Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma.
Device Story
Elecsys® Estradiol Assay is an in vitro diagnostic immunoassay for quantitative estradiol measurement in human serum and plasma. Device uses competition principle; 50 µL sample incubated with biotinylated estradiol-specific antibody; streptavidin-coated microparticles and ruthenium-labeled estradiol derivative added. Microparticles magnetically captured on electrode surface; voltage application induces chemiluminescent emission measured by photomultiplier. Results calculated via instrument-specific calibration curve and reagent bar code master curve. Used in clinical laboratory settings on Elecsys® 2010 instrument. Provides quantitative estradiol levels to assist clinicians in endocrine assessment and patient management.
Clinical Evidence
Bench testing only. Precision evaluated using modified NCCLS protocols (N=60 per level); Elecsys showed improved precision (Total %CV 5.0-9.0%) compared to predicate (Total %CV 9.1-16.1%). Linearity established from 10-4600 pg/mL. Method comparison against predicate (N=74) yielded y=0.902x + 42.94, r=0.976. Interference testing performed for bilirubin, hemoglobin, lipemia, and biotin. Specificity cross-reactivity profiles provided for various steroids.
Technological Characteristics
Electrochemiluminescence immunoassay; uses streptavidin-coated microparticles and ruthenium complex (Tris(2,2'-bipyridyl)ruthenium(II)) label. Requires Elecsys® 2010 instrument. Calibration via 2-point calibration and reagent bar code master curve. Assay duration 18 minutes at 37 °C.
Indications for Use
Indicated for the in vitro quantitative determination of estradiol in human serum and plasma for clinical diagnostic purposes.
Regulatory Classification
Identification
An estradiol test system is a device intended to measure estradiol, an estrogenic steroid, in plasma. Estradiol measurements are used in the diagnosis and treatment of various hormonal sexual disorders and in assessing placental function in complicated pregnancy.
{0}
FEB 18 1997
K9105109
BOEHRINGER MANNHEIM CORPORATION
510(k) Summary
Introduction
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
1. Submitter name, address, contact
Boehringer Mannheim Corporation
2400 Bisso Lane
P.O. Box 4117
Concord, CA 94524-4117
(510) 674 - 0690 extension 8413
FAX (510) 687 - 1850
Contact Person: Yvette Lloyd
Date Prepared: December 13, 1996
2. Device name
Proprietary name: Elecsys® Estradiol Assay
Common name: Electrochemiluminescence assay for the determination of estradiol.
Classification name: System, Test, Estradiol
3. Predicate device
We claim substantial equivalence to the Enzymun-Test® Estradiol Assay (K916132).
Continued on next page
{1}
510(k) Summary, Continued
## 4. Device Description
Competition principle. Total duration of assay: 18 minutes, 37 °C.
- 1st incubation (9 minutes): By incubating the sample (50 µL) with an estradiol-specific biotinylated antibody (65 µL), an immunocomplex is formed, the amount of which is dependent upon the analyte concentration in the sample.
- 2nd incubation (9 minutes): After addition of streptavidin-coated microparticles (35 µL) and an estradiol derivative labeled with a ruthenium complex** (65 µL), the still-vacant sites of the biotinylated antibodies become occupied, with the formation of an antibody-hapten complex. The entire complex becomes bound to the solid phase via interaction of biotin and streptavidin.
- The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
- Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.
**Tris(2,2'-bipyridyl)ruthenium(II) complex (Ru(bpy)³⁺)
## 5. Intended use
Immunoassay for the in vitro quantitative determination of estradiol in human serum and plasma.
Continued on next page
{2}
510(k) Summary, Continued
6. Comparison to predicate device
The Boehringer Mannheim Elecsys® Estradiol Assay is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Enzymun-Test® Estradiol Assay (K916132).
The following table compares the Elecsys® Estradiol Assay with the predicate device, Enzymun-Test® Estradiol Assay. Specific data on the performance of the test have been incorporated into the draft labeling in attachment 5. Labeling for the predicate device in provided in attachment 6.
Similarities:
- Intended Use: Immunoassay for the in vitro quantitative determination of estradiol
- Sample type: Serum and plasma
- Antibody: Same polyclonal anti-estradiol antibody
- Solid phase binding principle: Streptavidin/Biotin
- Assay Standardization: ID-GC/MS
Differences:
| Feature | Elecsys® Estradiol | Enzymun® Estradiol |
| --- | --- | --- |
| Reaction test principle | Electrochemiluminescence | ELISA/1-step sandwich assay. |
| Instrument required | Elecsys® 2010 | ES 300 |
| Calibration Stability | A calibration is recommended every 7 days if kit is not consumed; 4 weeks with same reagent lot if reagent is consumed within 7 days. | Full calibration required every 2 weeks. One-point calibration required every run. |
Continued on next page
{3}
510(k) Summary, Continued
6. Comparison to predicate device, cont.
Performance Characteristics:
| Feature | Elecsys® Estradiol | | | Enzymun® Estradiol | | |
| --- | --- | --- | --- | --- | --- | --- |
| Precision | Modified NCCLS (pg/mL): | | | Modified NCCLS (pg/mL): | | |
| Level | Low | Mid | High | Low | Mid | High |
| N | 60 | 60 | 60 | 60 | 60 | 60 |
| Within-Run: Mean | 53.3 | 410 | 3021 | 44.0 | 277.6 | 1217.0 |
| %CV | 6.5 | 2.7 | 3.0 | 14.3 | 8.5 | 7.7 |
| Total: Mean | 53.3 | 410 | 3021 | 44.0 | 277.6 | 1217.0 |
| %CV | 9.0 | 5.0 | 5.5 | 16.1 | 9.1 | 10.3 |
| Lower Detection Limit | 10 pg/mL | | | 10 pg/mL | | |
| Linearity | 10-4600 pg/mL (with a deviation from a linear line of ±10%) | | | 10-1300 pg/mL (with a deviation from a linear line of ±10%) | | |
| Method Comparison | Vs Enzymun-Test® Estradiol
Least Squares
y = 0.902x + 42.94
r = 0.976
SEE = 91.96
N = 74
Passing/Bablok
y = 0.962x + 12.51
r = 0.976
SEE = 38.71
N = 74 | | | Vs Enzymun-Test® Estradiol
Least Squares
y = 1.027x + 6.32
r = 0.995
SEE = 31.03
N = 64 | | |
| Interfering substances
Bilirubin
Hemoglobin
Lipemia
Biotin | No interference at:
25 mg/dL
1.0 g/dL
1500 mg/dL
24 ng/mL | | | No interference at:
64.5 mg/dL
200 mg/dL
1250 mg/dL
20 ng/mL | | |
Continued on next page
{4}
page 31
# 510(k) Summary, Continued
## 6. Comparison to predicate device, cont.
### Performance Characteristics:
| Feature | Elecsys® Estradiol | | Enzymum® Estradiol | |
| --- | --- | --- | --- | --- |
| Specificity | Level tested (ng/mL) | % Cross-reactivity | Level tested | % Cross-reactivity |
| 17 hydroxy-Progesterone | 100 | < 0.5 | --- | < 0.001 |
| Ethisterone | 100 | < 0.5 | --- | --- |
| Norethindrone-Acetate | 100 | < 0.5 | --- | --- |
| 2-Methoxy-estradiol | 100 | < 1.0 | --- | 0.07 |
| 6-Hydroxy-estradiol | 0.78 | < 600 | --- | 62 |
| Estriol | 160 | < 1.0 | --- | 0.3 |
| Prednisolone | 1000 | < 0.25 | --- | --- |
| Danazol | 100000 | <0.1 | --- | 0.0006 |
| Testosterone | 100000 | <1.0 | --- | 0.03 |
| Di-Hydro-Testosterone | 100000 | < 1.0 | --- | --- |
| 5-Androsten-3β-17β-Diol | 100000 | < 0.4 | --- | --- |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.