ACTIVE ULTRASENSITIVE UNCONJUGATED ESTRIOL EIA
Device Facts
| Record ID | K962875 |
|---|---|
| Device Name | ACTIVE ULTRASENSITIVE UNCONJUGATED ESTRIOL EIA |
| Applicant | Diagnostic Systems Laboratories, Inc. |
| Product Code | CGI · Clinical Chemistry |
| Decision Date | Aug 23, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1265 |
| Device Class | Class 1 |
Indications for Use
The DSL ULTRA-SENSITIVE UNCONJUGATED ESTRIOL EIA assay is intended for the quantitative determination of Unconjugated Estriol in human serum. The measurement of Unconjugated Estriol is used as a diagnostic aid in the diagnosis and treatment of fetoplacental distress.
Device Story
Competitive binding enzyme immunoassay (EIA) for quantitative measurement of unconjugated estriol in human serum; utilizes horseradish peroxidase-labeled unconjugated estriol competing with sample estriol for antibody binding sites; substrate tetramethylbenzidine (TMB) added post-incubation/wash; enzymatic turnover measured via dual-wavelength absorbance; used in clinical laboratory settings by trained personnel; results aid clinicians in assessing fetoplacental status to guide management of pregnancy complications.
Clinical Evidence
Bench testing only. Comparative study of 106 patient serum samples using both the subject EIA kit and the predicate RIA method. Linear regression analysis showed r = 0.93, Y = 0.85(X) - 0.11.
Technological Characteristics
Competitive binding enzyme immunoassay (EIA); horseradish peroxidase label; tetramethylbenzidine (TMB) substrate; dual-wavelength absorbance detection; manual or automated plate reader format.
Indications for Use
Indicated for quantitative measurement of unconjugated estriol in human serum as a diagnostic aid in the diagnosis and treatment of fetoplacental distress.
Regulatory Classification
Identification
An estriol test system is a device intended to measure estriol, an estrogenic steroid, in plasma, serum, and urine of pregnant females. Estriol measurements are used in the diagnosis and treatment of fetoplacental distress in certain cases of high-risk pregnancy.
Predicate Devices
- DSL ULTRA-SENSITIVE UNCONJUGATED ESTRIOL RIA
Related Devices
- K955356 — ACTIVE ULTRA-SENSITIVE UNCONJUGATED ESTRIOL · Diagnostic Systems Laboratories, Inc. · May 15, 1996
- K974721 — UNCONJUGATED ESTRIOL ASSAY FOR THE BAYER IMMUNO 1 SYSTEM (IN VITRO DIAGNOSTIC SYSTEM) · Bayer Corp. · Mar 3, 1998
- K973901 — 17 B-ESTRADIOL (ELISA) · Kmi Diagnostics, Inc. · Nov 24, 1997
- K041349 — LUMINESCENT IMMUNOASSAY KIT FOR THE DETECTION OF ESTRADIOL IN SALIVA AND SERUM · Ibl GmbH · Sep 24, 2004
- K965109 — ELECSYS ESTRADIOL ASSAY · Boehringer Mannheim Corp. · Feb 18, 1997