21 CFR 862.1493 — Breast Milk Macronutrients Test System

Clinical Chemistry (CH) · Part 862 Subpart B—Clinical Chemistry Test Systems · § 862.1493

Identification

A breast milk macronutrients test system is a device intended to quantitatively measure fat, protein, and total carbohydrate content in human breast milk. These measurements, in conjunction with other clinical assessments, may be used to aid in the nutritional management of infants.

Classification Rationale

Class II (special controls). The special controls for this device are:

Product Codes

Product CodeDevice NameClassDevicesAttributes
QEIBreast Milk Macronutrients Test System23

Special Controls

QEI — Breast Milk Macronutrients Test System

A breast milk macronutrient test system must comply with the following special controls:

De Novo Order DEN180007

QEI — Breast Milk Macronutrients Test System

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following: (i) An appropriate traceability plan, as determined by FDA, to minimize the risk of drift in the breast milk macronutrient test system results over time. (ii) Data that demonstrate appropriate precision, as determined by FDA, of the breast milk macronutrients test system. Precision studies must include assessment of a minimum of three breast milk specimens containing different concentrations (low, medium, and high levels) of fat, carbohydrates, and protein. Precision data must include breast milk specimen measurements that are collected at a minimum of three laboratory sites. (iii) Data that demonstrate appropriate measurement accuracy, as determined by FDA, of fat, carbohydrates, and protein in breast milk. Measurement accuracy data must include breast milk specimen measurements that are collected at a minimum of one laboratory site. (iv) Data from studies appropriate, as determined by FDA, to demonstrate that the device is free from significant interference from substances that could be present in human milk, including hemoglobin, and medications that are used by breastfeeding subjects. (2) The labeling required under § 809.10 of this chapter must include a limiting statement indicating that the results should be used only as an aid in the nutritional management of infants and not as the sole basis for making nutrition decisions.

eCFR

QEI — Breast Milk Macronutrients Test System

(1) Design verification and validation must include the following: (i) An appropriate traceability plan, as determined by FDA, to minimize the risk of drift in the breast milk macronutrient test system results over time. (ii) Data that demonstrate appropriate precision, as determined by FDA, of the breast milk macronutrients test system. Precision studies must include assessment of a minimum of three breast milk specimens containing different concentrations (low, medium, and high levels) of fat, carbohydrates, and protein. Precision data must include breast milk specimen measurements that are collected at a minimum of three laboratory sites. (iii) Data that demonstrate appropriate measurement accuracy, as determined by FDA, of fat, carbohydrates, and protein in breast milk. Measurement accuracy data must include breast milk specimen measurements that are collected at a minimum of one laboratory site. (iv) Data from studies appropriate, as determined by FDA, to demonstrate that the device is free from significant interference from substances that could be present in human milk, including hemoglobin, and medications that are used by breastfeeding subjects. (2) The labeling required under § 809.10 of this chapter must include a limiting statement indicating that the results should be used only as an aid in the nutritional management of infants and not as the sole basis for making nutrition decisions.

Ecfr Llm

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...