The Breathe Technologies™ Ventilator, with accessories, is a volume assist ventilator intended to aid adult patients with respiratory insufficiency. It is designed for patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occupational therapy, respiratory therapy, and other rehabilitation efforts in an institutional environment. The device is intended for operation by trained personnel under the direction of a physician.
Device Story
Breathe Technologies Ventilator (BT-V) is a battery-powered, wearable volume assist ventilator; augments spontaneous breathing in tracheostomy patients. Device inputs: patient's spontaneous inspiratory effort. Device outputs: physician-prescribed volume delivered via Transtracheal Tube (BT-TT). Used in institutional settings; operated by trained personnel under physician direction. Small, lightweight (3.1 lb) form factor allows patient mobility, ambulation, and participation in physical/occupational therapy. Benefits include increased patient movement and rehabilitation potential compared to bulkier, full-featured ventilators. Healthcare providers use device to supplement inspiratory efforts, facilitating patient activity while maintaining respiratory support.
Clinical Evidence
Bench testing only. Biocompatibility testing performed in compliance with ISO 10993. Bench testing demonstrated compliance with medical community expectations, product labeling, and relevant guidance and standards. No clinical data provided.
Technological Characteristics
Battery-powered, wearable volume assist ventilator. Dimensions/weight: 3.1 lbs. Connectivity: None stated. Sterilization: Not specified. Software: Embedded firmware for volume assist. Biocompatibility: ISO 10993 compliant. Sensing/Actuation: Volume-based ventilation triggered by patient spontaneous breathing.
Indications for Use
Indicated for adult patients with respiratory insufficiency and a tracheostomy capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. Intended for continuous use in institutional settings to support ambulation and rehabilitation.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
K100528 — BREATHE TECHNOLOGIES VENTILATOR AND ACCESSORIES BT-V2S · Breathe Technologies, Inc. · Apr 23, 2010
K131562 — BREATHE NIOV VENTILATION SYSTEM · Breathe Technologies · Mar 6, 2014
K103345 — BREATHE TECHNOLOGIES VENTILATOR AND ACCESSORIES (BT-V2S) · Breathe Technologies, Inc. · Feb 11, 2011
K141943 — BREATHE TECHNOLOGIES LIFE2000 VENTILATION SYSTEM · Breathe Technologies · Jun 12, 2015
K102525 — BREATHE TECHNOLOGIES VENTILATOR AND ACCESSORIES (BT-V2S) · Breathe Technologies, Inc. · Dec 2, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K0829982
# JUL 2009
| Category | Comments |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Sponsor: | Breathe Technologies<br>3089 Skyway Court<br>Fremont, CA 94539-5909<br>Tel: 510.360.9966<br>Fax: 510.360.9967 |
| Correspondent Contact<br>Information: | Suzon Lommel<br>Breathe Technologies<br>4000 Executive Parkway Suite 190<br>San Ramon, CA 94583 |
| Device Common Name: | Continuous Ventilator, Facility Use |
| Device Classification & Code: | Class II, CBK |
| Device Classification Name: | 21 CFR 868.5895 |
| Device Proprietary Name: | Ventilator |
## Section 5: 510(k) Summary
#### Predicate Device Information:
| Predicate Device: | LTV 1000 & 800 Models |
|-----------------------------------------|-------------------------------------|
| Predicate Device Manufacturer: | Pulmonetic Systems |
| Predicate Device Common Name: | Continuous Ventilator, Facility Use |
| Predicate Device Classification: | 21 CFR 868.5895 |
| Predicate Device Classification & Code: | Class II, CBK |
| Premarket Notification Number: | K981371 |
| Predicate Device: | Legendair XL2 |
|-----------------------------------------|-------------------------------------|
| Predicate Device Manufacturer: | Puritan Bennett |
| Predicate Device Common Name: | Continuous Ventilator, Facility Use |
| Predicate Device Classification: | 21 CFR 868.5895 |
| Predicate Device Classification & Code: | Class II, CBK |
| Premarket Notification Number: | K070899 |
## b. Date Summary Prepared
10 June 2009
#### c. Description of Device
## c.1. Intended Patient Populations and Medical Conditions
Breathe Technologies has identified a group of ventilator dependent, or highly oxygen dependent patients, in institutional settings that are being poorly served by current ventilators. This group needs ventilation support but only at low levels. They need mobility within the facility. "Mobility" includes modest movement such as free motion in
{1}------------------------------------------------
bed, standing and sitting to use a commode, and movement to the bathroom. It may also, at the physicians' discretion, allow a level of ambulation to support participation in physical therapy, but the weight and bulk of current ventilators greatly inhibit their potential for physical movement. The ease of carrying the 3.1 lb ventilator, while their inspiratory efforts are properly supplemented, may allow the potential for ambulation for these patients. It is intended only for institutional use, not for home use.
### c.2. General Description of the Breathe Technologies Ventilator, Transtracheal Tube and accessories
The Breathe Technologies™ Ventilator (BT-V) is a battery powered, wearable, volume ventilator that augments the patient's spontaneous breathing.
The BT-V administers this physician-prescribed volume to the patient via the attached Breathe Technologies Transtracheal Tube (BT-TT). The end of the BT-TT is inserted into the patient's tracheostomy tube.
The ventilator is small and light enough to be worn on a patient's belt, or slung over their shoulder, because it has only a subset of the features of a full-featured ventilator.
#### d. Intended Use
The Breathe Technologies™ Ventilator (BT-V), with accessories, is a volume assist ventilator intended to aid adult patients with respiratory insufficiency. It is designed for patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The BT-V is designed for continuous applications in an institutional environment. The BT-V is a restricted medical device intended for operation by trained personnel under the direction of a physician.
#### e. Comparison to Predicate Device
The Breathe Technologies Ventilation System (Ventilator and Accessories) is substantially equivalent to the predicate Pulmonetic Systems' LTV 1000 & 800 Models (K981371) and the Puritan Bennett Legendair XL2 (K070899) in intended use, classification code, technology design and physician's use in that it conforms to a subset of the intended use, technology, physician use, features, performance and alarms of the predicate devices.
By conforming to a subset of design and use of the predicate devices, the Breathe Technologies Ventilation System can be used by some spontaneously breathing, ventilator dependent patients, to ambulate in an institutional setting.
Breath Technologies concludes that the predicate devices and the Breathe Technologies Ventilator System (Ventilator and all accessories) are substantially equivalent.
{2}------------------------------------------------
## f. Summary of Supporting Data
Biocompatibility data demonstrates that the device is in compliance with ISO 10993.
Bench testing has demonstrated that the device is in compliance with the medical community's expectations, the product labeling and the pertinent sections of the guidance's and standards.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three diagonal lines that curve at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus symbol. The text is in all capital letters and is evenly spaced around the circle.
#### SEP 9 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Suzon Lommel Vice President, Regulatory and Quality Affairs Breathe Technologies, Incorporated 4000 Executive Parkway, Suite 190 San Ramon, California 94583
Re: K082982
Trade/Device Name: Breathe Technologies Ventilator & Accessories Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: ONZ Dated: June 25, 2009 Received: June 30, 2009
Dear Ms. Lommel:
This letter corrects our substantially equivalent letter of July 20, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Lommel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Susan Runser
Susan Runner, DDS, MA Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Breathe Technologies. The word "Breathe" is in large, bold letters on the top line. Below that, in smaller letters, is the word "TECHNOLOGIES". Underneath the words is a curved line.
# Section 4: Indications for Use Statement
510(k) Number K082982:
Breathe Technologies Ventilator & Accessories Device Name:
## Indications For Use:
The Breathe Technologies™ Ventilator, with accessories, is a volume assist ventilator intended to aid adult patients with respiratory insufficiency. It is designed for patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occupational therapy, respiratory therapy, and other rehabilitation efforts in an institutional environment. The device is intended for operation by trained personnel under the direction of a physician.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) ห Schuts
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of
: 510(k) Number: K082982
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.