BREATHE TECHNOLOGIES VENTILATOR AND ACCESSORIES BT-V2S

K100528 · Breathe Technologies, Inc. · ONZ · Apr 23, 2010 · Anesthesiology

Device Facts

Record IDK100528
Device NameBREATHE TECHNOLOGIES VENTILATOR AND ACCESSORIES BT-V2S
ApplicantBreathe Technologies, Inc.
Product CodeONZ · Anesthesiology
Decision DateApr 23, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Breathe Technologies™ Ventilator BT-V2S, with accessories, is a volume assist ventilator intended to aid patients with respiratory insufficiency. It is designed for patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occupational therapy, respiratory therapy, and other rehabilitation efforts in an institutional environment. The device is intended for operation by trained personnel under the direction of a physician

Device Story

BT-V2S is a battery-powered, wearable volume ventilator for institutional use; designed to augment spontaneous breathing in tracheostomy patients. Device inputs include patient inspiratory effort, which triggers volume delivery via the Breathe Technologies Patient Circuit (BT-PC). Lightweight (1 lb) form factor allows patients to wear the device on a belt or shoulder strap, facilitating ambulation and participation in physical, occupational, and respiratory therapies. Operated by trained personnel under physician supervision. Output consists of physician-prescribed tidal volume delivered to the patient. Benefits include increased patient mobility and rehabilitation participation compared to standard stationary ventilators. Substantially equivalent to the BT-VS predicate; utilizes similar volume assist ventilation principles but features updated mechanical components for reduced weight and size.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing and biocompatibility testing in compliance with ISO 10993 and relevant medical device standards.

Technological Characteristics

Battery-powered, wearable volume ventilator. Materials comply with ISO 10993. Patient circuit is single-use and ethylene oxide sterilized. Operates via patient-triggered inspiration to deliver volume assist. Weight: 1 lb. Connectivity/software details not specified beyond general mention of software equivalence to predicate.

Indications for Use

Indicated for adult patients with respiratory insufficiency and a tracheostomy capable of spontaneous breathing (min tidal volume 3.5cc/kg predicted body weight). Intended for institutional use during ambulation and rehabilitation therapies. Operated by trained personnel under physician direction.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ # APR 2 3 2011 ### Breathe Technologies Inc. Breathe Technologies BT-V2S Special 510(k) Premarket Notification | Device Information: | | |------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | Category | Comments | | Sponsor: | Breathe Technologies<br>4000 Executive Parkway, Ste. 190<br>San Ramon, CA 94583<br>Tel: 925-359-1500 | | Correspondent Contact<br>Information: | Suzon Lommel<br>Breathe Technologies<br>4000 Executive Parkway, Ste. 190<br>San Ramon, CA 94583<br>Tel: 925-359-1508<br>Fax: 925-886-8622 | | | Device Common Name: | | Device Classification Number: | 21 CFR 868.5895 | | Device Classification &<br>Product Code: | Class II,<br>ONZ | | Device Proprietary Name: | Ventilator (BT-V2S) | ### Section 5: 510(k) Summary #### Predicate Device Information: [repeat table for each predicate] | Predicate Device: | BT-VS | |-------------------------------------------|-------------------------------------| | Predicate Device Manufacturer: | Breathe Technologies Inc. | | Predicate Device Common Name: | Continuous Ventilator, Facility Use | | Predicate Device Premarket Notification # | K082982 | | Predicate Device Classification: | 21 CFR 868.5895 | | Predicate Device Classification & | Class II, ONZ | | Product Code: | ONZ | #### b. Date Summary Prepared 22 February 2010 #### c. Description of Device #### c.1. Intended Patient Populations with Medical Condition Breathe Technologies™ has identified a group of ventilator dependent, or highly oxygen dependent patients, in institutional settings that are being poorly served by standard ventilators excluding the predicate Breathe Technologies BT-VS. This group needs ventilation support but only at low levels. They need to ambulate within the facility and participate in respiratory, physical and occupational therapy. The ease of carrying the 1 pound ventilator, while their inspiratory efforts are properly supplemented, may allow for ambulation for these patients. It is intended only for institutional use. {1}------------------------------------------------ ### c.2. General Description of the Breathe Technologies Ventilator and Patient Circuit (BT-V2S) The Breathe Technologies™ Ventilator (BT-V2S) is a battery powered (which may be charged during use) wearable, volume ventilator that augments the patient's spontaneous breathing. The BT-V2S administers this physician-prescribed volume to the patient via the attached Breathe Technologies Patient Circuit (BT-PC). The end of the BT-PC is inserted into the patient's tracheostomy tube. The ventilator is small and light enough to be worn on a patient's belt, or slung over their shoulder. ### d. Intended Use The Breathe Technologies™ Ventilator BT-V2S, with accessories, is a volume assist ventilator intended to aid patients with respiratory insufficiency. It is designed for patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occupational therapy, respiratory therapy, and other rehabilitation efforts in an institutional environment. The device is intended for operation by trained personnel under the direction of a physician ### e. Comparison to Predicate Device The BT-V2S is substantially equivalent in design, performance, software, materials, and intended use to the predicate device BT-VS, K#0802982 cleared July 20, 2009. Both devices provide volume assist ventilation through a patient circuit inserted into a trans-tracheal tube to aid patients with respiratory insufficiency. Both devices supply oxygen supplemented by entrained ambient air through the patient circuit. Both devices are triggered by a patients' inspiration. The devices use different mechanical components allowing for a lighter weight and smaller footprint. The predicate device weighs 3.1lbs while the subject device weighs 11b. The Company device may be more appropriate for patients as it is easier and lighter when worn. The testing described below demonstrates that the differences in the devices do not raise any unresolved issues of safety or efficacy. {2}------------------------------------------------ ## Breathe Technologies Inc. Both ventilators may be cleaned using the same method and both patient circuits are single use only and ethylene oxide sterilized. Company concludes that the devices are substantially equivalent. #### f. Summary of Supporting Data Biocompatibility data demonstrates that the device is in compliance with ISO 10993. Bench testing has demonstrated that the device is in compliance with the medical community's expectations, the product labeling and pertinent sections of the guidance's and standards. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Suzon Lommel Vice President, Regulatory and Quality Affairs Breathe Technologies, Incorporated 4000 Executive Parkway, Suite 190 San Ramon, California 94583 APR 2 3 2010 Re: K100528 Trade/Device Name: Ventilator, BT-V2S with Accessories Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: ONZ Dated: April 20, 2010 Received: April 21, 2010 Dear Ms. Lommel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2- Mr. Lommel Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. h for Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Section 4: Indications for Use Statement 510(k) Number (if known): Device Name: Breathe Technologies Inc. Ventilator, BT-V2S with accessories The Breathe Technologies™ Ventilator BT-V2S, with accessories, is a volume assist ventilator intended to aid adult patients with respiratory in u vouln'is assist patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occurational therapy, respiratory therapy, and other rehabilitation efforts in an institutional environment. The device is intended for operation by trained personnel under the direction of a plysician Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of L. Schulten (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number:
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...