOB was determined to be 0.10 mIU/mL. Limit of Detection For determination of limit of detection (LoD), five serum samples containing low – analyte concentration (approximately 1-5 times LoB) were tested with three lots of reagents on one Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) in four runs with five replicates per run over three days. In total 60 measured values of samples with low analyte concentrations were obtained per lot. The LoD was determined to be 0.20 mIU/mL. Limit of Quantitation For determination of limit of quantitation (LoQ), five human serum samples covering the range between 0.30 and 70 mIU/mL were tested with three lots of reagents on one Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) in four runs with five replicates per run over three days. Limit of quantitation (LoQ) is the lowest concentration which is the design requirements of TEa%(Total Error Allowance) of ≤30%. The LoQ was determined to be 0.50 mIU/mL. 7. Assay Cut-Off: Not applicable; this is a quantitative assay. 8. Accuracy (Instrument): Not applicable. 9. Carry-Over: Samples with high hCG (1,250,000 mIU/mL) and low hCG (3.03 mIU/mL) hCG were tested alternating five times (i.e. HH,LL,HH,LL, etc.) in replicates of three, using one reagent lot on one Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C). No significant sample carryover effect was observed. B Comparison Studies: 1. Method Comparison with Predicate Device: Method comparison studies were conducted following recommendations in CLSI EP09-A3 Guideline. A total of 110 female human serum samples were tested in singlicate using iFlash-HCG Assay on Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) {9} (candidate device) and Elecsys HCG+β (comparator device). The regression results are shown in the following table: | Parameter | Passing Bablok | | --- | --- | | n | 110 | | Slope | 0.9860 | | 95% CI | 0.970-1.008 | | Intercept | -0.0470 | | 95% CI | 3.161 - 0.312 | | Correlation Coefficient (R2) | 0.9980 | | Sample Range | 0.531-9717 | # 2. Matrix Comparison: A matrix comparison study was performed to evaluate matrices suitable for use with the iFlash-HCG assay. Serum samples were compared to matched Lithium Heparin, Sodium Heparin, and $\mathrm{K}_2$ -EDTA plasma samples using three lots of reagents on one Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C). The results were assessed using Passing/Bablok regression analysis. Serum/Lithium-Heparin Plasma Summary Results | Parameter | Passing Bablok | | --- | --- | | n | 97 | | Sample Range (mIU/mL) | 0.62 - 8398 | | Slope | 0.988 | | Intercept | 0.189 | | Correlation Coefficient (R2) | 0.999 | Serum/Sodium Heparin Plasma Summary Results | Parameter | Passing Bablok | | --- | --- | | n | 97 | | Sample Range (mIU/mL) | 0.65 - 8952 | | Slope | 1.011 | | Intercept | -0.319 | | Correlation Coefficient (R2) | 0.999 | Serum/K2-EDTA Plasma Summary Results | Parameter | Passing Bablok | | --- | --- | | n | 97 | | Sample Range (mIU/mL) | 0.65 - 8952 | | Slope | 1.0005 | | Intercept | -0.0331 | | Correlation Coefficient (R2) | 0.990 | K223690 - Page 10 of 11 {10} K223690 - Page 11 of 11 C Clinical Studies: 1. Clinical Sensitivity: Not applicable. 2. Clinical Specificity: Not applicable. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable. D Clinical Cut-Off: Not applicable. E Expected Values/Reference Range: Serum samples from 130 healthy, non-pregnant premenopausal women and 125 healthy, postmenopausal women were tested using the iFlash-HCG on the Chemiluminescence Immunoassay Analyzer (Model: iFlash 3000-C) according to CLSI EP28-A3c. The reference interval was calculated using the nonparametric method according to CLSI C28-A3, and was determined to be 0.05 - 0.85 mIU/mL for women ≤ age 50, and 0.1 - 4.4 mIU/mL for women over age 50. The central 95% reference interval was established to be 0.6 mIU/mL for premenopausal women and 5.4 mIU/mL for post-menopausal women. It is recommended that each laboratory establish its own expected reference range for the specific population. F Other Supportive Instrument Performance Characteristics Data: Not applicable. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...