WIENER LAB. CK-NAC UV UNITEST, WIENER LAB. CK-NAC UV AA, WIENER LAB. CK-NAC UV AA, WIENER LAB.NAC UV AA LIQUIDA
Applicant
Wiener Laboratories Saic
Product Code
CGS · Clinical Chemistry
Decision Date
Jul 3, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
The "Wiener lab. CK-NAC UV" test system is a device intended to measure the activity of the enzyme creatine phosphokinase in plasma and serum with manual methodology and automated clinical chemistry analyzers. Measurements of creatine phosphokinase and isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Device Story
The Wiener Lab. CK-NAC UV is an in vitro diagnostic test system for measuring creatine phosphokinase (CK) activity in human serum and heparinized plasma. The device utilizes a kinetic UV method based on the IFCC-recommended reaction sequence: CK catalyzes the conversion of creatine phosphate and ADP to creatine and ATP; hexokinase (HK) then uses ATP and glucose to produce glucose-6-phosphate; finally, glucose-6-phosphate dehydrogenase (G-6-PDH) reduces NADP+ to NADPH, which is measured spectrophotometrically at 340 nm. N-acetylcysteine (NAC) is included as an activator. The test is performed manually or on automated clinical chemistry analyzers in a laboratory setting. Healthcare providers use the resulting enzyme activity levels to diagnose and monitor myocardial infarction and various muscle diseases. The device provides a quantitative assessment of CK levels, assisting in clinical decision-making regarding cardiac and muscular health.
Clinical Evidence
Bench testing only. Performance evaluation included precision studies (within-run and run-to-run) and linearity assessment. Within-run precision for normal serum was 2.33% CV and 1.33% CV for abnormal serum. Run-to-run precision was 2.12% CV for normal serum and 1.53% CV for abnormal serum. Linearity was established up to 2,000 U/l at 37°C. No clinical prospective or retrospective studies were required for this 510(k) submission.
Technological Characteristics
In vitro diagnostic reagent kit. Principle: Kinetic UV spectrophotometry (334-366 nm). Components: Creatine phosphate, ADP, Glucose, HK, NADP, G-6-PDH, NAC. Form factor: Reagent powder/solution. Connectivity: Compatible with manual and automated clinical chemistry analyzers. Sterilization: Not applicable (reagent kit).
Indications for Use
Indicated for the quantitative determination of creatine phosphokinase activity in human serum and heparinized plasma to aid in the diagnosis and treatment of myocardial infarction and muscle diseases, including progressive Duchenne-type muscular dystrophy.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
DMA CK NAC test system (Cat. Nº 1380-200)
RANDOX CK NAC-activated kit (Cat. Nº CK522)
Related Devices
K122083 — ELITECH CLINICAL SYSTEMS CK NAC SL · Elitechgroup · Aug 22, 2012
{0}------------------------------------------------
3 2002
CK-NAC UV Wiener lab.
Image /page/0/Picture/2 description: The image is a circular seal or logo. The text "WIENER lab." is arranged around the top of the circle, and "SISTEMA DE CALIDAD CERTIFICADO" is arranged around the bottom. In the center of the circle, there is a stylized "W" logo, followed by the text "ISO 9001" and a "TUV CERT" logo.
Image /page/0/Picture/3 description: The image shows the logo for Wiener lab. The logo consists of a circular emblem with the letter W inside, followed by the text "Wiener lab." in a bold, sans-serif font. Below the company name, there is the text "Especialidades para Laboratorios Clínicos" in a smaller font.
**WIENER LABORATORIOS S.A.I.C.** - Riobamba 2944 - 2000 Rosario - Argentina
Phone +54 (341) 432-9191/6 - Fax +54 (341) 432-5454/5555
Internet: http://www.wiener-lab.com.ar
## Section 6 - Summary
510(k) Summary "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92"
"The assigned 510(k) number is: _2 3 3 3 3 3 3 3 3 3 3 2 3 3 2 3 2 3 2
- According to the requirements of 21 CFR 862.1215, the follow-Introduction ing information provides sufficient details to understand the basis of a determination of substantial equivalence.
Wiener Lab Group 6-1 Submitter Name, Address. Riobamba 2944 2000 - Rosario - Argentina Contact Contact person: Viviana Cétola Date Prepared: March 28, 2002.
- Proprietary name: WIENER LAB. CK-NAC UV 6-2 Device Name Common name: Creatine Phosphokinase/Creatine Kinase test svstem. Classification name: NAD Reduction/NADH Oxidation, CPK or lsoenzymes. Device Class II
- 6-3 Predicate De-We claim substantial equivalence to the currently marketed DMA CK NAC test system (Cat. Nº 1380-200) and RANDOX vice CK NAC-activated kit (Cat. Nº CK522).
{1}------------------------------------------------
Image /page/1/Figure/1 description: This image describes the device and intended use of the WIENER LAB. CK-NAC UV test system. The reaction system involves creatine phosphate, ADP, ATP, glucose, glucose-6-P, and NADP+. The test system measures creatine phosphokinase activity in plasma and serum and is substantially equivalent to other products like DMA CK NAC and RANDOX CK NAC.
currently marketed DMA CK NAC test system.
{2}------------------------------------------------
| | OMA<br>Test System | WIENER LAB. Test<br>System | |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|--|
| Intended use | Quantitative determina-<br>tion of Creatine Kinase<br>in human serum. | Quantitative determina-<br>tion of Creatine Kinase<br>in human serum and<br>heparinized plasma. | |
| | The reaction system is as follows: | | |
| Test principle | Creatine phosphate + ADP | ск<br>Creatine + ATP | |
| | HK<br>ATP + Glucose<br>Glucose-6-P + NADP+ | ADP + Glucose-6-P<br>G-6-PDH | |
| | Gluconate-6-P + NADPH + H*<br>In the reaction system. NAC works as activator of<br>the Creatine Kinase, recommended by the IFCC. | | |
| | CK: Creatine Kinase<br>HK: Hexokinase<br>G-6-PDH: Glucose-6-phosphate dehydrogenase | | |
| Essential Compo-<br>nents | Creatine phosphate -<br>ADP - Glucose - HK -<br>NAD - G-6PDH - NAC | Creatine phosphate -<br>ADP - Glucose - HK -<br>NADP - G-6PDH - NAC | |
| Reagent Deterio-<br>ration | Reagent must be a<br>white powder<br>Reagent blank > 0.600<br>O.D. at 340 nm | Reagent blank > 0.800<br>O.D. at 340 nm | |
| Working Reagent<br>Stability | Stable 21 days at 2-<br>10°C | Stable 20 days at 2-<br>10°C or 3 days at room<br>temperature. | |
| Sample | Human serum | Human serum and hepa-<br>rinized plasma | |
| Continued on next page | | | |
| | DMA<br>Test System | WIENER LAB. Test<br>System | |
| Working Tem-<br>peratures | 30°C or 37°C | 25°C, 30°C or 37°C | |
| Wavelength of<br>reading. | 340 nm | 334 nm - 340 nm - 366<br>nm | |
| Linearity | 1,500 U/l (30°C) -<br>2,300 U/l (37°C) | 2,000 U/l (37°C) | |
| Instructions for<br>samples exceed-<br>ing linearity | Dilution in saline and correction of result | | |
| Expected values | 7-114 U/l (30°C)<br>25-160 U/l (37°C) | Male:<br>≤80 U/l (25°C)<br>≤130 U/l (30°C)<br>≤195 U/l (37°C)<br>Female:<br>≤70 U/l (25°C)<br>≤110 U/l (30°C)<br>≤170 U/l (37°C) | |
| Within-run preci-<br>sion | Normal Serum:<br>CV = 2.7%<br>Abnormal Serum:<br>CV = 3.9% | Normal Serum:<br>CV = 2.33%<br>Abnormal Serum:<br>CV = 1.33% | |
| Run-to-run preci-<br>sion | Normal Serum:<br>CV = 2.9%<br>Abnormal Serum:<br>CV = 4.2% | Normal Serum:<br>CV = 2.12%<br>Abnormal Serum:<br>CV = 1.53% | |
| | | Continued on next page | |
{3}------------------------------------------------
{4}------------------------------------------------
The following table illustrates the similarities and differences between the WIENER LAB. CK-NAC UV test system and the currently marketed RANDOX CK NAC-activated system.
| | RANDOX<br>Test System | WIENER LAB. Test<br>System |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Quantitative determina-<br>tion of Creatine Kinase<br>in human serum, hepa-<br>rinized or EDTA plasma. | Quantitative determina-<br>tion of Creatine Kinase<br>in human serum and<br>heparinized plasma. |
| Test principle | The reaction system is as follows:<br>$Creatine phosphate + ADP \xrightarrow{CK} Creatine + ATP$ $ATP + Glucose \xrightarrow{HK} ADP + Glucose-6-P$ $Glucose-6-P + NADP^+ \xrightarrow{G-6-PDH} Gluconate-6-P + NADPH + H^+$ CK: Creatine Kinase<br>HK: Hexokinase<br>G-6-PDH: Glucose-6-phosphate dehydrogenase | |
| Essential Compo-<br>nents | Creatine phosphate – ADP – Glucose – HK –<br>NADP – G-6PDH – NAC | |
| Working Reagent<br>Stability | Stable 3 weeks at 2-8°C<br>or 3 days at 15-25°C | Stable 20 days at 2-<br>10°C or 3 days at room<br>temperature. |
| Sample | Human serum, heparin-<br>ized or EDTA plasma | Human serum and hepa-<br>rinized plasma |
| | RANDOX<br>Test System | WIENER LAB. Test<br>System |
| Working Reagent<br>Stability | Stable 3 weeks at 2-8ºC<br>or 3 days at 15-25ºC | Stable 20 days at 2-<br>10℃ or 3 days at room<br>temperature. |
| Sample | Human serum, heparin-<br>ized or EDTA plasma | Human serum and hepa-<br>rinized plasma |
| Working Tem-<br>peratures | 25°C, 30°C or 37°C | |
| Wavelength of<br>reading. | 334 nm - 340 nm - 365<br>nm | 334 nm - 340 nm - 366<br>nm |
| Linearity | O.D. 0.25 at 340 nm /<br>334 nm<br>O.D. 0.14 at 365 nm | 2,000 Ull (37°C) |
| Instructions for<br>samples exceed-<br>ing linearity | Dilution in saline and correction of result | |
| Expected values | Male:<br>10-80 U/I (25°C)<br>15-130 UN (30°C)<br>24-195 UI (37°C)<br>Female:<br>10-70 UJI (25°C)<br>15-110 U/I (30ºC)<br>24-170 U/l (37°C) | Male:<br><80 Ull (25°C)<br><130 Ull (30°C)<br><195 Ull (37°C)<br>Female:<br><70 Ull (25°C)<br><110 Ull (30°C)<br><170 U/I (37°C) |
| Within-run preci-<br>sion | Not stated in insert | Normal Serum:<br>CV = 2.33%<br>Abnormal Serum:<br>CV = 1.33% |
| Run-to-run preci-<br>sion | Not stated in insert | Normal Serum:<br>CV = 2.12%<br>Abnormal Serum:<br>CV = 1.53% |
{5}------------------------------------------------
{6}------------------------------------------------
- 6-7 Conclusion Based on the above mentioned data, we believe that the extended claims continue to support substantial equivalence to other products in commercial distribution intended for similar use
{7}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image is a seal or logo for the Department of Health & Human Services (CA). The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (CA)" arranged around the perimeter. In the center of the seal is a stylized graphic that resembles a bird in flight or a flowing design, possibly representing movement or progress. The overall design is simple and uses a monochrome color scheme.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
## 3 2002 ប្រជ
JUL. 3 2003
Dr. Viviana Cetola OC/OA Manager Weiner Laboratorios S.A. I.C. Riobamba 2944 Rosario, Santa Fe Argentina
Re: k021332
> Trade/Device Name: CK-NAC UV Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: CGS Dated: April 17, 2002 Received: April 26, 2002
Dear Dr. Cetola:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdch/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
CDRH ODE
Page
K021332
DUPLICATE
510(k) Number (if known):
Device Name: Wiener lab.
. UV
Indications For Use:
The "Wiener lab. CK-NAC UV" test system is a device" Theended to measure the activity of the enzyme creatine phosphokinase in Masma and serum with manual methodology and automated clining chemistry analyzers. Measurements of creatine phosphokinase and isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Cooper
(Division 3.2-01)
Division of Clinical Laboratory
510(k) Number K021332
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDBO)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.