5.0 Fr Arndt Endobronchial Blocker Set

K160542 · Cook Incorporated · CBI · Aug 2, 2016 · Anesthesiology

Device Facts

Record IDK160542
Device Name5.0 Fr Arndt Endobronchial Blocker Set
ApplicantCook Incorporated
Product CodeCBI · Anesthesiology
Decision DateAug 2, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5740
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The 5.0 Fr Arndt Endobronchial Blocker Set is intended to differentially intubate a patient's bronchus in order to isolate the left or right lung for procedures that require one-lung ventilation. The 5.0 Fr Arndt Endobronchial Blocker is indicated for pediatric populations, in children 1 year and older. The 9.0 Fr Arndt Endobronchial Blocker Set is intended to differentially intubate a patient's bronchus in order to isolate the left or right lung for procedures that require one-lung ventilation. The 9.0 Fr Arndt Endobronchial Blocker is indicated for adult use only.

Device Story

Device consists of blocker catheter with distal silicone balloon, proximal one-way valve, and pilot balloon assembly. Nylon guide loop at distal end attaches to bronchoscope tip for precise placement. Includes Arndt Multiport Airway Adapter, proximal straight connector, suction adapter, and syringe. Used in clinical settings (e.g., OR) by physicians to isolate lungs during procedures requiring one-lung ventilation. Bronchoscope assists in navigation. Output is physical isolation of lung via balloon inflation. Benefits include controlled lung collapse for surgical access.

Clinical Evidence

Bench testing only. Testing included tensile strength, balloon shear loading, pressurization, inflation diameter, and burst testing. All tests performed at zero-time and after three-year accelerated aging. Results met all established acceptance criteria.

Technological Characteristics

Materials: Nylon catheter shaft, silicone balloon/pilot assembly, polyamide/acetal components, PEEK reinforcement. Sensing/Actuation: Mechanical balloon inflation via air. Dimensions: 5.0 Fr and 9.0 Fr diameters; 50-78 cm lengths. Connectivity: None. Sterilization: Ethylene Oxide (EtO).

Indications for Use

Indicated for differential intubation of a patient's bronchus to isolate the left or right lung for one-lung ventilation. 5.0 Fr version indicated for pediatric patients 1 year and older; 9.0 Fr version indicated for adult use only.

Regulatory Classification

Identification

A tracheal/bronchial differential ventilation tube is a device used to isolate the left or the right lung of a patient for anesthesia or pulmonary function testing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 2, 2016 Cook Incorporated Kotei Aoki Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404 Re: K160542 Trade/Device Name: 5.0 Fr Arndt Endobronchial Blocker Set 9.0 Fr Arndt Endobronchial Blocker Set Regulation Number: 21 CFR 868.5740 Regulation Name: Tracheal/Bronchial Differential Ventilation Tube Regulatory Class: Class II Product Code: CBI Dated: July 1, 2016 Received: July 5, 2016 Dear Kotei Aoki: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, *Tejashri Purohit-Sheth, M.D.* Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160542 #### Device Name 5.0 Fr Arndt Endobronchial Blocker Set 9.0 Fr Arndt Endobronchial Blocker Set #### Indications for Use (Describe) The 5.0 Fr Arndt Endobronchial Blocker Set is intended to differentially intubate a patient's bronchus in order to isolate the left or right lung for procedures that require one-lung ventilation. The 5.0 Fr Arndt Endobronchial Blocker is indicated for pediatric populations, in children 1 year and older. The 9.0 Fr Arndt Endobronchial Blocker Set is intended to differentially intubate a patient's bronchus in order to isolate the left or right lung for procedures that require one-lung ventilation. The 9.0 Fr Arndt Endobronchial Blocker is indicated for adult use only. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| | | X | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in large, white, sans-serif font at the top, and below it, the word "MEDICAL" is in a smaller, white, sans-serif font. # 510(k) Summary ## 5.0 Fr Arndt Endobronchial Blocker Set 9.0 Fr Arndt Endobronchial Blocker Set 21 CFR §868.5740 Date Prepared: August 1, 2016 #### Submitted By: | Applicant: | Cook Incorporated | |-----------------------|--------------------------------------------------| | Contact: | Kotei Aoki | | Applicant Address: | Cook Incorporated | | | 750 Daniels Way | | | Bloomington, IN 47404 | | Contact Phone Number: | (812) 335-3575 x102630 | | Contact Fax Number: | (812) 332-0281 | | Device Information: | | | Trade Name: | 5.0 Fr Arndt Endobronchial Blocker Set | | | 9.0 Fr Arndt Endobronchial Blocker Set | | Common Name: | Endobronchial Blocker | | Classification Name: | Tracheal/bronchial differential ventilation tube | | Regulation | 21 CFR §868.5740 | | Product Code | CBI | ## Predicate Devices: - K962167, Bronchial Blocker ● - . K002288, Arndt Pediatric Endobronchial Blocker ## Device Description: The 5.0 Fr Arndt Endobronchial Blocker Set and the 9.0 Fr Arndt Endobronchial Blocker Set are comprised of a blocker catheter with a silicone balloon near the distal end. The proximal end of the blocker catheter is bonded with a one-way valve and a pilot balloon assembly. A nylon guide loop at the distal end of the blocker catheter is wrapped around the tip of the bronchoscope and tightened, in order for the bronchoscope to assist in the precise placement of the blocker catheter. An Arndt Multiport Airway Adapter, a proximal straight connector, a suction adapter, and a syringe are also provided with each set. {4}------------------------------------------------ ### Intended Use: The 5.0 Fr Arndt Endobronchial Blocker Set is intended to differentially intubate a patient's bronchus in order to isolate the left or right lung for procedures that require onelung ventilation. The 5.0 Fr Arndt Endobronchial Blocker is indicated for pediatric populations, in children 1 year and older. The 9.0 Fr Arndt Endobronchial Blocker Set is intended to differentially intubate a patient's bronchus in order to isolate the left or right lung for procedures that require onelung ventilation. The 9.0 Fr Arndt Endobronchial Blocker is indicated for adult use only. #### Comparison to Predicates: The proposed 5.0 Fr Arndt Endobronchial Blocker Set is substantially equivalent to the predicate device the Arndt Pediatric Endobronchial Blocker (K002288) in that these devices have the same intended use, fundamental technological characteristics, method of operation, and materials of construction. The proposed 9.0 Fr Arndt Endobronchial Blocker Set is substantially equivalent to the predicate device the Bronchial Blocker (K962167) in that these devices have the same intended use, fundamental technological characteristics, method of operation, and similar materials of construction. The substantial equivalence of the modified devices to the predicate device is supported by testing. A substantial equivalence comparison of each proposed device is provided in Tables 1 and 2, respectively. {5}------------------------------------------------ | | | Predicate Device | Subject Device | |-----------------------------------------------|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| | | | Arndt Pediatric Endobronchial<br>Blocker K002288 | 5.0 Fr Arndt Endobronchial<br>Blocker Set | | Regulation Number | | 21 CFR §868.5740 | Identical | | Product Code | | CBI | Identical | | Classification Name | | Tracheal/bronchial differential<br>ventilation tube | Identical | | Class | | II | Identical | | Intended Use | | Intended for use to differentially<br>intubate a patient's bronchus in<br>order to isolate the left or right<br>lung for procedures which require<br>one-lung ventilation. | Identical* | | | Material | Nylon | Identical | | | Centimeter Marker | Yes | Identical | | Catheter | Inner Diameter (in) | 0.028 | Identical | | Shaft | Outer Diameter (Fr) | 5.0 | Identical | | | Length (cm) | 50 | 50, 65 | | | Tip Shape | Straight | Identical | | | Number of Sideports | None | Identical | | | Snare Material | Nylon | Identical | | | Adapter Material | Polyamide | Identical | | Guide | Cap Material | Acetal | Identical | | Loop<br>Snare | Usable Loop length along<br>Semi-Major Axis (mm) | 5 | 10 | | | May Be Reinserted<br>(reinforcement material) | No | Identical | | | Material | Silicone | Identical | | | Length (mm) | 8 | Identical | | Distal | Shape | Elliptical, Spherical | Spherical | | Balloon | Inflation Media | Air | Identical | | | Inflation Volume (cc) | 2 | 0.5 – 2 | | | Inflation Diameter (in) | 0.150 | Identical | | Manifold | Y-shape fitting Material | Polyamide | Identical | | Assembly | Pilot Balloon Assembly<br>Material | Silicone | Identical | | | Used in Conjunction with<br>Endotracheal Tube | Yes | Identical | | Used in Conjunction with<br>Bronchoscope | | Yes | Identical | | | Accessory Components | Arndt Multiport Airway Adapter,<br>Proximal Straight Connector,<br>Syringe | Arndt Multiport Airway<br>Adapter, Proximal Straight<br>Connector, Syringe, Suction<br>Adapter | | Packaging | | Tyvek/polyethylene film pouch | Identical | | Sterilization | | EtO | Identical | | | | Predicate Device<br>Bronchial Blocker<br>K962167 | Subject Device<br>9.0 Fr Arndt Endobronchial<br>Blocker Set | | Regulation Number | | 21 CFR §868.5740 | Identical | | Product Code | | CBI | Identical | | Classification Name | | Tracheal/bronchial differential<br>ventilation tube | Identical | | Class | | II | Identical | | Intended Use | | Intended for use to differentially<br>intubate a patient's bronchus in<br>order to isolate the left or right<br>lung for procedures which require<br>one-lung ventilation. | Identical | | Catheter<br>Shaft | Material | Polyethylene | Nylon | | | Centimeter Marker | Yes | Identical | | | Inner Diameter (in) | 0.068 | Identical | | | Outer Diameter (Fr) | 9.0 | Identical | | | Length (cm) | 78 | 65, 78 | | | Tip Shape | Straight | Identical | | | Number of Sideports | None | 2 | | Guide<br>Loop<br>Snare | Snare Material | Nylon | Identical | | | Adapter Material | Acetal | Polyamide | | | Cap Material | Acetal | Identical | | | Usable Loop length along<br>Semi-Major Axis (mm) | 15 | Identical | | | May Be Reinserted<br>(reinforcement material) | No | Yes (Polyetheretherketone) | | Distal<br>Balloon | Material | Silicone | Identical | | | Length (mm) | 23 | Identical | | | Shape | Elliptical, Spherical | Spherical | | | Inflation Media | Air | Identical | | | Inflation Volume (cc) | 6 | 4 - 8 | | | Inflation Diameter (in) | 0.350 - 0.450 | Identical | | Manifold<br>Assembly | Y-shape fitting Material | Acetal | Polyamide | | | Pilot Balloon Assembly<br>Material | Silicone | Identical | | Used in Conjunction with<br>Endotracheal Tube | | Yes | Identical | | Used in Conjunction with<br>Bronchoscope | | Yes | Identical | | Accessory Components | | Arndt Multiport Airway Adapter,<br>Proximal Straight Connector,<br>Syringe | Arndt Multiport Airway<br>Adapter, Proximal Straight<br>Connector, Syringe, Suction<br>Adapter | | Packaging | | Tyvek/polyethylene film pouch | Identical | | Sterilization | | EtO | Identical | Table 1 Substantial Equivalence Comparison – 5.0 Fr Arndt Endobronchial Blocker Set * The 5.0 French Arndt Endobronchial Blocker is indicated for pediatric populations, in children 1 year and older. The predicate (K002288) was also indicated for pediatric use. {6}------------------------------------------------ {7}------------------------------------------------ ### Technological Characteristics: The following tests have been conducted to ensure reliable design and performance under the specified design requirements: - Tensile Tests (the Catheter Shaft, the Distal Sideport, the Snare-to-Proximal . Manifold Bond, the Hub-to-Shaft Bond, and the Pilot Balloon Assembly-to-Manifold) (Zero-Time, Three-Year Accelerated Aged) - The peak load value of the test articles was demonstrated to be greater than or equal to the established acceptance criteria. - Analysis of Re-insertion Method of Snare (Zero-Time) This test was performed . for characterization only. - . Balloon Shear Loading Test (Zero-Time, Three-Year Accelerated Aged) – The test articles do not completely separate from the catheter shaft (both proximal and distal balloon bonds fail, i.e.) or, if the balloon ruptures, not fragment. The acceptance criteria were met. - . Pressurization Testing (Zero-Time, Three-Year Accelerated Aged) - The test articles do not exhibit signs of leakage or burst while pressurized. The acceptance criteria were met. - Balloon Inflation Test (Zero-Time. Three-Year Accelerated Aged) The inflation . diameter of the test articles were demonstrated to be greater than the established acceptance criteria. - Balloon Burst Testing (Zero-Time, Three-Year Accelerated Aged) The . minimum failure volume of the test articles were demonstrated to be greater than the established acceptance criteria. The results of these tests support a conclusion that the 5.0 Fr Arndt Endobronchial Blocker Set and the 9.0 Fr Arndt Endobronchial Blocker Set met the design input requirements based on the intended use and support the conclusion that these devices do not raise new questions of safety or effectiveness as compared to the predicate device. The subject devices are therefore substantially equivalent to the predicate devices, the Bronchial Blocker (Cook Incorporated, K962167) and the Arndt Pediatric Endobronchial Blocker (Cook Incorporated, K002288).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...