K153577 · Nanoentek USA, Inc. · CDZ · Oct 14, 2016 · Clinical Chemistry
Device Facts
Record ID
K153577
Device Name
FREND Testosterone Test System
Applicant
Nanoentek USA, Inc.
Product Code
CDZ · Clinical Chemistry
Decision Date
Oct 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1680
Device Class
Class 1
Indications for Use
The FREND™ Testosterone test is a fluorescent nanoparticle immunoassay designed for in vitro quantitative measurement of total testosterone in human serum and plasma (K3EDTA and lithium heparin). Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes. The FREND™ Testosterone microfluidic flow cartridge is designed for use in the FREND™ System fluorescent immunoassay reader. The FREND™ Testosterone Test System is intended for use in clinical laboratories. For in vitro diagnostic use only. The test is not intended for use in point-of-care settings.
Device Story
FREND™ Testosterone Test System is a rapid, two-step competitive fluorescence immunoassay for quantitative measurement of total testosterone in human serum and plasma. The system consists of a disposable microfluidic cartridge and a benchtop fluorescence reader. Operation involves manual incubation of a 70 µL patient sample with a pretreatment tube, followed by loading 35 µL into the cartridge. The cartridge uses microfluidic lateral flow where testosterone in the sample competes with testosterone-biotin fluorescent bead conjugates for binding sites in the test zone. The reader measures laser-induced fluorescence in the test and reference zones; the ratio of these intensities is inversely proportional to testosterone concentration. Results are calculated via a lot-specific calibration curve provided by a code-chip. Used in clinical laboratories by trained personnel to assist physicians in diagnosing androgen-related disorders. The system provides automated analysis, displaying results on a touchscreen, aiding clinical decision-making regarding hormone-related conditions.
Clinical Evidence
Bench testing only. Precision study (CLSI EP5-A3) showed CVs 5.4-11.8%. Linearity established 20-1500 ng/dL (R²=0.994). LoQ 20 ng/dL. Interference and cross-reactivity testing performed per CLSI EP7-A2. Method comparison with predicate (n=155) yielded Passing-Bablok slope 0.983, intercept -2.353, r=0.977. Matrix comparison (serum vs. K3EDTA/Li-Heparin plasma) showed high correlation (r=0.989-0.993). Reference intervals established using 488 healthy adults.
Technological Characteristics
Fluorescence immunoassay using microfluidic lateral flow technology. Disposable plastic cartridge contains reagents; reader uses laser-induced fluorescence detection. Analyte quantification based on ratio of test zone to reference zone fluorescent intensity. Calibration via lot-specific code-chip. Benchtop form factor with touchscreen interface. Software controls hardware, data acquisition, and analysis. No specific ASTM material standards cited.
Indications for Use
Indicated for in vitro quantitative measurement of total testosterone in human serum and plasma (K3EDTA and lithium heparin) to aid in diagnosis/treatment of androgen-related disorders, including primary/secondary hypogonadism and impotence in males, and hirsutism/virilization in females. For prescription use only in clinical laboratories; not for point-of-care settings.
Regulatory Classification
Identification
A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.
K021972 — FASTPACK TOTAL TESTOSTERONE IMMUNOASSAY · Qualigen, Inc. · Aug 21, 2002
K101388 — FASTPACK TESTO IMMUNOASSAY · Qualigen, Inc. · Feb 18, 2011
K151529 — Dimension Vista LOCI Total Testosterone Flex reagent cartridge, Dimension Vista Testosterone Calibrator · Siemens Healthcare Diagnostics, Inc. · Feb 11, 2016
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k153577
B. Purpose for Submission:
New Device
C. Measurand:
Testosterone
D. Type of Test:
Quantitative, Fluorescence Immunoassay
E. Applicant:
NanoEnTek, Inc.
F. Proprietary and Established Names:
FREND™ Testosterone Test System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1680 Testosterone Test System
2. Classification:
Class I, Reserved
3. Product code:
CDZ
4. Panel:
Clinical Chemistry (75)
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below
2. Indication(s) for use:
The FREND™ Testosterone test is a fluorescent nanoparticle immunoassay designed for in vitro quantitative measurement of total testosterone in human serum and plasma (K3EDTA and lithium heparin). Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.
The FREND™ Testosterone microfluidic flow cartridge is designed for use in the FREND™ System fluorescent immunoassay reader. The FREND™ Testosterone Test System is intended for use in clinical laboratories. For in vitro diagnostic use only. The test is not intended for use in point-of-care settings.
3. Special conditions for use statement(s):
For prescription use only
Not for use in Point-of-Care settings
4. Special instrument requirements:
NanoEnTek FREND™ System
I. Device Description:
The FREND™ Testosterone Test System includes the following in the kit:
- FREND™ Testosterone cartridges (20)
- Testosterone Gold Antibody Pretreatment Tubes (20)
- Disposable pipette tips (30)
- FREND™ Testosterone Code Chip (1)
- FREND™ Testosterone Package Insert (1)
One cartridge contains:
- Monoclonal mouse anti-testosterone antibody: 32±3.2 ng
- Testosterone-Bovine Serum Albumin (BSA): 482.0±48.2 ng
- Fluorescent nanoparticles: 3.6±0.36 µg
{2}
One pretreatment tube contains:
- Gold antibody conjugates: 30±3.0 µg
The FREND Testosterone Code Chip contains all the calibration information stored electronically, and is specific for each lot of FREND Testosterone cartridge.
The FREND™ System was previously cleared in k124056 and is not provided with the kit but is required for the use of the FREND™ Testosterone test cartridge. The FREND System is a portable, bench top fluorescence reader.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Abbot ARCHTECT 2nd Generation Testosterone
2. Predicate 510(k) number(s):
k120009
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: FRENDTM Testosterone Test System (k153577) | Predicate Device: Abbott ARCHITECT 2nd Generation Testosterone (k120009) |
| Intended Use | In vitro diagnostic test for the quantitative measurement of total testosterone in human serum and plasma (K_{3}EDTA and Lithium heparin) | Same |
| Sample Type | Human serum and plasma (K_{3}EDTA and Lithium heparin) | Same |
| Type of Test | Immunoassay | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device: FRENDTM Testosterone Test System (k153577) | Predicate Device: Abbott ARCHITECT 2nd Generation Testosterone (k120009) |
| Methodology | Fluorescent | Chemiluminescent |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device: FRENDTM Testosterone Test System (k153577) | Predicate Device: Abbott ARCHITECT 2^{nd} Generation Testosterone (k120009) |
| Sample Size | 70 μL for the first incubation step and 35 μL for running the test | 150 μL for the first Testosterone test and 100 μL for each additional test from the same test cup |
| Test Cartridge | Disposable single-use cartridge | No single-use cartridge |
| Measuring Range | 20 – 1500 ng/dL | 4.33 – 1500 ng/dL |
## K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3, Evaluation of precision Performance of Quantitative Measurement Methods; Approved Guideline
CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures; A Statistical Approach; Approved Guideline
CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline
CLSI EP09-A3, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline
CLSI EP14-A3, Evaluation of Commutability of Processed Samples; Approved Guideline
CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guidelines
CLSI EP28-A3C, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline
## L. Test Principle:
The FREND™ Testosterone Test is a single-use rapid “competitive” immunoassay utilizing fluorescent nanoparticle in microfluidic flow to capture and quantify total testosterone levels in human serum and plasma (K3EDTA and lithium heparin) specimens using the FREND™ system.
The FREND™ Testosterone Test System is a two-step competitive immunoassay with gold nanoparticles labeled with mouse monoclonal anti-testosterone antibody, testosterone-biotin
{4}
labeled with fluorescence nanoparticles and fluorescence detection by FREND™ System.
The FREND™ Testosterone test utilizes microfluidic technology and detects immune-complexes bound to testosterone. Firstly, a $70~\mu \mathrm{L}$ patient sample is incubated in the Testosterone Gold Antibody pretreatment tube, where the sample interacts with a proprietary mix of a pretreatment solution. Secondly, the Test Cartridge is placed on the warming platform of the heating block and $35~\mu \mathrm{L}$ of the mixture from Step 1 is manually loaded into the inlet of the cartridge. The cartridge remains on the warming platform for 30 seconds, while the sample hydrates the testosterone-biotin fluorescent bead conjugate and migrates along the test strip. During migration the bound testosterone in the sample and the testosterone-biotin fluorescent bead conjugates compete to form antigen-antibody complexes in the test zone. Unbound testosterone-biotin fluorescent conjugates flow through and bind to the anti-testosterone antibody that is immobilized on the surface in the reference zone. The cartridge is inserted into the FREND instrument for analysis where fluorescent signals in the test and reference zones are measured.
Testosterone quantification is based upon the ratio of the intensity of the test and reference zones. The magnitude of the fluorescent ratio is inversely proportional to the amount of testosterone in the sample.
There is no calibration needed by the user because each cartridge is coded with the calibration information generated by the manufacturer.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
A single lot imprecision study was performed internally as described in the CLSI protocol EP5-A3. Three serum pools with low, intermediate and high testosterone levels were assayed in duplicate twice per day for 20 days (80 total measurements per level). The results are summarized below:
| Sample Pool | Mean Conc. (ng/dL) | Repeatability | | Within Laboratory | |
| --- | --- | --- | --- | --- | --- |
| | | SD | CV% | SD | CV% |
| 1 | 39.723 | 4.5 | 11.2 | 4.7 | 11.8 |
| 2 | 202.965 | 16.7 | 8.2 | 17.3 | 8.5 |
| 3 | 1012.208 | 54.7 | 5.4% | 57.5 | 5.7 |
# b. Linearity/assay reportable range:
Linearity was established according to CLSI-EP6-A in three reagent lots using seven levels of serum testosterone tested in quadruplicate. The range for analyte concentration was tested from $17\mathrm{ng / dL}$ to $1650\mathrm{ng / dL}$ using low and high serum pools. The low and high pools were mixed to make five intermediate levels. A total of seven concentrations were tested (17.5, 180.8, 527.2, 850.2, 1203.6, 1509.0, and $1736.6\mathrm{ng / dL}$ ). The following linear regression equation was obtained:
{5}
$$
\mathrm {Y} = 1. 0 3 8 \mathrm {X} - 4. 8 1 8, \mathrm {R} ^ {2} = 0. 9 9 4
$$
Linearity data supports the FREND™ Testosterone reportable range of 20 ng/dL to 1500 ng/dL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The standards/calibrators are prepared gravimetrically and confirmed by measurement on the ARCHITECT i 2nd Generation Testosterone assay (k120009), which is traceable to the USP testosterone. The calibrators are factory calibration.
There is no need for calibration by the operator as the calibration information is coded in the individual cartridge.
Controls: The sponsor recommends the use of commercially available quality control material that contains Testosterone as the measurand.
d. Detection limit:
LoB, LoD and LoQ studies were performed according to the CLSI EP17-A guideline using one lot of reagent.
LoB was performed with five different steroid depleted serum samples with 12 replicates each, which generated a total of 60 replicates.
The Limit of Detection (LoD) for the FREND™ Testosterone was established using five different low level samples. All samples were each assayed 12 times over a period of five days using a single reagent lot. Each sample has a total of 60 replicates.
LoQ was determined based on five low samples with concentration between 17 to 20 ng/dL. The sponsor obtained a LoQ value based on <20%CV at sample concentration of 20 ng/dL.
| LoB | LoD | LoQ |
| --- | --- | --- |
| 9.46 ng/dL | 14.33 ng/dL | 20.0 ng/dL |
The claimed measuring range of this assay is 20 ng/dL to 1500 ng/dL.
e. Analytical specificity:
Interference Studies
Potential interference from common endogenous substances was evaluated in accordance with CLSI EP7-A2. Interferents were added to normal human sera spiked with two levels of testosterone, and controls with no interferent were included. The sponsor defines significant interference as > ±10% difference. No significant interference was found if recoveries were between 90% to 110% of the expected
{6}
testosterone value. Results with non-significant interference substances are summarized in the table below.
| Class of substances tested | Interferent (Concentration Tested) |
| --- | --- |
| Endogenous Substances | Hemoglobin (500 mg/dL) |
| | Bilirubin, conjugated (30 mg/dL) |
| | Bilirubin, unconjugated (30 mg/dL) |
| | Triglyceride (3000 mg/dL) |
| | Total protein (12 g/dL) |
| | Biotin (30 ng/mL) |
| | Sex Hormone Binding Globulin (100 nmol/L) |
| Pharmaceuticals | Acetylcystein (415 μg/mL) |
| | Ampicillin-Na (50.3 μg/mL) |
| | Ascorbic acid (60 μg/mL) |
| | Ca-Dobesilate (40 μg/mL) |
| | Cyclosporine (3 μg/mL) |
| | Cefoxitin (66 μg/mL) |
| | Heparin (3,000 U/L) |
| | Levodopa (4 mg/mL) |
| | Methyldopa (15 μg/mL) |
| | Metronidazole (120 μg/L) |
| | Doxycycline (30 μg/mL) |
| | Acetylsalicylic Acid (250 μ/g/mL) |
| | Rifampin (640 μg/mL) |
| | Acetaminophen (200 μg/mL) |
| | Ibuprofen (250 μg/mL) |
| | Theophylline (400 μg/mL) |
| Heterophilic Antibodies | Rheumatoid Factor (RF) (1075 IU/mL) |
| | Human Anti-mouse antibodies (HAMA) (70 ng/mL) |
Sponsor has included the following limitation in the labeling:
"Specimens from patients with heterophilic antibodies, such as anti-mouse (HAMA), anti-goat (HAGA), or anti-rabbit (HARA) antibodies, may show falsely elevated or depressed values or may result in an incomplete test. Patients routinely exposed to animals or animal serum products can be prone to these types of heterophilic interference."
## Cross Reactivity
The following substances were evaluated for potential cross-reactivity with the FREND™ Testosterone at two concentrations (low and high testosterone levels). Testing was done according to the CLSI protocol EP07-A2. Results of the percent cross reactivity are summarized in the table below.
{7}
| Cross-reactant | Cross-reactant Conc. (ng/dL) | % Cross-reactivity | |
| --- | --- | --- | --- |
| | | Testosterone Low | Testosterone High |
| Androstenedione | 50 nmol/L | 0.0032 | 0.0337 |
| Androsterone | 1,000 nmol/L | 0.0004 | 0.0158 |
| Cortisone | 1,000 nmol/L | 0.0077 | 0.0300 |
| Danazol | 1,000 nmol/L | 0.0007 | 0.0496 |
| Estradiol | 200 nmol/L | 0.0033 | 0.5160 |
| Estrone | 500 nmol/L | 0.0144 | 0.0378 |
| 17a-Ethynyl estradiol | 1,000 ng/mL | 0.0033 | 0.0081 |
| Progesterone | 2,000 nmol/L | 0.0007 | 0.0167 |
| Dexamethasone | 5 μmol/L | 0.0255 | 0.0122 |
| Ethisterone | 20 nmol/L | 0.0437 | 0.3840 |
| D(-) Norgestrel, | 20 ng/mL | 0.0125 | 0.1667 |
| Prednisolone | 2,000 nmol/L | 0.0001 | 0.0175 |
| Prednisone | 2,000 nmol/L | 0.0031 | 0.0040 |
| Spironolactone | 500 ng/mL | 0.0039 | 0.0172 |
| Cortisol | 10,000 nmol/L | 0.0004 | 0.0070 |
| DHEA | 50 nmol/L | 0.0178 | 1.0463 |
| DHEAS | 50 μmol/L | 0.0002 | 0.0009 |
| Dihydrotestosterone | 40 nmol/L | 0.2040 | 0.4134 |
| Epitestosterone | 100 nmol/L | 0.0467 | 0.1502 |
| Ethynodiol diacetate | 50 ng/mL | 0.0274 | 0.1280 |
% Cross-reactivity = 100 x ((Spiked value – unspiked value)/concentration of Cross-reactant)
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Comparison studies using 155 serum samples were performed externally. The comparator method was the Abbott ARCHITECT 2nd Generation Testosterone Assay (predicate device) run on the Abbott ARCHITECT i System. The samples spanned the measuring range of the FREND Testosterone Test System (20.84 to 1420.33 ng/dL). All samples were assayed using serum samples split between the candidate device and the comparator method.
Results from the FREND™ Testosterone on the FREND™ System (y) were compared with the comparator results (x) by Passing-Bablok regression analysis, giving a slope of 0.983 (95%CI: 0.944 to 1.00), an intercept of -2.353 (95%CI: -12.0 to 14.36) and a correlation coefficient of 0.977.
{8}
b. Matrix comparison:
The matrix comparison study was performed internally. Testosterone concentrations in the serum, lithium heparin plasma and K3EDTA plasma from 40 individuals were measured using the FREND™ Testosterone Test System. Regression analysis results when compared against the serum samples and are summarized in the table below.
| Parameter | K_{3}EDTA Plasma | Lithium Heparin Plasma |
| --- | --- | --- |
| Slope: (95% CI) | 1.006 (0.962 to1.066) | 1.010 (0.968 to 1.051) |
| y-Intercept: (95% CI) | -8.185 (-46.842 to 9.627) | 1.264 (-14.078 to 22.488) |
| Correlation, r | 0.989 | 0.993 |
| Sample range tested | 23.5 to 1356.2 ng/dL | 20.3 to 1343.6 ng/dL |
Sponsor has the following recommendation in the labeling:
Patient samples are "recommended to use the same specimen matrix when following patients because the results may not be interchangeable".
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Serum samples from a total of 488 normal, apparently healthy adult male and female individuals were assayed on 3 lots of the FREND™ Testosterone assay using two FREND™ Systems according to CLSI C28-A3 guidelines. The central 95th percentile range of each sample category was calculated by finding the 2.5th
{9}
ranked sample on the low end and the 97.5th ranked sample on the high end. The reference interval using 95% interval for the FREND™ Testosterone Test System, stratified by age range and gender, is provided in the table below.
| Gender (Age Range) | Testosterone, ng/dL | Number Tested |
| --- | --- | --- |
| F (21 – 49) | <20 – 107.5 | 120 |
| F (50 – 90) | <20 – 150.3 | 124 |
| M (21 – 49) | 170.1 – 1263.6 | 123 |
| M (50 – 88) | 152.4 – 1095.2 | 121 |
It is recommended that each laboratory establish its own expected values for testosterone as performed on the FREND System.
## N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.