QAV · High Flow/High Velocity Humidified Oxygen Delivery Device

Anesthesiology · 21 CFR 868.5454 · Class 2

Overview

Product CodeQAV
Device NameHigh Flow/High Velocity Humidified Oxygen Delivery Device
Regulation21 CFR 868.5454
Device ClassClass 2
Review PanelAnesthesiology

Identification

A high flow humidified oxygen delivery device is a prescription device that delivers high flow oxygen with humidification for patients who are suffering from respiratory distress and/or hypoxemia.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the high flow humidified oxygen delivery device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, including the following: (i) Alarm testing must be performed; (ii) Continuous use thermal stability testing must be performed; (iii) Humidity output testing must be performed; and (iv) Blender performance testing must evaluate fraction of inspired oxygen ( *Fi* O2 ) blending accuracy.(3) Performance data must validate cleaning instructions for any reusable components of the device. (4) Electrical safety, thermal safety, mechanical safety, electromagnetic compatibility, and radiofrequency identification testing must be performed. (5) Software verification, validation, and hazard analysis must be performed. (6) Labeling must include: (i) A description of available *Fi* O2 ranges for different flowrates and inlet gas pressures;(ii) Instructions for applicable flowrates for all intended populations; (iii) A warning that patients on high flow oxygen are acute and require appropriate monitoring, to include pulse oximetry; (iv) A warning regarding the risk of condensation at low set temperatures and certain flows; and (v) A description of all alarms and their functions.

Recent Cleared Devices (3 of 3)

RecordDevice NameApplicantDecision DateDecision
K221338F&P Airvo 3Fisher & Paykel HealthcareJan 27, 2023SESE
K221318HVT 2.0Vapotherm, Inc.Dec 13, 2022SESE
DEN170001Precision Flow® HVNIVapotherm, Inc.Apr 10, 2018DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...