IMX FREE TRIIODOTHYRONINE (FREE T3)
Device Facts
| Record ID | K963833 |
|---|---|
| Device Name | IMX FREE TRIIODOTHYRONINE (FREE T3) |
| Applicant | Abbott Laboratories |
| Product Code | CDP · Clinical Chemistry |
| Decision Date | Oct 30, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1710 |
| Device Class | Class 2 |
Indications for Use
IMx Free T₃ is a microparticle enzyme immunoassay for the quantitative determination of free T₃ in human serum or plasma (lithium heparin, sodium heparin, or dipotassium EDTA).
Device Story
Microparticle enzyme immunoassay (MEIA) for quantitative free T₃ measurement; utilizes Abbott IMx Analyzer; input: human serum or plasma; output: quantitative free T₃ concentration (pg/mL); used in clinical laboratory settings by trained personnel; results assist clinicians in evaluating thyroid status; dynamic range 1.1-30.0 pg/mL.
Clinical Evidence
Bench testing only. Correlation study comparing IMx Free T₃ to Ciba Corning ACS Free T₃ using 767 specimens. Results: r=0.942, slope=1.14, standard error of estimate=0.783, Y-intercept=-0.47 pg/mL.
Technological Characteristics
Microparticle enzyme immunoassay (MEIA) technology; utilizes Abbott IMx Analyzer; calibrated with Abbott calibrators; verified with Abbott controls.
Indications for Use
Indicated for quantitative determination of free T₃ in human serum or plasma (lithium heparin, sodium heparin, or dipotassium EDTA) for clinical assessment of thyroid function.
Regulatory Classification
Identification
A total triiodothyronine test system is a device intended to measure the hormone triiodothyronine in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases such as hyperthyroidism.
Predicate Devices
- Ciba Corning Automated Chemiluminescence System (ACS) Free T₃ Immunoassay
Related Devices
- K963827 — AXSYM FREE TRIIODOTHYRONINE (FREE T3) · Abbott Laboratories · Oct 30, 1996
- K983439 — ABBOTT ARCHITECT FREE T3 · Abbott Laboratories · Nov 20, 1998
- K960913 — ACCESS FREE T3 ASSAY · Bio-Rad Laboratories, Inc. · Mar 29, 1996
- K983434 — ABBOTT ARCHITECT TOTAL T3 · Abbott Laboratories · Nov 20, 1998
- K962355 — FREE TRIIODOTHYRONINE(FT3) ASSAY FOR THE TECHNICON IMMUNO 1 SYSTEM (IN-VITRO DIAGNOSTIC SYSTEM) · Bayer Corp. · Jul 12, 1996