← Product Code [CBP](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBP) · K955634

# MONTGOMERY TRACHEOSTOMY SPEAKING VALVE (K955634)

_Boston Medical Products, Inc. · CBP · Jun 10, 1996 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBP/K955634

## Device Facts

- **Applicant:** Boston Medical Products, Inc.
- **Product Code:** [CBP](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBP.md)
- **Decision Date:** Jun 10, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5870
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

For use as a one-way speaking valve between a tracheostomy tube and ventilation equipment.

## Device Story

Montgomery Tracheostomy Speaking Valve; in-line device between tracheostomy tube and ventilator; features thin silicone diaphragm; opens on inspiration, closes on expiration; allows ventilator-dependent patients to speak during both phases of ventilation cycle without waiting for ventilator cycling; used in clinical settings; operated by clinicians/caregivers; output is restored speech capability for patient.

## Clinical Evidence

Bench testing only; visual inspection for defects; testing with ISO 5356-1 test gauges for 15mm and 22mm connections.

## Technological Characteristics

Silicone diaphragm; 15mm and 22mm connections; ISO 5356-1 compliant connections; mechanical one-way valve operation.

## Regulatory Identification

A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.

## Predicate Devices

- Tracheostomy Speaking Valve (15mm / 22mm), Product No. 007 (Passy and Passy)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

JUN 10 1996

K955634

# SMDA Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Product: Montgomery Tracheostomy Speaking Valve (15mm / 22mm)

Description: Montgomery® Tracheostomy Speaking Valves are designed to offer ventilator dependent patients the opportunity to speak without having to wait for the ventilator to cycle, allowing patients to speak during inspiration and expiration without interruption. The valve is placed in-line between the tracheostomy tube and the ventilator and features a thin silicone diaphragm which opens on inspiration and closes on expiration.

Indications For Use: For use as a one-way speaking valve between a tracheostomy tube and ventilation equipment.

Predicate Device: Tracheostomy Speaking Valve (15mm / 22mm), Product No. 007 marketed by Passy and Passy (4521 Campus Drive, Irvine, CA).

Testing: Each valve is visually inspected for defects and tested with ISO 5356-1 test gauges for 15mm and 22mm connections prior to packaging.

Submitted by: Stuart K. Montgomery, President

Date: April 9, 1996

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBP/K955634](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBP/K955634)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
