ADVIA CHEMISTRY GLUCOSE HEXOKINASE_3 (GLUH_3) REAGENT, MODELS REF 05001429

K101854 · Siemens Healthcare Diagnostics · CFR · Mar 7, 2011 · Clinical Chemistry

Device Facts

Record IDK101854
Device NameADVIA CHEMISTRY GLUCOSE HEXOKINASE_3 (GLUH_3) REAGENT, MODELS REF 05001429
ApplicantSiemens Healthcare Diagnostics
Product CodeCFR · Clinical Chemistry
Decision DateMar 7, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

For in vitro diagnostic use in the quantitative determination of glucose in human serum, plasma, urine, and cerebrospinal fluid (CSF) on the ADVIA 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and insulin overdose.

Device Story

ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) is a two-component enzymatic reagent system for the ADVIA 1650 Chemistry System. It measures glucose in human serum, plasma (Li-heparin, K-EDTA), urine, and CSF. Principle: Hexokinase phosphorylates glucose with ATP; glucose-6-phosphate dehydrogenase oxidizes the product, reducing NAD to NADH. The system measures NADH absorbance at 340/410 nm as an endpoint reaction. The difference in absorbance between Reagent 1 (buffer, ATP, NAD) and Reagent 2 (enzymes, buffer, ATP, NAD) is proportional to glucose concentration. Used in clinical laboratories by trained personnel. Output is a quantitative glucose concentration (mg/dL). Results assist clinicians in diagnosing and managing carbohydrate metabolism disorders. The device includes automated dilution for samples exceeding the upper reportable range.

Clinical Evidence

Bench testing only. Precision studies (N=80 per serum/urine control, N=40 per CSF control) showed CVs ranging from 0.3% to 2.4%. Linearity confirmed across 3-833 mg/dL range. Method comparison against predicate (N=51-113 samples per matrix) showed high correlation (r=1.000) and slopes near 1.0. Matrix comparison for K-EDTA and Li-Heparin plasma confirmed equivalence. No clinical data required.

Technological Characteristics

Two-component liquid reagent containing hexokinase, glucose-6-phosphate dehydrogenase, ATP, NAD, and sodium azide. Photometric detection at 340/410 nm. Operates on ADVIA 1650 Chemistry System. Traceable to NIST SRM 965a. Storage: 2-8°C; 18-month shelf life; 60-day on-board stability. Standards: CLSI EP5-A2 (precision), CLSI EP17-A (LoD/LoB).

Indications for Use

Indicated for quantitative glucose measurement in human serum, plasma, urine, and CSF for diagnosis and treatment of carbohydrate metabolism disorders (diabetes mellitus, neonatal/idiopathic hypoglycemia, insulin overdose).

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k101854 B. Purpose for Submission: Change in reagent formulation to a previously cleared device (k042015) and adding another anticoagulant (K-EDTA plasma) as a sample type. C. Measurand: Glucose D. Type of Test: Quantitative enzymatic test E. Applicant: Siemens Healthcare Diagnostics Inc. F. Proprietary and Established Names: ADVIA Chemistry Glucose Hexokinase (GLUH_3) Reagent G. Regulatory Information: 1. Regulation section: 21 CFR 862.1345 Glucose Test System 2. Classification: Class II 3. Product code: CFR - Hexokinase, Glucose {1} 4. Panel: Clinical Chemistry - 75 H. Intended Use: 1. Intended use(s): Refer to indications for use, below. 2. Indication(s) for use: For in vitro diagnostic use in the quantitative determination of glucose in human serum, plasma, urine, and cerebrospinal fluid (CSF) on the ADVIA 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and insulin overdose. 3. Special conditions for use statement(s): For in vitro diagnostic use only. For prescription use only. 4. Special instrument requirements: ADVIA 1650 Chemistry System I. Device Description: The ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) device consists of two-component reagents. Reagent 1 contains buffer, ATP, and NAD. Reagent 2 contains glucose-6-phosphate dehydrogenase, hexokinase, buffer, ATP, and NAD. Reagents are ready to use and require no additional preparation. Reagents 1 and 2 contain sodium azide. J. Substantial Equivalence Information: 1. Predicate device name: ADVIA Chemistry Glucose Hexokinase II Reagent 2. Predicate 510(k) number: k042015 3. Comparison with predicate: 2 {2} | Comparison Table | | | | --- | --- | --- | | Item | New Device (k101854) | Predicate (k042015) | | Intended Use | For in vitro diagnostic use in the quantitative determination of glucose in human serum, plasma, urine, and CSF on the ADVIA 1650 Chemistry System. | same | | System Principle | Enzymatic method based on use of hexokinase and glucose-6-phosphate dehydrogenase enzymes. | same | | Detection method | Photometric – Hexokinase | same | | Calibration | Siemens calibrators | same | | Measurement Range | 4-700 mg/dL | 0-700 mg/dL | | Specimen Type | Human serum, Li-heparin plasma, K-EDTA plasma, urine, and CSF | Human serum, Li-heparin plasma, urine, and CSF | | Format | Liquid | Concentrate | | Reagents | R1 and R2 | R1, R2, R2 mix | | Storage Conditions | On-board: 60 days at 2-8°C Shelf-life: 18 months at 2-8°C | same | K. Standard/Guidance Document Referenced (if applicable): 1. CLSI EP5-A2. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline. 2. CLSI EP17-A. Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline. L. Test Principle: This enzymatic method is based on the method by Slein using hexokinase and glucose-6-phosphate dehydrogenase enzymes. The ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) method is a two-component reagent. Sample is added to Reagent 1, which contains the buffer, ATP, and NAD. Absorbance readings of the sample in Reagent 1 are taken and are used to correct for interfering substances in the sample. Reagent 2 is added, which initiates the reaction. Glucose is phosphorylated by adenosine triphosphate (ATP) in the presence of hexokinase. The glucose-6-phosphate that forms is oxidized in the presence of glucose-6-phosphate dehydrogenase causing the reduction of NAD to NADH. The absorbance of NADH is measured as an endpoint reaction at $340/410\mathrm{nm}$. The difference between the absorbance in Reagent 1 and Reagent 2 is proportional to the glucose concentration. M. Performance Characteristics (if/when applicable): 1. Analytical performance: {3} a. Precision/Reproducibility: Precision studies were performed using commercial quality controls and native samples. Studies were carried out in duplicate, twice per day, over a period of 10 to 20 days using two ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) reagent lots on two different ADVIA 1650 Chemistry Systems. A low level urine sample was assayed in duplicate on one system over a period of nine days. The mean, standard deviation (SD), and coefficients of variation (CV) were determined at each level as summarized below: | Precision Data Summary | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Sample | N | Mean (mg/dL) | Within Run | | Total | | | | | | SD (mg/dL) | CV (%) | SD (mg/dL) | CV (%) | | Serum Control 1 | 80 | 87 | 0.39 | 0.5 | 0.66 | 0.8 | | Serum Control 2 | 80 | 297 | 1.35 | 0.5 | 2.02 | 0.7 | | Serum Pool 1 | 80 | 51 | 0.18 | 0.4 | 0.46 | 0.9 | | Serum Pool 2 | 80 | 605 | 3.55 | 0.6 | 6.52 | 1.1 | | Urine Control 1 | 80 | 18 | 0.12 | 0.7 | 0.21 | 1.2 | | Urine Control 2 | 80 | 301 | 1.29 | 0.4 | 5.63 | 1.9 | | Urine Pool 1 | 80 | 23 | 0.13 | 0.6 | 0.26 | 1.2 | | Urine Pool 2 | 80 | 49 | 0.27 | 0.6 | 0.51 | 1.1 | | Urine Sample | 57 | 7 | 0.14 | 2.2 | 0.15 | 2.4 | | CSF Control 1 | 40 | 56 | 0.24 | 0.4 | 0.54 | 1.0 | | CSF Control 2 | 40 | 97 | 0.33 | 0.3 | 0.81 | 0.8 | | CSF Control 3 | 40 | 619 | 1.81 | 0.3 | 4.97 | 0.8 | b. Linearity/assay reportable range: Two linearity studies were carried out using eleven different levels of human urine samples and human serum samples. Low pools consisted of a normal human urine sample and a glucose-depleted normal human serum sample. High pools consisted of low pools spiked with glucose to concentrations of approximately 800 mg/dL. The glucose concentrations ranged from 3.5-796 mg/dL. In addition, a linearity study was performed using a CSF pool of human samples at eleven different levels covering the glucose range from 3.0-833 mg/dL. Each sample was measured in duplicate using one ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) reagent lot on one ADVIA 1650 Chemistry System. Linearity was evaluated by least-square regression analysis and comparison of observed recovery versus expected values. {4} | Linearity Data Summary | | | | | | --- | --- | --- | --- | --- | | Serum | | | | | | Linear Regression Data | Recovery Data (mg/dL) | | | | | | | Expected | Observed | % Difference | | Number of samples (n) | 11 | 3.5 | 3.5 | 0.0% | | Range (mg/dL) | 4-791 | 28.1 | 27.0 | -4.1% | | Slope | 1.005 (0.998-1.012) | 52.7 | 50.1 | -5.0% | | Intercept | -2.336 (-5.316-0.644) | 101.9 | 101.7 | -0.2% | | Correlation coefficient | 1.000 | 200.4 | 196.9 | -1.7% | | | 298.8 | 295.9 | -1.0% | | | | | 397.2 | 397.2 | 0.0% | | | | 495.7 | 492.7 | -0.6% | | | | 594.1 | 593.0 | -0.2% | | | | 692.6 | 691.3 | -0.2% | | | | 791.0 | 798.2 | 0.9% | | Urine | | | | | | Linear Regression Data | Recovery Data (mg/dL) | | | | | | | Expected | Observed | % Difference | | Number of samples (n) | 11 | 4.2 | 4.2 | 0.0% | | Range (mg/dL) | 4-796 | 29.0 | 29.4 | 1.5% | | Slope | 0.995 (0.993-0.998) | 53.7 | 55.0 | 2.4% | | Intercept | 1.146 (0.198-2.094) | 103.2 | 104.3 | 1.1% | | Correlation coefficient | 1.000 | 202.2 | 202.3 | 0.0% | | | 301.2 | 301.6 | 0.1% | | | | | 400.2 | 400.2 | 0.0% | | | | 499.2 | 498.3 | -0.2% | | | | 598.3 | 596.6 | -0.3% | | | | 697.3 | 696.1 | -0.2% | | | | 796.3 | 792.0 | -0.5% | | Cerebro Spinal Fluid (CSF) | | | | | | Linear Regression Data | Recovery Data (mg/dL) | | | | | | | Expected | Observed | % Difference | | Number of samples (n) | 11 | 3.0 | 3.0 | 0.0% | | Range (mg/dL) | 3-833 | 28.9 | 28.5 | -1.5% | | Slope | 0.994 (0.989-0.999) | 54.9 | 54.0 | -1.6% | | Intercept | 0.1937 (-2.000-2.3874) | 106.8 | 106.6 | -0.1% | | Correlation coefficient | 1.000 | 210.6 | 209.8 | -0.4% | | | 314.4 | 314.3 | 0.0% | | | | | 418.2 | 418.2 | 0.0% | | | | 522.1 | 519.7 | -0.4% | | | | 625.9 | 617.5 | -1.3% | | | | 729.7 | 727.6 | -0.3% | | | | 833.5 | 829.7 | -0.5% | {5} The claimed reportable range for the ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) reagent on the ADVIA 1650 Chemistry System is 4 to 700 mg/dL. Data was provided to demonstrate that samples resulting in glucose values above the upper range limit will be automatically diluted and reported after applying dilution correction factor. c. Traceability, Stability, Expected values (controls, calibrators, or methods): ## Traceability The ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) method is traceable to a standard reference serum (SRM 965a) glucose material from the National Institute of Standards and Technology (NIST). A second set of standard reference material, a serum-based sample, a spiked serum sample, and an aqueous glucose sample are analyzed on the ADVIA 1650 Chemistry System and recovered values must fall within acceptance criteria based on the target value of the NIST standard. ## Stability The stability of the ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) reagent was evaluated with real-time stability studies. The reagent shelf life claim is 18 months when stored at the recommended temperature of 2 to 8°C. An on-board stability study confirmed that the reagent is stable for 60 days on-board the ADVIA 1650 Chemistry System at the recommended temperature of 2 to 8°C. The manufacturer does not recommend removing open reagents from the system. ## Calibrators and Controls A calibration frequency study with the ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) method showed that the system should be calibrated with the Siemens Chemistry Calibrator at a frequency of at least every 60 days. The performance of the calibrator for the ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) method has been demonstrated in a previous 510(k) submission. The manufacturer states in the labeling that individual laboratory quality control programs and procedures may require more frequent calibration. The manufacturer also recommends the use of a commercial control material with at least two glucose levels tested according to workflow, system experience, and government regulation. d. Detection limit: The Limit of Blank (LoB) and Limit of Detection (LoD) for the ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) method were evaluated based on CLSI EP17-A guidelines. Studies were performed using N ≤ 2 reagent lots on one ADVIA 1650 Chemistry System. 6 {6} 7 | Detection Limit Data Summary | | | | --- | --- | --- | | | Limit | Protocol | | Serum | LoB | A human serum glucose-depleted sample pool was evaluated. A total of 6 runs (10 replicates per run) were completed, N=120, over a period of three days. | | | LoD | A low serum sample pool with a glucose concentration of 3-4 mg/dL was evaluated in duplicate. A total of 58 runs were over a period of nine days. | | Urine | LoB | A normal human urine pool was evaluated. A total of 6 runs (10 replicates per run) were completed, N=120, over a period of nine days. | | | LoD | A low urine sample with a glucose concentration of 6 mg/dL was evaluated in duplicate. A total of 57 runs were completed over a period of nine days. | | CSF | LoB | A human CSF glucose-depleted sample pool was evaluated. A total of 3 runs (10 replicates per run) were completed, N=60, over a period of three days. | | | LoD | A low CSF sample pool with a glucose concentration of 1.5 mg/dL was evaluated. A total of 3 runs (10 replicates per run) were completed, N=60, over a period of three days. | | Claimed Values | | | | LoB | 0.30 mg/dL | | | LoD | 4.0 mg/dL | | e. Analytical specificity: Interferences studies were performed by adding potentially interfering substances to samples and calculating bias relative to a control (no interferent). Five equally diluted samples using either human serum or human urine pools were tested by adding different interferent concentrations. Samples were measured in duplicate using one ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) reagent lot on one ADVIA 1650 Chemistry System. No significant interference defined as no more than $\pm 10\%$ was observed up to the levels shown in the table below for the following interfering substances: | Interferences Data Summary | | | | | --- | --- | --- | --- | | | No Interference at Listed Level (mg/dL) | | | | Substance | Low Serum | Mid Serum | High Serum | | | ~50 mg/dL | ~130 mg/dL | ~200 mg/dL | | Hemoglobin | 1000 | 1000 | 1000 | | Bilirubin (Unconjugated) | 30 | 30 | 30 | | Bilirubin (Conjugated) | 30 | 30 | 30 | | Triglycerides (Intralipid) | 1000 | 1000 | 1000 | {7} 8 | Interferences Data Summary | | | | --- | --- | --- | | No Interference at Listed Level (mg/dL) | | | | Substance | Low Urine | High Urine | | | ~30 mg/dL | ~100 mg/dL | | Ascorbic Acid | 200 | 200 | | Caffeine | 50 | 50 | | Creatinine | 500 | 500 | | Salicylate (Na) | 50 | 50 | | Urea Nitrogen | 500 | 500 | f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: In a method comparison study, glucose was measured in serum, urine, and CSF samples by one or two trained operators on both, the predicate device and the ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) reagent on ADVIA 1650 Chemistry Systems. Samples were analyzed in duplicate using one lot of each reagent. A number of samples were altered to cover the claimed measuring range for the device. Data were analyzed by means of least-squares linear regression analysis. Study results are summarized below. | Method Comparison Data Summary | | | | | --- | --- | --- | --- | | | Serum | Urine | CSF | | Number of samples (n) | 99 | 51 | 113 | | Range (mg/dL) | 5-676 | 4-664 | 20-654 | | Slope (95% CI) | 1.001 (0.999-1.003) | 0.989 (0.987-0.990) | 1.005 (1.004-1.006) | | Intercept (95% CI) | 0.264 (-0.171-0.698) | -0.316 (-0.648-0.015) | -0.097 (-0.313-0.120) | | Correlation coefficient | 1.000 | 1.000 | 1.000 | b. Matrix comparison: A matrix comparison study was performed by testing K-EDTA and Li-Heparin samples on both the predicate device and the ADVIA Chemistry Glucose Hexokinase_3 (GLUH_3) reagent on ADVIA 1650 Chemistry Systems, by two trained operators. A number of samples were altered to cover the claimed measuring range for the device. Data were analyzed by means of least-squares linear regression analysis. Study results are summarized below. {8} | Matrix Comparison Data Summary | | | | --- | --- | --- | | | K-EDTA plasma | Li-Heparin plasma | | Number of samples (n) | 87 | 88 | | Range (mg/dL) | 6-676 | 5-686 | | Slope (95% CI) | 1.002 (1.000-1.003) | 1.001 (0.999-1.00) | | Intercept (95% CI) | 0.002 (-0.298-0.301) | 0.184 (-0.173-0.541) | | Correlation coefficient | 1.000 | 1.000 | # 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable # 4. Clinical cut-off: Not applicable # 5. Expected values/Reference range: Expected reference ranges for the method are as follows: | Expected Values | | | | --- | --- | --- | | Sample Type | | Reference Range (mg/dL) | | Serum/plasma | Adult | 74-106 | | | Newborn 1 day | 40-60 | | | Newborn > 1 day | 50-80 | | | Child | 60-100 | | Urine | | < 0.5 g/day | | CSF | Adult | 40-70 | | | Infant/Child | 60-80 | Reference: Wu AHB. Tietz Clinical Guide to Laboratory Tests. $4^{\text{th}}$ ed. St. Louis, MO: WB Saunders Company: 2006: 444-450. {9} N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 10
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...