← Product Code [CHH](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CHH) · K251091

# Lipids (K251091)

_Truvian Health · CHH · Dec 30, 2025 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CHH/K251091

## Device Facts

- **Applicant:** Truvian Health
- **Product Code:** [CHH](/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CHH.md)
- **Decision Date:** Dec 30, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1175
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

Lipids is part of the TruWellness Panel™ and is intended for use on the TruVerus™. Lipids (part of the TruWellness Panel™) is an in vitro diagnostic device and intended to be used for the quantitative determination of Total Cholesterol (TChol) and Triglycerides (TRIG) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From the TRIG determination, Very Low-Density Lipoprotein Cholesterol (VLDL) is calculated by the analyzer. Lipids (part of the TruWellness Panel™) is an in vitro diagnostic test system that aids the physician in the diagnosis and treatment of the following disorders in adults 18 years of age or older: Total Cholesterol (TChol): Excess cholesterol in the blood and lipid and lipoprotein disorders. Triglyceride (TRIG): Diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism; various endocrine disorders.

## Device Story

TruVerus™ benchtop analyzer uses Single-Use Consumable Kit (TruWellness Panel™) to process lithium-heparinized venous whole blood. System automates sample handling via high-precision pipettor, onboard centrifugation, and motion control. Detection modules use photometric/colorimetric analysis (TChol 510nm, TRIG 540nm) with xenon-arc stroboscopic lamp and CMOS array detector. Instrument performs self-tests and QC checks using optical/electronic feedback. Results displayed on touchscreen for clinician review; VLDL calculated from TRIG. Used in clinical labs or point-of-care settings to aid diagnosis of lipid metabolism disorders. Benefits include rapid, integrated testing without specialized skills or external calibration.

## Clinical Evidence

Bench testing only. Precision/reproducibility (CLSI EP05-A3) across 3 sites/3 instruments showed total CV% 1.1-3.0%. Linearity (EP06-Ed2) confirmed 20-520 mg/dL (TChol) and 20-700 mg/dL (TRIG). Method comparison (N=326 TChol, N=313 TRIG) against Roche Cobas showed high correlation (R=0.995, 0.998; Slope 1.00). Interference testing (EP07/EP37) evaluated numerous endogenous/exogenous substances.

## Technological Characteristics

Enzymatic colorimetric assay; reagents include cholesterol esterase, cholesterol oxidase, glycerol kinase, glycerol phosphate oxidase, and peroxidase. Single-use plastic consumable kit with integrated dilution wells and waste containment. Photometric detection at 510 nm (TChol) and 540 nm (TRIG). Standalone operation on TruVerus analyzer. Traceability to NIST SRM1950 (TChol) and IFCC (TRIG).

## Regulatory Identification

A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

## Predicate Devices

- Piccolo Total Cholesterol Test System ([K023642](/device/K023642.md))
- Piccolo Triglycerides Test System ([K023639](/device/K023639.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K251091

B Applicant

Truvian Health

C Proprietary and Established Names

Lipids

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  CHH | Class I, meets the limitations of exemptions in 862.9(c)(4) | 21 CFR 862.1175 - Cholesterol (Total) Test System | CH - Clinical Chemistry  |
|  JGY | Class I, meets the limitations of exemptions in 862.9(c)(4) | 21 CFR 862.1705 - Triglyceride test system | CH - Clinical Chemistry  |

## II Submission/Device Overview:

A Purpose for Submission:

New Device

B Measurand:

Cholesterol

Triglycerides

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K251091 - Page 2 of 12

C Type of Test:

Quantitative, photometric/colorimetric

III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

Lipids is part of TruWellness Panel™ and is intended for use on the TruVerus. Lipids (part of the TruWellness Panel) is an in-vitro diagnostic device and intended to be used for the quantitative determination of Total Cholesterol (TChol) and Triglycerides (TRIG) in lithium-heparinized venous whole blood in clinical laboratory or point-of-care settings. From TRIG determination, Very Low-Density Lipoprotein (VLDL) is calculated by the analyzer. Lipids (part of the TruWellness Panel) is an in vitro diagnostic test system that aids the physician in diagnosis and treatment of the following disorders:

Total Cholesterol (TChol): Excess cholesterol in the blood and lipid and lipoprotein disorders.

Triglyceride (TRIG): Diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism; various endocrine disorders.

C Special Conditions for Use Statement(s):

Rx- For Prescription Use Only

Lipids (part of the TruWellness panel) is intended for use with adults aged 18 and older.

The Triglyceride analyte is sensitive to Icterus and is not for use with patients with hyperbilirubinemia (elevated bilirubin levels).

The instrument was validated up to an altitude of 2,000 m (6562 ft). Lipids should not be used above 2,000 m (6562 ft).

D Special Instrument Requirements:

TruVerus was cleared under K251058 with the trade name Tru Analyzer.

IV Device/System Characteristics:

A Device Description:

Lipids contains the following reagents:

|  Component | Quantity per Kit  |
| --- | --- |
|  4-Aminoantipyrine | 5.6 μg  |
|  Adenosine-5'-Triphosphate | 17.5 μg  |

{2}

|  Component | Quantity per Kit  |
| --- | --- |
|  Catalase | 21.5 U  |
|  Cholesterol Esterase | ≥ 21.9 mU  |
|  Cholesterol Oxidase | ≥ 15.4 mU  |
|  Glycerol Kinase | ≥ 19.1 mU  |
|  Glycerol Phosphate Oxidase | ≥ 57.3 mU  |
|  HDAOS | 4.7 μg  |
|  Peroxidase | > 975.0 mU  |
|  p-Hydroxybenzenesulfonate | 31.6 μg  |
|  Sodium N-Ethyl-N-(3-Sulfopropyl)-M-Anisidine | 5.6 μg  |

The Single-Use Consumable Kit houses all the components needed to process as well as analyze samples on the TruVerus, including dried reagents, internal process control solutions, barcodes that manage the identity of the kit lot (e.g., Disc and Support Pack ID), calibration information, dilution buffers, and single-use plastic pipette tips. It also serves as a waste container which the user discards of at the end of the run. The Support Pack contains a feature to accept a standard 4 mL blood tube. The Support Pack also houses 22 pipette tips for transferring and mixing samples and reagents, 11 dilution wells to support reagent processing activities within the test system (e.g., sample dilution, reagent dilution, rehydration of dried reagents), and 6 x 2 mL tubes that contain additional wet and dry reagents.

## B Principle of Operation:

### Total cholesterol

This assay is based on the work of Allain et al. and measures the total level of cholesterol by using a multi-step enzymatic system. Cholesterol esters are hydrolyzed by cholesterol esterase to cholesterol and fatty acids. Cholesterol is oxidized by cholesterol oxidase to Δ4-cholestenone with the simultaneous production of an equimolar amount of hydrogen peroxide. In the presence of peroxidase, hydrogen peroxide oxidizes to p-hydroxybenzenesulfonate (p-HBS) and 4-aminoantipyrine (4-AAP) to give a quinoneimine dye colored red, the absorbance of which can be measured at 510 nm. The absorbance is proportional to the concentration of cholesterol in the sample.

### Triglycerides

This assay measures triglyceride concentration in plasma by a series of enzymatic reactions. After hydrolysis by microbial lipases, the triglycerides yield glycerol and free fatty acids (FFA). Glycerol is phosphorylated by adenosine-5'-triphosphate (ATP) to glycerol-1-phosphate (G-1-P) in a reaction catalyzed by glycerol kinase (GK). G-1-P is oxidized to dihydroxyacetone phosphate (DAP) in a reaction catalyzed by glycerol phosphate oxidase (GPO). In this reaction, hydrogen peroxide is produced in equimolar concentration to the level of triglycerides present in the sample. In the indicating Trinder type reaction, hydrogen peroxide reacts with 4-aminoantipyrine (4-AAP) and n-ethyl-n-(3-sulfopropyl)-m-anisidine (ESPAS) in a reaction catalyzed by peroxidase (HPOD). The result of this oxidative coupling is a quinoneimine red-colored dye the absorbance of which can be measured at 540 nm. The absorbance is proportional to the concentration of triglycerides in the sample.

### Very low-density lipoprotein (VLDL) cholesterol

VLDL cholesterol is calculated using the following equation that utilizes the triglyceride measurement: VLDL Cholesterol = Triglyceride/5 (if units in mg/dL).

K251091 - Page 3 of 12

{3}

V Substantial Equivalence Information:

A Predicate Device Name(s):

Piccolo Total Cholesterol Test System – K023642
Piccolo Triglycerides Test System – K023639

B Predicate 510(k) Number(s):
K023642 and K023639

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K251091 Candidate | K023642 Predicate  |
| --- | --- | --- |
|  Device Trade Name | Lipids | Piccolo Total Cholesterol Test System  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | In vitro quantitative determination of Total Cholesterol and Triglycerides (TRIG) in heparinized whole blood in clinical laboratory setting or point-of-care location. | Same  |
|  Measuring Range | TChol: 20 - 520 mg/dL | Same  |
|  General Device Characteristic Differences |  |   |
|  Sample type | Lithium-heparinized venous whole blood | Heparinized whole blood, heparinized plasma, and serum  |
|  Device & Predicate Device(s): | K251091 Candidate | K023639 Predicate  |
| --- | --- | --- |
|  Device Trade Name | Lipids | Piccolo Triglycerides Test System  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | In vitro quantitative determination of Total Cholesterol and Triglycerides (TRIG) in heparinized whole blood in clinical laboratory setting or | Same  |

K251091 - Page 4 of 12

{4}

|   | point-of-care location. |   |
| --- | --- | --- |
|  General Device Characteristic Differences |  |   |
|  Sample type | Lithium-heparinized venous whole blood | Heparinized whole blood, heparinized plasma, and serum  |
|  Measuring Range | 20 - 700 mg/dL | 20 - 500 mg/dL  |

VI Standards/Guidance Documents Referenced:

Clinical and Laboratory Standards Institute (CLSI) EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition

CLSI EP06 Ed.2, Evaluation of the Linearity of Quantitative Measurement Procedures

CLSI EP07 Ed.3, Interference Testing in Clinical Chemistry

CLSI EP09c Ed.3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples

CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition

CLSI EP25 Ed.2, Evaluation of Stability of In Vitro Medical Laboratory Test Reagents

CLSI EP37 Ed.1, Supplemental Tables for Interference Testing in Clinical Chemistry

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

A. Reproducibility with control materials.

This study was designed according to CLSI EP05-A3. The study was conducted at 3 external point-of-care (POC) sites, with 3 analyzers per site and at least 3 POC operators per site. Testing was conducted over 5 days using a single lot of consumables and consisted of 3 levels of control samples at low, medium, and high concentrations, with 6 replicates run per day, 3 run in the morning and 3 run in the afternoon (each replicate is run on an individual instrument). Overall, there were at least 90 replicates for each precision control sample level (3 sites x 1 sample x 3 replicates per run x 2 runs per day x 5 days = 90 replicates). The total precision as well as within run, between day, and between site precision were estimated. Results are displayed below.

K251091 - Page 5 of 12

{5}

|  Analyte | Level | N | Mean | Within - Run |   | Between-Run |   | Between-Day |   | Between-Site |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |   |  SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV%  |
|  TChol | Low | 92 | 82 | 2 | 2.6 | 0 | 0.0 | 0 | 0.0 | 1 | 1.5 | 2 | 3.0  |
|  TChol | Med | 91 | 189 | 4 | 2.3 | 1 | 0.4 | 0 | 0.0 | 2 | 1.0 | 5 | 2.6  |
|  TChol | High | 92 | 228 | 3 | 1.5 | 0 | 0.0 | 0 | 0.0 | 2 | 1.1 | 4 | 1.9  |
|  TRIG | Low | 92 | 88 | 2 | 1.8 | 0 | 0.0 | 0 | 0.0 | 1 | 0.9 | 2 | 2.0  |
|  TRIG | Med | 91 | 136 | 3 | 2.1 | 1 | 0.4 | 1 | 0.6 | 0.4 | 0.3 | 3 | 2.2  |
|  TRIG | High | 92 | 452 | 5 | 1.1 | 0 | 0.0 | 1 | 0.3 | 1 | 0.2 | 5 | 1.1  |

B. Whole blood precision

This study was evaluated at five sites collecting samples from intended use population with normal and abnormal analyte levels. At least 20 subjects were enrolled at each site and for each test subject, eight replicates were measured across four instruments and two operators. The SD and CV% were calculated by pooling subject standard deviations and/or % CV for predefined Low, Medium, and High sub-intervals selected to represent normal and abnormal regions of the analytical measuring range and encompassing medical decision levels. Pooled imprecision is considered a representative estimate of reproducibility that include variability of sites, instruments, operators, days and repeatability.

|  Analyte | Range (mg/dL) | Subjects (N) | Mean (mg/dL) | SD | CV%  |
| --- | --- | --- | --- | --- | --- |
|  TChol | 20–200 | 93 | 139 | 3.5 | 2.5  |
|  TChol | 200–240 | 10 | 214 | 4.8 | 2.3  |
|  TChol | 240–520 | 9 | 313 | 8.2 | 2.6  |
|  TRIG | 20–150 | 69 | 91 | 2.2 | 2.9  |
|  TRIG | 150–250 | 25 | 184 | 3.6 | 1.9  |
|  TRIG | 250–700 | 9 | 339 | 5.7 | 1.6  |

C. Whole blood between-lot precision.

22 subjects consisting of 6 healthy individuals and 16 individuals in various disease states were enrolled at a single point-of-care clinical site and tested across 3 lots of the candidate panel on 3 individual analyzers by 2 operators. A total of 6 replicates per subject participant, 2 replicates per lot, were tested by 1 of the 2 operators. Results below present the number of subjects evaluated, the % difference from the mean and pooled %CV by lot and by analyte.

|  Analyte | N-Subjects Per Lot | Mean relative difference % |   |   | Pooled %CV  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Lot-A | Lot-B | Lot-C | Lot-A | Lot-B | Lot-C  |
|  TCHOL | 22 | -0.5% | -0.5% | 1.0% | 2.6 % | 2.1 % | 1.9 %  |
|  TRIG | 21 | 0.7% | -0.7% | 0.0% | 2.5 % | 2.0 % | 1.8%  |

D. Control between-lot reproducibility.

To evaluate the lot-to-lot reproducibility for the assays, three (3) control samples (Low, Medium, High) were tested on 3 analyzers with 3 lots of the candidate test. Each sample level was tested with at least five (5) replicates on one analyzer each day with each single-use consumable kit lot

K251091 - Page 6 of 12

{6}

over three (3) days to achieve a minimum of fifteen (15) replicates per sample level and kit lot. Results are displayed below:

|  Analyte | Sample | Units | N | Mean | SD |   |   | CV%  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |   |   |  Lot 1 | Lot 2 | Lot 3 | Lot 1 | Lot 2 | Lot 3  |
|  TChol | Low | mg/dL | 45 | 84.5 | 2.25 | 1.40 | 2.13 | 2.7% | 1.7% | 2.5%  |
|  TChol | Medium | mg/dL | 45 | 193.8 | 2.32 | 3.77 | 3.33 | 1.2% | 1.9% | 1.7%  |
|  TChol | High | mg/dL | 45 | 231.7 | 3.27 | 5.56 | 3.07 | 1.4% | 2.4% | 1.3%  |
|  Trig | Low | mg/dL | 45 | 89.3 | 1.43 | 1.83 | 1.84 | 1.6% | 2.1% | 2.0%  |
|  Trig | Medium | mg/dL | 45 | 138.3 | 1.59 | 1.81 | 1.67 | 1.2% | 1.3% | 1.2%  |
|  Trig | High | mg/dL | 45 | 454.4 | 4.29 | 12.42 | 3.73 | 0.9% | 1.0% | 0.8%  |

## 2. Linearity:

Linearity testing was performed in accordance with CLSI document EP06-Ed2. Linearity was assessed using multiple panels of samples. The linear range for total cholesterol and triglycerides were determined by testing nine (9) whole blood samples containing varying concentrations of analyte. Each sample was tested in quadruplicate, using one (1) kit lot, across four (4) analyzers, on a single day. The results of the study supported the claimed linear interval for TChol and Trig. Within the claimed analytical measuring range, the maximum observed deviation from linearity for TChol and TRIG were &lt;2% and &lt;4% respectively.

## 3. Analytical Specificity/Interference:

Potential interference from endogenous and exogenous substances with the assays was evaluated following CLSI EP07-A3 and EP37 guidelines. Testing was performed with contrived lithium heparin whole blood samples from healthy donor population using two (2) analyte concentration levels targeted based on the recommendation in CLSI EP07. Endogenous and exogenous substances were spiked into samples at two targeted analyte concentration levels: approximately 150 mg/dL and 220 mg/dL for TChol, 100 mg/dL and 250 mg/dL for TRIG. For substances identified as an interferent, dose response testing was conducted to assess the highest concentration limit below which no significant interference was observed. Interference is defined as the analyte result shifted by more than 10% between the test and control samples*.

### Endogenous Substances

|  Substance | Maximum Concentration without Interference (mg/dL)  |   |
| --- | --- | --- |
|   |  Total Cholesterol | Triglyceride  |
|  Hemoglobin | 122 | 154  |
|  Conjugated | 4.1 | < 200  |
|  Bilirubin |   | > 200  |
|  Unconjugated | 5.0 | < 200  |
|  Bilirubin |   | > 200  |
|  Triglycerides | 1814 | N/A  |

* Low level Conjugated Bilirubin testing showed ≤11% bias

K251091 - Page 7 of 12

{7}

Exogenous Substances

|  Substance | Max Concentration without Interference (mg/dL)  |   |
| --- | --- | --- |
|   |  Total Cholesterol | Triglycerides  |
|  Acetaminophen | 15.6 | 15.6  |
|  Acetylsalicylic acid | 3 | 3  |
|  Ampicillin | 7.5 | 7.5  |
|  Cefoxitin | 495 | 660  |
|  Cyclosporine | 0.18 | 0.18  |
|  Doxycycline | 1.8 | 1.8  |
|  Heparin | 3,300 U/L | 3,300 U/L  |
|  Ibuprofen | 10.95 | 21.9  |
|  Levodopa (L-Dopa) | 0.75 | 0.3  |
|  Methyldopa | 1.12 | 0.78  |
|  Metronidazole | 12.3 | 12.3  |
|  Phenylbutazone | 16.05 | 24.08  |
|  Rifampicin | 4.8 | 4.8  |
|  Theophylline | 6 | 6  |
|  Acetylcysteine | 15 | 5  |
|  Ascorbic Acid | 5.25 | 1.75  |
|  Calcium Dobesilate | 3 | 1.5  |
|  Acetoacetate | 20 | 20  |
|  Atorvastatin | 0.075 | 0.075  |
|  Beta-Hydroxybutyrate | 333 | 750  |
|  Caffeine | 10.8 | 10.8  |
|  Cefotaxime | 52.8 | 52.8  |
|  Cephalothin (Keflin) | 180 | 180  |
|  Cimetidine | 3 | 3  |
|  Creatinine | 15 | 15  |
|  Cysteine | 10 | 5  |
|  Digoxin | 0.0039 | 0.0039  |
|  Dipyrone | 3.3 | 3.3  |
|  Dobutamine | 0.121 | 0.121  |
|  Fenofibrate | 4.5 | 4.5  |
|  Fructose | 18 | 18  |
|  Gemfibrozil (Lopid) | 13.8 | 13.8  |
|  Glutathione | 15 | 3.75  |
|  Isoniazide | 6 | 6  |
|  Lactate Lithium | 90 | 90  |
|  Lactose | 100 | 100  |
|  Lidocaine | 1.5 | 1.5  |
|  Lovastatin (Mevacor) | 0.021 | 0.021  |
|  Methotrexate | 102 | 54.4  |
|  Nicotinic Acid (Niacin) | 10 | 10  |

K251091 - Page 8 of 12

{8}

|  Substance | Max Concentration without Interference (mg/dL)  |   |
| --- | --- | --- |
|   |  Total Cholesterol | Triglycerides  |
|  Phenytoin | 6 | 6  |
|  Pravastatin | 0.0207 | 0.0207  |
|  Rosuvastatin | 0.0111 | 0.0111  |
|  Salicylic Acid | 2.86 | 2.86  |
|  Simvastatin | 0.168 | 0.168  |
|  Sodium Methicillin | 7.5 | 7.5  |
|  Urea | 120 | 120  |
|  Uric Acid | 23.5 | 23.5  |

The sponsor provided adequate information to support hemolysis, icterus and lipemia flags will detect samples subject to these interferences.

## 4. Assay Reportable Range:

The TruVerus Analyzer will report results within the AMR as listed in the table below.

|  Analyte | AMR  |
| --- | --- |
|  Total Cholesterol (mg/dL) | 20–520  |
|  Triglycerides (mg/dL) | 20–700  |

## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

Traceability

The sponsor claims the following regarding traceability:

|  Assay | Traceable Material  |
| --- | --- |
|  Total Cholesterol | NIST SRM1950  |
|  Triglycerides | IFCC through commercial materials  |

## Sample Stability

Sample stability studies performed demonstrated that samples measured on the candidate tests are stable for 60-minutes post-collection.

## Open Pouch Stability and Open Disc Stability

The open disc stability study supports an open disc stability of 4 minutes when used on TruVerus at the following operating condition limits: 15°C - 30°C and 8% - 60% RH.

## 6. Detection Limit:

Detection limits were determined in accordance with recommendations in the CLSI EP17-A2 guideline.

K251091 - Page 9 of 12

{9}

The Limit of Blank (LoB) corresponds to the highest measurement result that is likely to be observed for a blank sample. The assay is designed to have a LoB ≤ Limit of Detection (LoD).

The Limit of Detection (LoD) corresponds to the lowest concentration of analyte that can be detected with a probability of 95%. The assay is designed to have an LoD ≤ Limit of Quantitation (LoQ).

The Limit of Quantitation (LoQ) corresponds to the lowest concentration of analyte in a sample that had a CV ≤ 20%. Detection limits are provided in the table below.

LoD and LoQ were established in Li-Hep whole blood samples.

|  Analyte | LoB | LoD | LoQ  |
| --- | --- | --- | --- |
|  Total Cholesterol (mg/dL) | 3.76 | 6 | 7  |
|  Triglycerides (mg/dL) | 12.45 | 14 | 16  |

7. Assay Cut-Off:

Not applicable.

B Comparison Studies:

1. Method Comparison with Predicate Device:

Whole blood specimens were collected from subjects at five external sites to evaluate the accuracy of the Lipids test. Specimens comprised of apparently healthy patient samples and samples from patients with acute or chronic health conditions were collected from 5 point-of-care sites, tested by at least 3 point-of-care operators per site, on at least 4 analyzers per site, using 10 lots of reagents. Results (using Passing Bablok regression) compared against a legally marketed comparator are displayed below. The data demonstrated that accuracy was consistent across sites.

|  Analyte | N | Range (mg/dL) | Slope | Intercept | R  |
| --- | --- | --- | --- | --- | --- |
|  TChol | 326 | 23 – 475 | 1.00 (0.98, 1.01) | -4.00 (-5.0, -1.43) | 0.995  |
|  Trig | 313 | 29-683 | 1.00 (0.99, 1.01) | -6.00 (-6.05, -4.26) | 0.998  |

Predicted Differences at Medical Decision Levels (MDLs)

|   |   | All  |   |
| --- | --- | --- | --- |
|   | MDL (mg/dL) | Bias | %Bias  |
|  Total CHOL | 90 | -4.0 | -4.4%  |
|   |  200 | -4.0 | -2.0%  |
|   |  240 | -4.0 | -1.7%  |

K251091 - Page 10 of 12

{10}

|   |   | All  |   |
| --- | --- | --- | --- |
|   | MDL (mg/dL) | Bias | %Bias  |
|  TRIG | 150 | -6.0 | -4.0%  |
|   |  400 | -6.0 | -1.5%  |
|   |  500 | -6.0 | -1.2%  |

2. Matrix Comparison:

Not applicable.

## C Clinical Studies:

1. Clinical Sensitivity:

Not applicable

2. Clinical Specificity:

Not applicable

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

High Altitude Study

An altitude study was conducted to evaluate the Lipids test at sea-level and a high altitude of 2000 meters. The results demonstrated acceptable performance at both tested altitude conditions.

## D Clinical Cut-Off:

Not applicable

## E Expected Values/Reference Range:

The sponsor provided the following risk classification intervals for TChol and TRIG based on scientific literature¹.

|  Analyte |   | Reference Range (mg/dL)  |
| --- | --- | --- |
|  Total Cholesterol | Desirable | <200  |
|   |  Borderline | 200 - 239  |
|   |  High | ≥ 240  |
|  Triglycerides | Normal | < 150  |
|   |  Borderline High | 150 – 199  |
|   |  High | 200 – 499  |

K251091 - Page 11 of 12

{11}

|  Analyte |   | Reference Range (mg/dL)  |
| --- | --- | --- |
|   | Very High | ≥ 500  |

$^{1}$ National Cholesterol Education Program. Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III). National Institutes of Health, National Heart, Lung, and Blood Institute, NIH Publication No. 01-3670, May 2001.

## VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K251091 - Page 12 of 12

---

**Source:** [https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CHH/K251091](https://fda.innolitics.com/submissions/CH/subpart-b%E2%80%94clinical-chemistry-test-systems/CHH/K251091)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
